Dear customer, Thank you for choosing one of our products. Our name stands for high-quality, thoroughly tested prod- ucts for applications in the areas of heat, gentle therapy, blood pressure/diagnosis, weight, massage, beauty, air and babycare. Read these instructions for use carefully and keep them for later use, be sure to make them accessible to other users and observe the information they contain.

With kind regards, Your Beurer team

1. Getting to know your device

Check that the packaging of the Beurer BM 95 blood pressure monitor with ECG function has not been tampered with and make sure that all the required contents are present. Before use, ensure that there is no visible damage to the device or accessories and that all packaging material has been re- moved. If you have any doubts, do not use the device and contact your retailer or the specified Cus- tomer Services address. The blood pressure monitor with ECG function is used to carry out non-invasive measurement and monitoring of the arterial blood pressure values in adults and to calculate the heart rhythm. This allows you to quickly and easily measure your blood pressure, save the measured values and display the development and average values of the measured values taken (possible only in Beurer CardioExpert software and app). The recorded values are classified and evaluated graphically. The ECG stick is used to calculate the heart rhythm. The device provides information about your av- erage pulse value as well as any divergence from a normal ECG. The Beurer CardioExpert software and app allow the recorded results to be presented graphically and printed for your doctor.

  • The BM 95 blood pressure monitor with ECG function provides the following functions:
  • Integration of blood pressure measurement and ECG measurement into a single device.
  • Measurement of systolic and diastolic pressure.
  • Practical ECG measurement with handy ECG stick.
  • ECG recording of 30 seconds.
  • Transfer of the data via USB and Bluetooth ® Smart.
Page 3 of the Beurer BM 95 user manual — 2. Important notes
Seite 3 · Zum Vergrößern klicken
Seite 3

2. Important notes

Signs and symbols

The following symbols are used in these instructions for use, on the packaging and on the type plate for the device and the accessories:

Beurer BM 95 — 2. Important notes diagram (page 3)
Abb. 1 · 2. Important notes

The CE labelling certifies that the product complies with the es- sential requirements of Directive 93/42/EEC on medical products.

Disposal in accordance with the Waste Electrical and Electronic Equipment EC Directive – WEEE

  • In order to ensure comparable values, always measure your blood pressure at the same time of day.
  • Before every measurement, relax for about five minutes.
  • If you want to perform several measurements on the same person, wait five minutes between each measurement.
  • Do not take a measurement within 30 minutes of eating, drinking, smoking or exercising.
  • Repeat the measurement if you are unsure of the measured value.
Page 4 of the Beurer BM 95 user manual — 2. Important notes
Seite 4 · Zum Vergrößern klicken
Seite 4
  • The measurements taken by you are for your information only – they are no substitute for a medical examination! Discuss the measured values with your doctor and never base any medical decisions on them (e.g. medicines and their administration).
  • Do not use the blood pressure monitor on newborns or patients with preeclampsia. We recommend consulting a doctor before using the blood pressure monitor during pregnancy.
  • Cardiovascular diseases may lead to incorrect measurements or have a detrimental effect on measurement accuracy. The same also applies to very low blood pressure, diabetes, circulatory disorders and arrhythmias as well as chills or shaking.
  • The blood pressure monitor must not be used in connection with a high-frequency surgical unit.
  • Only use the unit on people that have the specified upper arm measurement for the unit.
  • Please note that when inflating, the functions of the limb in question may be impaired.
  • During the blood pressure measurement, the blood circulation must not be stopped for an unnecessarily long time. If the device malfunctions remove the cuff from the arm.
  • Avoid any mechanical restriction, compression or bending of the cuff line.
  • Do not allow sustained pressure in the cuff or frequent measurements. The resulting restriction of the blood flow may cause injury.
  • Make sure that the cuff is not placed on an arm in which the arteries or veins are undergoing medical treatment, e.g. intravascular access or intravascular or therapy, or an arteriovenous (AV) shunt.
  • Do not use the cuff on people who have undergone a mastectomy.
  • Do not place the cuff over wounds as this may cause further injury.
  • The blood pressure monitor can only be operated with batteries. Please note that data transfer and data storage is only possible when your blood pressure monitor is supplied with power. As soon as the batteries are empty, the blood pressure monitor loses the date and time.
  • To conserve the batteries, the blood pressure monitor switches off automatically if you do not press any buttons for two minutes.
  • The device is only intended for the purpose described in these instructions for use. The manufacturer is not liable for damage resulting from improper or careless use.
  • General notes on using the ECG stick
  • The ECG measuring device is a single-channel electrocardiogram (ECG) device that you can use to rapidly record an electrocardiogram (ECG). In addition, the device provides a clear evaluation of the recording, particularly as regards any disruption of the heart rhythm.
  • The ECG measuring device displays changes in the heart rhythm, which may have various different causes. These may be harmless, but may also be triggered by illnesses or diseases of differing degrees of severity. Please consult a medical specialist if you believe you may have an illness or disease.
Page 5 of the Beurer BM 95 user manual — 2. Important notes
Seite 5 · Zum Vergrößern klicken
Seite 5
  • Electrocardiograms recorded using the ECG measuring device show the heart's function at the time of the measurement. As such, any changes occurring before or after the measurement is taken may not necessarily be detected.
  • ECG measurements, such as those taken with the ECG measuring device, cannot identify all heart diseases. Regardless of the measurement taken, you should consult your doctor immediately if you experience symptoms that could indicate acute heart disease. Such symptoms could include (but are not limited to): - pain or pressure on the left of the chest area or abdomen, - radiating pain in the area of the mouth/jaw/face, or in the shoulders, arm or hands, - back pain, - nausea, - burning sensation in the chest, - tendency to collapse, - breathlessness, - rapid heartbeat or irregular heart rhythm - or, in particular, a combination of these symptoms.
  • ALWAYS consult a doctor IMMEDIATELY if you experience any of these symptoms. If you have any doubts, seek an emergency medical examination.
  • Do not self-diagnose or self-medicate on the basis of the measurement without consulting your doctor. In particular, do not start taking any new medication or change the type and/or dosage of any existing medication without prior approval.
  • The ECG measuring device is not a substitute for a medical examination of your heart function or for medical electrocardiogram recordings, which require more complex measurements.
  • It is not possible to use the ECG measuring device to diagnose illnesses or diseases that may be causing an ECG change. This is exclusively the responsibility of your doctor.
  • We recommend that you record the ECG curves and provide them to your doctor if required. This applies in particular if the status messages of the ECG measuring device do not show the “OK” symbol.
  • Important safety notes on using the ECG stick
  • We recommend against using the device if you have a pacemaker or other implanted devices. Follow the advice given by your doctor, if applicable.
  • Do not use the device with a defibrillator.
  • Do not use the device during an MRT examination.
Page 6 of the Beurer BM 95 user manual — 2. Important notes
Seite 6 · Zum Vergrößern klicken
Seite 6
  • Do not expose the device to static electricity. Always ensure that you are free of static electricity before operating the device.
  • Never submerge the device in water or other liquids. Do not clean the device with acetone or other volatile solutions. Clean the device with a cloth dampened with water or a mild cleaning liquid. Then dry the device using a dry cloth.
  • Do not place the device in pressure vessels or gas sterilisation devices.
  • Do not allow the device to fall and do not step on or shake the device.
  • Do not dismantle the device, as this could cause damage or malfunctions or impede the operation of the device.
  • Do not use the device with persons with sensitive skin or allergies.
  • This device is not intended for use by people (including children) with restricted physical, sensory or mental skills or a lack of experience and/or a lack of knowledge, unless they are supervised by a person who is responsible for their safety or are instructed by such a person in how to use the device. Supervise children around the device to ensure they do not play with it.
  • This device is not approved for use with children weighing less than 10 kg.
  • Do not allow the electrodes of the device to come into contact with other conductive parts (including earth).
  • Do not store the device in the following locations: locations in which the device is exposed to direct sunlight, high temperatures or levels of moisture, or heavy contamination; locations near to sources of water or fire; or locations that are subject to strong electromagnetic influences.

Instructions for storage and maintenance • The blood pressure monitor with ECG function is made from precision and electronic components. The accuracy of the measured values and service life of the device depend on its careful handling: – Protect the device from impacts, humidity, dirt, marked temperature fluctuations and direct

sunlight. – Do not drop the device. – Do not use the device in the vicinity of strong electromagnetic fields and keep it away from radio

systems or mobile telephones. – Only use the cuff included with the delivery or original replacement parts. Otherwise incorrect

measured values will be recorded. • We recommend that the batteries be removed if the device will not be used for a prolonged peri- od of time.

Page 7 of the Beurer BM 95 user manual — 2. Important notes
Seite 7 · Zum Vergrößern klicken
Seite 7
  • Choking hazard ! Small children may swallow and choke on batteries. Store the batteries out of the reach of small children.
  • Observe the plus (+) and minus (-) polarity signs.
  • If a battery has leaked, put on protective gloves and clean the battery compartment with a dry cloth.
  • Protect batteries from excessive heat.
  • Risk of explosion! Never throw batteries into a fire.
  • Do not charge or short-circuit batteries.
  • If the device is not to be used for a relatively long period, take the batteries out of the battery compartment.
  • Use identical or equivalent battery types only.
  • Always replace all batteries at the same time.
  • Do not use rechargeable batteries.
  • Do not disassemble, split or crush the batteries.
Page 8 of the Beurer BM 95 user manual — 3.1 Main unit
Seite 8 · Zum Vergrößern klicken
Seite 8

3. Device description

3.1 Main unit

1. Connection for cuff connector
2. Setting buttons (for date and time/alarm)
3. START/STOP button (blood pressure measurement)
4. Memory button M (to access stored measured values, change user memory)
5. Connection for ECG stick or USB transfer cable
Page 9 of the Beurer BM 95 user manual — 3.3 Display description
Seite 9 · Zum Vergrößern klicken
Seite 9

3.2 ECG stick

Beurer BM 95 — 3.2 ECG stick diagram (page 9)
Abb. 2 · 3.2 ECG stick
1. Upper electrode
2. Lower electrode
3. Activation switch

3.3 Display description

1. Heart rate display: flashes in sync with your heartbeat during the measurement. 2. Heart rate: shows the average heart rate during the recording process. 3. ECG result display 4. Systolic blood pressure display 5. Diastolic blood pressure display 6. Classification of measured values 7. Standby mode = time, ECG measurement = 30 sec. countdown; memory mode = shows the number of measurements/measurement time 8. Alarm symbol 9. Low battery status symbol 10. USB connection symbol 11. Bluetooth® connection symbol 12. Current user memory 13. Measurement mode (blood pressure = “BP”, ECG = “ECG”

Beurer BM 95 — 3.3 Display description diagram (page 9)
Abb. 3 · 3.3 Display description

7 8 9 10 11 12 13

Page 10 of the Beurer BM 95 user manual — 4.2 Set the date and time
Seite 10 · Zum Vergrößern klicken
Seite 10

4. Initial use

4.1 Insert the batteries

1. Slide off the battery compartment lid on the rear of the main unit. 2. Insert four AAA type (LR03) batteries into the battery compartment. Make sure that the batteries are inserted the correct way round. 3. Then close the battery compartment lid until you see and hear it click into place.

The batteries must be replaced when appears on the display.

4.2 Set the date and time

After inserting the batteries, the date and time must be set. The device automatically stores all measured values with the date and time.

The date and time must be reset after replacing the batteries.

The following parameters are set one after the other: Year -> Month -> Day -> Hour -> Minute

1. Press and hold the + button on the side for 5 seconds. The year flashes on the display.
2. Set the year using the +/- buttons on the side ( you can select a year between 2015 and 2031). To confirm the year, press the memory button M .
Page 11 of the Beurer BM 95 user manual — 4.3 Setting the alarm
Seite 11 · Zum Vergrößern klicken
Seite 11
3. The month now flashes on the display. Set the month using the +/- buttons. To confirm the month, press the memory button M .
4. The day now flashes on the display. Set the day using the +/- buttons. To confirm the day, press the memory button M .
5. The hour now flashes on the display. Set the hour using the +/- buttons. To confirm the hour, press the memory button M .
6. The minute now flashes on the display. Set the minute using the +/- buttons. To confirm the minute, press the memory button M .
7. The set year appears on the display again. To complete the operation, press the START/STOP button . The device then switches off.

4.3 Setting the alarm

An optional alarm can be set on the device. To set the alarm, proceed as follows:

1. Press and hold the - button on the side for 5 seconds. “ oFF ” flashes on the display if the alarm is deactivated. “ ” flashes on the display if the alarm is active. Activate or deactivate the alarm using the +/- buttons. Press the memory button M to confirm the selection.
2. The alarm hour flashes on the display when the alarm has been activated. Set the alarm hour using the +/- buttons. To confirm the alarm hour, press the memory button M .
3. The alarm minute now flashes on the display. Set the alarm minute using the +/- buttons. To confirm the alarm minute, press the memory button M .
4. “ ” appears on the display. To complete the operation, press the START/STOP button . The device then switches off.
Page 12 of the Beurer BM 95 user manual — 5.1 Attaching the cuff
Seite 12 · Zum Vergrößern klicken
Seite 12

4.4 Select the user memory

The device has two user memories. Each user memory can store up to 60 measured values. Once all memory spaces in a user memory have been used, the oldest measured values are replaced by the most recent. • Briefly press the memory button M. The last set user is displayed. Press and hold the memory but- ton M for 5 seconds to switch between the user memories.

5. Measuring blood pressure

Before commencing blood pressure measurement, observe the “Notes on using the blood pressure monitor” in chapter “2. Important notes”.

5.1 Attaching the cuff

Place the cuff on to the bare left upper arm. The circulation of the arm must not be hindered by tight clothing or similar.

The cuff must be placed on the upper arm so that the bottom edge is posi- tioned 2 – 3 cm above the elbow and over the artery. The line should point to the centre of the palm.

Now tighten the free end of the cuff, but make sure that it is not too tight around the arm and close the hook-and-loop fastener. The cuff should be fastened so that two fingers can fit under the cuff.

Now insert the cuff line into the connection for the cuff connector.

Page 13 of the Beurer BM 95 user manual — 5.3 Start the blood pressure measurement
Seite 13 · Zum Vergrößern klicken
Seite 13

Blood pressure may vary between the right and left arm, which may mean that the measured blood pres- sure values are different. Always perform the measurement on the same arm. Blood pressure may vary between the right and left arm, which may mean that the measured blood pressure values are different. If the values between the two arms are significantly different, please consult your doctor to determine which arm should be used for the measurement.

5.2 Adopting the correct posture

  • Before every measurement, relax for about five minutes. Otherwise deviations can occur.
  • You can take the measurement while sitting or lying. Always make sure that the cuff is at heart level.
  • To take your blood pressure, make sure you are sitting comfortably with your arms and back leaning on something. Do not cross your legs. Place your feet flat on the ground.
  • To avoid falsifying the measurement, it is important to remain still during the measurement and not to speak.

5.3 Start the blood pressure measurement

1. As described above, attach the cuff and adopt the correct posture.
2. Press the START/STOP button to start the blood pressure measurement. The blood pressure symbol

appears on the display and the cuff inflates. The heart symbol begins to flash in sync as soon as the device detects a pulse.

Page 14 of the Beurer BM 95 user manual — 5.4 Evaluating measurement results
Seite 14 · Zum Vergrößern klicken
Seite 14

You can cancel the blood pressure measurement at any time by pressing the START/STOP button again. Cancelled blood pressure measurements are not stored.

3. The results are displayed on the display as soon as the blood pressure measurement is complete.
4. To switch off the device, press the START/STOP button or wait for 2 minutes until the device switches off automatically.

Wait at least 5 minutes before taking another measurement!

5.4 Evaluating measurement results

The measured values can be classified and evaluated in accordance with the following table. However, these standard values serve only as a general guideline, as the individual blood pressure varies in different people and different age groups etc. It is important to consult your doctor regularly for advice. Your doctor will tell you your individual val- ues for normal blood pressure as well as the value above which your blood pressure is classified as dangerous. The bar chart on the display shows which category the recorded blood pressure values fall into. If the values of systole and diastole fall into two different categories (e.g. systole in the 'High normal' cate- gory and diastole in the 'Normal' category), the graphical classification on the device always shows the higher category; for the example given this would be 'High normal'.

Blood pressure value cate- gory

Level 3severe hypertension ≥180 ≥110 seek medical attention
Level 2moderate hypertension 160 – 179 100 – 109 seek medical attention
Level 1mild hypertension 140 – 159 90 – 99 regular monitoring by doctor

High normal 130 – 139 85 – 89 regular monitoring by doctor

Normal 120 – 129 80 – 84 self-monitoring

Optimal <120 <80 self-monitoring SourceWHO, 1999 (World Health Organization)

Systole (in mmHg) · Diastole (in mmHg) · Action

Page 15 of the Beurer BM 95 user manual — M
Seite 15 · Zum Vergrößern klicken
Seite 15

5.5 Accessing measured blood pressure values

The device has two user memories. Each user memory can store up to 60 measured values. Once all memory spaces in a user memory have been used, the oldest measured values are replaced by the most recent.

You can access stored measured values on the device at any time. Proceed as follows:

1. With the device switched off, press the memory button M . The last stored measured value in the selected user memory is displayed.

To access the measured values in the second user memory, press and hold the memory button M for 5 seconds. 2. To navigate between the individual measured values, press the mem- ory button M.

M

6. ECG measurement

6.1 Preparing the ECG measurement

  • Observe the following before beginning the ECG measurement:
  • Never use the ECG stick on top of clothing.
  • If the electrode surfaces of the ECG stick are dirty, clean them with a cotton bud soaked in rubbing alcohol.
  • If your skin or hands are dry, moisten them using a damp cloth before the measurement.
  • Please note that there must be no skin contact between your right and left hand (measurement method C) or hand and chest (measurement methods A/B). Otherwise, the measurement cannot be taken correctly.
  • Ensure that your right hand does not come into contact with your body during the measurement. To ensure an accurate measurement, press only lightly on the upper and lower electrodes of the ECG stick.
  • Never use the ECG stick upside down.
  • Do not speak or move while carrying out the ECG measurement as this can cause inaccuracies during the measurement.
Page 16 of the Beurer BM 95 user manual — 6.2 Taking the ECG measurement
Seite 16 · Zum Vergrößern klicken
Seite 16

6.2 Taking the ECG measurement

There are three different methods of taking the measurement. Start with measurement method A, “right index finger–chest”. If this method does not deliver measurements, or delivers only unstable measure- ments (“EE” is frequently displayed), switch to method B “left index finger–chest” and, if necessary, to method C “left hand–right hand”. The most suitable method/procedure depends on the heart configuration (shape of the heart) of each individual user. If it is not possible to take stable measurements using a given measurement method, this could have a harmless cause such as the shape of the heart. However, the cause may also be an illness or disease.

Measurement method C offers maximum comfort, but offers a much lower measurement sta- bility than methods A or B.

1. Insert the ECG stick cable into the connection for the ECG stick on the main unit.
2. Briefly press the activation switch on the ECG stick to switch on the device.
3. Press and hold the memory button M for 3 seconds to select the desired user memory ( or ).
Page 17 of the Beurer BM 95 user manual — 6.2 Taking the ECG measurement
Seite 17 · Zum Vergrößern klicken
Seite 17

“right index finger-chest” (approximately corresponds to “lead 2”)

Top Electrodes Place your right index finger on the upper electrode on the device and hold the device vertically in your hand.

  • You can determine the correct position for placing the lower electrode of the device against your chest using the following methods:
  • Draw an imaginary line from the front of your armpit downwards. At the same time, draw an imaginary line going up 10 cm from the lowest rib on the left side of your body. Place the lower electrode of the device at the point where these two lines meet. or
  • Draw an imaginary line from the bottom of the centre of your breastbone (sternum) to the left. At the same time, draw an imaginary line from the front of your armpit downwards. Place the lower electrode of the device at the point where these two lines meet.

Lightly press the electrode against your chest until you hear or feel a click.

Measurement method A · 5 cm

Page 18 of the Beurer BM 95 user manual — 6.2 Taking the ECG measurement
Seite 18 · Zum Vergrößern klicken
Seite 18

“left index finger-chest” (approximately corresponds to “lead 3”)

Top Electrodes Place your left index finger on the upper electrode on the device and hold the device vertically in your hand.

You can determine the correct position for placing the lower electrode of the device against your chest using the following methods: • Draw an imaginary line from the front of your armpit downwards. At the same time, draw an imaginary line going up 10 cm from the lowest rib on the left side of your body. Place the lower electrode of the device at the point where these two lines meet.

or • Draw an imaginary line from the bottom of the centre of your breastbone (sternum) to the left. At the same time, draw an imaginary line from the front of your armpit downwards. Place the lower electrode of the device at the point where these two lines meet.

Lightly press the electrode against your chest until you hear or feel a click. Important: Do not exert excessive pressure on the electrodes with your fingers.

Measurement method B · 5 cm

Page 19 of the Beurer BM 95 user manual — 6.2 Taking the ECG measurement
Seite 19 · Zum Vergrößern klicken
Seite 19

“left hand–right hand” (approximately corresponds to “lead 1”)

Place your right index finger on the upper electrode on the device. Place a finger of your left hand on the lower electrode.

Press the lower electrode until you hear or feel a click.

Please note that there must be no skin contact between your right and left hand (measurement method C) or hand and chest (measurement methods A/B). Otherwise, the measurement can- not be taken correctly. Stay still during the measurement, do not speak and hold the device still. Movements of any kind will falsify the measurements. Maintain constant pressure during the measurement. Do not press the electrodes too firmly against the skin as this can cause the muscles to tense and give rise to inaccurate meas- ured values.

4. A 30-second countdown appears at the bottom left of the display, the current heart rate is displayed in real time. At the same time, the heart symbol ( ) flashes in sync with your heartbeat.

The average heart rate is displayed after the 30-second countdown has elapsed.

5. A summary of the ECG measurement appears on the display after the 30-second countdown has elapsed.
6. To repeat the ECG measurement, press the activation switch again. To switch off the device, press the START/STOP button . Alternatively, the device will switch off automatically after 2 minutes.

Measurement method C

Page 20 of the Beurer BM 95 user manual — 6.3 Incorrect ECG measuring methods
Seite 20 · Zum Vergrößern klicken
Seite 20

6.3 Incorrect ECG measuring methods

Beurer BM 95 — 6.3 Incorrect ECG measuring methods diagram (page 20)
Abb. 4 · 6.3 Incorrect ECG measuring methods

The right index finger is not in sufficient contact with the upper electrode.

You are holding the ECG stick in your left hand.

Once you have taken the measurement, the following results may appear on the LCD display.

Indicates one or more pauses in the car- diac cycle, each of which is longer than 2 seconds.

Indicates a decreased heart rate (bradycardia) of less than 55 [bpm].

Indicates a disruption to the rhythm during the ECG recording.

Indicates an increased heart rate (tachycardia) of greater than 100 [bpm].

A flashing heart rate on the display indicates unstable or weak ECG signals. In this case, please repeat the measurement.

Page 21 of the Beurer BM 95 user manual — 7.1 System requirements
Seite 21 · Zum Vergrößern klicken
Seite 21

Background information and medical statistics for your doctor can be found in the “Annex for attend- ing doctor” document included with your device.

6.5 Accessing ECG measured values

The device has two user memories. Each user memory can store up to 60 measurements. Once all memory spaces in a user memory have been used, the oldest measurements are replaced by the most recent.

You can access stored measurements on the device at any time. Proceed as follows:

1. With the device switched off, press the memory button M . The last stored measured value in the selected user memory is displayed.

To access the measured values in the second user memory, press and hold the memory button M for 3 seconds. 2. To navigate between the individual measurements, press the mem- ory button M.

M

7. Beurer CardioExpert

To obtain a detailed representation of your recorded data, you can install the PC version of “Beurer CardioExpert” from the supplied CD or download the app, which is available as a free download from the Apple App Store or Google Play Store. The data can be transferred via USB interface or Blue- tooth® Smart.

7.1 System requirements

  • PC version
  • Operating systems: Windows XP, Windows Vista, Windows 7, Windows 8
  • Processor: Intel Core 3-3220 3.3 GHz or higher
  • RAM: min. 1 GB
  • Hard drive: min. 1 GB with 1 GB free memory
  • Screen resolution: min. 1280 x 1024
  • CD-ROM drive, USB interface
Page 22 of the Beurer BM 95 user manual — 8. What if there are problems?
Seite 22 · Zum Vergrößern klicken
Seite 22
  • App requirements
  • Bluetooth® 4.0, iOS version 7.0 or higher
  • Android™ devices - version 4.3 or higher with Bluetooth ® Smart Ready

8. What if there are problems?

Problem Possible cause Solution

The ECG measurement flash- es on the display.

The ECG signal is unstable or too weak.

Repeat the measurement as per the instructions in these Instruc- tions for use.

The ECG measurement shows “EE” on the display.

  • Contact pressure on the skin is too weak.
  • The measurement was interrupted.
  • The sound level was too high during the measurement.

Repeat the measurement as per the instructions in these Instruc- tions for use.

The blood pressure measure- ment shows “EE_0-4” on the display.

The cuff was not attached correctly.

Re-attach the cuff in accordance with the instructions in chapter “5.1 Attaching the cuff”.

The batteries are flat (“EE 4”). Replace the batteries.

Page 23 of the Beurer BM 95 user manual — 8. What if there are problems?
Seite 23 · Zum Vergrößern klicken
Seite 23

Problem Possible cause Solution

The device will not switch on. The batteries are flat. Replace the batteries.

The batteries have been in- serted incorrectly.

Re-insert the batteries, observing the correct polarity (-/+).

The cuff does not inflate. The cuff line is not correctly inserted into the device.

Ensure that the cuff line is correctly inserted into the device.

The cuff is ruptured. Replace the cuff. Contact Customer Services.

The blood pressure measure- ments are very high/low.

The cuff was not attached correctly.

Re-attach the cuff.

You moved or talked during the measurement.

Do not move or talk during the measurement.

The measurement was ob- structed by clothing.

Ensure that clothing cannot obstruct the arm cuff during the measure- ment.

Stored measured values can no longer be retrieved from the memory.

Old measured values have been overwritten by more recent values as the memory is full.

Download the stored measured values to your computer every so often.

The ECG measurement does not start even though there is contact with the skin.

Insufficient contact pressure. Ensure that the lower electrode is pressed firmly against the skin.

Bluetooth® connection failed. Connection problems be- tween the smartphone/tablet and the app.

Switch off the main unit, close the app and deactivate Bluetooth® on your smartphone/tablet. Establish the connection again.

Page 24 of the Beurer BM 95 user manual — 10. Disposal
Seite 24 · Zum Vergrößern klicken
Seite 24

Problem Possible cause Solution

The device ID does not ap- pear under “Setting” in the “CardioExpert” app.

Data transfer problem on ini- tial connection.

Switch off the main unit, close the app and deactivate Bluetooth® on your smartphone/tablet. Establish the connection again.

9. Maintenance and cleaning

  • Clean the main unit, ECG stick and cuff carefully using a slightly damp cloth only.
  • Do not use any corrosive cleaning agents or solvents.
  • If the electrode surfaces of the ECG stick are dirty, clean them with a cotton bud soaked in rubbing alcohol.
  • When you are not using the ECG stick, disconnect it from the main unit.
  • Under no circumstances should you hold the main unit, ECG stick or cuff under water, as this can cause liquid to enter and damage the device internally.
  • No heavy objects should be placed on top of the device when in storage. Remove the batteries. The cuff line should not be bent sharply.

10. Disposal

Empty, completely flat batteries must be disposed of through specially designated collection boxes, recycling points or electronics retailers. You are legally required to dispose of the batteries. The codes below are printed on batteries containing harmful substances: Pb = Battery contains lead, Cd = Battery contains cadmium, Hg = Battery contains mercury.

For environmental reasons, do not dispose of the device in the household waste at the end of its useful life. Dispose of the device at a suitable local collection or recycling point. Dispose of the device in accordance with EC Directive – WEEE (Waste Electrical and Electronic Equipment). If you have any questions, please contact the local authorities re- sponsible for waste disposal.

Pb Cd Hg

Page 25 of the Beurer BM 95 user manual — 11. Technical specifications
Seite 25 · Zum Vergrößern klicken
Seite 25

11. Technical specifications

Model No. BM 95

Measurement method Oscillometric, non-invasive blood pressure measurement on the upper arm. Single-channel ECG in freely selectable front positions / ECG sig- nal ground (earth) referenced.

Measurement range Blood pressure30 – 280 mmHg, Pulse: 40 – 199 beats/minute

ECG range/sample rate 0.05 to 40 Hz/256 Hz

Display accuracy Blood pressure: ± 3 mmHg or 2% of the value shown Pulse:

Measurement inaccuracy Max. permissible standard deviation according to clinical testing: Systolic 8 mmHg/diastolic 8 mmHg

Memory 2x 60 memory spaces

Dimensions Main unit: L 128 mm x W 128 mm x H 40 mm ECG stick: D 25 mm x H 125 mm

Weight Main unitApproximately 300 g (without batteries) ECG stick: Approximately 40 g

Cuff size 22 to 42 cm

Permissible operating conditions +10 °C to +40 °C, 30-85 % relative humidity (non-condensing)

Permissible storage conditions -20 °C to +50 °C, 10-85 % relative humidity (non-condensing)

Power supply 4x AAA batteries

Battery life For approx. 300 measurements, depending on levels of blood pressure and pump pressure

Accessories Cuff, Instructions for use, 4x AAA batteries, ECG stick, USB cable, storage pouch

Classification Internal supply, IPX0, no AP or APG, continuous operation Blood pressure: Application part, type BF ECG stick: Application part type CF

Page 26 of the Beurer BM 95 user manual — 11. Technical specifications
Seite 26 · Zum Vergrößern klicken
Seite 26

Patent TWM474484, TW201233370 / CN102631193A, CN203539335U / US20130345575 / EP2676599 / JP2014039800

  • Technical information is subject to change without notification to allow for updates.
  • This blood pressure monitor complies with European Standard EN60601-1-2 and is subject to particular precautions with regard to electromagnetic compatibility. Please note that portable and mobile HF communication systems may interfere with this unit. More details can be requested from the stated Customer Services address or found at the end of the instructions for use.
  • The blood pressure monitor complies with the EU Medical Devices Directive 93/42/EC, the German Medical Devices Act (Medizinproduktgesetz) and the standards EN1060-1 (non-invasive sphygmomanometers – Part 1: General requirements), EN1060-3 (non-invasive sphygmomanometers – Part 3: Supplementary requirements for electro-mechanical blood pressure measuring systems) and IEC80601-2-30 (Medical electrical equipment – Part 2 – 30: Particular requirements for the basic safety and essential performance of automated non-invasive sphygmomanometers).
  • The ECG stick complies with the EU Medical Devices Directive 93/42/EC, the German Medical Devices Act (Medizinproduktgesetz) and the standards IEC 60601-2-25 (Medical electrical equipment - part 2-25: Particular requirements for the safety of electrocardiographs) and IEC 60601-2-47 (Medical devices - part 2-47: Particular requirements for the safety, including essential performance characteristics of ambulatory electrocardiographic systems).
  • The safety class of the ECG stick is CF.
  • The accuracy of this blood pressure monitor has been carefully checked and developed with regard to a long useful life. If using the device for commercial medical purposes, it must be regularly tested for accuracy by appropriate means. Precise instructions for checking accuracy may be requested from the service address.

Subject to errors and changes

Page 27 of the Beurer BM 95 user manual — ELECTROMAGNETIC COMPATIBILITY INFORMATION
Seite 27 · Zum Vergrößern klicken
Seite 27

ELECTROMAGNETIC COMPATIBILITY INFORMATION

Guidance and manufacturer’s declaration - electromagnetic emissions

The BM95 is intended for use in the electromagnetic environment specified below. The customer or the user of the BM95 should assure that it is used in such an environment.

Emissions test Compliance Electromagnetic environment - guidance

RF emissions CISPR 11

Group 1 The BM95 uses RF energy only for its internal function. Therefore, its RF emissions are very low and are not likely to cause any interference in nearby electronic equipment.

RF emissions CISPR 11

Class B The BM95 is suitable for use in all establishments other than domestic and those directly connected to the pub- lic low-voltage power supply network that supplies buil- dings used for domestic purposes. Harmonic emissions IEC 61000-3-2

Voltage fluctuations/ flicker emissions

Class A · Complies

Page 28 of the Beurer BM 95 user manual — ELECTROMAGNETIC COMPATIBILITY INFORMATION
Seite 28 · Zum Vergrößern klicken
Seite 28

Guidance and manufacturer’s declaration - electromagnetic immunity

The BM95 is intended for use in the electromagnetic environment specified below. The customer or the user of BM95 should assure that it is used in such an environment.

IMMUNITY test IEC 60601 test level Compliance level Electromagnetic environment

Electrostatic discharge (ESD) IEC 61000-4-2

±6 kV contact ±8 kV air

±6 kV contact ±8 kV air

Floors should be wood, concrete or ceramic tile. If floor are covered with synthetic material, the relative humidity should be at least 30%.

Electrical fast transient/burst IEC 61000-4-4

±2 kV for power supply lines

±2kV for power supply lines

Mains power quality should be that of a typical commercial or hospital environment.

Surge IEC 61000-4-5 ±1 kV line(s) to line(s) ±2 kV line(s) to earth

±1 kV line(s) to line(s) ±2 kV line(s) to earth

Mains power quality should be that of a typical commercial or hospital environment.

Voltage dips, short interruptions and voltage vari- ations on power supply input lines IEC 61000-4-11

95% dip in UT) for 0.5 cycle

95% dip in UT) for 0.5 cycle

Mains power quality should be that of a typical commercial or hospital environment. If the user of the BM95 requires continued operation during power mains interruptions, it is recommen- ded that the BM95 be powered from an uninterruptible power supply or a battery.

40% UT (60% dip in UT) for 5 cycles

40% UT (60% dip in UT) for 5 cycles

70% UT (30% dip in UT) for 25 cycles

70% UT (30% dip in UT) for 25 cycles

95% dip in UT) for 5 sec

95% dip in UT) for 5 sec

- guidance

Page 29 of the Beurer BM 95 user manual — ELECTROMAGNETIC COMPATIBILITY INFORMATION
Seite 29 · Zum Vergrößern klicken
Seite 29

Power frequency (50Hz) magnetic field IEC 61000-4-8

3 A/m Not applicable Not applicable

Guidance and manufacturer’s declaration - electromagnetic immunity

The BM95 is intended for use in the electromagnetic environment specified below. The customer or the user of the BM95 should assure that it is used in such an environment.

IMMUNITY test IEC 60601 test level Compliance level Electromagnetic environment -

Portable and mobile RF commu- nications equipment should be used no closer to any part of the BM95 including cables, than the recommended separation dis- tance calculated from the equa- tion applicable to the frequency of the transmitter. Recommended separation dis- tance d = 1,2 P

Conducted RF IEC 61000-4-6

3 Vrms 150 kHz to 80 MHz

Radiated RF IEC 61000-4-3

3 V/m 80 MHz to 2.5 GHz

d = 1.2 P 80 MHz to 800 MHz

d = 2.3 P 800 MHz to 2.5 GHz

where P is the maximum output power rating of the transmitter in watts (W) according to the trans- mitter manufacturer and d is the recommended separation Dis- tance in metres (m).

guidance · 3 Vrms · 3 V/m

Page 30 of the Beurer BM 95 user manual — W
Seite 30 · Zum Vergrößern klicken
Seite 30

a.Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telepho- nes and land mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted theoretically with accuracy. To assess the electromagnetic environment due to fixed RF transmitters, an electromagnetic site survey should be considered. If the measured field strength in the location in which the BM95 is used exceeds the applicable RF compliance level above, the BM95 should be observed to verify normal operation. If abnormal performance is observed, additio- nal measures may be necessary, such as re-orienting or relocating the BM95.

Recommended separation distances between portable and mobile RF

communications equipment and the BM95.

The BM95 is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled. The customer or the user of the BM95 can help prevent electromagnetic interference by maintaining a minimum distance between portable and mobile RF communications equipment (transmitters) and the BM95 as recommended below, according to the maximum output power of the communications equipment.

Separation distance according to frequency of transmitter

W

150 kHz to 80 MHz d = 1.2 P

80 MHz to 800 MHz d = 1.2 P

800 MHz to 2,5 GHz d = 2.3 P

0,01 0.12 0.12 0.23

0,1 0.38 0.38 0.73

1 1.2 1.2 2.3

10 3.8 3.8 7.3

100 12 12 23

Rated maximum output · power of transmitter · m

Page 31 of the Beurer BM 95 user manual — ELECTROMAGNETIC COMPATIBILITY INFORMATION
Seite 31 · Zum Vergrößern klicken
Seite 31

For transmitters rated at a maximum output power not listed above, the recommended separation distance d in meters (m) can be estimated using the equation applicable to the frequency of the transmitter, where P is the maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer. NOTE 1 At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies. NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures, objects and people.

Page 32 of the Beurer BM 95 user manual — ELECTROMAGNETIC COMPATIBILITY INFORMATION
Seite 32 · Zum Vergrößern klicken
Seite 32

751.876 - 0915

Reparatur

Fehlerbehebung

Häufige Probleme, wahrscheinliche Ursachen und Lösungen.

ProblemMögliche UrsacheAbhilfe
The ECG measurement fal sh- es on the display.The ECG signal is unstable or too weak.Repeat the measurement as per the instructions in these Instruc- tions for use. p.22
The ECG measurement shows “EE” on the display.• Contact pressure on the skin is too weak. • The measurement was in- terrupted. • The sound level was too high during the measure- ment.Repeat the measurement as per the instructions in these Instruc- tions for use. p.22
The blood pressure measure- ment shows “EE 0-4” on the _ display.The cuff was not attached correctly.Re-attach the cuf fin accordance with the instructions in chapter “5.1 Attaching the cuff”. p.22
The blood pressure measure- ment shows “EE 0-4” on the _ display.The batteries are fal t (“EE 4”).Replace the batteries. p.22
The device will not switch on.The batteries are fal t.Replace the batteries. p.23
The device will not switch on.The batteries have been in- serted incorrectly.Re-insert the batteries, observing the correct polarity (-/+). p.23
The cuf fdoes not infal te.The cuf fline is not correctly inserted into the device.Ensure that the cuf fline is correctly inserted into the device. p.23
The cuf fdoes not infal te.The cuf fis ruptured.Replace the cuf.f Contact Customer Services. p.23
The blood pressure measure- ments are very high/low.The cuff was not attached correctly.Re-attach the cuf.f p.23
The blood pressure measure- ments are very high/low.You moved or talked during the measurement.Do not move or talk during the measurement. p.23
The blood pressure measure- ments are very high/low.The measurement was ob- structed by clothing.Ensure that clothing cannot obstruct the arm cuff during the measure- ment. p.23
Stored measured values can no longer be retrieved from the memory.Old measured values have been overwritten by more recent values as the memory is full.Download the stored measured values to your computer every so often. p.23
The ECG measurement does not start even though there is contact with the skin.Insufcfi ient contact pressure.Ensure that the lower electrode is pressed frimly against the skin. p.23
Bluetooth® connection failed.Connection problems be- tween the smartphone/tablet and the app.Switch of fthe main unit, close the app and deactivate Bluetooth® on your smartphone/tablet. Establish the connection again. p.23
The device ID does not ap- pear under “Setting” in the “CardioExpert” app.Data transfer problem on ini- tial connection.Switch of fthe main unit, close the app and deactivate Bluetooth® on your smartphone/tablet. Establish the connection again. p.24

Frage zu Beurer BM 95 stellen

Besitzer und unser Team antworten in der Regel innerhalb eines Tages – und Ihre Frage hilft auch anderen weiter.

Add your email to get notified when it’s answered — no spam, ever.