Dear customer, Thank you for choosing one of our products. Our name stands for high-quality, thoroughly tested products for applications in the areas of heat, gentle therapy, blood pressure/diag- nosis, weight, massage and air. Please read these instructions for use carefully and observe the information they contain.
With kind regards, Your Beurer Team
Use This clinical thermometer is intended for non-contact measurement of forehead, object and room temperature. The thermometer is suitable for domestic use. It allows you to quickly and easily measure body temperature Values are saved automatically.


Normal temperature range with various thermometers
Measurements Thermometer used Forehead temperature 35.8 °C to 37.6 °C (96.4 °F to 99.7 °F)
Ear temperature 36.0 °C to 37.8 °C (96.8 °F to 100.0 °F)
Oral temperature 36.0 °C to 37.4 °C (96.8 °F to 99.3 °F)
Rectal temperature 36.3 °C to 37.8 °C (97.3 °F to 100.0 °F)
Holding the thermometer for too long in the hand before taking a measurement can cause the device to warm up. This means the measurement could be incorrect.

Beurer tip Taking the body temperature provides a current measurement of a person’s tem- perature. If you are uncertain about interpreting the results or if the values are abnormal (e.g. fever), please consult your doctor. This also applies in the case of slight temperature changes if there are other symptoms of illness such as agita- tion, severe sweating, flushed skin, fast pulse rate, tendency to collapse, etc.

7 Battery compartment lock
8 Battery compartment lid
Buttons Functions Turns the device on and off. SCAN Starts the temperature measurement.
MODE/MEM MODE Sets the measurement mode. MEM Displays stored measurements.
LIGHT/SET LIGHT Manual switch on of the illuminated display. SET Sets the basic functions.

1 Object temperature mode 1 2 3 4

2 Room temperature mode
3 Forehead temperature mode
4 Acoustic signal symbol
9 Temperature/memory space number display
10 Celsius/Fahrenheit measurement unit
11 Measurement ≥ 38.0°C “fever”
12 Measurement < 38.0°C “no fever”
13 Illuminated display symbol

Briefly press the On/Off button . After a brief self-test and two short beeps the thermometer is ready for forehead measure- ment. The device always starts up in forehead temperature mode . 5.1 Setting the basic functions This menu allows you to set the following functions individually, one after another.

In the 24h format, the date is displayed as day/
month. In the 12h format, it is displayed as month/ day.

The time flashes in the display (fig. 4). • Use the MODE/MEM button to set the time and con- firm with the LIGHT/SET button.
In the 12h format, the time is displayed as AM/PM.

Temperature meas. unit

Time · Acoustic signal



You can also activate the illuminated display manually by briefly pressing the LIGHT/ SET button. The display lights up for 5 seconds.
The automatic and manual illuminated display cannot be used while measure-
ment is in progress.
Briefly press the On/Off button . After a brief self-test and two short beeps the device is ready for forehead temperature measurement. The device is in forehead temperature mode. This is indicated by the symbol.


Hold the thermometer 2 to 3 cm from the measuring point. Press the SCAN button and move the thermometer from side to side over the forehead area (fig. 8). During measurement you will hear short beeps (only if the acoustic sig- nal is activated), which signal that the thermometer has found a new highest measurement. End of measurement is signalled by a long beep. Release the SCAN button. You can now read the measured value. In addition to the temperature, the fever or no fever symbols also appear in the dis- play: The no fever symbol indicates that the body temperature is within normal range; the fever symbol indicates a measurement equal to or higher than 38.0°C (100.4°F), which triggers a temperature alarm. If the acoustic signal is switched on, three beeps will sound once the measurement is com- plete if the temperature is equal to or higher than 38.0°C (100.4 °F). The measuring time is usually 2 seconds but may take up to 30 seconds. When two short beeps sound and the forehead symbol stops flashing, the device is ready to take another measurement. The measurement is automatically saved with the date/time and the “fever” / “no fever” classification. 6.1 Displaying saved measurements The device only stores measurements in forehead temperature mode . The device auto- matically stores the values from the last 60 measurements. When 60 storage places are exceeded the oldest value is deleted.

Measurements taken in object temperature mode are not stored.
Measurements taken in room temperature mode are not stored. Fig. 10

Measurement during self-test, device not yet ready for measure- ment.
Object temperature mode · Room temperature mode

Large variation in room tempera- ture.
Store the device for at least 30 min- utes in a room with room temperature (10 – 40 °C / 50 –104 °F).
Room temperature below 10 °C or over 40 °C (104 °F).
Room temperature must be between 10 °C and 40 °C (104 °F).
Faultless functionality cannot be guaranteed.
Remove the battery for approx. 1 min- ute and then reinsert. In the event of a repeated error message, contact your retailer or Customer Services.
(1) Forehead temperature mode:
Operate the thermometer only between the specified temperature ranges. In the event of a repeated error message, con- tact your retailer or Customer Services.
The temperature recorded is higher than 42.2 °C (108 °F). (2) Object temperature mode: The
temperature recorded is higher than 80 °C (176 °F).
(1) Forehead temperature mode:
Operate the thermometer only between the specified temperature ranges. In the event of a repeated error message, con- tact your retailer or Customer Services.
The temperature recorded is lower than 34 °C (93.2 °F). (2) Object temperature mode: The
temperature recorded is lower than -22 °C (-7.6 °F).
Self-test unsuccessful. Replace the batteries.
The batteries are empty. Replace the batteries.
Error message · Problem Solution

Used batteries should not be disposed of in normal household waste. You are legally required to dispose of the batteries. Dispose of them via your electronics retailer or your local recycling point. Note: The codes below are printed on batteries containing harmful substances: Pb = battery contains lead, Cd = battery contains cadmium, Hg = battery contains mercury. The batteries in this device do not contain any pollutants.
Clean the measuring sensor after each use. Use a clean cloth or cotton bud that can be moistened with disinfectant or 70 % alcohol. To clean the entire device, please use a soft cloth slightly moistened with a mild soapy solu- tion. Under no circumstances may liquid enter the device. Do not use the device again until it is completely dry.
Pb Cd Hg

13. Technical Data If the device is used other than in accordance with the specifications, perfect functioning cannot be guaranteed! The accuracy of this thermometer has been carefully checked and developed with regard to a long useful life. If using the device for commercial medical purposes, it must be regularly tested for accu- racy by appropriate means. Precise instructions for checking accuracy may be requested from the service address. We reserve the right to make technical changes to improve and develop the product.
Measurement method Non-contact infrared measurement Basic functions Forehead temperature measurement Object temperature measurement Room temperature measurement Measurement units Celsius (°C ) and Fahrenheit (°F) Operating conditions 10 °C to 40 °C (50 °F to 104 °F) with a relative humidity of < 95 % Storage conditions -20 °C to 50 °C (-4 °F to 122 °F) with a relative humidity of < 95 %

Measurement distance 2 to 3 cm from the measuring point Measurement range and accuracy of forehead temperature measure- ment
Forehead temperature measurement 34 °C to 42.2 °C (93.2 °F to 108 °F) Measurement accuracy 36 °C to 39 °C: ± 0.2 °C
(96.8 °F to 102 °F: ± 0.4 °F) In the remaining measurement range ± 0.3 °C (± 0.5 °F) Clinical repeat precision 0.23 °C (0.41 °F) Measurement range and accuracy Object temperature measurement
Object temperature measurement -22 °C to 80 °C (-7.6 °F to 176 °F) Measurement accuracy ± 4% or ± 2 °C (± 4 °F) Memory function Body temperature measurement
Automatically stores the last 60 measurements.
Acoustic signal With device activation, during measurement, when measurement is complete (can be set). Display LCD display Energy saving functions Device automatically switches off after 1 minute. Dimensions Width x depth x height approx. 47.6 mm x 29.0 mm x 188.0 mm Weight 82 g (without batteries) 2 x AAA (LR03) batteries Batteries last for approx. 3000 measurements Activated functions such as acoustic signal or illumi- nated display also reduce the battery life. Signs and symbols Device classification type BF Please read the instructions for use.
Protect from moisture

This device meets the requirements of the EU Directive 93/42/EC concerning medical devices, as well as those of the Medizinproduktegesetz (German Medical Devices Act), the ASTM E 1965-98 and the European Standard EN60601-1-2, and is subject to particular pre- cautions with regard to electromagnetic compatibility. Please note that portable and mobile HF communication systems may interfere with this unit. More details can be requested from the stated Customer Service address or found at the end of the instructions for use.

Guidance and manufacturer’s declaration - electromagnetic emissions
The JFT90 is intended for use in the electromagnetic environment specified below. The customer or the user of the JFT90 should assure that it is used in such an environment.
Emission test Compliance Electromagnetic environment - guidance
The JFT90 uses RF energy only for its internal function. Therefore, its RF emissions are very low and are not likely to cause any interference in nearby electronic equipment.
RF emissions CISPR 11 Group 1
RF emissions CISPR 11 Class B The JFT90 is suitable for use in all establish- ments, including domestic establishments and those directly connected to the public low- voltage power supply network that supplies buildings used for domestic purposes.
Harmonic emissions IEC 61000-3-2 Not applicable
Voltage fluctuations/ flicker emissions IEC 61000-3-3
Not applicable

Guidance and manufacturer’s declaration – electromagnetic immunity
The JFT90 is intended for use in the electromagnetic environment specified below. The customer or the user of the JFT90 should assure that it is used in such an environment.
Electromagnetic environment – guidance
Portable and mobile RF communications equip- ment should be used no closer to any part of the JFT90, including cables, than the recommended separation distance calculated from the equation applicable to the frequency of the transmitter.
Conducted RF IEC 61000- 4-6
3 Vrms 150 kHz to 80 MHz
Recommended separation distance d = 1.2 P d = 1.2 P 80 MHz to 800 MHz d = 2.3 P 800 MHz to 2.5 GHz
where P is the maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer and d is the recommen- ded separation Distance in metres (m).
Radiated RF IEC 61000- 4-3
3 V/m 80 MHz to 2.5 GHz
Field strengths from fixed RF transmitters, as determined by an electromagnetic site survey, a should be less than the compliance level in each frequency range.b
Interference may occur in the vicinity of equip- ment marked with the following symbol:

a. Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted theoretically with accuracy. To assess the electromagnetic environment due to fixed RF transmitters, an electromagnetic site survey should be considered. If the measured field strength in the location in which the JFT90 is used exceeds the applicable RF compliance level above, the JFT90 should be observed to verify normal operation. If abnormal performance is observed, additional measures may be necessary, such as re-orienting or relocating the JFT90.
b. Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3 V/m.

Guidance and manufacturer’s declaration - electromagnetic immunity
The JFT90 is intended for use in the electromagnetic environment specified below. The customer or the user of the JFT90 should assure that it is used in such an environment.
Immunity test IEC 60601 test level Compliance level Electromagnetic
Floors should be wood, concrete or ceramic tile. If floors are covered with synthetic material, the relative humidity should be at least 30 %.
6 kV contact 8 kV air 6 kV contact 8 kV air
Electrical fast transient/ burst
2 kV for power supply lines 1 kV for input/output lines Not applicable
Mains power quality should be that of a typi- cal commercial or hospital environment.
Mains power quality should be that of a typi- cal commercial or hospital environment.
1 kV line(s) to line(s) 2 kV line(s) to earth Not applicable

95 % dip in UT) for 0,5 cycle
Mains power quality should be that of a typical commercial or hospital environment. If the user of the JFT90 requires continued operation during power mains interruptions, it is recommended that the JFT90 be powered from an uninterruptible power supply or a battery.
40 % UT (60 % dip in UT) for 5 cycles
interruptions and voltage variations on power supply input lines
70 % UT (30 % dip in UT)
for 25 cycles 95 % dip in UT) for 5 sec
Power frequency ma- gnetic fields should be at levels characteristic of a typical location in a typical commercial or hospital environment.
Power frequency (50/60 Hz) magnetic field
3 A/m 3 A/m
Not applicable

Recommended separation distances between portable and mobile RF communications equipment and the JFT90
The JFT90 is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled. The customer or the user of the JFT90 can help prevent electromagnetic interference by maintaining a minimum distance between portable and mobile RF communications equipment (trans- mitters) and the JFT90 as recommended below, according to the maximum output power of the com- munications equipment.
Separation distance according to frequency of transmitter
150 kHz to 80 MHz d = 1.2 P
80 MHz to 800 MHz d = 1.2 P
800 MHz to 2,5 GHz d = 2.3 P
0.01 0.12 0.12 0.23
0.1 0.38 0.38 0.73
1 1.2 1.2 2.3
10 3.8 3.8 7.3
100 12 12 23
For transmitters rated at a maximum output power not listed above, the recommended separation distance d in metres (m) can be estimated using the equation applicable to the frequency of the trans- mitter, where P is the maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer. NOTE 1 At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies. NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures, objects and people.
Rated maximum output · power of transmitter · m

752.343 – 0114 Irrtum und Änderungen vorbehalten
In dieser Anleitung genannte Seiten – Hersteller, Support und Downloads.
Häufige Probleme, wahrscheinliche Ursachen und Lösungen.
| Problem | Mögliche Ursache | Abhilfe | |
|---|---|---|---|
| Large variation in room tempera- ture. | Store the device for at least 30 min- utes in a room with room temperature (10 – 40 °C / 50 –104 °F). | p.13 | |
| Room temperature below 10 °C or over 40 °C (<50 °F, >104 °F). | Room temperature must be between 10 °C and 40 °C (<50 °F, >104 °F). | p.13 | |
| Faultless functionality cannot be guaranteed. | Remove the battery for approx. 1 min- ute and then reinsert. In the event of a repeated error message, contact your retailer or Customer Services. | p.13 | |
| (1) Forehead temperature mode: The temperature recorded is higher than 42.2 °C (108 °F). (2) Object temperature mode: The temperature recorded is higher than 80 °C (176 °F). | Operate the thermometer only between the specified temperature ranges. In the event of a repeated error message, con- tact your retailer or Customer Services. | p.13 | |
| The batteries are empty. | Replace the batteries. | p.13 |