Teledyne Analytical Instruments HIGHL Y TO XIC AND OR FLAMMABLE LIQUIDS OR GASES MA Y BE PRESENT IN THIS MONIT ORING SYSTEM. PERSONAL PRO TECTIVE EQUIPMENT MA Y BE REQUIRED WHEN SER VICING THIS SYSTEM. HAZARDOUS V OL T AGES EXIST ON CER T AIN COMPONENTS INTERNALL Y WHICH MA Y PERSIST FOR A TIME EVEN AFTER THE PO WER IS TURNED OFF AND DISCONNECTED .
INSTRUCTION MANUAL FOR TRACE OXYGEN ANALYZER MODEL 311XL TELEDYNE ANALYTICAL INSTRUMENTS 16830 CHESTNUT STREET CITY OF INDUSTRY, CA 91749 TELEPHONE: (888) 789-8168 (626) 934-1500 (626) 961-9221 FAX: (626) 961-2538 (626) 934-1651 Web: www.
WARNING The sensor(s) used in this instrument uses electrolytes which contain substances that are extremely harmful if touched, swallowed, or i nhaled. Avoid contact w ith ANY fluid or powder in or around the unit. What may appear to be plain water could contain one of these toxic substances.
TABLE OF CONTENTS 311XL SERIES 1. INTRODUCTION 1 1.1 Description 1 1.2 Method of Analysis 1 1.3 Outstanding Features 1 1.3.1 Micro-Fuel Cell 1 1.3.2 Reliable Calibration 2 1.3.3 Integral Power Supply 2 1.3.4 Accuracy and Response 3 1.3.5 Compact Packaging 3 2.
311XL SERIES TRACE OXYGEN ANALYZER 1. INTRODUCTION 1.1 Description : The Teledyne Analytical Instruments (TAI) Model 311XL and 311TCXL are portable, intrinsically safe trace oxygen analyzer which can be operated without an external power source. These analyzers w ill be referred to as Model 311XL or 311XL Series analyzers in this manual.
1.3.2 Reliable Calibration . Although this instrument can be calibrated with air, the use of a calibration gas gives better results on the lower PPM ranges.
A current limiting resistor (R1 and R2) is potted into the end of each battery. This assures that under no circumstances can more than 25 m illiamp eres (100 m illiwatts) be switched or drawn from either battery supply. This means that the Model 311XL m ay be used in explosive atmospheres where arcs of 100 milliwatts or less can be tolerated.
2. SUPPORTING EQUIPMENT AND SERVICES 2.1 Sampling Equipment . The customer must provide a m eans of controlling the pressure and flowrate of the sample gas. For positive pressure applications, TAI suggests a simple throttle valve installed in the sample line between the sample point and the analyzer.
3.1 Introduction . The Model 311XL is supplied completely assembled and ready for instant use. The Micro-Fuel Cell must be installed prior use. When a sensor is first installed, the instrument should be purged for 12-16 hours with zero gas (an inert gas such as nitrogen or argon with less than 0.
Span gas: An inert gas (nitrogen or argon) with a known trace amount of oxygen. The oxygen content should be 70 – 90% of the full scale on the range of interest (working or primary range).
2) Establish a flowrate in the sample line of from 0.2 to 5 liters/m in. Allow the sample to flow long enough to allow a stable low reading to be achieved. When disconnecting the instrument, reverse the procedure: source f itting first, and then vent fitting.
that an overnight recharge be accomplished ev ery three (3) weeks of continuous use. To recharge the batteries, place the range switch in the “OFF” position and connect the power cord to a convenient outlet. NOTE: The amber charge lamp (on the rear of the case) will be lit during charging process.
The spare cell should be carefully stored in an ar ea that is not subject to large variations in ambient temperature (75 Deg. F nom inal), and in such a way as to obviate any possibility of damage. Under no circumstances, disturb the integr ity of the cell package until the cell is to be actually used.
5. TRANSDUCER AND TEMPERATURE COMPENSATION The Micro-Fuel Cell has an inherent positive temper ature coeff icient, the effects of which have been minimized through the im plementation of a thermistor com pensation circuit. Internal electronic calibration is accomplished by TAI.
-11- 7. PRODUCT SPECIFICATION DATA RANGES: 0-2, 10, 100, 1000 PPM Oxygen and Cal range for air calibration ACCURACY:.
Accuracy at 25 o C and constant pressure after thermal equilibrium has been reached. Basic accuracy +/- 2% of full scale for all ranges except 0-2 PPM range. Accuracy for 0-2 PPM range (+/- 5% of full scale). Accuracy over the temperature range 0-35C at constant pressure after thermal equilibrium has been reached.
1 F-51 MICRO-FUSE, 0.5 AMP., FAST-BLOW (FOR 100-115 VAC) 1 C6689 -B2CXL MICRO-FUEL CELL, CLASS B-2CXL 2 B-9905 BATTERY 1 L-79 LAMP 1 M70 METER. 1 B-70283 DIAL-RANGE OF 0-2, 0-10 ppm O2 & CAL. (USE WITH M70 METER) 1 A36289 CALIBRATION KIT 1 A205 OP-AMP A3 (LF441) 1 A246 OP-AMP A1 (MAX 430) A MINIMUM CHARGE IS APPLICABLE TO SPARE PARTS ORDERS.
C-70246 SCHEMATIC D-70268 OUTLINE C-14627 W IRING DIAGRAM (110 VAC) C-10470 W IRING DIAGRAM (220VAC) DRAWING LIST MODEL 311TCXL C-70248 SCHEMATIC D-70270 OUTLINE C-41661 W IRING DIAGRAM (110 VAC) C-37.
Ein wichtiger Punkt beim Kauf des Geräts Teledyne ECO#02-0227 (oder sogar vor seinem Kauf) ist das durchlesen seiner Bedienungsanleitung. Dies sollten wir wegen ein paar einfacher Gründe machen:
Wenn Sie Teledyne ECO#02-0227 noch nicht gekauft haben, ist jetzt ein guter Moment, um sich mit den grundliegenden Daten des Produkts bekannt zu machen. Schauen Sie zuerst die ersten Seiten der Anleitung durch, die Sie oben finden. Dort finden Sie die wichtigsten technischen Daten für Teledyne ECO#02-0227 - auf diese Weise prüfen Sie, ob das Gerät Ihren Wünschen entspricht. Wenn Sie tiefer in die Benutzeranleitung von Teledyne ECO#02-0227 reinschauen, lernen Sie alle zugänglichen Produktfunktionen kennen, sowie erhalten Informationen über die Nutzung. Die Informationen, die Sie über Teledyne ECO#02-0227 erhalten, werden Ihnen bestimmt bei der Kaufentscheidung helfen.
Wenn Sie aber schon Teledyne ECO#02-0227 besitzen, und noch keine Gelegenheit dazu hatten, die Bedienungsanleitung zu lesen, sollten Sie es aufgrund der oben beschriebenen Gründe machen. Sie erfahren dann, ob Sie die zugänglichen Funktionen richtig genutzt haben, aber auch, ob Sie keine Fehler begangen haben, die den Nutzungszeitraum von Teledyne ECO#02-0227 verkürzen könnten.
Jedoch ist die eine der wichtigsten Rollen, die eine Bedienungsanleitung für den Nutzer spielt, die Hilfe bei der Lösung von Problemen mit Teledyne ECO#02-0227. Sie finden dort fast immer Troubleshooting, also die am häufigsten auftauchenden Störungen und Mängel bei Teledyne ECO#02-0227 gemeinsam mit Hinweisen bezüglich der Arten ihrer Lösung. Sogar wenn es Ihnen nicht gelingen sollte das Problem alleine zu bewältigen, die Anleitung zeigt Ihnen die weitere Vorgehensweise – den Kontakt zur Kundenberatung oder dem naheliegenden Service.