i Teledyne Analytical Instruments OPERA TING INSTR UCTIONS FOR Model 3300TB T race Oxygen Analyz er HIGHLY TOXIC AND OR FLAMMABLE LIQUIDS OR GASES MAY BE PRESENT IN THIS MONITORING SYSTEM. PERSONAL PROTECTIVE EQUIPMENT MAY BE REQUIRED WHEN SERVICING THIS SYSTEM.
ii Teledyne Analytical Instruments Copyright © 1999 Teledyne Analytical Instruments All Rights Reserved. No part of this manual may be reproduced, transmitted, tran- scribed, stored in a retrieval sy.
iii Teledyne Analytical Instruments Contents Introduction 1. 1 Overview ........................................................................ 1-1 1.2 Main Features of the Analyzer ....................................... 1-1 1.3 F ront P anel Descr iption .
iv Teledyne Analytical Instruments 4.4.1 Alarm 1 .................................................................. 4-3 4.4.2 Alarm 2 .................................................................. 4-3 4.4.3 Sensor F ail Alar m ....................
v Teledyne Analytical Instruments Model 3300TB complies with all of the requirements of the Commonwealth of Europe (CE) for Radio Frequency Interference, Electromagnetic Interference (RFI/EMI), and Low Voltage Directive (LVD).
vi Teledyne Analytical Instruments COMBUSTIBLE GAS USA GE W ARNING D ANGER This is a general purpose instrument, the Control Unit is de- signed for usage in a nonhazardous area. It is the customer's responsibility to ensure safety especially when combustible gases are being analyzed since the potential of gas leaks al- ways exist.
1-1 Trace Oxygen Analyzer Introduction 1 Teledyne Analytical Instruments Introduction 1.1 Overview The Teledyne Electronic Technologies Analytical Instruments (TET/AI) Model 3300TB is a microprocessor-based trace oxygen analyzer for real-time measurement of the parts per million of oxygen in inert gases, or in a wide variety of gas mixtures.
1-2 1 Introduction Model 3300 TB Teledyne Analytical Instruments • Three analog outputs: two for measurement (0–10 V dc, and negative ground 4–20 mA dc) and one for range identification (0-10 V dc). • Compact and rugged Control Unit, wall mounted NEMA-4 rated enclosure.
1-3 Trace Oxygen Analyzer Introduction 1 Teledyne Analytical Instruments • Set HI Range Set the high analysis range for the instrument (up to 0-9999ppm). • Set LO Range Set the low analysis range for the instrument (down to 0-10ppm). • Span Span calibrate the analyzer.
1-4 1 Introduction Model 3300 TB Teledyne Analytical Instruments • Power Connection AC version: 100–240 VAC, at 50/60Hz. The connector housing includes the fuse holder and the power switch. Fuse Holder: Replacing the fuse is described in Chapter 5, Maintenance .
2-1 Trace Oxygen Analyzer Operational Theory 2 Teledyne Analytical Instruments Operational Theory 2. 1 Introduction The analyzer is composed of two subsystems: 1.
2-2 2 Operational Theory Model 3300 TB Teledyne Analytical Instruments Micro-Fuel Cell is therefore a hybrid between a battery and a true fuel cell. (All of the reactants are stored externally in a true fuel cell.
2-3 Trace Oxygen Analyzer Operational Theory 2 Teledyne Analytical Instruments conductor connecting them, through some electrical circuitry, to one of the external contacts in the connector receptacle, which is on the top of the cell. 2.2.3 Electrochemical Reactions The sample gas diffuses through the Teflon membrane.
2-4 2 Operational Theory Model 3300 TB Teledyne Analytical Instruments external current, will increase, even though the proportion of oxygen has not changed. Fortunately, Dalton's Law confirms that every gas in a mixture contributes the same pressure to the mixture that it would exert if it were alone in the same amount in that same volume.
2-5 Trace Oxygen Analyzer Operational Theory 2 Teledyne Analytical Instruments 2. 3 Electronics 2.3.1 General The signal processing uses an Intel microcontroller with on-board RAM and ROM to control all signal processing, input/output, and display functions for the analyzer.
2-6 2 Operational Theory Model 3300 TB Teledyne Analytical Instruments In the presence of oxygen the cell generates a current. The sensor has an internal thermistor compensation network. The output of the sensor is converted to voltage millivolt range.
3-1 Trace Oxygen Analyzer Installation 3 Teledyne Analytical Instruments Installation of the analyzer includes: 1. Unpacking the system. 2. Mounting the Control Unit to a wall 3. Installing the Micro-Fuel Cell 4. Making the electrical connections. 5. Making the gas connections.
3 Installation Model 3300 TB 3-2 Teledyne Analytical Instruments 3 . 2 Location and Mounting 3.2.1 Control Unit Installation The 3300TB Control Unit is designed to be wall-mounted, in a general purpose, area. The unit should be installed at viewing level in a sheltered area.
3-3 Trace Oxygen Analyzer Installation 3 Teledyne Analytical Instruments 3 . 3 Electrical Connections Figure 3-1 shows the Model 3300TB rear panel of the control unit. For detailed pinouts, see the wiring/interconnection drawings in the Draw- ings section at the rear of this manual.
3 Installation Model 3300 TB 3-4 Teledyne Analytical Instruments Analog Outputs: There are three DC output signal connectors with screw terminals on the panel.
3-5 Trace Oxygen Analyzer Installation 3 Teledyne Analytical Instruments The specific descriptions for each type of alarm are as follows: Alarm #1 Programmable as high or low alarm (actuates when concentration is above threshold). Can be set anywhere between 0 and 9999ppm, but must be set ABOVE the threshold set for the LO Alarm.
3 Installation Model 3300 TB 3-6 Teledyne Analytical Instruments 3.4.1 Vacuum Service Option For vacuum service pluming reference figure 3-4. The vacuum service option is recommended for applications in which the sample source is not pressurized. The customer must supply a pump and by-pass system to complete the sample system.
3-7 Trace Oxygen Analyzer Installation 3 Teledyne Analytical Instruments 3 . 5 Installation Checklist Before connecting the instrument to the power source and turning it on, make sure you have: • Correctly installed the Sample and Exhaust gas lines • Checked for leaks • Set the sample pressure to 5–10 psig (34.
3 Installation Model 3300 TB 3-8 Teledyne Analytical Instruments.
Trace Oxygen Analyzer Operation 4 4-1 Teledyne Analytical Instruments Operation 4. 1 Introduction Once the analyzer has been mounted, the gas lines connected and the electrical connections made, the Analyzer can be configured for your appli- cation. This involves setting the system parameters: • Defining the user selectable analysis ranges.
4 Operation Model 3300 TB 4-2 Teledyne Analytical Instruments 4 . 2 Using the Function and Data Entry Buttons When no buttons on the Analyzer are being pressed, the instrument is in the Analyze mode. It is monitoring the anount of oxygen in the sample gas that is flowing through the sampling system.
Trace Oxygen Analyzer Operation 4 4-3 Teledyne Analytical Instruments 4.3.1 HI Range Setting the HI Range fullscale value defines the LEAST sensitive analysis range to be used.
4 Operation Model 3300 TB 4-4 Teledyne Analytical Instruments until the display reads the desired percent concentration. (if within 5 seconds no key is pressed, the instrument will return to the sample mode and display oxygen concentration).
Trace Oxygen Analyzer Operation 4 4-5 Teledyne Analytical Instruments 4 . 5 Selecting a Fixed Range or Autoranging The Model 3300TB can operate in fixed high, fixed low, or autoranging mode. To change modes: 1. Press and then release the SET HI RANGE and the SET LO RANGE buttons simultaneously.
4 Operation Model 3300 TB 4-6 Teledyne Analytical Instruments 4 . 7 Displa ying P er cent & PPM on the LED Display The analyzer displays the concentration in percent whenever the read- ing is over 9999 ppm. When the reading changes to percent, the LED display will alternate between flashing “PC” and the oxygen concentration.
5-1 Trace Oxygen Analyzer Maintenance 5 Teledyne Analytical Instruments Maintenance Aside from normal cleaning and checking for leaks at the gas connections, the Model 3300TB should not require any maintenance beyond replacement of expended Micro-Fuel Cells, and perhaps a blown fuse.
5-2 5 Maintenance Model 3300 TB Teledyne Analytical Instruments fuse holder will slide out. There are two fuses in use and are visible in the clip. 4. Remove the bad fuse and replace it with a 5x20mm 0.5 A, 250 VAC, IEC time lag (T) fuse (P/N F1128). 5.
5-3 Trace Oxygen Analyzer Maintenance 5 Teledyne Analytical Instruments WARNING: THE SENSOR USED IN THE MODEL 3300TB CON- TAINS AN ELECTROLYTE WHICH INCLUDES SUBSTANCES THAT ARE EXTREMELY HARMFUL IF TOUCHED, SWALLOWED, OR INHALED. AVOID CONTACT WITH ANY FLUID OR POWDER IN OR AROUND THE UNIT.
5-4 5 Maintenance Model 3300 TB Teledyne Analytical Instruments 4. Slowly withdraw the cap from the block. The cell should come out with the cap . 5. Carefully pull the cell off of the cap. DO NOT TOUCH THE SCREENED END OF THE CELL OR ANY FLUID THAT MAY BE LEAKING FROM IT.
5-5 Trace Oxygen Analyzer Maintenance 5 Teledyne Analytical Instruments If you have a warranty claim, you must return the cell in question to the factory for evaluation. If it is determined that failure is due to faulty workmanship or material, the cell will be replaced at no cost to you.
5-6 5 Maintenance Model 3300 TB Teledyne Analytical Instruments.
A-1 Trace Oxygen Analyzer Appendix Teledyne Analytical Instruments Appendix A.1 Specifications Ranges : Two user selectable ranges can be set between 0-10 ppm and 0-9999 ppm oxygen, and a 0-25 % (nominal) Air Cali- bration Range. Signal Output : Voltage: 0–10 V dc, negative ground Current: 4-20 mA, negative ground Range ID: 0-10 V dc.
A-2 Appendix Model 3300 TB Teledyne Analytical Instruments A . 2 Spare P arts List QTY P/N DESCRIPTION 1 C-65220-A PC Board, Main 1 C-64586 PC Board, Power Supply 1 C-6689-B-2C Micro-Fuel Cell, class .
A-3 Trace Oxygen Analyzer Appendix Teledyne Analytical Instruments A . 3 Reference Drawing D-70361 Outline Diagram D-70362 Final Assembly A.4 Miscellaneous The symbol: ~ is used on the rear panel of the model 3300TB to signify volts alternating current (VAC).
A-4 Appendix Model 3300 TB Teledyne Analytical Instruments.
A-5 Trace Oxygen Analyzer Appendix Teledyne Analytical Instruments A.5 Material Safety Data Sheet Section II – Ph ysical and Chemical Data Chemical and Common Names: Potassium Hydroxide (KOH), 15% (.
A-6 Appendix Model 3300 TB Teledyne Analytical Instruments Section III – Ph ysical Hazards Section IV – Health Hazard Data Potential for fire and explosion: The electrolyte in the Micro-Fuel Cells is not flammable. There are no fire or explosion hazards associated with Micro-Fuel Cells.
A-7 Trace Oxygen Analyzer Appendix Teledyne Analytical Instruments Section V – Emergency and First Aid Procedures NOT E: The oxygen sensors are sealed, and under normal circumstances, the contents of the sensors do not present a health hazard. The following information is given as a guide in the event that a cell leaks.
Ein wichtiger Punkt beim Kauf des Geräts Teledyne 3300TB (oder sogar vor seinem Kauf) ist das durchlesen seiner Bedienungsanleitung. Dies sollten wir wegen ein paar einfacher Gründe machen:
Wenn Sie Teledyne 3300TB noch nicht gekauft haben, ist jetzt ein guter Moment, um sich mit den grundliegenden Daten des Produkts bekannt zu machen. Schauen Sie zuerst die ersten Seiten der Anleitung durch, die Sie oben finden. Dort finden Sie die wichtigsten technischen Daten für Teledyne 3300TB - auf diese Weise prüfen Sie, ob das Gerät Ihren Wünschen entspricht. Wenn Sie tiefer in die Benutzeranleitung von Teledyne 3300TB reinschauen, lernen Sie alle zugänglichen Produktfunktionen kennen, sowie erhalten Informationen über die Nutzung. Die Informationen, die Sie über Teledyne 3300TB erhalten, werden Ihnen bestimmt bei der Kaufentscheidung helfen.
Wenn Sie aber schon Teledyne 3300TB besitzen, und noch keine Gelegenheit dazu hatten, die Bedienungsanleitung zu lesen, sollten Sie es aufgrund der oben beschriebenen Gründe machen. Sie erfahren dann, ob Sie die zugänglichen Funktionen richtig genutzt haben, aber auch, ob Sie keine Fehler begangen haben, die den Nutzungszeitraum von Teledyne 3300TB verkürzen könnten.
Jedoch ist die eine der wichtigsten Rollen, die eine Bedienungsanleitung für den Nutzer spielt, die Hilfe bei der Lösung von Problemen mit Teledyne 3300TB. Sie finden dort fast immer Troubleshooting, also die am häufigsten auftauchenden Störungen und Mängel bei Teledyne 3300TB gemeinsam mit Hinweisen bezüglich der Arten ihrer Lösung. Sogar wenn es Ihnen nicht gelingen sollte das Problem alleine zu bewältigen, die Anleitung zeigt Ihnen die weitere Vorgehensweise – den Kontakt zur Kundenberatung oder dem naheliegenden Service.