Victoreen ® 05-437 Primalert ® 35 Area Radiation Monitor Operators Manual March 2005 Manual No. 126011 Rev. 3 © 2003, 2005 Fluke Corporat i on, All rights reserved .
Fluke Biomedical Radiation Management Services 6045 Cochran Road Cleveland, Ohio 44139 440.498.2564 www.flukebiomedical.com/rms.
i Table of Contents Section 1: Introductio n ................................................................................................ 1-1 1.1 Product Descr iption ................................................................................
(Blank page).
Introduction Product Description 1 1-1 Section 1 Introduction 1.1 Product Description The Primalert® 35 Area Ra diation Monitor provides reliable, continu ous monitoring of radioacti ve materials. Six front panel, color-coded LED's (1, 2, 4, 8,16, and 32 mR/h) indicate an increase or decrease in radiation level.
Victoreen 05-437 Operators Manual 1-2 1.2 Specifications Detector Energy -compensated G-M tube Energy Dependence 10% to 40% from 50 KeV to 2 MeV, as referenced to 137 Cs Overload Protection Alarms in fields higher than 100 R/h wh en detector is completely blocked fo r pulse operation.
Introduction Specifications 1 1-3 Input 230 VAC, 50 Hz, 13.8 W Output 12 VDC, 580 mA Part 14-417, Part 14-416 (Australia) Dimensions 5.9 in. high x 3.4 in.
Victoreen 05-437 Operators Manual (Blank page).
Getting Started Receiving Inspection 2 2-1 Section 2 Getting Started 2.1 Receiving Inspection Upon receipt of the unit: 1. Inspect the carton(s) and contents for d amage . If damage is evident, file a claim with the carrier and notify Fluke Biomedical, Ra di ation Management Services at 440.
Victoreen 05-437 Operators Manual 2-2 2.4 Optional Power The Primapak II storage battery sup plies a source of 12-16 Volt DC powe r to the Primalert using the converter plug conne ction. The plug is a Switchcraft #760, or e quivalent, with the sleeve negative and pin positive.
Operation Set Up 3 3-1 Section 3 Operation 3.1 Set Up The only set-up required for the Primalarm is the att a chment of one end of the power cord (converter) to the unit and the other end into a properly gr ounded AC outlet as noted in Section 2. When power is first applied to the unit, the alarm and one or more of the LED's may light.
Victoreen 05-437 Operators Manual (Blank page).
Maintenance Maintenance 4 4-1 Section 4 Maintenance 4.1 Maintenance This instrument contains CMOS integrated circuits. No service should ever be attempted unless by a qualified service technici an thoroughly familiar with these devices.
Victoreen 05-437 Operators Manual 4-2 Table 4-2. PC Board Assembly (126021) Parts List Item No. Ref. No. Part No. Description Qty 6 V1 010054 Tube Shield Assembly 1 7 U1, 4 630001 IC, Quad Nor Gate, 4.
Maintenance Maintenance 4 4-3 46 R22, 23 302041 Resistor, Film, 200k, .25W, 5% 2 47 R20, 26,30, 32,33,38, 341041 Resistor, Film, 100k, .25W, 5% 6 49 R27, 29,31, 34,40 341032 Resistor, Film, 10k, .25W, 5% 5 50 R13-18, 28 185-1448 Resistor, Film, 200, .
Victoreen 05-437 Operators Manual 4-4 Figure 4-1 Primalert 35 Assembly (126028).
Maintenance Maintenance 4 4-5 Figure 4-2 Primalert 35 Schematic (126015).
Victoreen 05-437 Operators Manual 4-6 Figure 4-3 PC Assembly (126021).
.
Fluke Biomedical Radiation Management Services 6045 Cochran Road Cleveland, Ohio 44139 440.498.2564 www.flukebiomedical.com/rms.
Ein wichtiger Punkt beim Kauf des Geräts Fluke 05-437 (oder sogar vor seinem Kauf) ist das durchlesen seiner Bedienungsanleitung. Dies sollten wir wegen ein paar einfacher Gründe machen:
Wenn Sie Fluke 05-437 noch nicht gekauft haben, ist jetzt ein guter Moment, um sich mit den grundliegenden Daten des Produkts bekannt zu machen. Schauen Sie zuerst die ersten Seiten der Anleitung durch, die Sie oben finden. Dort finden Sie die wichtigsten technischen Daten für Fluke 05-437 - auf diese Weise prüfen Sie, ob das Gerät Ihren Wünschen entspricht. Wenn Sie tiefer in die Benutzeranleitung von Fluke 05-437 reinschauen, lernen Sie alle zugänglichen Produktfunktionen kennen, sowie erhalten Informationen über die Nutzung. Die Informationen, die Sie über Fluke 05-437 erhalten, werden Ihnen bestimmt bei der Kaufentscheidung helfen.
Wenn Sie aber schon Fluke 05-437 besitzen, und noch keine Gelegenheit dazu hatten, die Bedienungsanleitung zu lesen, sollten Sie es aufgrund der oben beschriebenen Gründe machen. Sie erfahren dann, ob Sie die zugänglichen Funktionen richtig genutzt haben, aber auch, ob Sie keine Fehler begangen haben, die den Nutzungszeitraum von Fluke 05-437 verkürzen könnten.
Jedoch ist die eine der wichtigsten Rollen, die eine Bedienungsanleitung für den Nutzer spielt, die Hilfe bei der Lösung von Problemen mit Fluke 05-437. Sie finden dort fast immer Troubleshooting, also die am häufigsten auftauchenden Störungen und Mängel bei Fluke 05-437 gemeinsam mit Hinweisen bezüglich der Arten ihrer Lösung. Sogar wenn es Ihnen nicht gelingen sollte das Problem alleine zu bewältigen, die Anleitung zeigt Ihnen die weitere Vorgehensweise – den Kontakt zur Kundenberatung oder dem naheliegenden Service.