Benutzeranleitung / Produktwartung 1210 des Produzenten 3M
Zur Seite of 15
Description 2 Each unit includes the f ollowing items: • 1 Electronic Pipettor • 1 Cha r ging Stand • 1 A C Adapter • 1 T est Performance Certi f icate • 1 Instruction Manual • 1 Quick Reference Card • Replacemen t T ip Cone Filters (50/bag) Make sure that all items are included and that no damage has occurred during shipment .
Pr ecautions 4 • Always follo w Good Laboratory Practices. • A void contamination of the tip cone. Do not rest the pipettor on its side. Always place the pipettor in the charging stand when not in use. • Ensure consistent technique to maintain precise results.
Instructions F or Use 6 1. Connect the AC adapter into the connector jack at the rear of the char ging stand's base. 2. Plug the AC adapter into an A C electrical outlet. 3. The green light at the top of the char ging stand indicates that the stand is in operation.
Po wer “ON” Display and Startup The display of a switched “ON” pipettor indicates that programming has been reset to default settings. 1. Switch the unit “ON”. Display: 2. Press the Start Button twice. The display sho ws that the unit is ready for the default 1:5 dilution (Mode d1, 5mL) or for program changes.
1:10 Dilution (5mL T otal V olume) 1. Press repeatedly until the display reads: 2. Press to enter selection. 3. Attach a pipette tip to the tip cone. The pipettor is now ready to be gin the 1:10 dilution (5mL total volume). The display indicates that the pipettor is ready to aspirate 4.
1:5 Dilution (1mL T otal V olume) 1. Press repeatedly until the display reads: 2. Press to enter selection. 3. Attach a pipette tip to the tip cone. The pipettor is now ready to be gin the 1:5 dilution (1mL total volume). The display indicates that the pipettor is ready to aspirate 0.
1:10 Dilution (1mL T otal V olume) 1. Press repeatedly until the display reads: 2. Press to enter selection. 3. Attach a pipette tip to the tip cone. The pipettor is now ready to be gin the 1:10 dilution (1mL total volume). The display indicates that the pipettor is ready to aspirate 0.
7. Position the pipette tip to dispense and press the Start Button. The unit will dispense the second 1.00mL aliquot. The follo wing display indicates that unit is ready to dispense the third of the three 1.00mL aliquots. 8. Position the pipette tip to dispense and press the Start Button.
Pipettor Pr ogramming 18 Open Dilution Programming 1. Press repeatedly until the display reads: 2. Press to enter selection. The display sho ws the default setting for the diluent aspiration volume. 3. Press or to select a new diluent v olume (0.10 - 5.
Pipettor Pr ogramming 20 Speed Selection Speed changes are made using these four keys: Speed "1" is slo west and Speed "5" is fastest. 1. Press to display the aspiration speed. 2. Press or to select a new aspiration speed. 3. Press when the desired aspiration speed is displayed.
22 T r oubleshooting Liquid in Tip Cone If liquid is aspirated beyond the tip cone f ilter , clean the tip cone as follo ws. 1. Switch the unit “OFF”. 2. Remov e the tip ejector collar: hold down the tip ejector le ver and twist of f the tip ejector collar .
Care and Maintenance The 3M Electronic Pipettor - 5mL requires regular cleaning to ensure trouble-free operation. Use a soft cloth lightly moistened with a mild detergent to clean the outer surface of the pipettor . DO NO T A UTOCLA VE. Cleaning the Tip Cone Monthly tip cone cleaning is recommended.
Car e and Maintenance 26 1:5 Dilution T est V olume> 4.00 mL 1.00 mL 5.00 mL V olume (mL) T olerance Range > 3.968-4.032 mL 0.979-1.021 mL 4.945- 5.055 mL W eight (g) W eight (g) W eight (g) T emperature T olerance Range* T olerance Range* T olerance Range* 19°C 3.
Pipettor T ips 28 Genuine 3M Pipettor T ips are strongly recommended for use with the 3M Electronic Pipettor - 5mL. Other tips may adversely af fect the accuracy of the instrument and may damage the tip cone and ejector mechanism.
Limited W arranty 29 3M warrants that - e xcept for depletion of the battery power or where due to product abuse, product use outside the instructions or intended applications, or reasonable wear and .
Ein wichtiger Punkt beim Kauf des Geräts 3M 1210 (oder sogar vor seinem Kauf) ist das durchlesen seiner Bedienungsanleitung. Dies sollten wir wegen ein paar einfacher Gründe machen:
Wenn Sie 3M 1210 noch nicht gekauft haben, ist jetzt ein guter Moment, um sich mit den grundliegenden Daten des Produkts bekannt zu machen. Schauen Sie zuerst die ersten Seiten der Anleitung durch, die Sie oben finden. Dort finden Sie die wichtigsten technischen Daten für 3M 1210 - auf diese Weise prüfen Sie, ob das Gerät Ihren Wünschen entspricht. Wenn Sie tiefer in die Benutzeranleitung von 3M 1210 reinschauen, lernen Sie alle zugänglichen Produktfunktionen kennen, sowie erhalten Informationen über die Nutzung. Die Informationen, die Sie über 3M 1210 erhalten, werden Ihnen bestimmt bei der Kaufentscheidung helfen.
Wenn Sie aber schon 3M 1210 besitzen, und noch keine Gelegenheit dazu hatten, die Bedienungsanleitung zu lesen, sollten Sie es aufgrund der oben beschriebenen Gründe machen. Sie erfahren dann, ob Sie die zugänglichen Funktionen richtig genutzt haben, aber auch, ob Sie keine Fehler begangen haben, die den Nutzungszeitraum von 3M 1210 verkürzen könnten.
Jedoch ist die eine der wichtigsten Rollen, die eine Bedienungsanleitung für den Nutzer spielt, die Hilfe bei der Lösung von Problemen mit 3M 1210. Sie finden dort fast immer Troubleshooting, also die am häufigsten auftauchenden Störungen und Mängel bei 3M 1210 gemeinsam mit Hinweisen bezüglich der Arten ihrer Lösung. Sogar wenn es Ihnen nicht gelingen sollte das Problem alleine zu bewältigen, die Anleitung zeigt Ihnen die weitere Vorgehensweise – den Kontakt zur Kundenberatung oder dem naheliegenden Service.