Provide and explain § 3, 7, 8, and 9 to the Patient. 1. INCLUDED ITEMS



A. Description
The Plié® 3 MPC Knee is a single axis prosthetic knee joint system providing microprocessor control of both the swing and stance phases of gait. The microprocessor monitors an embedded load sensor and an angle sensor to precisely control the transitions between the stance and swing phases of gait. Three manual settings allow the hydraulic cylinder to provide adjustable resistance for Stance Flexion, Swing Flexion, and Swing Extension. The hydraulic cylinder also provides non- adjustable stance extension resistance. The Plié Control software allows the knee function to be optimized for each individual user’s gait, including the stumble recovery parameters. The Gait Lab software provides the prosthetist with access to recorded data files of the microprocessor for analysis and documentation.
Device Information



The expected lifetime of the Plié 3 MPC Knee is defined by the warranty period and contingent on adherence to the service plan. 3. INTENDED USE/INDICATIONS
The Plié 3 MPC Knee is appropriate for users who would benefit from the safety inherent in the stability of a microprocessor- controlled knee. This medical device is supplied to healthcare professionals (prosthetists) who will train the patient in its use. The prescription is made by a doctor together with the prosthetist, who assess the patient's ability to use it.
This device is for SINGLE PATIENT use. It should not be reused on another patient.
The Plié® 3 MPC Knee is intended for use as a component in a prosthetic leg for individuals with lower-limb loss including:

The Plié 3 MPC Knee is appropriate for users who would benefit from the safety inherent in the stability of a microprocessor-controlled knee. These users should also have the ability or have the potential to:
The device incorporates a male pyramid link that allows it to be used with female pyramid connectors for the Pyramid Top version and a M36 x 1.5 thread for the Threaded Top version. 6. ASSEMBLY AND PATIENT FITTING


To open the battery cap, press down latch, this will cause the cap to spring open. To close the battery cap, press firmly on entire cap until latch engages. Do not submerse the knee in water when the battery cap is open at any time.
Cap is Closed, Locked and Watertight Cap is Unlocked and Opened


The following picture show the proper steps to open and close the battery cap.
Steps to Open or Close the Cap

To Open: Press down latch (the cap will spring open). To Close: Press firmly on entire cap until latch engages.
The following pictures shows the proper steps to utilizing the air pump.




Follow the Prosthetist Setup Guide to properly align and adjust the device. Two installation programs, Plié Control and Plié Gait Lab, are provided to optimize the computer settings for each patient. Verify the computer system minimum requirements before software installation (Microsoft Windows 7 or later operating system, 1.0 MHz Processor, 256K RAM and USB Port).

Careful attention to the alignment of the socket in relation to the Plié® 3 MPC Knee and the prosthetic foot is essential for a successful user outcome. The prosthesis alignment should account for the range of motion (ROM), voluntary control, and balance of individual users. Proper alignment and user voluntary control are essential to the optimal function of the prosthesis.
Prior to assembly of the prosthesis, the prosthetist should measure the patient’s hip joint range of motion (ROM) on the prosthetic side to determine if the user has a hip flexion contracture (Figure 1). If present, the user’s hip flexion contracture should be accommodated by attaching the socket to the prosthesis with an appropriate amount of flexion (Figure 2). Failure to sufficiently accommodate a hip flexion contracture can compromise the patient’s function during standing and ambulation (Figures 3 and 4).
During standing, an unaccommodated hip flexion contracture may prevent the patient from standing straight, compromising the patient’s balance (Figure 5). Additionally, an unaccommodated hip flexion contracture can cause excessive lumbar lordosis, compromising the structural integrity of the patient’s spinal column (Figure 4).
During ambulation, an unaccommodated hip flexion contracture can cause an excessively asymmetrical gait pattern.
The Plié® 3 MPC Knee should be in a fully extended position during stance phase for level ground ambulation. An inherently stable trochanter-knee-ankle (TKA) alignment is essential to a successful user outcome with the Plié MPC Knee. A plumbline dropped from the trochanter reference point on the lateral socket should fall at or up to 5mm anterior to the knee joint axis (Figure 5). Align the foot as recommended by the foot manufacturer. An inherently unstable trochanter-knee-ankle (TKA) alignment can cause a rapid extension moment at the knee joint after mid stance. The non-adjustable stance extension resistance could be insufficient to dampen the extension moment if the user does not exert sufficient voluntary control and/or if the alignment recommendations are not implemented. The recommended inherently stable alignment of the Plié 3 MPC Knee may differ from other knee joints. Consequently, if a new socket is not fabricated for use with a Plié 3 MPC Knee, the socket attachment component may require re-lamination to the socket to achieve the recommended alignment

Figure 1 Figure 2 Figure 3 Figure 4 Figure 5 • Ensure the transverse rotation of the knee and foot is appropriate.
The Plié Control 6 software is required to program a Plié 3 MPC Knee. Download software from the following website to install the program at https://www.pliesupport.com/download This software is for use by prosthetists to program the knee. This software should not be used by the patient.

Thoroughly read and follow the instructions as stated in the Prosthetist Setup Guide and Plié Setup Wizard to set up and program the knee. Utilize the “Help” function in the Plié Control software to assist with issues that may arise.
A discontinuous, two-piece cover is recommended for cosmetic finishing. Care should be taken to ensure that the battery compartment and manual adjustments are accessible. 7. DETECTION OF MALFUNCTIONS
If you notice any abnormal behavior or feel any changes in the characteristics of the device, or if the device has received a severe impact, consult your prosthetist. 8. WARNINGS, CONTRAINDICATIONS, AND SIDE EFFECTS
Inappropriate use of the device, in relation to the recommendations of your prosthetist, can cause the degradation of parts of the device (carrying heavy loads for example, excessive stress, exceeding the service life, etc.). Never attempt to loosen the bolts affixing the pyramid connector. If unusual movement or product wear is detected in a structural part of a prosthesis at any time, immediately discontinue use of the device and consult a clinical specialist. Water resistant: The device is safe for occasional submersion in fresh water up to 1 m (3 ft) for up to 30 minutes. The device should not be used for bathing and should not be immersed in salt water or chlorinated water as these may cause corrosion.
Failure to follow the safety precautions can result in device malfunctioning and risk injury to the user.
The device is inappropriate for users with:
There are no known negative side effects. Any serious incident that has occurred in relation to the device should be reported to the manufacturer and the competent authority. 9. MAINTENANCE, STORAGE, DISPOSAL, AND DURABILITY
No maintenance operation such as lubrication, work on the screws, or other parts is required. Inspect the device every six months. If the user is more active, more frequent inspection may be necessary. Service as necessary. After use in water:
Battery temperature range for charging: 0°C to 45°C (32°F to 113°F) Storage and shipping temperature range: -20°C to 80°C (-4°F to 176°F) Operating temperature range: -5°C to 45°C (23°F to 113°F) Storage and operating relative humidity range: 0% to 100%, including condensation Storage and operating pressure range: 500 hPa to 1060 hPa (7.3 psi to 15.4 psi) Water and Dust Resistance: Rated to IP67 per IEC 60529 (Protected against dust and the effect of submersion in water up to 1m for 30 minutes)
The different items of the device are special wastes and must be handled according to local laws.
For Technical Support in the USA, please contact PROTEOR USA toll-free at 1-855-450-7300. For Technical Support in the EU, please contact PROTEOR France at +33 3 80 78 42 08.
The Plié® 3 MPC Knee purchase includes a 36-month warranty covering all manufacturer defects effective only if the product is used according to manufacturer recommendations. The batteries, battery charger, and accessories are provided with a 12- month warranty. An extended warranty is available. See product catalog for details.

For the warranty to remain in effect, the knee must be serviced 12 and 24 months after purchase. For knee service or repair, please contact PROTEOR. A loaner knee will be provided upon request. For service and/or repair, please ship the knee in the provided shipping case to the following address:
PROTEOR France Rue du Cheffin 21250 Seurre France +33 3 80 78 42 08 cs@proteor.com 10. DESCRIPTION OF SYMBOLS
PROTEOR USA 425 East 400 North (PO Box 50) Gunnison, UT 84634 USA Toll-free: 1-855-450-7300 support@proteorusa.com
Warning
CE marking and year of 1st declaration
Type BF Applied Part
FCC Declaration of Conformity
WEEE Directive on waste electrical and electronic equipment, which should not be disposed of in regular wastes at the end of its usable life
This product is tested and certified to comply with the MDR 2017/745 (EN 55011 Class B and EN60601-1 and EN60601-1-2), STSI EN 300-328 under R&TTE Directive 1999/5/EC and, ISO 10328 - P6 - 125 kg. The device complies with Part 15 of the FCC Rules and carries the CE mark