INFRARED EAR THERMOMETER

Instructions and Guarantee

This equipment needs to be installed and put into service in accordance

with the information provided in this manual

TE-150-EU

Page 2 of the Salter TE-150 user manual — IMPORTANT INFORMATION BEFORE USE
Seite 2 · Zum Vergrößern klicken
Seite 2

INTRODUCTION

This Thermometer is intended to be used as the intermittent measurement of human body temperature for people of all ages in the home environment. Utilizing infrared technology, this thermometer takes temperatures in seconds by measuring heat generated by the ear. This product conforms to the provisions of the EC directive MDD(93/42/EEC). Its advantages include: 1. Ear measurement 2. Probe cover not required 3. Waterproof probe design 4. One-second reading 5. Auto power off for power saving 6. Low-battery Indicator 7. Warning indication - indications for battery condition and measuring range. 8. Jumbo LCD Display 9. Memory function

IMPORTANT INFORMATION BEFORE USE

When using this product, please be sure to follow all the notes listed below. Any action against these notices may cause injury or affect the accuracy.

1. Do not disassemble, repair, or remodel the thermometer. 2. Be sure to clean the thermometer lens each time after usage. 3. Avoid direct finger contact with the lens. 4. No modification of this equipment is allowed. 5. It is recommended that user may take 3 temperatures. If they are different, use the highest reading. 6. Do not expose the thermometer to extreme temperature, very high humidity, or direct sunlight. 7. Avoid extreme shock or dropping the device. 8. Before the measurement, users and thermometer should stay in steady state room condition for at least 30 minutes. 9. Avoid measuring temperature for 30 minutes after exercise, bathing, or returning from outdoors.

Page 3 of the Salter TE-150 user manual — PRODUCT IDENTIFICATION
Seite 3 · Zum Vergrößern klicken
Seite 3

10. To protect the environment, dispose of empty batteries at appropriate collection sites according to national or local regulations. 11. It is ill-advised to disassemble the thermometer. 12. Please use the thermometer solely for its intended purpose. 13. Carefully hold the device when in use to avoid dropping the device. 14. Allow one minute between successive measurements as slight variations may occur if measurements are taken over a short period of time. Use average temperatures instead. 15. There are no absolute body temperature standards. Keep reliable records of your personal temperature to serve as a reference for judging a fever. 16. Under any circumstances, the temperature taking result is ONLY for reference. Before taking any medical action, please consult your doctor. 17. It is recommended calibrate the device every 1 year.

PRODUCT IDENTIFICATION

Show 6 more labels / short text on this page
Probe tip · Power & Measure · button · Memory Recall button · LCD display · Battery compartment lid
Page 4 of the Salter TE-150 user manual — TIPS FOR MEASURING HUMAN TEMPERATURE
Seite 4 · Zum Vergrößern klicken
Seite 4

DESCRIPTION OF LCD DISPLAY

Memory symbol and sets of records

Celcius scale Fahrenheit scale

Smiling face Measurement

Scan in progress Frowning face Measurement ≥38.0°C(≥ 100.4°F) Fever

TIPS FOR MEASURING HUMAN TEMPERATURE

  • Bear in mind that the thermometer needs to have been in the room in which the measurement is taken for at least 30 minutes before use. NOTE:
  • Some people produce different readings in their left and right ear. In order to record temperature changes, always measure a person’s temperature in the same ear.
  • The ear thermometer may be used by children only under adult supervision. Measurement is usually possible over the age of 6 months. In infants under 6 months, the ear canal is still very narrow so the temperature of the eardrum often cannot be recorded and the result displayed is often too low.
  • The measurement must not be taken in an ear affected by inflammatory diseases (e.g. discharging pus or secretion), after possible ear injuries (e.g. eardrum damage) or in the healing period after operative procedures. In all of these cases, please consult your doctor.
  • Use of the thermometer on different persons can be inappropriate in the event of certain acute infectious diseases because of the possible spread of germs despite cleaning and disinfection. If you have any doubts, please consult your doctor.
  • This thermometer may only be used without a disposable protective cover.
  • If you have been lying on one ear for some time, the temperature is slightly raised. Wait 60 seconds or measure in the other ear.
  • As ear wax can affect the measurement, you should clean the ear before measuring if necessary.

Temperature in degrees · Low battery warning · No Fever

Page 5 of the Salter TE-150 user manual — MEASURING BODY TEMPERATURE IN THE EAR
Seite 5 · Zum Vergrößern klicken
Seite 5

MEASURING BODY TEMPERATURE IN THE EAR

1. Press the Power & Measure Button for 1 second to switch the thermometer on. All screen data will display on screen. Then wait for beep sound to let you know the appliance is ready and followed by last temperature reading.
2. Make sure that the sensor tip and also the ear canal are clean. As the ear canal is slightly curved, you have to pull the ear slightly up and backwards before inserting the sensor tip. This is important so that the sensor tip can be pointed directly at the eardrum.

Under 1 year Have child lay flat with the head sideways, so that the ear is facing upwards. Gently pull the ear straight back.

1 year + Stand behind and slightly to the side of the child/adult. Gently pull the ear up and back to straighten the ear canal.

3. Insert the probe into the ear canal (the probe must insert well to ensure an accurate temperature measurement), then press the Power & Measure Button for about 1 second, release the button and you will hear 1 short beep sound means complete the measurement. 4. You may take out the device to read the measurement result. 5. The backlight display will remain light for approximately 5 seconds each time the Power & Measure Button is pressed, after each reading is completed, and each time the Memory button is pressed to recall a previous reading. 6. The thermometer will automatically shut off after one minute without use.

Salter TE-150 — MEASURING BODY TEMPERATURE IN THE EAR diagram (page 5)
Abb. 1
Salter TE-150 — MEASURING BODY TEMPERATURE IN THE EAR diagram (page 5)
Abb. 2
Page 6 of the Salter TE-150 user manual — MEMORY FUNCTION
Seite 6 · Zum Vergrößern klicken
Seite 6

The readings If the temperature measurement is below 38°C (100.4°C), a smiling face will be appear next to the reading.

If the reading is 38°C (100.4°C) or above, a frowning face will be displayed.

CHANGING THE MEASUREMENT UNIT (ºC/ºF)

Changing °C/ °F, please make sure the device is in power on condition. Press the Power & Measure and Memory Recall buttons together to toggle between °C and °F.

MEMORY FUNCTION

You can recall up to 10 stored measurements in memory to share with your doctor or trained healthcare professional. 1. Press the Memory Recall button. The first reading displayed is the latest measurement stored in memory. 2. Continue to press the Memory Recall button button to view the next previously stored measurement. 3. Any new measurement will be recorded and the oldest memory deleted without you having to do anything.

Page 7 of the Salter TE-150 user manual — APPLIED STANDARDS
Seite 7 · Zum Vergrößern klicken
Seite 7

CLEANING AND DISINFECTING

For home use device disinfection, 70% Ethanol or Isopropyl alcohol (available in the pharmacy) can be used. Clean the measuring sensor after each use. Use a clean cloth or cotton bud that can be moistened with 70 % alcohol.

Lens/ Measurement Sensor Gently clean with an alcohol swab. Do not use water to wash the thermometer lens directly.

Thermometer Clean with a soft, dry cloth. Do not use water to rinse the device.

APPLIED STANDARDS

This product conforms to the provisions of the EC directive MDD(93/ 42/ EEC). The following standards apply to design and/or manufacture of the products:

ISO 80601-2-56 Medical electrical equipment -- Part 2-56: Particular requirements for basic safety and essential performance of clinical thermometers for body temperature measurement

IEC/EN 60601-1 Medical electrical equipment- Part 1: General requirement for safety

IEC/EN 60601-1-2 Medical electrical equipment- Part 2: Collateral standard: Electromagnetic compatibility - Requirements and tests

IEC/EN 60601-1-11 Medical electrical equipment - Part 1-11: General requirements for basic safety and essential performance - Collateral standard: Requirements for medical electrical equipment and medical electrical systems used in the home healthcare environment

Page 8 of the Salter TE-150 user manual — CLEANING AND DISINFECTING
Seite 8 · Zum Vergrößern klicken
Seite 8
Salter TE-150 — CLEANING AND DISINFECTING diagram (page 8)
Abb. 3

Low battery warning When the batteries get weak, the battery warning symbol appears . It is still possible to measure temperature. The batteries must be replaced. When the battery symbol flashes and Lo appears in the display, the batteries must be replaced. If the batteries are too flat, then the thermometer will switch off automatically.

  • NOTE:
  • It is recommended to remove the batteries if the unit will not be used for an extended period of time.
  • Please store batteries away from small children and heat.
  • Do not use rechargeable batteries.
  • Dispose of used batteries in accordance with the applicable legal regulations. Never dispose of batteries in the normal household waste.

Replacing the Battery 1. Use a Phillips head screwdriver to loosen battery cover screw. Remove the battery cover. 2. Remove used battery. 3. Replace with a lithium 3V CR2032 battery in designated area. The + sign on the battery should be facing upwards. 4. Place the battery cover on the thermometer and tighten the screw to secure it in place.

Salter TE-150 — CLEANING AND DISINFECTING diagram (page 8)
Abb. 4
Page 9 of the Salter TE-150 user manual — ERROR CODES
Seite 9 · Zum Vergrößern klicken
Seite 9

CLINICAL ACCURACY VALIDATION METHOD

This device is an adjusted mode clinical thermometer the validated information for clinical accuracy in each adjusted mode are: GroupA1: ∆cb=-0.01°C, LA=0.18, δr = ±0.08°C GroupA2: ∆cb= 0.06°C, LA=0.22, δr = ±0.08°C GroupB : ∆cb=-0.01°C, LA=0.20, δr = ±0.07°C GroupC : ∆cb=-0.01°C, LA=0.18, δr = ±0.07°C

∆cbCLINICAL BIAS L A : LlMITS OF AGREEMENT δ r: CLINICAL REPEATABILITY

ERROR CODES

When a malfunction or incorrect temperature measurement occurs, an error message will appear as described below.

LCD DisplayCauseSolution The temperature measured is lower than 34°C (93.2°F)Operate the thermometer only between the specifeid temperature ranges. If necessary, clean the sensor tip. In the event of a repeated error message, contact your retailer or Customer Services. The temperature measured is higher than 43°C (109.4°F) The operating temperature is not in the range 15°C~35°C (59°F~95°F)Operate the thermometer only between the specifeid temperature ranges.
Page 10 of the Salter TE-150 user manual — EMC TABLES
Seite 10 · Zum Vergrößern klicken
Seite 10

TECHNICAL SPECIFICATION

  • Measuring range: Human Body: 34°C ~43°C (93.2°F ~ 109.4°F)
  • Laboratory Accuracy: 34°C ~43°C±0.2°C (93.2°F ~ 109.4°F ±0.4°F)
  • Measuring site: Ear
  • Reference body site: Oral
  • Calibration Accuracy: ±0.2°C(±0.4°F): from 35°C ~42°C (95°F ~ 107.6°F) ±0.3°C(±0.5°F): out of range
  • Display resolution: 0.1°C/°F
  • Operating environment: 15°C~35°C (59°F ~ 95°F) with relative humidity up to 95% (non condensing)
  • Storage/transportation environment: -25 to 55 °C(-13 to 131 °F) with relative humidity up to 95% (non condensing)
  • Power supply: A lithium 3V CR2032 battery
  • Memory: 10 sets
  • Weight: approx. 49g (with battery)
  • Dimensions: approx. 112.8mm×34.4mm×44.5mm (L×W×H)
  • Shelf life: 3 years

EMC TABLES

Medical Electrical Equipment needs special precautions regarding EMC and needs to be installed and put into service according to the EMC information.

1. Electromagnetic/Radio Frequency Interference (EMC/RFI): Readings may be affected if the unit is operated within a radio frequency electromagnetic field strength of approximately 3 volts per metre, but the performance of the instrument will not be permanently affected. Care should be taken to keep the thermometer at least 15cm/6 inches away from R/C transmitters to avoid radio frequency interference. 2. Avoid keeping the thermometer too close to objects that continuously generate high heat (like a hot plate) for long periods of time, which can cause overheating of the thermometer.

Page 11 of the Salter TE-150 user manual — TECHNICAL SPECIFICATION
Seite 11 · Zum Vergrößern klicken
Seite 11

Table 1 For all ME EQUIPMENT and ME SYSTEMS

Guidance and manufacturer’s declaration-electromagnetic emissions The Ear Thermometer is intended for use in the electromagnetic environment specifeid below. The customer or the user of the Ear Thermometer should assure that it is used in such an environment. Emissions testComplianceElectromagnetic environment guidance RF emissions CISPR 11CISPR 11The Ear Thermometer uses RF energy only for its internal function. Therefore, its RF emissions are very low and are not likely to cause any interference in nearby electronic equipment. RF emissions CISPR 11Class BThe Ear Thermometer is suitable for use in all establishments other that domestic and those directly connected to the public low-voltage power supply network that supplies buildings used for domestic purposes. Harmonic emissions IEC 61000-3-2Not applicable Voltage fulctuations Flicker emissions IEC 61000-3-3Not applicable
Page 12 of the Salter TE-150 user manual — TECHNICAL SPECIFICATION
Seite 12 · Zum Vergrößern klicken
Seite 12

Table 2 For all ME EQUIPMENT and ME SYSTEMS

Guidance and manufacturer’s declaration-electromagnetic immunity The Ear Thermometer is intended for use in the electromagnetic environment specifeid below. The customer or the user of the Ear Thermometer should assure that it is used in such an environment. Immunity testIEC 60601 test levelCompliance levelElectromagnetic environmentguidance Electrostatic discharge (ESD) IEC 61000-4-2±6kV contact ±8kV air±6kV contact ±8kV airFloor should be wood, concrete or ceramic tile. If folors are covered with shythetic material, the relative humidity should be at least 30%. Electrical fast transient/burst IEC 61000-4-4±2 kV for power supply lines ±1 kV air for input/ output linesNot applicableMains power quality should be that of a typical commercial or hospital environment. Surge IEC 61000-4-5±1kV difefrential mode ±2kV common modeNot applicableMains power quality should be that of a typical commercial or hospital environment. Voltage dips, short interruptions and voltage variations on power supply input lines IEC 61000-4-11<5% UT (>95% dip in UT for 0.5 cycle 40% UT (60% dip in UT for 5 cycles 70% UT (30% dip in UT for 25 cycles 5% UT (>95% dip in UT for 5 secNot applicableMains power quality should be that of a typical commercial or hospital environment. If the use of the Ear Thermometer requires continued operation during power mains interruptions, it is recommended that the Ear Thermometer be powered from an uninterruptible power supply or a battery. Power frequency (50/60 Hz) magnetic feild IEC 61000-4-83 A/m3A/mPower frequency magnetic feilds should be at levels characteristic of a typical location in a typical commercial or hospital environment. NOTE UT is the a.c. mains voltage prior to application of the test level.
Page 13 of the Salter TE-150 user manual — TECHNICAL SPECIFICATION
Seite 13 · Zum Vergrößern klicken
Seite 13

Table 3 For EQUIPMENT and SYSTEMS that are not LIFE-SUPPORTING

Guidance and manufacturer’s declaration-electromagnetic immunity

The Ear Thermometer is intended for use in the electromagnetic environment specified below. The customer or the user of the Ear Thermometer should assure that it is used in such an environment.

Immunity test IEC 60601 test level

Compliance level Electromagnetic environment-guidance

Conducted RF IEC 61000-4-6

3Vrms 150kHz to 80 MHz

Portable and mobile RF communications equipment should be used no closer to any part of the Ear Thermometer, including cables, than the recommended separation distance calculated from the equation applicable to the frequency of the transmitter. Recommended separation distance. d=1.2 d=1.2 80 MHz to 800 MHZ d=2.3 800 MHz to 2,5 Ghz where P is the maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer and d is the recommended separation distance in meters (m). Field strengths from fixed RF transmitters, as determined by an electromagnetic site survey,” should be less than the compliance level in each frequency range.” Interference may occur in the vicinity of equipment marked with the following symbol:

Radiated RF IEC 61000-4-3

3V/m 80MHz to 2.5 GHz

Field strengths from fixed transmitters, such as base stations from radio (cellular/cordless) telephones and land mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast can not be predicted theoretic call with accuracy. To assess the electromagnetic environment due to fixed RF transmitters, an electromagnetic site survey should be considered. If the measured field strength in the location in which the Ear Thermometer is used exceeds the applicable RF compliance level above, the Ear Thermometer should be observed to verify normal operation. If abnormal performance is observed, additional measures may be necessary, such as reorienting or relocating the Ear Thermometer. Over the frequency range 150kHz to 80MHz, field strengths should be less than [V1] V/m.

3 Vrms · 3 V/m

Page 14 of the Salter TE-150 user manual — TECHNICAL SPECIFICATION
Seite 14 · Zum Vergrößern klicken
Seite 14

Table 4 For EQUIPMENT and SYSTEMS that are not LIFE-SUPPORTING

Recommended separation distances between portable and mobile RF communications equipment and the Ear Thermometer

The Ear Thermometer is intended for use in an electromagnetic environment in which radiated RF distances are controlled. The customer or the user of the Ear Thermometer can help prevent electromagnetic interference by maintaining a minimum distance between portable and mobile RF communications equipment (transmitters) and the Ear Thermometer as recommended below according to the maximum output power of the communications equipment.

Rated maximum output power of

Separation distance according to frequency of transmitter / m

150 kHz to 80 MHZ d=1.2

80 MHz to 800 MHZ d=1.2

800 MHz to 2.5 GHz d=2.3

0.01 0.12 0.12 0.23

0.1 0.38 0.38 0.73

1 1.2 1.2 2.3

10 3.8 3.8 7.3

100 12 12 23

2 YEAR GUARANTEE FKA Brands Ltd guarantees this product from defect in material and workmanship for a period of 2 years from the date of purchase, except as noted below. This FKA Brands Ltd product guarantee does not cover damage caused by misuse or abuse; accident; the attachment of any unauthorised accessory; alteration to the product; or any other conditions whatsoever that are beyond the control of FKA Brands Ltd. This guarantee is effective only if the product is purchased and operated in the UK / EU. A product that requires modification or adaptation to enable it to operate in any country other than the country for which it was designed, manufactured, approved and / or authorised, or repair of products damaged by these modifications is not covered under this guarantee. FKA Brands Ltd shall not be responsible for any type of incidental, consequential or special damages. To obtain guarantee service on your product, return the product post-paid to your local service centre along with your dated sales receipt (as proof of purchase). Upon receipt, FKA Brands Ltd will repair or replace, as appropriate, your product and return it to you, post-paid. Guarantee is solely through Salter Service Centre. Service of this product by anyone other than the Salter Service Centre voids the guarantee. This guarantee does not affect your statutory rights. For your local Salter Service Centre, go to www.salterhousewares.co.uk/servicecentres

transmitter / W

Page 15 of the Salter TE-150 user manual — REACH
Seite 15 · Zum Vergrößern klicken
Seite 15

EXPLANATION OF SYMBOLS

The CE marking with the Registration Number of the Notified Body. This denotes the compliance of European Medical Device Directive 93/42/EEC

Consult the instruction for use

Disposal information: Should you wish to dispose of the article, do so in accordance with current regulations. Details are available from your local authority

Type of protection of applied part against electric shock, body floating

This product meets the basic safety and essential performance requirements indicated in the IP22 conditioning test (protection against solid foreign objects of 12.5mm Ø and greater and against vertically falling water drops when enclosure tilted up to 15°)

Manufacturer’s name and address

SN YYMWWWXXXXX SN: Product Serial Number YY: year, MM: month, WWW: working sheet, XXXXX: serial no.

The empty, completely flat batteries must be disposed of through specially designated collection boxes, recycling points or electronics retailers. You are legally required to dispose of the batteries.

RoHS This product fulfilling the requirements of the RoHS Directive 2011/65/EU.

This product fulfilling the requirements of the REACH Directive EC 1907/2006 and its amendments, do not contain Substances of Very High Concern in concentration above the limit of 0.1 %. No substance(s) is/are present in the parts of the product above the concentration of 0.1 % weight by weight.

REACH

Temperature limits · European Authorized Representative

Frage zu Salter TE-150 stellen

Besitzer und unser Team antworten in der Regel innerhalb eines Tages – und Ihre Frage hilft auch anderen weiter.

Add your email to get notified when it’s answered — no spam, ever.

0 questions
Noch keine Fragen zu Salter TE-150

Stellen Sie die erste Frage — andere Nutzer oder unser Team antworten Ihnen.

Originaldokument

Salter TE-150 PDF-Anleitung

72 Seiten·7.7 MB·Thermometer·PDF
Klicken Sie oben auf eine Seite, um sie hier zu öffnen
Seite 1 / …
100%
Seiten werden geladen…