1. COMPONENTS INCLUDED


A. Description
Blade comprising the following components:

Design height: 220 mm.
This medical device is supplied to health professionals (orthopaedic prosthetists), who will train the patient in their use. The prescription is drawn up by a doctor, alongside the orthopaedic prosthetist, to assess whether the patient is suited to using it. This device is for SINGLE-PATIENT use. It must not be reused on another patient.

This device is designed to be mounted preferably on the tibial socket or agenesis, but can still be mounted on the femoral socket for individuals with a high activity level (4). Maximum weight (including load borne): 110 kg
The device will enable an amputee to exercise (jogging or running), thanks to its shock absorption and energy return functions.

USER MANUAL Page 2 of 4

(15 mm above the articular interline)
Preparation:
Trace the load lines and knee joint centre on the socket. Alignment in sagittal plane:


The load line must run through the knee joint centre marked on the socket, and 60 to 70 mm back from the blade contact point with the ground. The knee-ground mounting height must be 30 to 50 mm more than the height on the patient (including shoe). No flexum. If the patient has genuine flexum, this must be respected. Alignment in frontal plane:
Knee height from ground with shoe +30 to 50 mm

The construction line must run via the middle of the socket and the middle of the blade (the blade must have a few degrees’ external rotation). Finishing:

After setting and alignment, glue the screws (XC047 medium threadlock), and torque tighten to 6 Nm.
1. Jumping from one foot to the other: the blade must press flat on the ground (frontal view). 2. Hopping: the blade must drive upwards (not to the front or rear). If necessary, modify the blade angle. In every case, the patient must feel that their knee is free-moving and easy to control (no hyperextension caused when the blade is pressed on the ground). 3. Running tests: the objective is to limit muscle and ligament stress, and reduce the patient’s energy consumption. To do so, you can adjust the height, and the frontal and sagittal angles, according to how the patient feels. The aim is to obtain stride symmetry at the patient’s running speed.
Action: replace the module. For a fine adjustment, you can move the blade forward in relation to the socket. • Patient feeling impacts, discomfort: the blade module is probably too stiff.
Action: replace the module. To fine-tune, you can move the blade backward in relation to the prothesis. • Patient feeling outward or inward tilting of the prothesis: the blade is probably too far inward or outward.
Action: incline it in the direction of the valgum/varum, or shift it across. • Patient experiences falling sensation before contact (shoulder dropping): The prothesis is probably too short.
Action: increase the height of the prothesis. • Patient feeling difficulty striding: the prothesis is probably too long.

4 x M6 countersunk screws
Modifying the orientation of the pyramid, or replacing the connector:
Connector plate (th. 1 mm)
Shim (th. 2 mm)
If you observe abnormal behaviour or feel modifications in the device’s properties, or if it undergoes a heavy impact, please consult your orthopaedic prosthetist.

USER MANUAL Page 3 of 4
A. Warnings
Using the device in a way that goes against the advice given by your orthopaedic prosthetist may cause deterioration of its components (e.g. bearing heavy loads, excessive strain, using beyond the service life, etc.).
No greasing, adjustments to the fasteners or any other operation is required, except by the orthopaedic prosthetist.
There are specific risks due to lower limb amputation (falls, trophic disorders of the residual limb through conflict with the prothesis socket, etc.). Running training with an orthopaedic prosthetist, physiotherapist or parasports club is essential.
Running on a conveyor belt may is only permissible with in-person monitoring, and not as part of solo training.
Check the whole prothesis whenever it has been used (screwed joint, no perceptible gap).
Do not modify the product (trimming or drilling, etc.)
Do not heat up the blade with a hot air gun, do not machine the blade to reduce its cross-section, and do not sand the blade too heavily when replacing the sole (risk of removing the upper carbon layer).
Do not expose the prothesis to a flame or powerful heat source (radiator, etc.)
The device is resistant to soft, salty and chlorinated water, provided that the care instructions are followed (§9.A).
This device is not designed for activities with an excessive overload risk or jumps from significant heights, for example.
Adhere to the exercise instructions (running, jogging) when using the product
Do not use the product without the sole provided, or if it is detached or damaged.
There are no side effects directly linked to the device. Any serious incident arising in relation to the device must be reported to the manufacturer and to the competent authority of the Member State.
A. Care/ cleaning
Clean the envelope with a damp cloth or sponge.
No greasing, adjustments to the fasteners or any other operation is required for this blade, except by the orthopaedic prosthetist.
The presence of sand, particles or foreign bodies inside the foot could modify its behaviour and damage it. After use in a very dirty environment:
The sole and under-layer may require replacement by the orthopaedic prosthetist: green kit (1A92013-1) or black kit (1A92013-2)

Detach the sole, and clean the glued surface (do not use any solvent, to avoid ruining the aesthetics of the strip).
Manually sand the surfaces before gluing the under-layer onto the strip, and do the same to glue the sole on the under-layer (XC090 neoprene glue).
The various components of the blade are special waste: Elastomer, epoxy resin-based composite, titanium, stainless steel. They must be treated in accordance with the locally applicable legislation. D. Service life
It is advisable to arrange for an orthopaedic prosthetist to perform an annual check.

USER MANUAL Page 4 of 4
CE marking and year of 1st declaration
This product is a CE marked medical device, and certified compliant with Regulation (EU) 2017/745
PROTEOR SAS 6 rue de la Redoute – 21850 Saint-Apollinaire – France Tel.: +33 3 80 78 42 42 – Fax: +33 3 80 78 42 15 cs@proteor.com – www.proteor.com
Manufacturer · Risk identified
In dieser Anleitung genannte Seiten – Hersteller, Support und Downloads.
Besitzer und unser Team antworten in der Regel innerhalb eines Tages – und Ihre Frage hilft auch anderen weiter.
Stellen Sie die erste Frage — andere Nutzer oder unser Team antworten Ihnen.