Benutzeranleitung / Produktwartung CS512 des Produzenten Campbell Manufacturing
Zur Seite of 16
Campbell Scientific CS512 Oxyguard Type III Dissolved Oxygen Probe Revision: 12/06 Copyright © 2001-2006 Campbell Scientific, Inc..
Warranty and Assistance The CS512 OXYGUARD TYPE III DISSOLVED OXYGEN PROBE is warranted by CAMPBELL SCIENTIFIC, INC. to be free from defects in materi als and workmanship under norm al use and service for t welve (12) months from date of shipm ent unless specified otherwi se.
i CS512 Table of Contents PDF viewers note: These page numbers refer to the printed version of this document. Use the Adobe Acrobat® bookmarks tab for links to specific sections. 1. General Information ................................................
CS512 Table of Contents ii Examples 1. Sample Program using P1 .......................................................................... 5 2. Sample Program using P2 .......................................................................... 5 3. Sample Program using VoltSE .
1 Campbell Scientific CS512 Oxyguard Type III Dissolved Oxygen Probe An accurate and reliable sensor is a critical element in any measurement system. The Oxyguard Stationary Probe meets these criteria for the measurement of dissolved oxygen.
Campbell Scientific CS512 Oxyguard Type III Dissolved Oxygen Probe 2 2. Specifications Principle of Measurem ent: Membrane covered galvanic oxygen probe Output Signal: 2.
Campbell Scientific CS512 Oxyguard Type III Dissolved Oxygen Probe 3 3. Optional Probe Accessories PT4-L — Agitator for stagnant condi tions 18026 — DO electrolyte, 1 liter 18027 — Membrane kit , 10 mem branes 4.
Campbell Scientific CS512 Oxyguard Type III Dissolved Oxygen Probe 4 FIGURE 4-2. CS512 with PT4 Agitator Optional Repeat Cycle Timer for Agitators : Reciprocating act ion may be controlled by the opti onal solid state Repeat Cycle Tim er. It sends 12 VDC pulses to the agitator coil.
Campbell Scientific CS512 Oxyguard Type III Dissolved Oxygen Probe 5 7. Programming In the CR510, CR10X, and CR 23X datalogger, Instruction P1 or P2 can be used. Example 1 uses a single-ended m easurement for the CS512 probe; exampl e 2 uses a differential measurement.
Campbell Scientific CS512 Oxyguard Type III Dissolved Oxygen Probe 6 8. Calibration The multiplier is used to calibrate the CS512 probe. To calculate the m ultiplier: 1) Program the datalogger using a multiplier of one. 2) Place the OxyGuard probe in the air, shad ed from the sun.
Campbell Scientific CS512 Oxyguard Type III Dissolved Oxygen Probe 7 9. Maintenance The CS512 probe needs little maintenance. Regular cleaning of the m embrane is all that is required. The membrane is ve ry durable and can be cleaned with a cloth or soft paper.
Campbell Scientific CS512 Oxyguard Type III Dissolved Oxygen Probe 8 EXAMPLE 8. Sample Program using P86 and P22 Instructions 45: Do (P86) 1: 41 Set Port 1 Hi gh 46: Excitation with Delay (P22 ) 1: 1 Ex Channel 2: 20 Delay W/Ex (uni ts = 0.01 sec) 3: 0 Delay After Ex (uni ts = 0.
A-1 Appendix A. Dissolved Oxygen in Fresh Water Table.
Appendix A. Dissolved Oxygen in Fre sh Water Table A-2.
This is a bl ank page ..
Campbell Scientific Companies Campbell Scientific, Inc. (CSI) 815 West 1800 Nort h Logan, Utah 84 321 UNITED STATES www.campbellsci.com info@campbellsci.com Campbell Scientific Afri ca Pty. Lt d. (CSAf) PO Box 2450 Somerset West 7129 SOUTH AFRICA www.
Ein wichtiger Punkt beim Kauf des Geräts Campbell Manufacturing CS512 (oder sogar vor seinem Kauf) ist das durchlesen seiner Bedienungsanleitung. Dies sollten wir wegen ein paar einfacher Gründe machen:
Wenn Sie Campbell Manufacturing CS512 noch nicht gekauft haben, ist jetzt ein guter Moment, um sich mit den grundliegenden Daten des Produkts bekannt zu machen. Schauen Sie zuerst die ersten Seiten der Anleitung durch, die Sie oben finden. Dort finden Sie die wichtigsten technischen Daten für Campbell Manufacturing CS512 - auf diese Weise prüfen Sie, ob das Gerät Ihren Wünschen entspricht. Wenn Sie tiefer in die Benutzeranleitung von Campbell Manufacturing CS512 reinschauen, lernen Sie alle zugänglichen Produktfunktionen kennen, sowie erhalten Informationen über die Nutzung. Die Informationen, die Sie über Campbell Manufacturing CS512 erhalten, werden Ihnen bestimmt bei der Kaufentscheidung helfen.
Wenn Sie aber schon Campbell Manufacturing CS512 besitzen, und noch keine Gelegenheit dazu hatten, die Bedienungsanleitung zu lesen, sollten Sie es aufgrund der oben beschriebenen Gründe machen. Sie erfahren dann, ob Sie die zugänglichen Funktionen richtig genutzt haben, aber auch, ob Sie keine Fehler begangen haben, die den Nutzungszeitraum von Campbell Manufacturing CS512 verkürzen könnten.
Jedoch ist die eine der wichtigsten Rollen, die eine Bedienungsanleitung für den Nutzer spielt, die Hilfe bei der Lösung von Problemen mit Campbell Manufacturing CS512. Sie finden dort fast immer Troubleshooting, also die am häufigsten auftauchenden Störungen und Mängel bei Campbell Manufacturing CS512 gemeinsam mit Hinweisen bezüglich der Arten ihrer Lösung. Sogar wenn es Ihnen nicht gelingen sollte das Problem alleine zu bewältigen, die Anleitung zeigt Ihnen die weitere Vorgehensweise – den Kontakt zur Kundenberatung oder dem naheliegenden Service.