Indicator Light Button description



Operation description · LCD display



This device can both warm and thaw breast milk or supplement in a hospital environ- ment and operates by circulating heated air in an enclosed chamber which contains either a breast milk storage container or a syringe. Calesca heats liquids in most com- monly used containers to approximately 34°C (34°C ± 4°C) – be they frozen, out of the fridge or at room temperature or thaws it to a refrigerated temperature. The slow warming process helps preserve nutrients and vitamins.
The device‘s functions, features, and design are especially developed to fit into NICUs (Neonatal Intensive Care Units) as a single bedside unit. However, the device can also be used in the maternity, on the pediatric ward, etc.
When using electrical devices, basic safety precautions should always be followed to reduce risk of burns, electrocution, fire or injury to persons.

NOTE: This equipment has been tested and found to comply with the limits for a Class A digital device, pursuant to Part 15 of the FCC Rules. These limits are designed to provide reasonable protection against harmful interference when the equipment is operated in a commercial environment. This equipment generates, uses, and can radiate radio frequency energy and, if not installed and used in accordance with the instruction manual, may cause harmful interference to radio communications. Operation of this equipment in a residential area is likely to cause harmful interference in which case the user will be required to correct the interfer- ence at his own expense.

Remove Calesca from packaging. Allow unit to reach room temperature before use. Review instructions and assembly. Place sticker with display explanation in your language on Calesca if needed.

Connect the power cord provided (choose the correct plug for your country) with the device in the socket located on the back of the unit. Push the wire cord retainer located around the device power socket down over the plug until secure.
Installation on a counter space or other hard surface
Make sure that surface is clean and dry. Place the device on a solid surface. Ensure the device is secure on the surface to avoid tipping or falling. Place the device on a surface low enough for staff to see into the device cavity.
Do not block vents on bottom or rear of device. Do not place on a soft surface.
Remove IV pole adapter from pack- aging (sold separately). Locate the two IV positioning slots on either side of the device. Verify the pin on the IV pole adapter is engaged (see picture). If pin is not engaged, turn the round knob clockwise until the pin engages.

While securely holding the device in one hand, carefully slide the U-shaped portion of the adapter into the desired IV slot located on the device until it clicks into place. Position the adapter onto the desired IV pole. Make sure the device does not interfere with any other equipment on the IV pole. To secure, turn the triangular knob clockwise until the clamp is tight onto the IV pole.
Do not block vents on bottom or rear of unit.

To remove the device from the IV pole bracket, pull pin out, turn ¼ turn, and release. The pin will remain disengaged and allow for device to slide off the bracket. IV pole bracket does not interfere with countertop use of device if left attached.
Pin not engaged · Pin engaged

Acceptable breast milk containers:

Plug the hospital grade plug into a wall outlet. Press the power button on the device display panel. The unit is setup with both a visual indicator located on the front of the unit and a low volume audible indicator. To mute the audible indicator, press the mute button. To confirm that the audible indicator is muted, the symbol will appear in the upper right portion of the LCD Display.

NEW IMAGE Remove Medela Disposable Inserts from plastic bag. For proper identification, patient labels can be placed on the bag portion of the insert; or using a permanent marker patient information can be written on the hard surface of the insert, including date and time.

Open the lid of the device. Expand the insert prior to installing it into the device. Bend the insert at the hinge and place it into position. Securely install insert using the tabs located on the back and front of the device cavity.

Evenly spread liner within warmer cavity.

Place the tray, flat-side down, into the bottom of the device.
Insert installation

Warming or thawing in a container

Always use a lid on the milk storage container. Make sure insert is installed and insert tray is properly seated in device. Place the container on the center of the tray in the device and close the lid. The device will not operate if the insert is not installed correctly. Note the volume of milk to be warmed.
Warming or thawing in a container

Always use a cap on the syringe. Make sure insert is installed and insert tray is properly seated in device. Close the lid of the device. The device will not operate if the insert is not installed correctly. Note the volume of milk to be warmed.


Press On/Off button to turn on the device. Look at the LCD display to confirm the device is in WARM mode.

Select the milk starting temperature (refrigerated, room, or frozen) by using the arrow keys / to highlight the desired starting temperature.Press the enter button enter to confirm starting temperature.

Use the arrows to enter the volume of milk to be warmed. Press the enter button enter to confirm volume.

The LCD display will flash the information that was entered into the device and display the completion time. Confirm that the information is correct and press the start button start to begin warming.
stop and repeat programming steps above.

There are two timers located on the LCD display. The lower left display will show the total cycle time and will count down to zero. The upper left display shows the total elapsed time the breast milk or supplement has been in the device. If the lid is opened during the warming cycle, the unit will go into a pause mode for 60 seconds. The visual indicator on the front of the unit will flash orange. If the syringe or con- tainer is placed back into the unit, the start button
start can be pressed, and the cycle will continue to completion. If the 60 seconds is exceeded, the unit will stop and reset.
Warming mode


When the warming cycle is complete, the visual indicator on the unit will flash green and the audible indicator (if not muted) will chime. When the warming cycle is complete, the device will automatically go into a holding cycle for up to 30 minutes. During the holding cycle, warm air that is maintained at body temperature will continue to circulate in the unit to ensure that the milk does not cool prior to feeding. The visual indicator will continue to flash green. The audible indicator (if not muted) will chime every 5 minutes until hold cycle time expires. Remove the milk before the hold cycle expires.

If the lid is opened during the holding cycle, the visual indica- tor will flash orange, and the unit will go into a pause mode for 60 seconds. If the syringe or container is placed back into the unit, the start button
start can be pressed, and the holding cycle will continue to completion. Once the hold cycle has expired, the unit will no longer main- tain temperature. It will beep every 15 seconds (if not muted) until the lid is opened or unit is stopped. Gently mix breast milk or supplement and press stop upon removal.
Press On/Off button to turn on the device. The device defaults to the warm mode . Press the right arrow key to highlight the thaw mode . Press the enter button
enter to confirm thaw mode. The device defaults to frozen when in thaw mode .

Use the arrows to enter the volume of milk to be thawed. Press the enter button
enter to confirm volume.
The LCD display will flash the information that was entered into the device and display the completion time. Confirm that the information is correct and press the start button
start to begin thawing.
stop and repeat programming steps above.
Thawing mode

There are two timers located on the LCD display. The lower left display will show the total cycle time and will count down to zero. The upper left display shows the total elapsed time the milk or supplement has been in the device. If the lid is opened during the thawing cycle, the unit will go into a pause mode for 60 seconds.The visual indicator on the front of the unit will flash orange. If the syringe or container is placed back into the unit, the start button
start can be pressed, and the cycle will continue to completion. If the 60 seconds is exceeded, the unit will stop and reset.

When the thawing cycle is complete, the warming action will automatically shut off. The visual indicator will continue to flash green and the audible indicator (if not muted) will chime every 15 seconds. Remove milk immediately while ice crys- tals are present.
Hygiene Wash hands before operating the device, opening inserts, and/or installing inserts.
Acceptable cleaning detergents*


Preparation Calesca should be cleaned prior to first use and between patients. This product’s inserts are sin- gle-patient use items and should not be shared. Use of the device by more than one user without proper cleaning may pose a health risk. Wash hands before operating the warmer, opening inserts, and/or installing inserts.
Remove Medela Disposable Insert and clean with an acceptable cleaning agent using wipes or a lint-free cloth. Do not spray or pour liquid directly into or onto the device. Allow to air dry before use.
The device should be wiped down with an acceptable cleaning agent prior to use and between patients. If milk is spilled on the device, remove insert and wipe down unit.
If the device is used for more than one patient, inserts should stay with the patient with clear and legible indicators of patient name, date and time. Inserts are single patient use items. It is recom- mended that the outer surface of the device is wiped down with an acceptable cleaning agent between patients.

In use Medela Disposable Inserts are a single-patient and limited use item. Medela recommends changing the inserts every 12 hours. Inserts should be changed if human breast milk or supplement comes in contact with the inserts. Inserts should also be changed if visible dam- age has occurred that would offer the potential for human breast milk or supplement to come in direct contact with the device.
Insert care Inspect insert after each warming and/or thawing session for condensation and/or milk. Inserts are single patient use items. Remove and discard the Disposable Insert if patient identified on label no longer requires feed warming.
If con- densation appears in
Condensation may appear in the insert if the fro- zen bottle placed and thawed in the device had frost on its outside. Leave the lid to the warmer open to allow the insert to air dry in between uses.
If milk appears in
Discard insert immediately. Install a new insert prior to next use.

It milk appears on
Inspect the device for any signs of milk and clean if necessary. Install a new insert in the device for next use.
Inspect the inserts for visible damage (ripos or tears). If the insert is damaged, immediately discard and replace it to avoid breast milk or supplement coming in contact with the device.
The warranty for the unit is 2 years from the date of delivery, excluding disposable liners and accessories. The manufacturer is not liable for any damage or consequential damage caused by incorrect operation, inappropriate usage or use by unauthorized persons. Routine checks and servicing may only be carried out by Medela authorized personnel.
The unit is made of technical plastics and metals that cannot be separated. It must be disposed of in accordance with local regulations.
the device · Damaged · inserts

Light indicator is flashing red Error indicator and insert indicator on LCD panel are illuminated
Light indicator is flashing red Error indicator and close lid indicator on the LCD panel is illuminated
Unit will not power up.
IV pole clamp slips out of the device after installed into the pole mount slot on the side of the device
Milk too warm Hot to the touch
Thawed milk has no ice
Cannot resume from pause mode “Pause” indicator
Unexpected pause mode during opera- tion “Pause” indicator
Error indicator and error code in the lower left section of the LCD panel where the large timer would be
Light indicator is flashing red
Problem Indicators · Milk too cool

Ensure insert is installed. Open the lid to the device and properly install the insert. Press start. Refer to Insert Installation section on page 9.
Close the lid and press start.
Check to see if plug in the back of the device is pushed completely in. Unplug the device from the wall outlet and try another outlet.
Make sure that the IV pole adapter pin is engaged prior to installation and snaps into place when the adapter is slid into the pole mount slot.
Allow milk to cool before feeding. Ensure proper settings are used. Note: High temperatures may damage some active immunological properties of breast milk.
Ensure proper settings are used. Retrieve milk promptly when thawing cycle is com- plete.
Ensure lid is closed and insert is fully seated. If using a syringe, ensure it is replaced before pressing start.
Ensure insert is fully seated and syringe is centered. If unit is near an intense infrared emitting light source such as sunlight or heat lamps, move to another location.
Hold power button for at least 4 seconds or unplug the device, wait ten seconds, plug the device back into the wall, and press the power button. Contact Medela customer service.
Return for service.
Ensure fan is not blocked.
Turn off and unplug warmer, allow unit to cool down for up to 30 minutes. Warmer should resume operation. If Error Code 13 is frequently generated, contact Medela customer service.
Ensure unit is in an area with adequate room temperature ventilation. Ensure vents on bottom and rear of unit are not blocked.
Possible Solution