Benutzeranleitung / Produktwartung 00825-0100-4593 des Produzenten Emerson
Zur Seite of 22
www.rosemount.com ¢00825-0100-4360N¤ Quick Installation Guid e 00825-0100-4593, Rev BA July 2009 Rosemount 1 151 Step 1: Mount the T ransmitter Step 2: Consider Housing Rotation Step 3: Connect the .
Quick Installation Guide 00825-0100-4593, Rev BA July 2009 Rosemount 1 151 2 W ARNING Explosions could result in death or serious injury: Installation of this tr ansmitter in a n explosive env iron ment must be in accordan ce wit h t he approp riat e local, national, and inter national st andards, code s, and pract ices.
Quick Installation Guid e 00825-0100-4593, Rev BA July 2009 Rosemount 1 151 3 S TE P 1: M OUNT THE T RANSMI TTER A. Applications Liquid Flow Application s 1. Place taps to the side of the line. 2. Mount beside or below the taps. Gas Flow Applicatio ns 1.
Quick Installation Guide 00825-0100-4593, Rev BA July 2009 Rosemount 1 151 4 S TE P 1 CONTI NUED ... B. Optional Mounting Brackets When installing the transmitter to one of the optional mounting brackets, torque the bracket bolts to 125 in.-lbs. (0,9 N-m).
Quick Installation Guid e 00825-0100-4593, Rev BA July 2009 Rosemount 1 151 5 S TE P 1 CONTI NUED ... C. O-rings with Flange Adapters W ARNING Whenever the flanges or adapters are removed, vi su ally inspect the o-rings. Replace them if there are any signs of damage, such as nicks or cuts.
Quick Installation Guide 00825-0100-4593, Rev BA July 2009 Rosemount 1 151 6 S TE P 2: C ON SIDER H OUSING R OT ATION T o improve field access or to better view the optional LCD d isplay: 1. Loosen the housing lock nut. 2. Rotate the housing clockwise to the desired posit ion – up to 90° from its o riginal position.
Quick Installation Guid e 00825-0100-4593, Rev BA July 2009 Rosemount 1 151 7 S TE P 3: C ON NECT THE W IRING AND P OWER U P Use the following steps to wire the transmitter: 1. Remove the housing cover on the side marked TERMINALS on the namepla te. 2.
Quick Installation Guide 00825-0100-4593, Rev BA July 2009 Rosemount 1 151 8 S TE P 3 CONTI NUED ... Power Supply The dc power supply should provide power with less than tw o percent ripple. The total resistance load is the sum of the resistance of the signal leads and the load resistance of the controller , indicator , and related pieces.
Quick Installation Guid e 00825-0100-4593, Rev BA July 2009 Rosemount 1 151 9 S TE P 4: C ON FIGURE THE T RANSMITTER Failure Mode Alarm Switch 1. Remove the housing cover . 2. Locate the failure mode switch (see Figure 3). 3. Move the switch to the desired alarm setting.
Quick Installation Guide 00825-0100-4593, Rev BA July 2009 Rosemount 1 151 10 S TE P 5: C AL IBRATE THE T RANSMITTER V erify T ransmitter Configuration NOTE: A check ( ✓ ) indicates the basic configuration para meters. At minimum, these parameters should be verified as part of the configuration and startup procedure.
Quick Installation Guid e 00825-0100-4593, Rev BA July 2009 Rosemount 1 151 11 S TE P 5 CONTI NUED ... Configure LCD Displa y Figure 4. Sample 1 151 LCD Display NOTE The LCD display time-out is approximately 16 se conds. If keys ar e not pressed within this period, the indicator reverts to reading the signal.
Quick Installation Guide 00825-0100-4593, Rev BA July 2009 Rosemount 1 151 12 S TE P 5 CONTI NUED ... Set the Display Equivalent to a 4 mA Sign al 1. Unscrew the retaining ring shown in Figure 4 and remove the LCD display cover . 2. Press the left button for two seconds.
Quick Installation Guid e 00825-0100-4593, Rev BA July 2009 Rosemount 1 151 13 S TE P 6: T RIM THE T RANSMITTE R NOTE T ransmitte rs are shipped fully calibrated per reques t or by the factory d efault of full scale (span = upper range limit).
Quick Installation Guide 00825-0100-4593, Rev BA July 2009 Rosemount 1 151 14 S TE P 6 CONTI NUED ... Using the T ransmitter Zero Adjustment Buttons Perform the following steps to perform a rerange using the zer o adjustment buttons (see Figure 3). 1.
Quick Installation Guid e 00825-0100-4593, Rev BA July 2009 Rosemount 1 151 15 P RODUCT C E RTIFICATI ONS Approved Manufacturi ng Locations Rosemount Inc.
Quick Installation Guide 00825-0100-4593, Rev BA July 2009 Rosemount 1 151 16 Canadian Certifications Canadian St andards Association (CSA) Approvals E6 Explosion Proof for Class I, Division 1, Gr oups C and D; Class II, Division 1, Group s E, F , and G; Class III, Division 1 Hazardous Loc ations.
Quick Installation Guid e 00825-0100-4593, Rev BA July 2009 Rosemount 1 151 17 N1 A TEX T y pe N and Dust Certification Certificate Number: BAS99A TEX3293X A TEX Marking: II 3 GD Ex nL IIC T5 (-40 °C.
Quick Installation Guide 00825-0100-4593, Rev BA July 2009 Rosemount 1 151 18 N7 T ype N Certificate Number: Aus Ex 122X Ex n IIC T5 (T a = 80 °C) / T6 IP66 Special Conditions for safe use (x): 1. It is a condition of safe use that a nomi nal voltage of 30 V for Ex n application shall not be exceeded.
Quick Installation Guid e 00825-0100-4593, Rev BA July 2009 Rosemount 1 151 19 DoC 4593_Rev_BA.fm Page 19 Friday, August 7, 2009 3:38 PM.
Quick Installation Guide 00825-0100-4593, Rev BA July 2009 Rosemount 1 151 20 4593_Rev_BA.fm Page 20 Friday, August 7, 2009 3:38 PM.
Quick Installation Guid e 00825-0100-4593, Rev BA July 2009 Rosemount 1 151 21 4593_Rev_BA.fm Page 21 Friday, August 7, 2009 3:38 PM.
Quick Installation Guide 00825-0100-4593, Rev BA July 2009 Rosemount 1 151 22 4593_Rev_BA.fm Page 22 Friday, August 7, 2009 3:38 PM.
Ein wichtiger Punkt beim Kauf des Geräts Emerson 00825-0100-4593 (oder sogar vor seinem Kauf) ist das durchlesen seiner Bedienungsanleitung. Dies sollten wir wegen ein paar einfacher Gründe machen:
Wenn Sie Emerson 00825-0100-4593 noch nicht gekauft haben, ist jetzt ein guter Moment, um sich mit den grundliegenden Daten des Produkts bekannt zu machen. Schauen Sie zuerst die ersten Seiten der Anleitung durch, die Sie oben finden. Dort finden Sie die wichtigsten technischen Daten für Emerson 00825-0100-4593 - auf diese Weise prüfen Sie, ob das Gerät Ihren Wünschen entspricht. Wenn Sie tiefer in die Benutzeranleitung von Emerson 00825-0100-4593 reinschauen, lernen Sie alle zugänglichen Produktfunktionen kennen, sowie erhalten Informationen über die Nutzung. Die Informationen, die Sie über Emerson 00825-0100-4593 erhalten, werden Ihnen bestimmt bei der Kaufentscheidung helfen.
Wenn Sie aber schon Emerson 00825-0100-4593 besitzen, und noch keine Gelegenheit dazu hatten, die Bedienungsanleitung zu lesen, sollten Sie es aufgrund der oben beschriebenen Gründe machen. Sie erfahren dann, ob Sie die zugänglichen Funktionen richtig genutzt haben, aber auch, ob Sie keine Fehler begangen haben, die den Nutzungszeitraum von Emerson 00825-0100-4593 verkürzen könnten.
Jedoch ist die eine der wichtigsten Rollen, die eine Bedienungsanleitung für den Nutzer spielt, die Hilfe bei der Lösung von Problemen mit Emerson 00825-0100-4593. Sie finden dort fast immer Troubleshooting, also die am häufigsten auftauchenden Störungen und Mängel bei Emerson 00825-0100-4593 gemeinsam mit Hinweisen bezüglich der Arten ihrer Lösung. Sogar wenn es Ihnen nicht gelingen sollte das Problem alleine zu bewältigen, die Anleitung zeigt Ihnen die weitere Vorgehensweise – den Kontakt zur Kundenberatung oder dem naheliegenden Service.