DB1 Liquid Level Measurement Sensor Revision: 12/07 Copyright © 1993-2007 Campbell Scientific, Inc..
Warranty and Assistance The DB1 LIQUID LEVEL MEASUREMENT SENSOR is warranted by CAMPBELL SCIENTIFIC, INC. to be free from defects in m aterials and workmanship under norm al use and service for twelve (12) m onths from date of shipment unl ess specified otherwise.
DB1 Table of Contents PDF viewers note: These page numbers refer to the printed version of this document. Use the Adobe Acrobat® bookmarks tab for links to specific sections. 1. Overview ...............................................................
This is a blank page..
DB1 Liquid Level Measurement Sensor 1. Overview The double bubbler is a self-cali brating system for l iquid level m easurement. This manual focuses on water level measurement in stilling wells o r open channels, but the inform ation applies to other l iquids as well.
DB1 Liquid Level Measurement Sensor 2. Specifications Measurement Range: Option - 5 (0 to 5 psi/0 to 11.5 ft.) Option - 15 (0 to 15 psi/0 to 34.5 ft.) Option - 30 (0 to 30 psi/0 to 69 ft .) Maximum Overpressure Range: Option - 5 20 psi Option - 10 45 psi Option - 15 60 psi Transducer Accuracy: ± 0.
DB1 Liquid Level Measurement Sensor fluctuates between 5 and 15 ft (2.2 a nd 6.5 psi ), set the pressure regulator above 7 psi (read 10 psi) to ensure adequate pressure. Open the nitrogen bottle valve, set the p ressure regulator, an d adjust the nitrogen flow by adjusting t he two needle valves.
DB1 Liquid Level Measurement Sensor FIGURES 1 and 2. The DB1 System 4.
DB1 Liquid Level Measurement Sensor CR10(X) , CR1000, CR 800 CR850, 21X FIGURE 3. Wiring Diagram for DB1 5.
DB1 Liquid Level Measurement Sensor 5. Programming Every 60 seconds, the following program examples calcul ate the pressure transducer multiplier, measure the offset, and measure the liquid level. The average liquid level i s stored in final m emory every 15 m inutes.
DB1 Liquid Level Measurement Sensor C a s e 3 P o r t S e t ( 3 , 1 ) 'Open valve to atmosphere EndSelect Delay(0,150,MSEC) S u m = 0 For j = 1 To 25 'Take 25 pressure readings for each tube.
DB1 Liquid Level Measurement Sensor 3: Excitation with Delay (P22 ) 1: 1 EX Chan 2: 0 Delay w/EX (units=.01sec) 3: 150 Delay after EX (unit s=.01sec) 4: 0 mV Excit ation Subroutine 1 measures the pres.
DB1 Liquid Level Measurement Sensor 2: Z=F (P30) 1: 0 F 2: 0 Exponent of 10 3: 7 Z Loc [:SUM ] 3: Beginning of Loop (P87) 1: 0 Delay 2: 25 Loop Count 4: Full Bridge (P6) 1: 1 Rep 2: 23 25 mV 60 Hz rej.
DB1 Liquid Level Measurement Sensor 13: Do (P86) 1: 1 Call Subrouti ne 1 14: Z=X (P31) 1: 8 X Loc AVG LEVEL 2: 9-- Z Loc [:m V/V #1 ] 15: Do (P86) 1: 51-- Set low Port 1 16: End (P95) 17: Z=X-Y (P35) .
DB1 Liquid Level Measurement Sensor 6.1 Periodic Maintenance • Inspect submerged bubbler t ubes for sediment or ot her debris. Remove debris and clean tubes as required. • Replace submerged bubbler tubes if cracked or dama ged. • Replace nitrogen bottle every 2 to 3 months dependi ng on bubble rate.
DB1 Liquid Level Measurement Sensor 12 7. Replacement Parts Part # Description Qty. in DB1 5199 SWAK THREAD SEALANT 1 7397 TUBE FITTING ELL 1/4 O.D.x 1/8NPT 4 7558 NEEDLE VALVE ASSEMBLY 2 7561 2-WAY NC SOLENOID VALVE 3 7571 PLUG PIPE 1/8NPT TEFCOAT 2 6078 MOUNTING SCREW (6/32 x 2.
This is a bl ank page ..
Campbell Scientific Companies Campbell Scientific, Inc. (CSI) 815 West 1800 Nort h Logan, Utah 84 321 UNITED STATES www.campbellsci.com info@campbellsci.com Campbell Scientific Afri ca Pty. Lt d. (CSAf) PO Box 2450 Somerset West 7129 SOUTH AFRICA www.
Ein wichtiger Punkt beim Kauf des Geräts Campbell Manufacturing DB1 (oder sogar vor seinem Kauf) ist das durchlesen seiner Bedienungsanleitung. Dies sollten wir wegen ein paar einfacher Gründe machen:
Wenn Sie Campbell Manufacturing DB1 noch nicht gekauft haben, ist jetzt ein guter Moment, um sich mit den grundliegenden Daten des Produkts bekannt zu machen. Schauen Sie zuerst die ersten Seiten der Anleitung durch, die Sie oben finden. Dort finden Sie die wichtigsten technischen Daten für Campbell Manufacturing DB1 - auf diese Weise prüfen Sie, ob das Gerät Ihren Wünschen entspricht. Wenn Sie tiefer in die Benutzeranleitung von Campbell Manufacturing DB1 reinschauen, lernen Sie alle zugänglichen Produktfunktionen kennen, sowie erhalten Informationen über die Nutzung. Die Informationen, die Sie über Campbell Manufacturing DB1 erhalten, werden Ihnen bestimmt bei der Kaufentscheidung helfen.
Wenn Sie aber schon Campbell Manufacturing DB1 besitzen, und noch keine Gelegenheit dazu hatten, die Bedienungsanleitung zu lesen, sollten Sie es aufgrund der oben beschriebenen Gründe machen. Sie erfahren dann, ob Sie die zugänglichen Funktionen richtig genutzt haben, aber auch, ob Sie keine Fehler begangen haben, die den Nutzungszeitraum von Campbell Manufacturing DB1 verkürzen könnten.
Jedoch ist die eine der wichtigsten Rollen, die eine Bedienungsanleitung für den Nutzer spielt, die Hilfe bei der Lösung von Problemen mit Campbell Manufacturing DB1. Sie finden dort fast immer Troubleshooting, also die am häufigsten auftauchenden Störungen und Mängel bei Campbell Manufacturing DB1 gemeinsam mit Hinweisen bezüglich der Arten ihrer Lösung. Sogar wenn es Ihnen nicht gelingen sollte das Problem alleine zu bewältigen, die Anleitung zeigt Ihnen die weitere Vorgehensweise – den Kontakt zur Kundenberatung oder dem naheliegenden Service.