D Kontaktloses Fieberthermometer Gebrauchsanleitung 2-17

Beurer JFT90 — JFT 90 diagram (page 1)
Abb. 2

GB Non-contact fever thermometer Instructions for use 18-33

F Thermomètre médical sans contact Mode d’emploi 34-49

E Termómetro clínico sin contacto Instrucciones de uso 50-65

I Termometro per febbre senza contatto Istruzioni per l’uso 66-82

TR Temassız termometre Kullanım kılavuzu 83-98

RUS Бесконтактный термометр Инструкция по применению 99-117

PL Bezdotykowy termometr lekarski Instrukcja obsługi 118-133

Electromagnetic Compatibility Information 134-139

  • BEURER GmbH
  • Söflinger Str. 218
  • 89077 Ulm (Germany)
  • Tel.: +49 (0)731 / 39 89-144
  • Fax: +49 (0)731 / 39 89-255 <a href="http://www.beurer.com" rel="nofollow noopener" target="_blank">www.beurer.com</a>
  • Mail: [email protected]
Page 2 of the Beurer JFT90 user manual — JFT 90
Seite 2 · Zum Vergrößern klicken
Seite 2

Dear customer, Thank you for choosing one of our products. Our name stands for high-quality, thoroughly tested products for applications in the areas of heat, gentle therapy, blood pressure/diag- nosis, weight, massage and air. Please read these instructions for use carefully and observe the information they contain.

With kind regards, Your Beurer Team

Use This clinical thermometer is intended for non-contact measurement of forehead, object and room temperature. The thermometer is suitable for domestic use. It allows you to quickly and easily measure body temperature Values are saved automatically.

1. Safety notes 19 2. Information about this thermometer 20 3. Unit description 22 4. Initial use 23 5. Switching on and setting the thermometer 24 6. Forehead measurement 26 7. Measuring object temperature / room temperature 28
8. Error messages 28 9. Batteries 30 10. Cleaning the device 30 11. Storing the device 31 12. Disposing of the device 31 13. Technical Data 31 14. Included in delivery 33
Page 3 of the Beurer JFT90 user manual — 1. Safety notes
Seite 3 · Zum Vergrößern klicken
Seite 3

1. Safety notes

  • 1.1 Risks to the user
  • Only use the device once you have read and understood these instructions for use.
  • Retain these instructions for use. The instructions for use must be accessible to all users. All instructions must be followed.
  • The thermometer needs to be in the room in which the measurement is taken for at least 30 minutes before use.
  • The JFT 90 thermometer is only designed for the measuring area on the human body stated in the instructions for use.
  • The device is only intended for the purpose stated in these instructions for use.
  • Children must not be allowed to use the device. Medical products are not toys.
  • Check before each use that the lens is intact. If it is damaged, please contact your retailer or the service address.
  • The device has been designed for practical use, but is not a substitute for a visit to the doctor.
  • Should you have any questions about using the device, please contact either your retailer or Customer Services.
  • 1.2 Risks to the device
  • Do not subject the device to mechanical impacts.
  • Do not expose the device to direct sunlight.
  • Do not expose the device to liquids. The device is not waterproof. Avoid all direct contact with water or other liquids.
  • Have the device repaired by authorised service centres only, otherwise its warranty is invalidated.
  • Portable and mobile HF communication systems may interfere with this device. For more details, please contact our customer service at the address indicated or see the enclosed information sheet.
Page 4 of the Beurer JFT90 user manual — 1. Safety notes
Seite 4 · Zum Vergrößern klicken
Seite 4
2. Information about this thermometer The temperature measurement varies depending on the part of the body where the measurement is taken. In a healthy person, the variance can be between 0.2 °C and 1 °C (0.4 °F and 1.8 °F) in different parts of the body.

Normal temperature range with various thermometers

Measurements Thermometer used Forehead temperature 35.8 °C to 37.6 °C (96.4 °F to 99.7 °F)

Ear temperature 36.0 °C to 37.8 °C (96.8 °F to 100.0 °F)

Oral temperature 36.0 °C to 37.4 °C (96.8 °F to 99.3 °F)

Rectal temperature 36.3 °C to 37.8 °C (97.3 °F to 100.0 °F)

  • Beurer tip
  • Temperatures measured with different thermometers should never be compared with one another.
  • Tell your doctor what type of thermometer you used to take your temperature and on what part of the body. Also bear this in mind if you are diagnosing yourself.

Holding the thermometer for too long in the hand before taking a measurement can cause the device to warm up. This means the measurement could be incorrect.

Show 4 more labels / short text on this page
Forehead thermometer · Ear thermometer · Conventional thermometer · Conventional thermometer
Page 5 of the Beurer JFT90 user manual — 1. Safety notes
Seite 5 · Zum Vergrößern klicken
Seite 5
  • Influences on body temperature
  • A person’s individual metabolism
  • Age Body temperature is higher in babies and toddlers than in adults. Greater temperature fluctuations occur faster and more often in children. Normal body temperature decreases with age.
  • Clothing
  • Outside temperature
  • Time of day Body temperature is lower in the morning and increases throughout the day towards evening.
  • Activities Physical and, to a lesser extent, mental activities increase body temperature.

Beurer tip Taking the body temperature provides a current measurement of a person’s tem- perature. If you are uncertain about interpreting the results or if the values are abnormal (e.g. fever), please consult your doctor. This also applies in the case of slight temperature changes if there are other symptoms of illness such as agita- tion, severe sweating, flushed skin, fast pulse rate, tendency to collapse, etc.

Page 6 of the Beurer JFT90 user manual — 3. Unit description
Seite 6 · Zum Vergrößern klicken
Seite 6

3. Unit description

7 Battery compartment lock

8 Battery compartment lid

Buttons Functions Turns the device on and off. SCAN Starts the temperature measurement.

MODE/MEM MODE Sets the measurement mode. MEM Displays stored measurements.

LIGHT/SET LIGHT Manual switch on of the illuminated display. SET Sets the basic functions.

Show 6 more labels / short text on this page
1 Display · 2 MODE/MEM button · 3 On/Off button · 4 LIGHT/SET button · 5 SCAN button · 6 Measuring sensor
Page 7 of the Beurer JFT90 user manual — 3.1 Display description
Seite 7 · Zum Vergrößern klicken
Seite 7

3.1 Display description

1 Object temperature mode 1 2 3 4

Beurer JFT90 — 3.1 Display description diagram (page 7)
Abb. 3 · 3.1 Display description

2 Room temperature mode

3 Forehead temperature mode

4 Acoustic signal symbol

9 Temperature/memory space number display

10 Celsius/Fahrenheit measurement unit

11 Measurement ≥ 38.0°C “fever”

12 Measurement < 38.0°C “no fever”

13 Illuminated display symbol

4. Initial use If provided, pull out the battery insulating strip at the battery compartment or remove the protective film from the battery and insert the battery with the correct polarity. w 9. Batteries After a brief self-test and two short beeps the thermometer is ready for forehead measurement.
Show 5 more labels / short text on this page
5 Year · 6 Memory function · 7 Date · 8 Battery level · 14 Year/date/time display
Page 8 of the Beurer JFT90 user manual — 5. Switching on and setting the thermometer
Seite 8 · Zum Vergrößern klicken
Seite 8

5. Switching on and setting the thermometer

Briefly press the On/Off button . After a brief self-test and two short beeps the thermometer is ready for forehead measure- ment. The device always starts up in forehead temperature mode . 5.1 Setting the basic functions This menu allows you to set the following functions individually, one after another.

Beurer JFT90 — 5. Switching on and setting the thermometer diagram (page 8)
Abb. 4 · 5. Switching on and setting the thermometer
  • With the thermometer switched on, press and hold the LIGHT/SET button for 5 seconds. The time format option flashes on the display (fig. 1).
  • Use the MODE/MEM button to set your preferred time format and confirm with the LIGHT/SET button.
  • The year flashes in the display (fig. 2).
  • Use the MODE/MEM button to set the year and confirm with the LIGHT/SET button. Day/month flashes in the display (fig. 3).
  • Use the MODE/MEM button to set the day and month, and confirm with the LIGHT/SET button.

In the 24h format, the date is displayed as day/

month. In the 12h format, it is displayed as month/ day.

Page 9 of the Beurer JFT90 user manual — 5. Switching on and setting the thermometer
Seite 9 · Zum Vergrößern klicken
Seite 9

The time flashes in the display (fig. 4). • Use the MODE/MEM button to set the time and con- firm with the LIGHT/SET button.

In the 12h format, the time is displayed as AM/PM.

  • The temperature measurement unit flashes in the display (fig. 5). You can set the device to display the temperature in degrees Celsius (°C) or degrees Fahrenheit (°F).
  • To display the temperature in Celsius, select °C using the MODE/MEM button and confirm with the LIGHT/SET button.
  • To display the temperature in Fahrenheit, select °F using the MODE/MEM button and confirm with the LIGHT/SET button.
Beurer JFT90 — 5. Switching on and setting the thermometer diagram (page 9)
Abb. 5 · 5. Switching on and setting the thermometer

Temperature meas. unit

  • The acoustic signal symbol flashes in the display (fig. 6). You can activate/deactivate the acoustic signals (device activation, measurement in progress, measurement completion).
  • To activate the acoustic signals, select using the MODE/MEM button and confirm with the LIGHT/ SET button.
  • To deactivate the acoustic signals, select using the MODE/MEM button and confirm with the LIGHT/SET button.
Beurer JFT90 — 5. Switching on and setting the thermometer diagram (page 9)
Abb. 6 · 5. Switching on and setting the thermometer

Time · Acoustic signal

Page 10 of the Beurer JFT90 user manual — 6. Forehead measurement
Seite 10 · Zum Vergrößern klicken
Seite 10
Beurer JFT90 — 6. Forehead measurement diagram (page 10)
Abb. 7 · 6. Forehead measurement
  • The light symbol flashes in the display (fig. 7). You can activate or deactivate the automatic illuminated display (appears following forehead temperature measurement).
  • To activate the automatic illuminated display, select using the MODE/MEM button and confirm with the LIGHT/SET button.
  • To deactivate the automatic illuminated display, select using the MODE/MEM button and confirm with the LIGHT/SET button.
Beurer JFT90 — 6. Forehead measurement diagram (page 10)
Abb. 8 · 6. Forehead measurement

You can also activate the illuminated display manually by briefly pressing the LIGHT/ SET button. The display lights up for 5 seconds.

The automatic and manual illuminated display cannot be used while measure-

ment is in progress.

6. Forehead measurement

Briefly press the On/Off button . After a brief self-test and two short beeps the device is ready for forehead temperature measurement. The device is in forehead temperature mode. This is indicated by the symbol.

Beurer JFT90 — 6. Forehead measurement diagram (page 10)
Abb. 9 · 6. Forehead measurement
  • Beurer tip Remember that
  • physical activity, increased perspiration on the forehead, taking vasoconstrictive medication and skin irritations can distort the result,
  • the forehead, or the temples, must be free from perspiration and cosmetics.
Page 11 of the Beurer JFT90 user manual — 6. Forehead measurement
Seite 11 · Zum Vergrößern klicken
Seite 11

Hold the thermometer 2 to 3 cm from the measuring point. Press the SCAN button and move the thermometer from side to side over the forehead area (fig. 8). During measurement you will hear short beeps (only if the acoustic sig- nal is activated), which signal that the thermometer has found a new highest measurement. End of measurement is signalled by a long beep. Release the SCAN button. You can now read the measured value. In addition to the temperature, the fever or no fever symbols also appear in the dis- play: The no fever symbol indicates that the body temperature is within normal range; the fever symbol indicates a measurement equal to or higher than 38.0°C (100.4°F), which triggers a temperature alarm. If the acoustic signal is switched on, three beeps will sound once the measurement is com- plete if the temperature is equal to or higher than 38.0°C (100.4 °F). The measuring time is usually 2 seconds but may take up to 30 seconds. When two short beeps sound and the forehead symbol stops flashing, the device is ready to take another measurement. The measurement is automatically saved with the date/time and the “fever” / “no fever” classification. 6.1 Displaying saved measurements The device only stores measurements in forehead temperature mode . The device auto- matically stores the values from the last 60 measurements. When 60 storage places are exceeded the oldest value is deleted.

  • The memory can be called up as follows:
  • With the thermometer switched on, press and hold the MODE/MEM button for 5 seconds. The most recent measurement is displayed.
  • With each subsequent pressing of the MODE/MEM button, first the storage space number is displayed followed by the measurement upon release.
Page 12 of the Beurer JFT90 user manual — 8. Error messages
Seite 12 · Zum Vergrößern klicken
Seite 12
  • In the upper line, the time and date are displayed alternately.

7. Measuring object temperature / room temperature

  • If you want to measure an object temperature with this thermometer, switch to object temperature mode.
  • With the thermometer switched on, briefly press the MODE/MEM button. The device switches to the object temperature mode .
  • Hold the thermometer 2 to 3 cm from the measuring point. Briefly press the SCAN button and read the temperature in the display (fig. 9).

Measurements taken in object temperature mode are not stored.

  • If you want to use the thermometer to measure the room temperature you need to switch to room temperature mode.
  • With the thermometer switched on, briefly press the MODE/MEM button twice. The device switches to the room temperature mode .
  • The room temperature is immediately displayed (fig. 10).

Measurements taken in room temperature mode are not stored. Fig. 10

8. Error messages

Beurer JFT90 — 8. Error messages diagram (page 12)
Abb. 10 · 8. Error messages

Measurement during self-test, device not yet ready for measure- ment.

Object temperature mode · Room temperature mode

Page 13 of the Beurer JFT90 user manual — 7. Measuring object temperature / room temperature
Seite 13 · Zum Vergrößern klicken
Seite 13

Large variation in room tempera- ture.

Store the device for at least 30 min- utes in a room with room temperature (10 – 40 °C / 50 –104 °F).

Room temperature below 10 °C or over 40 °C (104 °F).

Room temperature must be between 10 °C and 40 °C (104 °F).

Faultless functionality cannot be guaranteed.

Remove the battery for approx. 1 min- ute and then reinsert. In the event of a repeated error message, contact your retailer or Customer Services.

(1) Forehead temperature mode:

Operate the thermometer only between the specified temperature ranges. In the event of a repeated error message, con- tact your retailer or Customer Services.

The temperature recorded is higher than 42.2 °C (108 °F). (2) Object temperature mode: The

temperature recorded is higher than 80 °C (176 °F).

(1) Forehead temperature mode:

Operate the thermometer only between the specified temperature ranges. In the event of a repeated error message, con- tact your retailer or Customer Services.

The temperature recorded is lower than 34 °C (93.2 °F). (2) Object temperature mode: The

temperature recorded is lower than -22 °C (-7.6 °F).

Self-test unsuccessful. Replace the batteries.

The batteries are empty. Replace the batteries.

Error message · Problem Solution

Page 14 of the Beurer JFT90 user manual — 10. Cleaning the device
Seite 14 · Zum Vergrößern klicken
Seite 14
  • 9. Batteries The device requires two AAA (LR03) batteries.
  • Open the battery compartment. Use a pointed object to press down on the battery compartment lock and at the same time slide the battery compartment downwards.
  • Take the used batteries out of the battery compartment.
  • Insert new batteries. Make sure that the batteries are inserted the right way round.
  • Close the battery compartment.

Used batteries should not be disposed of in normal household waste. You are legally required to dispose of the batteries. Dispose of them via your electronics retailer or your local recycling point. Note: The codes below are printed on batteries containing harmful substances: Pb = battery contains lead, Cd = battery contains cadmium, Hg = battery contains mercury. The batteries in this device do not contain any pollutants.

10. Cleaning the device

  • The measuring sensor is the most sensitive part of the thermometer. Take great care of the measuring sensor when cleaning the device.
  • Do not use any harsh cleaning products.
  • Always observe all safety notes for user and device. Safety notes w Page 19.

Clean the measuring sensor after each use. Use a clean cloth or cotton bud that can be moistened with disinfectant or 70 % alcohol. To clean the entire device, please use a soft cloth slightly moistened with a mild soapy solu- tion. Under no circumstances may liquid enter the device. Do not use the device again until it is completely dry.

Pb Cd Hg

Page 15 of the Beurer JFT90 user manual — 10. Cleaning the device
Seite 15 · Zum Vergrößern klicken
Seite 15
11. Storing the device The device must not be stored or used at an excessively high or low temperature or humidity (see technical specifications), in sunlight, in association with an electrical current or in dusty locations. Otherwise inaccuracies can occur. If prolonged storage is planned, you should remove the batteries.
12. Disposing of the device Please dispose of the device in accordance with EC Directive – WEEE (Waste Electrical and Electronic Equipment). If you have any queries, please contact the appropriate local authorities.

13. Technical Data If the device is used other than in accordance with the specifications, perfect functioning cannot be guaranteed! The accuracy of this thermometer has been carefully checked and developed with regard to a long useful life. If using the device for commercial medical purposes, it must be regularly tested for accu- racy by appropriate means. Precise instructions for checking accuracy may be requested from the service address. We reserve the right to make technical changes to improve and develop the product.

Measurement method Non-contact infrared measurement Basic functions Forehead temperature measurement Object temperature measurement Room temperature measurement Measurement units Celsius (°C ) and Fahrenheit (°F) Operating conditions 10 °C to 40 °C (50 °F to 104 °F) with a relative humidity of < 95 % Storage conditions -20 °C to 50 °C (-4 °F to 122 °F) with a relative humidity of < 95 %

Page 16 of the Beurer JFT90 user manual — 10. Cleaning the device
Seite 16 · Zum Vergrößern klicken
Seite 16

Measurement distance 2 to 3 cm from the measuring point Measurement range and accuracy of forehead temperature measure- ment

Forehead temperature measurement 34 °C to 42.2 °C (93.2 °F to 108 °F) Measurement accuracy 36 °C to 39 °C: ± 0.2 °C

(96.8 °F to 102 °F: ± 0.4 °F) In the remaining measurement range ± 0.3 °C (± 0.5 °F) Clinical repeat precision 0.23 °C (0.41 °F) Measurement range and accuracy Object temperature measurement

Object temperature measurement -22 °C to 80 °C (-7.6 °F to 176 °F) Measurement accuracy ± 4% or ± 2 °C (± 4 °F) Memory function Body temperature measurement

Automatically stores the last 60 measurements.

Acoustic signal With device activation, during measurement, when measurement is complete (can be set). Display LCD display Energy saving functions Device automatically switches off after 1 minute. Dimensions Width x depth x height approx. 47.6 mm x 29.0 mm x 188.0 mm Weight 82 g (without batteries) 2 x AAA (LR03) batteries Batteries last for approx. 3000 measurements Activated functions such as acoustic signal or illumi- nated display also reduce the battery life. Signs and symbols Device classification type BF Please read the instructions for use.

Protect from moisture

Page 17 of the Beurer JFT90 user manual — 10. Cleaning the device
Seite 17 · Zum Vergrößern klicken
Seite 17

This device meets the requirements of the EU Directive 93/42/EC concerning medical devices, as well as those of the Medizinproduktegesetz (German Medical Devices Act), the ASTM E 1965-98 and the European Standard EN60601-1-2, and is subject to particular pre- cautions with regard to electromagnetic compatibility. Please note that portable and mobile HF communication systems may interfere with this unit. More details can be requested from the stated Customer Service address or found at the end of the instructions for use.

  • 14. Included in delivery
  • Clinical thermometer
  • 2 x AAA (LR03) batteries
  • Storage box
  • Instructions for use
Page 18 of the Beurer JFT90 user manual — ELECTROMAGNETIC COMPATIBILITY INFORMATION
Seite 18 · Zum Vergrößern klicken
Seite 18

ELECTROMAGNETIC COMPATIBILITY INFORMATION

Guidance and manufacturer’s declaration - electromagnetic emissions

The JFT90 is intended for use in the electromagnetic environment specified below. The customer or the user of the JFT90 should assure that it is used in such an environment.

Emission test Compliance Electromagnetic environment - guidance

The JFT90 uses RF energy only for its internal function. Therefore, its RF emissions are very low and are not likely to cause any interference in nearby electronic equipment.

RF emissions CISPR 11 Group 1

RF emissions CISPR 11 Class B The JFT90 is suitable for use in all establish- ments, including domestic establishments and those directly connected to the public low- voltage power supply network that supplies buildings used for domestic purposes.

Harmonic emissions IEC 61000-3-2 Not applicable

Voltage fluctuations/ flicker emissions IEC 61000-3-3

Not applicable

Page 19 of the Beurer JFT90 user manual — ELECTROMAGNETIC COMPATIBILITY INFORMATION
Seite 19 · Zum Vergrößern klicken
Seite 19

Guidance and manufacturer’s declaration – electromagnetic immunity

The JFT90 is intended for use in the electromagnetic environment specified below. The customer or the user of the JFT90 should assure that it is used in such an environment.

Electromagnetic environment – guidance

Portable and mobile RF communications equip- ment should be used no closer to any part of the JFT90, including cables, than the recommended separation distance calculated from the equation applicable to the frequency of the transmitter.

Conducted RF IEC 61000- 4-6

3 Vrms 150 kHz to 80 MHz

Recommended separation distance d = 1.2 P d = 1.2 P 80 MHz to 800 MHz d = 2.3 P 800 MHz to 2.5 GHz

where P is the maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer and d is the recommen- ded separation Distance in metres (m).

Radiated RF IEC 61000- 4-3

3 V/m 80 MHz to 2.5 GHz

Field strengths from fixed RF transmitters, as determined by an electromagnetic site survey, a should be less than the compliance level in each frequency range.b

Interference may occur in the vicinity of equip- ment marked with the following symbol:

Show 8 more labels / short text on this page
Immunity · IEC 60601 test · Compliance · test · level · level · Not applicable · 3 V/m
Page 20 of the Beurer JFT90 user manual — ELECTROMAGNETIC COMPATIBILITY INFORMATION
Seite 20 · Zum Vergrößern klicken
Seite 20

a. Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted theoretically with accuracy. To assess the electromagnetic environment due to fixed RF transmitters, an electromagnetic site survey should be considered. If the measured field strength in the location in which the JFT90 is used exceeds the applicable RF compliance level above, the JFT90 should be observed to verify normal operation. If abnormal performance is observed, additional measures may be necessary, such as re-orienting or relocating the JFT90.

b. Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3 V/m.

Page 21 of the Beurer JFT90 user manual — IEC 61000-4-5
Seite 21 · Zum Vergrößern klicken
Seite 21

Guidance and manufacturer’s declaration - electromagnetic immunity

The JFT90 is intended for use in the electromagnetic environment specified below. The customer or the user of the JFT90 should assure that it is used in such an environment.

Immunity test IEC 60601 test level Compliance level Electromagnetic

Floors should be wood, concrete or ceramic tile. If floors are covered with synthetic material, the relative humidity should be at least 30 %.

6 kV contact 8 kV air 6 kV contact 8 kV air

IEC 61000-4-2

Electrical fast transient/ burst

2 kV for power supply lines 1 kV for input/output lines Not applicable

Mains power quality should be that of a typi- cal commercial or hospital environment.

IEC 61000-4-4

Mains power quality should be that of a typi- cal commercial or hospital environment.

1 kV line(s) to line(s) 2 kV line(s) to earth Not applicable

IEC 61000-4-5

Show 4 more labels / short text on this page
environment - · guidance · Electrostatic discharge (ESD) · Surge
Page 22 of the Beurer JFT90 user manual — IEC 61000-4-8
Seite 22 · Zum Vergrößern klicken
Seite 22

95 % dip in UT) for 0,5 cycle

Mains power quality should be that of a typical commercial or hospital environment. If the user of the JFT90 requires continued operation during power mains interruptions, it is recommended that the JFT90 be powered from an uninterruptible power supply or a battery.

40 % UT (60 % dip in UT) for 5 cycles

interruptions and voltage variations on power supply input lines

70 % UT (30 % dip in UT)

IEC 61000-4-11

for 25 cycles 95 % dip in UT) for 5 sec

Power frequency ma- gnetic fields should be at levels characteristic of a typical location in a typical commercial or hospital environment.

Power frequency (50/60 Hz) magnetic field

3 A/m 3 A/m

IEC 61000-4-8

Not applicable

Page 23 of the Beurer JFT90 user manual — W
Seite 23 · Zum Vergrößern klicken
Seite 23

Recommended separation distances between portable and mobile RF communications equipment and the JFT90

The JFT90 is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled. The customer or the user of the JFT90 can help prevent electromagnetic interference by maintaining a minimum distance between portable and mobile RF communications equipment (trans- mitters) and the JFT90 as recommended below, according to the maximum output power of the com- munications equipment.

Separation distance according to frequency of transmitter

W

150 kHz to 80 MHz d = 1.2 P

80 MHz to 800 MHz d = 1.2 P

800 MHz to 2,5 GHz d = 2.3 P

0.01 0.12 0.12 0.23

0.1 0.38 0.38 0.73

1 1.2 1.2 2.3

10 3.8 3.8 7.3

100 12 12 23

For transmitters rated at a maximum output power not listed above, the recommended separation distance d in metres (m) can be estimated using the equation applicable to the frequency of the trans- mitter, where P is the maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer. NOTE 1 At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies. NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures, objects and people.

Rated maximum output · power of transmitter · m

Page 24 of the Beurer JFT90 user manual — IEC 61000-4-11
Seite 24 · Zum Vergrößern klicken
Seite 24

752.343 – 0114 Irrtum und Änderungen vorbehalten

Quellen

Referenzen & Links

In dieser Anleitung genannte Seiten – Hersteller, Support und Downloads.

Reparatur

Fehlerbehebung

Häufige Probleme, wahrscheinliche Ursachen und Lösungen.

ProblemMögliche UrsacheAbhilfe
Large variation in room tempera- ture.Store the device for at least 30 min- utes in a room with room temperature (10 – 40 °C / 50 –104 °F). p.13
Room temperature below 10 °C or over 40 °C (<50 °F, >104 °F).Room temperature must be between 10 °C and 40 °C (<50 °F, >104 °F). p.13
Faultless functionality cannot be guaranteed.Remove the battery for approx. 1 min- ute and then reinsert. In the event of a repeated error message, contact your retailer or Customer Services. p.13
(1) Forehead temperature mode: The temperature recorded is higher than 42.2 °C (108 °F). (2) Object temperature mode: The temperature recorded is higher than 80 °C (176 °F).Operate the thermometer only between the specified temperature ranges. In the event of a repeated error message, con- tact your retailer or Customer Services. p.13
The batteries are empty.Replace the batteries. p.13

Frage zu Beurer JFT90 stellen

Besitzer und unser Team antworten in der Regel innerhalb eines Tages – und Ihre Frage hilft auch anderen weiter.

Add your email to get notified when it’s answered — no spam, ever.

0 questions
Noch keine Fragen zu Beurer JFT90

Stellen Sie die erste Frage — andere Nutzer oder unser Team antworten Ihnen.

Originaldokument

Beurer JFT90 PDF-Anleitung

140 Seiten·2.3 MB·Thermometer·PDF
Klicken Sie oben auf eine Seite, um sie hier zu öffnen
Seite 1 / …
100%
Seiten werden geladen…