3. Warnings and safety notes

  • Never submerge the device in water or other liquids and do not use it in the bathroom. Under no circum - stances may liquid enter the device.
  • Protect the device from heavy impact.
  • Never touch the mains cable with wet hands, as you could get an electric shock.
  • Do not pull the mains plug out of the socket using the mains cable.
  • Do not crush or bend the mains cable, pull it over sharp-edged objects or leave it dangling down, and protect it from sources of heat.
  • We recommend that the power cable is completely unrolled to avoid dangerous overheating.
  • If the mains connection cable of this device is damaged, it must be disposed of. If it cannot be removed, the device must be disposed of.
  • If the device is opened, there is a risk of electric shock. Disconnection from the power supply network is only guaranteed if the mains plug is unplugged.
  • Making modifications to the device or accessories is not permitted.
  • If the device has been dropped, exposed to high levels of moisture or suffered any other damage, it must no longer be used. If in doubt, contact Customer Services or the retailer.
  • The IH 26 nebuliser may only be operated with compatible Beurer atomisers and with the appropriate Beurer accessories. The use of atomisers and accessories made by other companies may result in less efficient therapy and could damage the device.
  • Keep the device and accessories out of the reach of children and pets.

Warning

  • Before use, ensure that there is no visible damage to the device or accessories. If you have any doubts, do not use the device and contact your retailer or the specified Customer Services address.
  • In the event of device faults, please see Chapter “10. Troubleshooting”.
  • The device is not a substitute for medical consultation and treatment. Consult your doctor first if you are experiencing any pain or are suffering from an illness.
  • If you have health concerns of any kind, consult your GP.
  • Please note the general hygiene measures when using the atomiser.
  • You should always follow the instructions of your doctor regarding the type of medication to use, the dosage, and the frequency and duration of inhalation.
  • Only use medication prescribed or recommended by your doctor or pharmacist. Please note: For treatment, only use parts indicated by your doc-
  • tor according to the particular diagnosis.
  • Check whether there are contraindications for use with the usual systems for aerosol therapy on the medication instruction leaflet.
  • If the device does not work properly, or you feel unwell or experience pain, stop using it immediately.
  • Keep the device away from your eyes when it is in use, as the mist of medication could be harmful.
  • Never use the device near flammable gases, oxygen or nitrogen oxide.
  • This device is not intended for use by children or people with restricted physical, sensory (e.g. reduced sensitivity to pain) or mental skills or a lack of experience and/or lack of knowledge, unless they are supervised by a person who is responsible for their safety or are instructed by such a person in how to use the device.
  • The device must be switched off and the plug pulled out before every cleaning and/or maintenance procedure.
  • Keep packaging material away from children (risk of suffocation).
  • To avoid the risk of entanglement and strangulation, store cables and air lines out of the reach of small children.
  • Do not use any additional parts that are not recommended by the manufacturer.
  • The device must only be connected to the mains voltage that is specified on the type plate.
  • Power cuts, sudden interference or other unfavourable conditions could lead to the device becoming inoperable. We therefore recommend that you obtain a replacement device or medication (the latter should be agreed with your doctor).
  • Should you require an adapter or extension lead, this must meet the applicable safety requirements. The power limit and the maximum output specified on the adapter must not be exceeded.
  • Do not store the device or the power cable near to sources of heat.
  • Do not use the device in a room in which a spray has previously been used. Air the room before carrying out the therapy.
  • Do not allow objects to get into the cooling vents.
  • Never use the device if it is making an abnormal sound.
  • For hygiene reasons, it is essential that every user uses their own accessories.
Page 3 of the Beurer IH26 user manual — Repairs
Seite 3 · Zum Vergrößern klicken
Seite 3
  • Always disconnect the mains plug from the mains socket after use.
  • Store the device in a location protected against climatic influences. The device must be stored in the environmental conditions specified.

4. Description of the device and accessories Overview of nebuliser

Beurer IH26 — 4. Description of the device and accessories Overview of nebuliser diagram (page 3)
Abb. 1 · 4. Description of the device and accessories Overvi…

PROTECTION • The unit is fitted with an overcurrent protection de- vice. This may only be replaced by authorized spe- cialist personnel.

General notes

  • The device is only to be used: – on humans – for the purpose for which it is designed (aerosol
  • inhalation) and in the manner specified in these instructions for use.
  • Any form of improper use can be dangerous.
  • In the event of an acute emergency, the provision of first aid has top priority.
  • Apart from the medication, only use distilled water or saline solution. Other liquids may cause a fault in the nebuliser or atomiser.
  • This device is not intended for commercial or clinical use; it is designed exclusively for self-treatment in a private home.

1 Carrying handle 2 On/Off switch (back) 3 Mains connector cable 4 Hose connector 5 Filter cap with filter 6 Holder for atomiser

Overview of atomiser and accessories

Prior to initial use

Beurer IH26 — Overview of atomiser and accessories diagram (page 3)
Abb. 2 · Overview of atomiser and accessories
  • Remove all packaging material before using the device.
  • Protect the device against dust, dirt and humidity, never cover the device while it is in operation.
  • Do not operate the device in a very dusty area.
  • Switch the device off immediately if it is faulty or not working properly.
  • The manufacturer is not liable for damage resulting from improper or careless use.

Repairs

7 Nasal douche 8 Compressed air hose 9 Atomiser 10 Atomiser insert 11 Valve 12 Medication container 13 Mouthpiece 14 Adult mask 15 Children’s mask 16 Nosepiece 17 Replacement filters

  • Under no circumstances should you open or repair the device yourself, as faultless functionality can no longer be guaranteed thereafter. Failure to comply will result in voiding of the warranty.
  • For repairs, please contact Customer Services or an authorised retailer.
Page 4 of the Beurer IH26 user manual — Before using the device for the first time
Seite 4 · Zum Vergrößern klicken
Seite 4
5. Initial use Positioning the device Take the device out of the packaging. Place the device on a flat surface. Make sure that the vents are free from obstructions.
  • Arrange the mains connector cable so that no-one will trip over it.
  • To disconnect the nebuliser from the mains after inhalation, first switch off the device and then pull the mains plug out of the socket.

6. Operation

Before using the device for the first time

  • Clean and disinfect the atomiser and accessories before using them for the first time. See “Cleaning and disinfection” on page 17 .
  • Connect the compressed air hose [8] to the bottom of the medication container [12].
  • For hygiene reasons, it is essential to clean the atomiser [9] and the accessories after each treatment and to disinfect them after the last treatment of the day.
  • The accessories may be used only by one person; use by several people is not recommended.
  • If the therapy involves inhaling several different medications one after the other, please be aware that the atomiser [9] must be rinsed under warm tap water following every usage. See “Cleaning and disinfection” on page 17.
  • Please observe the notes on changing the filter in these instructions for use.
  • Check that hose connectors are firmly attached to the nebuliser [4] and the atomiser [9] before each use of the device.
  • Before use, check the device is working correctly, by briefly switching on the nebuliser (together with the connected atomiser, but without medication). If air comes out of the atomiser [9], the device is working.
Beurer IH26 — Before using the device for the first time diagram (page 4)
Abb. 3 · Before using the device for the first time
  • Connect the other end of the compressed air hose [8] to the nebuliser hose connection [4] by turning it slightly.
  • 6.1 Installing the atomiser insert
  • Open the atomiser [9] by twisting the top anticlockwise against the medication container [12]. Place the atomiser insert [10] into the medication container [12].
  • Ensure that the cone for administering medication fits well on the cone for the air duct inside the nebuliser.
Beurer IH26 — Before using the device for the first time diagram (page 4)
Abb. 4 · Before using the device for the first time
  • 6.2 Filling the atomiser
  • Fill with an isotonic saline solution or pour the medication directly into the medication container [12]. Avoid overfilling. The maximum recommended filling quantity is 8 ml.
  • Use medication only as instructed by your doctor and ask about the appropriate inhalation period and quantity for you.
  • If the prescribed quantity of medication is less than 2 ml, top this up to at least 4 ml with isotonic saline solution. Dilution is also necessary with viscous medications. Here too, please observe the instructions of your doctor.
  • Power supply The device must only be connected to the mains voltage that is specified on the type plate.
  • Insert the mains plug of the mains connector cable [3] into a suitable socket.
  • To connect to the mains, fully insert the mains plug into the socket.
  • Ensure that there is a mains socket close to the setup area.
Page 5 of the Beurer IH26 user manual — 6.8 Nasal douche
Seite 5 · Zum Vergrößern klicken
Seite 5

6.3 Closing the atomiser • Close the atomiser [9] by twisting the top clockwise against the medication container [12]. Ensure that the connection is correct.

Essential oils, cough syrups, gargling solutions and drops to be used as a rub or in a steam bath are wholly unsuitable for inhalation using a nebuliser. These ad- ditions are often viscous and can impair the correct functioning of the device and therefore the effectiveness of the application in the long-term. For individuals with a hypersensitive bronchial system, medications contain- ing essential oils may under certain conditions cause an acute bronchospasm (a sudden cramp-like restriction of the bronchi with shortness of breath). Consult your doctor or pharmacist in relation to this matter.

6.4 Inserting the valve • Press the blue valve [11] into the opening intended for this purpose on the upper side of the atomiser [9].

The valve system ensures better delivery of the medi- cation to the lungs. Only use the valve [11] with the mouthpiece [13], never with the masks [14, 15] or the nosepiece [16].

  • 6.7 Stopping inhalation Once the mist is only coming out in an irregular flow or if the sound changes when inhaling, you can stop the treatment.
  • Switch off the nebuliser after treatment using the On/ Off switch [2] and disconnect it from the mains.
  • Place the atomiser [9] back in its holder [6] after the treatment.

6.5 Connecting accessories to the atomiser • Connect the atomiser [9] to the desired accessory (mouthpiece [13], adult mask [14], children’s mask [15] or nosepiece [16]).

The most effective form of nebulisation is by using the mouthpiece. Nebulisation using a mask is only recom- mended if it is not possible to use a mouthpiece (e.g. for children who are not yet able to inhale medication using a mouthpiece). If you are using the adult mask [14] you can secure it to your head with the adult mask strap. There is also a strap for the children’s mask [15].

6.8 Nasal douche

19. Cover 2 0. Collection container for nasal secretion 21. Saline container 22. Cover hole (Starting the application)
  • Before the treatment, pull the atomiser upwards out of the holder [6].
  • Start the nebuliser using the On/Off switch [2].
  • Spray mist pouring out of the atomiser indicates that the device is operating correctly.

You can use the nasal douche to clean your nasal cavi- ties. The nasal douche creates a fine jet of liquid for treating respiratory tract problems or for moistening the nasal mucous membranes.

  • 6.6 Treatment
  • When inhaling, sit upright and relaxed at a table and not in an armchair to avoid compressing the airways and therefore impairing the effectiveness of the treatment.
  • Breathe in the atomised medication deeply.

We recommend using the nasal douche for colds, dry- ness inside the nose, dust exposure, pollen allergies, chronic inflammation of the sinuses or upon the advice of your doctor.

The device is not suitable for continuous operation; after 30 minutes of operation it must be switched off for 30 minutes.

  • The nasal douche should only be used with isotonic saline solutions.
  • Replace the nasal douche after a year.

During the treatment, hold the atomiser straight (ver- tically), otherwise the atomisation will not work and faultless functionality is not guaranteed.

The maximum volume of the nasal douche is 10 ml. If you are unsure about the correct dilution of the saline solution, please consult your doctor.

Page 6 of the Beurer IH26 user manual — Warning
Seite 6 · Zum Vergrößern klicken
Seite 6

6.9 Using the nasal douche To use the nasal douche correctly, proceed as follows:

10. Once your nose feels clear again, you can stop the application. The application should be stopped at the latest once the maximum filling quantity has been used.
1. Twist the cover [19] 90° anticlock - wise and remove it in an upward direction. 2. Lift the collection container [20] out.

6.10 Cleaning See “Cleaning and disinfection” on page 17.

7. Changing the filter In normal operating conditions, the air filter must be replaced after approx. 500 operating hours or one year. Please check the air filter regularly (after 10-12 nebulisation procedures). Replace the used filter if it is very dirty or clogged. If the filter has become damp, it must also be exchanged for a new filter.
3. Fill the saline container [21] with saline solution.
4. Place the collection container [20] into the saline container [21]. 5. Place the cover [19] on the saline container [21]. Twist the cover [19] in a clockwise direction until it clicks firmly into place.
  • Do not attempt to clean the used filter and reuse it.
  • Use only the manufacturer’s original filter, otherwise your nebuliser may become damaged and sufficiently effective therapy cannot be guaranteed.
  • Do not repair or maintain the air filter while it is in use.
  • Never operate the device without a filter.

To replace the filter, proceed as follows:

6. Connect the compressed air hose [8] to the saline container [21] from below. 8 6 4 2
  • First switch the device off and disconnect it from the mains.
  • Allow the device to cool down.
1. Pull off the filter cap [5] towards the front.
7. Position the cover [19] carefully inside one of your nostrils. Breathe through the nose.

If the filter remains in the device after the cap has been removed, take the filter out of the device, e.g. with tweezers or similar.

8. Press the On/Off switch [2] on the back of the nebuliser. To start the application, place one finger over the cover hole [22] on the container.
2. Re-insert the filter cap [5] with a new filter. 3. Ensure that it is securely in place.

8. Cleaning and disinfection Atomiser and accessories

During the application, breathe slowly and evenly through the nose. Tilt your head slightly in the opposite direction to the nostril being treated so that the saline solution can flow deep into the nose. 9.To pause or stop the application, take your finger off the cover hole [22] on the saline container [21].

Warning

Adhere to the following hygiene instructions to avoid health risks. • The atomiser [9] and accessories are designed for multiple use. Please note that different areas of ap-

Page 7 of the Beurer IH26 user manual — Cleaning
Seite 7 · Zum Vergrößern klicken
Seite 7

plication involve different requirements in terms of cleaning and hygienic preparation.

Use a soft, dry cloth and non-abrasive cleaning products to clean the device.

Do not use any abrasive cleaning products and never submerge the device in water.

  • Notes:
  • Do not clean the atomiser or the accessories mechanically using a brush or similar device, as this could cause irreparable damage and it will mean that the best treatment results can no longer be guaranteed.
  • Please consult your doctor about the additional requirements in terms of the hygienic preparation needed (hand care, handling of medication/inhalation solutions) for high-risk groups (e.g. patients with cystic fibrosis).
  • Ensure thorough drying after each cleaning or disinfection process. Residual moisture or wetness can represent an increased risk of bacterial growth.
  • Ensure that no water gets inside the device.
  • Do not clean the device or accessories in the dishwasher.
  • Do not touch the device with wet hands when it is plugged in and do not allow water to spray on the device. Only operate the device if it is completely dry.
  • Do not spray any liquid into the vents. If liquid penetrates the device this could cause damage to the electrics or other nebuliser parts and lead to a malfunction.
  • Preparation
  • Immediately after each treatment, all parts of the atomiser [9] and the used accessories must be cleaned of residual medication and contamination.
  • To do this, dismantle the atomiser [9] and the nasal douche [7] into their individual parts.
  • Remove the mouth piece [13], the mask [14, 15] or the nosepiece [16] from the atomiser.
  • Dismantle the atomiser by twisting the top anticlockwise against the medication container [12].
  • Remove the atomiser insert [10] from the medication container [12].
  • Remove the valve [11] by pulling it out of the atomiser [9].
  • Reassembly is carried out in reverse order.
  • Condensation, hose care Condensation may form in the hose depending on the ambient conditions. It is essential to remove the moisture to prevent the bacterial growth and ensure ideal therapy. To do so, proceed as follows:
  • Remove the compressed air hose [8] from the atomiser [9].
  • The hose must remain connected to the inhaler [4].
  • Operate the nebuliser until the moisture is removed by the air passing through.
  • In the event of heavy contamination, replace the hose.
  • Disinfection Please carefully observe the points below when disinfecting your atomiser and accessories. We recommend disinfecting the individual parts on a daily basis after the last usage as a minimum measure. (All you need for this is a little white vinegar and distilled water.)
  • First, clean the atomiser and accessories as described in the “Cleaning” section.
  • Place the disassembled nebuliser, mouthpiece, nosepiece and nasal douche in boiling water for five minutes.
  • For the remaining accessories, use a vinegar solution consisting of ¼ vinegar and ¾ distilled water. Make sure that the volume is sufficient to fully submerse the parts, such as the atomiser, mask and mouthpiece, in the solution.
  • Leave the parts in the vinegar solution for 30 minutes.
  • Rinse the parts with water and dry carefully with a soft cloth.

Cleaning

The device must be switched off, disconnected from the mains and allowed to cool down each time before cleaning.

The atomiser and the used accessories such as the mouthpiece, mask, nasal douche, etc. must be washed with hot but not boiling water after each use. Dry the parts carefully using a soft cloth. Put the parts together again if they are completely dry and place them in a dry, sealed container or disinfect them.

When cleaning, ensure that any residue is removed. Never use any substances for cleaning that could po- tentially be toxic if they came into contact with the skin or mucous membranes, or if they were swallowed or inhaled.

Do not boil or autoclave the compressed air hose or the masks.

Page 8 of the Beurer IH26 user manual — 10. Troubleshooting
Seite 8 · Zum Vergrößern klicken
Seite 8
  • Put the parts together again if they are completely dry and place them in a dry, sealed container.

10. Troubleshooting

Please ensure that the parts are completely dried af- ter cleaning, otherwise the risk of bacterial growth is increased.

The atom- iser does not produce any aerosol or produces too little aerosol.

1. Too much or too little medication in the atomiser. Minimum: 2 ml, Maximum: 8 ml. 2. Check nozzle for blockages. Clean nozzle if necessary (e.g. by rinsing out). Then start the atomiser again. IMPORTANT: Carefully pierce the

Sterilisation • Use cold sterilisation solution in accordance with the manufacturer’s instructions.

Drying • Place the individual parts on a clean, dry and absor- bent surface and leave them to dry completely (at least 4 hours).

fine holes from the underside of the nozzle only. 3. Atomiser not held vertically. 4. Unsuitable medication fluid added for nebulisation (e.g. too viscous). The medication fluid should be prescribed by the doctor. The output is too low.

  • Durability of materials
  • As with any plastic parts, atomisers and their accessories are affected by a certain amount of wear and tear when used and hygienically prepared on a frequent basis. Over time, this can lead to a change in the aerosol, which can have a negative effect on the efficiency of the therapy. We therefore recommend that you replace the atomiser and other accessories after a year.
  • The following point should be observed when selecting the cleaning product or disinfectant: only use a mild cleaning product or disinfectant in the concentration and volume prescribed by the manufacturer.

Kinked hose, clogged filter, too much inhalation solution. What medica- tions are suitable for inhaling?

Please consult your doctor in relation to this matter. As a rule, all medication that is suitable and approved for device inhalation, can be inhaled. There is inha- lation solution residue in the atomiser.

This is normal and is necessary for technical reasons. Stop inhalation once the atomiser starts to make a notably different sound. What should be taken into account when using the device with infants and children?

  • Storage
  • Do not store in damp conditions (such as in a bathroom) and do not transport with any damp items.
  • When storing and transporting, protect from prolonged direct sunlight.
  • The accessories can be stowed securely in the accessory compartment. Store the device in a dry place, ideally in the original packaging.
1. On infants and children, the mask should cover the mouth and nose to ensure effective inhalation.
9. Disposal For environmental reasons, do not dispose of the device in the household waste. Please dispose of the device in accordance with EC Directive – WEEE (Waste Electrical and Electronic Equipment). If you have any questions, please contact the local authorities responsible for waste disposal.
2. On children, the mask should also cover both the nose and mouth. It is not a good idea to carry out nebulisation on someone who is sleeping, as in this case not enough of the medication will reach the lungs. Note: Inhalation should only be carried out under the supervision of an adult and with their assistance and the child should not be left alone.

Problem/ question · Possible cause/remedy

Page 9 of the Beurer IH26 user manual — PP/PVC 601.28
Seite 9 · Zum Vergrößern klicken
Seite 9

Storage and transportation conditions

Temperature: 0°C to +60°C Relative humidity: 10% to 95% Ambient pressure: 700 – 1060 hPa Aerosol properties 1) Flow: 5.3 l/min

Inhalation using the mask takes longer.

This is necessary for technical rea- sons. You breathe less medication per breath through the holes of the mask than via the mouthpiece. The aerosol is mixed with ambient air via the holes. Why should the atomiser be replaced regularly?

2. Aerosol delivery: 0.326 ml 3) Aerosol delivery rate: 0.132 ml/min 4) Particle size (MMAD): 3.07 µm

Subject to technical changes.

There are two reasons for this: 1. To ensure a therapeutically effective particle spectrum, the nozzle hole must not exceed a specific diameter. Due to mechanical and thermal stresses, the plastic is subject to a certain degree of wear. The atomiser insert [10] is particularly sensitive. This can also alter the composition of the aerosol drop- lets, which directly affects the effectiveness of the treatment. 2. Regularly changing the atomiser is also recommended for hy- giene reasons. Should each person have their own atomiser?

Beurer IH26 — Replacement items diagram (page 9)
Abb. 5 · Replacement items

Measurements were performed using a sodium fluoride solution with a “Next Generation Impactor” (NGI). The diagram may therefore not be applicable to sus- pensions or highly viscous medications. You can obtain more detailed information from the manufacturer of your medication.

Yes, this is absolutely essential for hygiene reasons.

Replacement items

11. Technical specifications

Designation Material REF Nebulisation stopper PP 162.819 Year pack (contains mouth- piece, nosepiece, adult mask, children’s mask, atomiser, compressed air hose, filter)

PP/PVC 601.28

Type IH 25/2 Dimensions (W x H x D) 166 x 141 x 148 mm Weight 1.4 kg Operating pressure Approx. 0.8 to 1.45 bar Atomiser filling volume

Nasal douche PP 601.37

If the device is not used according to the instructions specified, perfect functionality cannot be guaranteed! We reserve the right to make technical changes to improve and develop the product. This device and its accessories comply with the European standards EN60601-1, EN60601-1-2 and EN13544-1 and are subject to particular precautions with regard to electro- magnetic compatibility. Please note that portable and mobile HF communication systems may interfere with this unit. More details can be requested from the stated Customer Services address or found at the end of the instructions for use. This device meets the require- ments of European Directive 93/42/EEC for medical products, as well as those of the Medizinproduktege- setz (German Medical Devices Act).

Min. 2 ml Max. 8 ml Medication flow rate Approx. 0.3 ml/min. Nasal douche flow rate

Min. 2 ml Max. 10 ml Sound pressure Max. 52 dBA (acc. to DIN EN 13544-1 section 26) Power supply 230 V~; 50 Hz; 230 VA UK: 240 V~; 50 Hz; 240 VA Saudi Arabia: 220V~; 60Hz; 220VA Expected service life 1000 h Operating conditions

Temperature+10°C to +40°C Relative humidity: 10% to 95%
Show 6 more labels / short text on this page
Problem/ question · Possible cause/remedy · Particle size diagram · Cumulative undersize (%) · Particle size (µm) · Model IH 26
Page 10 of the Beurer IH26 user manual — Replacement items
Seite 10 · Zum Vergrößern klicken
Seite 10
  • ELECTROMAGNETIC COMPATIBILITY
  • The device complies with the currently valid requirements relating to electromagnetic compatibility and is suitable for use in all buildings including those designated for private residential purposes. The radio frequency emissions of the device are extremely low and are very unlikely to cause interference with other devices in the vicinity.
  • We recommend never placing the device on top of or near to other devices. If interference does occur with your other electrical appliances, move the device or connect it to a different mains socket.
  • Radio devices can have an influence on the operation of the device.

Subject to errors and changes

Page 11 of the Beurer IH26 user manual — ELECTROMAGNETIC COMPATIBILITY INFORMATION
Seite 11 · Zum Vergrößern klicken
Seite 11

ELECTROMAGNETIC COMPATIBILITY INFORMATION

Guidance and manufacturer’s declaration - electromagnetic emissions

The [IH 25/2] is intended for use in the electromagnetic environment specified below. The customer or the user of the [IH 25/2] should assure that it is used in such an environment

Emissions Compliance Electromagnetic environment – guidance

RF emissions CISPR 11

Group 1 The [IH 25/2] uses RF energy only for its internal function. Therefore, its RF emissions are very low and are not likely to cause any interference in nearby electronic equipment.

RF emissions CISPR 11

Class B The [IH 25/2] is suitable for use in all establish- ments other than domestic, and may be used in domestic establishments and those directly connected to the public low-voltage power supply network that supplies buildings used for domestic purposes, provided the following warning is heeded: Warning: This equipment/system is intended for use by healthcare professionals only. This equipment/ system may cause radio interfer- ence or may disrupt the operation of nearby equipment. It may be necessary to take mitiga- tion measures, such as re-orienting or relocating the [IH 25/2] or shielding the location.

Harmonic emissions IEC 61000-3-2

Voltage fluctuations/flicker emissions IEC 61000-3-3

Class A · Complies

Page 12 of the Beurer IH26 user manual — ELECTROMAGNETIC COMPATIBILITY INFORMATION
Seite 12 · Zum Vergrößern klicken
Seite 12

Guidance and manufacturer’s declaration – electromagnetic immunity

The [IH 25/2] is intended for use in the electromagnetic environment specified below. The customer or the user of the [IH 25/2] should assure that it is used in such an environment

Immunity test EN 60601 test level

Compliance level Electromagnetic environment – guidance

Electrostatic discharge (ESD) IEC 61000-4-2

±6 kV contact ±8 kV air

±6 kV contact ±8 kV air

Floors should be wood, concrete or ceramic tile. If floors are covered with synthetic material, the relative humid- ity should be at least 30 %

Electrical fast transient/burst IEC 61000-4-4

±2 kV for power sup- ply lines ±1 kV for input/output lines

±2 kV for power sup- ply lines

Mains power quality should be that of a typical commercial or hospital environment. The electrical fast transient burst (EFT) is generated by the switching of inductive loads. Separation between the equipment and other loads shall be considered before installation. Mains filter is required, if necessary.

±1 kV line(s) to line(s) ±2 kV line(s) to earth

±1 kV line(s) to line(s) Mains power quality should be that of a typical commercial or hospital environment.

Voltage dips, short interruptions and voltage variations on power supply input lines IEC 61000-4-11

95 % dip in UT) for 0,5 cycle 40 % UT (60 % dip in UT) for 5 cycles 70 % UT (30 % dip in UT) for 25 cycles 95 % dip in UT) for 5s

95 % dip in UT) for 0,5 cycle 40 % UT (60 % dip in UT) for 5 cycles 70 % UT (30 % dip in UT) for 25 cycles 95 % dip in UT) for 5s

Mains power quality should be that of a typical commercial or hospital en- vironment. If the user of the [IH 25/2] requires continued operation during power mains interruptions, it is rec- ommended the [IH 25/2] be powered from an uninterruptible power supply or a battery.

Power frequency (50/60Hz) magnetic field IEC 61000-4-8

3 A/m 3 A/m Power frequency magnetic fields should be at levels characteristic of a typical location in a typical commer- cial or hospital environment.

Surge IEC 61000-4-5

Page 13 of the Beurer IH26 user manual — ELECTROMAGNETIC COMPATIBILITY INFORMATION
Seite 13 · Zum Vergrößern klicken
Seite 13

Guidance and manufacturer’s declaration – electromagnetic immunity

The [IH 25/2] is intended for use in the electromagnetic environment specified below. The customer or the user of the [IH 25/2] should assure that it is used in such an environment

Immunity test EN 60601 test level

Electromagnetic environment – guidance

Conducted RF IEC 61000-4-6

3 Vrms 150 kHz to 80 MHz

Portable and mobile RF communications equipment should be used no closer to any part of the [IH 25/2], including cables, than the recommended separation distance calculated from the equation applicable to the frequency of the transmitter.

Radiated RF IEC 61000-4-3

3 V/m 80 MHz to 2.5 GHz

where p is the maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer and d is the recom- mended separation distance in metres (m) Field strengths from fixed RF transmitters, as determined by an electromagnetic site survey,a should be less than the compliance level in each frequency range.b Interference may occur in the vicinity of equip- ment marked with the following symbol:

a Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted theoretically with accuracy. To assess the electromagnetic environment due to fixed RF transmitters, an electromagnetic site survey should be considered. If the measured field strength in the location in which the [IH 25/2] is used exceeds the applicable RF compliance level above, the [IH 25/2] should be observed to verify normal opera- tion. If abnormal performance is observed, additional measures may be necessary, such as re-orienting or relocating the [IH 25/2]. b Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3 V/m.

Compliance level · 3 V/m

Page 14 of the Beurer IH26 user manual — W
Seite 14 · Zum Vergrößern klicken
Seite 14

Recommended separation distances between portable and mobile RF communications equipment and the [IH 25/2]

The [IH 25/2] is intended for use in an electromagnetic environment in which radiated RF disturbances are con- trolled. The customer or the user of the [IH 25/2] can help prevent electromagnetic interference by maintaining a minimum distance between portable and mobile RF communications equipment (transmitters) and the [IH 25/2] as recommended below, according to the maximum output power of the communications equipment.

Rated maximum output power of transmitter

Separation distance according to frequency of transmitter

W

150 kHz to 80 MHz 80 MHz to 800 MHz 800 MHz to 2.5 GHz

0.01 0.12 0.12 0.23

0.1 0.37 0.38 0.73

1 1.2 1.2 2.3

10 3.8 3.8 7.3

100 12 12 23

For transmitters rated at a maximum output power not listed above, the recommended separation distance d in metres (m) can be estimated using the equation applicable to the frequency of the transmitter, where P is the maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer. NOTE 1 At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies. NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorp- tion and reflection from structures, objects and people.

m

Page 15 of the Beurer IH26 user manual — ELECTROMAGNETIC COMPATIBILITY INFORMATION
Seite 15 · Zum Vergrößern klicken
Seite 15
Page 15 of the Beurer IH26 user manual — ELECTROMAGNETIC COMPATIBILITY INFORMATION
Page 16 of the Beurer IH26 user manual — ELECTROMAGNETIC COMPATIBILITY INFORMATION
Seite 16 · Zum Vergrößern klicken
Seite 16

758.099 - 0616

  • Beurer GmbH
  • Söflinger Str. 218
  • 89077 Ulm, Germany
  • <a href="http://www.beurer.com" rel="nofollow noopener" target="_blank">www.beurer.com</a> <a href="http://www.beurer-gesundheitsratgeber.com" rel="nofollow noopener" target="_blank">www.beurer-gesundheitsratgeber.com</a>
  • <a href="http://www.beurer-healthguide.com" rel="nofollow noopener" target="_blank">www.beurer-healthguide.com</a> 0483
Quellen

Referenzen & Links

In dieser Anleitung genannte Seiten – Hersteller, Support und Downloads.

Reparatur

Fehlerbehebung

Häufige Probleme, wahrscheinliche Ursachen und Lösungen.

ProblemMögliche UrsacheAbhilfe
The atom- iser does not produce any aerosol or produces too little aerosol.1. Too much or too little medication in the atomiser. Minimum: 2 ml, Maximum: 8 ml. p.8
The atom- iser does not produce any aerosol or produces too little aerosol.2. Check nozzle for blockages. Clean nozzle if necessary (e.g. by rinsing out). Then start the atomiser again. IMPORTANT: Carefully pierce the fine holes from the underside of the nozzle only. p.8
The atom- iser does not produce any aerosol or produces too little aerosol.3. Atomiser not held vertically. p.8
The atom- iser does not produce any aerosol or produces too little aerosol.4. Unsuitable medication fluid added for nebulisation (e.g. too viscous). The medication fluid should be prescribed by the doctor. p.8
The output is too low.Kinked hose, clogged filter, too much inhalation solution. p.8
What medica- tions are suitable for inhaling?Please consult your doctor in relation to this matter. As a rule, all medication that is suitable and approved for device inhalation, can be inhaled. p.8
There is inha- lation solution residue in the atomiser.This is normal and is necessary for technical reasons. Stop inhalation once the atomiser starts to make a notably different sound. p.8
What should be taken into account when using the device with infants and children?1. On infants and children, the mask should cover the mouth and nose to ensure effective inhalation. p.8
What should be taken into account when using the device with infants and children?2. On children, the mask should also cover both the nose and mouth. It is not a good idea to carry out nebulisation on some- one who is sleeping, as in this case not enough of the medica- tion will reach the lungs. p.8
What should be taken into account when using the device with infants and children?Note: Inhalation should only be carried out under the supervision of an adult and with their assis- tance and the child should not be left alone. p.8
Inhalation using the mask takes longer.This is necessary for technical rea- sons. You breathe less medication per breath through the holes of the mask than via the mouthpiece. The aerosol is mixed with ambient air via the holes. p.9
Why should the atomiser be replaced regularly?There are two reasons for this: 1. To ensure a therapeutically effective particle spectrum, the nozzle hole must not exceed a specific diameter. Due to mechanical and thermal stresses, the plastic is subject to a certain degree of wear. The atomiser insert [10] is particularly sensitive. This can also alter the composition of the aerosol drop- lets, which directly affects the effectiveness of the treatment. 2. R egularly changing the atomiser is also recommended for hy- giene reasons. p.9
Should each person have their own atomiser?Yes, this is absolutely essential for hygiene reasons. p.9

Frage zu Beurer IH26 stellen

Besitzer und unser Team antworten in der Regel innerhalb eines Tages – und Ihre Frage hilft auch anderen weiter.

Add your email to get notified when it’s answered — no spam, ever.

0 questions
Noch keine Fragen zu Beurer IH26

Stellen Sie die erste Frage — andere Nutzer oder unser Team antworten Ihnen.

Originaldokument

Beurer IH26 PDF-Anleitung

88 Seiten·6 MB·Inhalator·PDF
Klicken Sie oben auf eine Seite, um sie hier zu öffnen
Seite 1 / …
100%
Seiten werden geladen…