Dear customer, Thank you for choosing one of our products. Our name stands for high-quality, thoroughly tested prod- ucts for applications in the areas of heat, gentle therapy, blood pressure/diagnosis, weight, massage, beauty, air and babycare. Read these instructions for use carefully and keep them for later use, be sure to make them accessible to other users and observe the information they contain.
With kind regards, Your Beurer team
Check that the packaging of the Beurer BM 95 blood pressure monitor with ECG function has not been tampered with and make sure that all the required contents are present. Before use, ensure that there is no visible damage to the device or accessories and that all packaging material has been re- moved. If you have any doubts, do not use the device and contact your retailer or the specified Cus- tomer Services address. The blood pressure monitor with ECG function is used to carry out non-invasive measurement and monitoring of the arterial blood pressure values in adults and to calculate the heart rhythm. This allows you to quickly and easily measure your blood pressure, save the measured values and display the development and average values of the measured values taken (possible only in Beurer CardioExpert software and app). The recorded values are classified and evaluated graphically. The ECG stick is used to calculate the heart rhythm. The device provides information about your av- erage pulse value as well as any divergence from a normal ECG. The Beurer CardioExpert software and app allow the recorded results to be presented graphically and printed for your doctor.

Signs and symbols
The following symbols are used in these instructions for use, on the packaging and on the type plate for the device and the accessories:

The CE labelling certifies that the product complies with the es- sential requirements of Directive 93/42/EEC on medical products.
Disposal in accordance with the Waste Electrical and Electronic Equipment EC Directive – WEEE



Instructions for storage and maintenance • The blood pressure monitor with ECG function is made from precision and electronic components. The accuracy of the measured values and service life of the device depend on its careful handling: – Protect the device from impacts, humidity, dirt, marked temperature fluctuations and direct
sunlight. – Do not drop the device. – Do not use the device in the vicinity of strong electromagnetic fields and keep it away from radio
systems or mobile telephones. – Only use the cuff included with the delivery or original replacement parts. Otherwise incorrect
measured values will be recorded. • We recommend that the batteries be removed if the device will not be used for a prolonged peri- od of time.




1. Heart rate display: flashes in sync with your heartbeat during the measurement. 2. Heart rate: shows the average heart rate during the recording process. 3. ECG result display 4. Systolic blood pressure display 5. Diastolic blood pressure display 6. Classification of measured values 7. Standby mode = time, ECG measurement = 30 sec. countdown; memory mode = shows the number of measurements/measurement time 8. Alarm symbol 9. Low battery status symbol 10. USB connection symbol 11. Bluetooth® connection symbol 12. Current user memory 13. Measurement mode (blood pressure = “BP”, ECG = “ECG”

7 8 9 10 11 12 13

The batteries must be replaced when appears on the display.
After inserting the batteries, the date and time must be set. The device automatically stores all measured values with the date and time.
The date and time must be reset after replacing the batteries.
The following parameters are set one after the other: Year -> Month -> Day -> Hour -> Minute

An optional alarm can be set on the device. To set the alarm, proceed as follows:

The device has two user memories. Each user memory can store up to 60 measured values. Once all memory spaces in a user memory have been used, the oldest measured values are replaced by the most recent. • Briefly press the memory button M. The last set user is displayed. Press and hold the memory but- ton M for 5 seconds to switch between the user memories.
Before commencing blood pressure measurement, observe the “Notes on using the blood pressure monitor” in chapter “2. Important notes”.
Place the cuff on to the bare left upper arm. The circulation of the arm must not be hindered by tight clothing or similar.
The cuff must be placed on the upper arm so that the bottom edge is posi- tioned 2 – 3 cm above the elbow and over the artery. The line should point to the centre of the palm.
Now tighten the free end of the cuff, but make sure that it is not too tight around the arm and close the hook-and-loop fastener. The cuff should be fastened so that two fingers can fit under the cuff.
Now insert the cuff line into the connection for the cuff connector.

Blood pressure may vary between the right and left arm, which may mean that the measured blood pres- sure values are different. Always perform the measurement on the same arm. Blood pressure may vary between the right and left arm, which may mean that the measured blood pressure values are different. If the values between the two arms are significantly different, please consult your doctor to determine which arm should be used for the measurement.
appears on the display and the cuff inflates. The heart symbol begins to flash in sync as soon as the device detects a pulse.

You can cancel the blood pressure measurement at any time by pressing the START/STOP button again. Cancelled blood pressure measurements are not stored.
Wait at least 5 minutes before taking another measurement!
The measured values can be classified and evaluated in accordance with the following table. However, these standard values serve only as a general guideline, as the individual blood pressure varies in different people and different age groups etc. It is important to consult your doctor regularly for advice. Your doctor will tell you your individual val- ues for normal blood pressure as well as the value above which your blood pressure is classified as dangerous. The bar chart on the display shows which category the recorded blood pressure values fall into. If the values of systole and diastole fall into two different categories (e.g. systole in the 'High normal' cate- gory and diastole in the 'Normal' category), the graphical classification on the device always shows the higher category; for the example given this would be 'High normal'.
Blood pressure value cate- gory
High normal 130 – 139 85 – 89 regular monitoring by doctor
Normal 120 – 129 80 – 84 self-monitoring
Systole (in mmHg) · Diastole (in mmHg) · Action

The device has two user memories. Each user memory can store up to 60 measured values. Once all memory spaces in a user memory have been used, the oldest measured values are replaced by the most recent.
You can access stored measured values on the device at any time. Proceed as follows:
To access the measured values in the second user memory, press and hold the memory button M for 5 seconds. 2. To navigate between the individual measured values, press the mem- ory button M.
6. ECG measurement
6.1 Preparing the ECG measurement

There are three different methods of taking the measurement. Start with measurement method A, “right index finger–chest”. If this method does not deliver measurements, or delivers only unstable measure- ments (“EE” is frequently displayed), switch to method B “left index finger–chest” and, if necessary, to method C “left hand–right hand”. The most suitable method/procedure depends on the heart configuration (shape of the heart) of each individual user. If it is not possible to take stable measurements using a given measurement method, this could have a harmless cause such as the shape of the heart. However, the cause may also be an illness or disease.
Measurement method C offers maximum comfort, but offers a much lower measurement sta- bility than methods A or B.

“right index finger-chest” (approximately corresponds to “lead 2”)
Top Electrodes Place your right index finger on the upper electrode on the device and hold the device vertically in your hand.
Lightly press the electrode against your chest until you hear or feel a click.
Measurement method A · 5 cm

“left index finger-chest” (approximately corresponds to “lead 3”)
Top Electrodes Place your left index finger on the upper electrode on the device and hold the device vertically in your hand.
You can determine the correct position for placing the lower electrode of the device against your chest using the following methods: • Draw an imaginary line from the front of your armpit downwards. At the same time, draw an imaginary line going up 10 cm from the lowest rib on the left side of your body. Place the lower electrode of the device at the point where these two lines meet.
or • Draw an imaginary line from the bottom of the centre of your breastbone (sternum) to the left. At the same time, draw an imaginary line from the front of your armpit downwards. Place the lower electrode of the device at the point where these two lines meet.
Lightly press the electrode against your chest until you hear or feel a click. Important: Do not exert excessive pressure on the electrodes with your fingers.
Measurement method B · 5 cm

“left hand–right hand” (approximately corresponds to “lead 1”)
Place your right index finger on the upper electrode on the device. Place a finger of your left hand on the lower electrode.
Press the lower electrode until you hear or feel a click.
Please note that there must be no skin contact between your right and left hand (measurement method C) or hand and chest (measurement methods A/B). Otherwise, the measurement can- not be taken correctly. Stay still during the measurement, do not speak and hold the device still. Movements of any kind will falsify the measurements. Maintain constant pressure during the measurement. Do not press the electrodes too firmly against the skin as this can cause the muscles to tense and give rise to inaccurate meas- ured values.
The average heart rate is displayed after the 30-second countdown has elapsed.
Measurement method C


The right index finger is not in sufficient contact with the upper electrode.
You are holding the ECG stick in your left hand.
Once you have taken the measurement, the following results may appear on the LCD display.
Indicates one or more pauses in the car- diac cycle, each of which is longer than 2 seconds.
Indicates a decreased heart rate (bradycardia) of less than 55 [bpm].
Indicates a disruption to the rhythm during the ECG recording.
Indicates an increased heart rate (tachycardia) of greater than 100 [bpm].
A flashing heart rate on the display indicates unstable or weak ECG signals. In this case, please repeat the measurement.

Background information and medical statistics for your doctor can be found in the “Annex for attend- ing doctor” document included with your device.
The device has two user memories. Each user memory can store up to 60 measurements. Once all memory spaces in a user memory have been used, the oldest measurements are replaced by the most recent.
You can access stored measurements on the device at any time. Proceed as follows:
To access the measured values in the second user memory, press and hold the memory button M for 3 seconds. 2. To navigate between the individual measurements, press the mem- ory button M.
7. Beurer CardioExpert
To obtain a detailed representation of your recorded data, you can install the PC version of “Beurer CardioExpert” from the supplied CD or download the app, which is available as a free download from the Apple App Store or Google Play Store. The data can be transferred via USB interface or Blue- tooth® Smart.

Problem Possible cause Solution
The ECG measurement flash- es on the display.
The ECG signal is unstable or too weak.
Repeat the measurement as per the instructions in these Instruc- tions for use.
The ECG measurement shows “EE” on the display.
Repeat the measurement as per the instructions in these Instruc- tions for use.
The blood pressure measure- ment shows “EE_0-4” on the display.
The cuff was not attached correctly.
Re-attach the cuff in accordance with the instructions in chapter “5.1 Attaching the cuff”.
The batteries are flat (“EE 4”). Replace the batteries.

Problem Possible cause Solution
The device will not switch on. The batteries are flat. Replace the batteries.
The batteries have been in- serted incorrectly.
Re-insert the batteries, observing the correct polarity (-/+).
The cuff does not inflate. The cuff line is not correctly inserted into the device.
Ensure that the cuff line is correctly inserted into the device.
The cuff is ruptured. Replace the cuff. Contact Customer Services.
The blood pressure measure- ments are very high/low.
The cuff was not attached correctly.
Re-attach the cuff.
You moved or talked during the measurement.
Do not move or talk during the measurement.
The measurement was ob- structed by clothing.
Ensure that clothing cannot obstruct the arm cuff during the measure- ment.
Stored measured values can no longer be retrieved from the memory.
Old measured values have been overwritten by more recent values as the memory is full.
Download the stored measured values to your computer every so often.
The ECG measurement does not start even though there is contact with the skin.
Insufficient contact pressure. Ensure that the lower electrode is pressed firmly against the skin.
Bluetooth® connection failed. Connection problems be- tween the smartphone/tablet and the app.
Switch off the main unit, close the app and deactivate Bluetooth® on your smartphone/tablet. Establish the connection again.

Problem Possible cause Solution
The device ID does not ap- pear under “Setting” in the “CardioExpert” app.
Data transfer problem on ini- tial connection.
Switch off the main unit, close the app and deactivate Bluetooth® on your smartphone/tablet. Establish the connection again.
Empty, completely flat batteries must be disposed of through specially designated collection boxes, recycling points or electronics retailers. You are legally required to dispose of the batteries. The codes below are printed on batteries containing harmful substances: Pb = Battery contains lead, Cd = Battery contains cadmium, Hg = Battery contains mercury.
For environmental reasons, do not dispose of the device in the household waste at the end of its useful life. Dispose of the device at a suitable local collection or recycling point. Dispose of the device in accordance with EC Directive – WEEE (Waste Electrical and Electronic Equipment). If you have any questions, please contact the local authorities re- sponsible for waste disposal.
Pb Cd Hg

Model No. BM 95
Measurement method Oscillometric, non-invasive blood pressure measurement on the upper arm. Single-channel ECG in freely selectable front positions / ECG sig- nal ground (earth) referenced.
ECG range/sample rate 0.05 to 40 Hz/256 Hz
Display accuracy Blood pressure: ± 3 mmHg or 2% of the value shown Pulse:
Measurement inaccuracy Max. permissible standard deviation according to clinical testing: Systolic 8 mmHg/diastolic 8 mmHg
Memory 2x 60 memory spaces
Dimensions Main unit: L 128 mm x W 128 mm x H 40 mm ECG stick: D 25 mm x H 125 mm
Cuff size 22 to 42 cm
Permissible operating conditions +10 °C to +40 °C, 30-85 % relative humidity (non-condensing)
Permissible storage conditions -20 °C to +50 °C, 10-85 % relative humidity (non-condensing)
Power supply 4x AAA batteries
Battery life For approx. 300 measurements, depending on levels of blood pressure and pump pressure
Accessories Cuff, Instructions for use, 4x AAA batteries, ECG stick, USB cable, storage pouch
Classification Internal supply, IPX0, no AP or APG, continuous operation Blood pressure: Application part, type BF ECG stick: Application part type CF

Patent TWM474484, TW201233370 / CN102631193A, CN203539335U / US20130345575 / EP2676599 / JP2014039800
Subject to errors and changes

Guidance and manufacturer’s declaration - electromagnetic emissions
The BM95 is intended for use in the electromagnetic environment specified below. The customer or the user of the BM95 should assure that it is used in such an environment.
Emissions test Compliance Electromagnetic environment - guidance
RF emissions CISPR 11
Group 1 The BM95 uses RF energy only for its internal function. Therefore, its RF emissions are very low and are not likely to cause any interference in nearby electronic equipment.
RF emissions CISPR 11
Class B The BM95 is suitable for use in all establishments other than domestic and those directly connected to the pub- lic low-voltage power supply network that supplies buil- dings used for domestic purposes. Harmonic emissions IEC 61000-3-2
Voltage fluctuations/ flicker emissions
Class A · Complies

Guidance and manufacturer’s declaration - electromagnetic immunity
The BM95 is intended for use in the electromagnetic environment specified below. The customer or the user of BM95 should assure that it is used in such an environment.
IMMUNITY test IEC 60601 test level Compliance level Electromagnetic environment
Electrostatic discharge (ESD) IEC 61000-4-2
±6 kV contact ±8 kV air
±6 kV contact ±8 kV air
Floors should be wood, concrete or ceramic tile. If floor are covered with synthetic material, the relative humidity should be at least 30%.
Electrical fast transient/burst IEC 61000-4-4
±2 kV for power supply lines
±2kV for power supply lines
Mains power quality should be that of a typical commercial or hospital environment.
Surge IEC 61000-4-5 ±1 kV line(s) to line(s) ±2 kV line(s) to earth
±1 kV line(s) to line(s) ±2 kV line(s) to earth
Mains power quality should be that of a typical commercial or hospital environment.
Voltage dips, short interruptions and voltage vari- ations on power supply input lines IEC 61000-4-11
95% dip in UT) for 0.5 cycle
95% dip in UT) for 0.5 cycle
Mains power quality should be that of a typical commercial or hospital environment. If the user of the BM95 requires continued operation during power mains interruptions, it is recommen- ded that the BM95 be powered from an uninterruptible power supply or a battery.
40% UT (60% dip in UT) for 5 cycles
40% UT (60% dip in UT) for 5 cycles
70% UT (30% dip in UT) for 25 cycles
70% UT (30% dip in UT) for 25 cycles
95% dip in UT) for 5 sec
95% dip in UT) for 5 sec
- guidance

Power frequency (50Hz) magnetic field IEC 61000-4-8
3 A/m Not applicable Not applicable
Guidance and manufacturer’s declaration - electromagnetic immunity
The BM95 is intended for use in the electromagnetic environment specified below. The customer or the user of the BM95 should assure that it is used in such an environment.
IMMUNITY test IEC 60601 test level Compliance level Electromagnetic environment -
Portable and mobile RF commu- nications equipment should be used no closer to any part of the BM95 including cables, than the recommended separation dis- tance calculated from the equa- tion applicable to the frequency of the transmitter. Recommended separation dis- tance d = 1,2 P
Conducted RF IEC 61000-4-6
3 Vrms 150 kHz to 80 MHz
Radiated RF IEC 61000-4-3
3 V/m 80 MHz to 2.5 GHz
d = 1.2 P 80 MHz to 800 MHz
d = 2.3 P 800 MHz to 2.5 GHz
where P is the maximum output power rating of the transmitter in watts (W) according to the trans- mitter manufacturer and d is the recommended separation Dis- tance in metres (m).
guidance · 3 Vrms · 3 V/m

a.Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telepho- nes and land mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted theoretically with accuracy. To assess the electromagnetic environment due to fixed RF transmitters, an electromagnetic site survey should be considered. If the measured field strength in the location in which the BM95 is used exceeds the applicable RF compliance level above, the BM95 should be observed to verify normal operation. If abnormal performance is observed, additio- nal measures may be necessary, such as re-orienting or relocating the BM95.
Recommended separation distances between portable and mobile RF
communications equipment and the BM95.
The BM95 is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled. The customer or the user of the BM95 can help prevent electromagnetic interference by maintaining a minimum distance between portable and mobile RF communications equipment (transmitters) and the BM95 as recommended below, according to the maximum output power of the communications equipment.
Separation distance according to frequency of transmitter
150 kHz to 80 MHz d = 1.2 P
80 MHz to 800 MHz d = 1.2 P
800 MHz to 2,5 GHz d = 2.3 P
0,01 0.12 0.12 0.23
0,1 0.38 0.38 0.73
1 1.2 1.2 2.3
10 3.8 3.8 7.3
100 12 12 23
Rated maximum output · power of transmitter · m

For transmitters rated at a maximum output power not listed above, the recommended separation distance d in meters (m) can be estimated using the equation applicable to the frequency of the transmitter, where P is the maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer. NOTE 1 At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies. NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures, objects and people.

751.876 - 0915
Häufige Probleme, wahrscheinliche Ursachen und Lösungen.
| Problem | Mögliche Ursache | Abhilfe | |
|---|---|---|---|
| The ECG measurement fal sh- es on the display. | The ECG signal is unstable or too weak. | Repeat the measurement as per the instructions in these Instruc- tions for use. | p.22 |
| The ECG measurement shows “EE” on the display. | • Contact pressure on the skin is too weak. • The measurement was in- terrupted. • The sound level was too high during the measure- ment. | Repeat the measurement as per the instructions in these Instruc- tions for use. | p.22 |
| The blood pressure measure- ment shows “EE 0-4” on the _ display. | The cuff was not attached correctly. | Re-attach the cuf fin accordance with the instructions in chapter “5.1 Attaching the cuff”. | p.22 |
| The blood pressure measure- ment shows “EE 0-4” on the _ display. | The batteries are fal t (“EE 4”). | Replace the batteries. | p.22 |
| The device will not switch on. | The batteries are fal t. | Replace the batteries. | p.23 |
| The device will not switch on. | The batteries have been in- serted incorrectly. | Re-insert the batteries, observing the correct polarity (-/+). | p.23 |
| The cuf fdoes not infal te. | The cuf fline is not correctly inserted into the device. | Ensure that the cuf fline is correctly inserted into the device. | p.23 |
| The cuf fdoes not infal te. | The cuf fis ruptured. | Replace the cuf.f Contact Customer Services. | p.23 |
| The blood pressure measure- ments are very high/low. | The cuff was not attached correctly. | Re-attach the cuf.f | p.23 |
| The blood pressure measure- ments are very high/low. | You moved or talked during the measurement. | Do not move or talk during the measurement. | p.23 |
| The blood pressure measure- ments are very high/low. | The measurement was ob- structed by clothing. | Ensure that clothing cannot obstruct the arm cuff during the measure- ment. | p.23 |
| Stored measured values can no longer be retrieved from the memory. | Old measured values have been overwritten by more recent values as the memory is full. | Download the stored measured values to your computer every so often. | p.23 |
| The ECG measurement does not start even though there is contact with the skin. | Insufcfi ient contact pressure. | Ensure that the lower electrode is pressed frimly against the skin. | p.23 |
| Bluetooth® connection failed. | Connection problems be- tween the smartphone/tablet and the app. | Switch of fthe main unit, close the app and deactivate Bluetooth® on your smartphone/tablet. Establish the connection again. | p.23 |
| The device ID does not ap- pear under “Setting” in the “CardioExpert” app. | Data transfer problem on ini- tial connection. | Switch of fthe main unit, close the app and deactivate Bluetooth® on your smartphone/tablet. Establish the connection again. | p.24 |