INSTRUCTION MANUAL FOR TRACE OXYGEN ANALYZER MODEL 311-D RANGES: 0-100 PPM; 0-1000 PPM; 1% AND 100% TELEDYNE ANALYTICAL INSTRUMENTS 16830 CHESTNUT STREET CITY OF INDUSTRY, CALIF. 91749 TELEPHONE: (626) 934-1500 FAX: (626) 961-2538 TWX: (910) 584-1887 TDYANYL COID EASYLINK: 628-31862 Web: www.
TABLE OF CONTENTS MODEL 311D 1. INTRODUCTION 1 1.1 Description 1 1.2 Method of Analysis 1 1.3 Outstanding Features 1 1.3.1 Micro-Fuel Cell 1 1.3.2 Reliable Calibration 1 1.3.3 Integral Power Supply 2 1.3.4 Accuracy and Response 2 1.3.5 Compact Packaging 3 1.
1. INTRODUCTION 1.1 Description : The Teledyne Analytical Instruments (TAI) Model 311-D is a portable trace oxygen analyzer which can be opera ted without an external power source, and reliably calibrated without the use of cumbersome, questionable, “certified” calibration gases.
After the electronics have been properly zeroed (a one-time factory operation), the instrum ent cannot produce an output indication in the absen ce of oxygen; therefore, the need for a “zero” standardization gas is obviated.
Further disassembly may be accom plished by rem oving the backplate assemb ly from its four mounting standoffs, and laying the two separated assem blies out as illustrated on the “Analyzer Wiring Diagram”. The diagram is included among the drawings at the rear of the manual.
necessary or desirable, it must have a metallic diaphragm. Regulators with organic or plastic diaphragms are permeable to oxygen, and if used in the sampling system, will lead to high oxygen readings.
dictated by the customer’s application. If the instrument were being used to certify the oxygen content of a product for delivery, then a calibrati on prior to certification would certainly be in order.
If, on the other hand, the instrument is not to be used imm ediately after calibration, and a low ppm oxygen gas is being employed as a purge, allo w the m anifold to be purged overnight, and then disconnect both male fittings. ALWAYS DISCONNECT THE SOURCE FITTING FIRST, AND IMMEDIATELY THEREAFTER, THE VENT FITTING .
every two (2) days of continuous use. To rech arge the batteries, place the range switch in the “OFF” position and connect the power cord to a convenience outlet.
No tools are required to replace the cell in the in strument. Sim ply unscrew (ccw) the plug at the bottom of the analyzer, and the cell will drop out of the m anifold cavity. Remove the new cell from its package, and carefu lly rem ove the shorting clip.
1. Disconnect cell. 2. Move range switch to “CAL” position. 3. Adjust P1 such that the output of A3, pin 6, measures between 0 and +0.5 mV, ideally +-.3 mV. 4. Adjust P2 for 0 ± 1 mV at output of A2, pin 6. 5. Verify that the offset is the same on all ranges.
OPERATING TEMPERATURE RANGE: 30 ° TO 125 ° F. RESPONSE AND RECOVERY: AT THE SPECIFIED FLOWRATE (0.25 SCFH), THE SENSOR OUTPUT SHOULD BE, IN TEN SECONDS, WITHIN ± 10% OF AN INTRODUCED GAS HAVING A C.
EFFECT OF CO2 ON B-2C CELL LIFE -11-.
RECOMMENDED SPARE PARTS LIST MODEL 311D QTY. P/N DESCRIPTION 5 F-39 MICRO FUSE, 0.25 AMP., FAST-BLOW (FOR 210-240 VAC) 5 F-51 MICRO FUSE, 0.50 AMP., FAST-BLOW (FOR 100-125 VAC) 1 C-6689-A2C MICRO-FUEL.
DRAWING LIST MODEL 311D B-26473 PICTORAL DIAGRAM C-28339 SCHEMATIC C-26606 W IRING DIAGRAM NOTE: The MSDS on this material is available upon request through the Teledyne Environmental Health and Safety Coordinator.
.
Date: 6/20/2006 Time : 01:45:07 PM Title: Indsrsvr01drawingsMASTER26000C26606-14.dwg.
Date: 7/20/04 Time : 10:56:54 AM Title: G:MASTER28000C28339-8.DWG.
Ein wichtiger Punkt beim Kauf des Geräts Teledyne 311-D (oder sogar vor seinem Kauf) ist das durchlesen seiner Bedienungsanleitung. Dies sollten wir wegen ein paar einfacher Gründe machen:
Wenn Sie Teledyne 311-D noch nicht gekauft haben, ist jetzt ein guter Moment, um sich mit den grundliegenden Daten des Produkts bekannt zu machen. Schauen Sie zuerst die ersten Seiten der Anleitung durch, die Sie oben finden. Dort finden Sie die wichtigsten technischen Daten für Teledyne 311-D - auf diese Weise prüfen Sie, ob das Gerät Ihren Wünschen entspricht. Wenn Sie tiefer in die Benutzeranleitung von Teledyne 311-D reinschauen, lernen Sie alle zugänglichen Produktfunktionen kennen, sowie erhalten Informationen über die Nutzung. Die Informationen, die Sie über Teledyne 311-D erhalten, werden Ihnen bestimmt bei der Kaufentscheidung helfen.
Wenn Sie aber schon Teledyne 311-D besitzen, und noch keine Gelegenheit dazu hatten, die Bedienungsanleitung zu lesen, sollten Sie es aufgrund der oben beschriebenen Gründe machen. Sie erfahren dann, ob Sie die zugänglichen Funktionen richtig genutzt haben, aber auch, ob Sie keine Fehler begangen haben, die den Nutzungszeitraum von Teledyne 311-D verkürzen könnten.
Jedoch ist die eine der wichtigsten Rollen, die eine Bedienungsanleitung für den Nutzer spielt, die Hilfe bei der Lösung von Problemen mit Teledyne 311-D. Sie finden dort fast immer Troubleshooting, also die am häufigsten auftauchenden Störungen und Mängel bei Teledyne 311-D gemeinsam mit Hinweisen bezüglich der Arten ihrer Lösung. Sogar wenn es Ihnen nicht gelingen sollte das Problem alleine zu bewältigen, die Anleitung zeigt Ihnen die weitere Vorgehensweise – den Kontakt zur Kundenberatung oder dem naheliegenden Service.