Benutzeranleitung / Produktwartung P10 des Produzenten ResMed
Zur Seite of 16
User Guide Eng lis h Na s al Pi llow s Sy st em.
Fitting / Anlegen / Mise en place / Applicazione 1 2 3 Adjustment / Anpassen / Ajustement / Regolazione.
Removal / Abnehmen / Retr ait / Rimozione Disassembly / Auseinanderbau / Démontage / Smont aggio 1 2.
Reassembly / Z usammenbau / Remont age / Riassemblaggio 1 2 3.
English 1 English AirFit ™ P10 N ASAL P ILLOWS S YSTEM Thank you for choosing the AirFit P10. This document provide s the user instructions for the AirFit P10 and AirFit P10 for Her mas ks referred to collectively as AirFit P10 throughout this manual.
2 the device is turned on and functioning pro perly, new air from the device flu shes the exhaled air out through the mas k vent holes. However, when the device is not op erating, insufficient fresh air will be pro vided through the mask, and the exhaled air may be rebreathed.
English 3 Refer to your CPAP or bilevel d evice manual for details o n settings and op erational information. Remove all packaging before using the mask.
4 3. Finally, connect the end of the ma sk tubing into your device air tubing. Your mask is now fitted an d ready to use. When the device is running, a light breeze should be felt through the vent. To inspect the vent place a finger close b ut not over the entire vent area.
English 5 Cleaning your mask at home Your mask and headgear shoul d only be gently handwashed. The mask vent may di scolour over time due to dust. Fo llow the cleaning instructions i n this guide to maintai n performance and safe ty of the mask. WARNING Always follow cleaning in structions and use mild soap.
6 When all components including the vent are dry, reassemble according to the reassembly instructions. Weekly: Handwash the headgear in w arm (approximately 30°C) water using mild soap. It may be wash ed without being disassembled. Rinse well and allow it to air dry out of direct sunlight.
English 7 Problem/possible ca use Solution Vent is blocked or dirty/disc olored. If the ven t requires clea ning, use a soft bristle brush . The vent may discolo r over time a nd this is not a safety conc ern. Follow the cl eaning instructions in this guide t o maintain performance and saf ety of the mask.
8 Resistance Drop in pressure measured (nominal) at 50 L/m in: 0.4 cm H 2 O at 100 L/min: 1.4 cm H 2 O Sound DECLARED DUAL-NUMBE R NOISE EMISSION VALUES in accordance with ISO 4871. The A- weighted sound power level of the mask is 21 dBA, with uncertainty of 3 dBA.
English 9 Symbols The following sym bols may appear on your produ ct or packaging: Catalog nu mber Batch code Humidity limitation Tempera ture lim itation Fragile, hand le with care Not made with natu.
Mask components / Maskenk omponenten / Composants du masque / Componenti della maschera AirFit P10 AirFit P10 for Her * A Headgear / Kopfband / Harnais / Copricapo 62935 62936 1 Split-strap headgear /.
AirFit P10 AirFit P10 for Her * B+C Pillows system / Nasenpolstermaske / Masque narinaire / Cuscinetti nasali 62920 (XS) 62921 (S) 62922 (M) 62923 (L) 62920 (XS) 62921 (S) 62922 (M) 62923 (L) A+B+C Co.
* 1 2 3 4 5 6 7 8 * Manufacturer: ResMed L td 1 Elizabeth Macar thur Drive Bella Vista NSW 2153 Australia. Distributedby: ResMed Corp 9001 Spectrum Center Boulevard San Diego CA 92123 US A. ResMed (UK) L td 96 Jubilee Ave Milton P ark Abingdon Oxfordshire O X14 4RW UK.
Ein wichtiger Punkt beim Kauf des Geräts ResMed P10 (oder sogar vor seinem Kauf) ist das durchlesen seiner Bedienungsanleitung. Dies sollten wir wegen ein paar einfacher Gründe machen:
Wenn Sie ResMed P10 noch nicht gekauft haben, ist jetzt ein guter Moment, um sich mit den grundliegenden Daten des Produkts bekannt zu machen. Schauen Sie zuerst die ersten Seiten der Anleitung durch, die Sie oben finden. Dort finden Sie die wichtigsten technischen Daten für ResMed P10 - auf diese Weise prüfen Sie, ob das Gerät Ihren Wünschen entspricht. Wenn Sie tiefer in die Benutzeranleitung von ResMed P10 reinschauen, lernen Sie alle zugänglichen Produktfunktionen kennen, sowie erhalten Informationen über die Nutzung. Die Informationen, die Sie über ResMed P10 erhalten, werden Ihnen bestimmt bei der Kaufentscheidung helfen.
Wenn Sie aber schon ResMed P10 besitzen, und noch keine Gelegenheit dazu hatten, die Bedienungsanleitung zu lesen, sollten Sie es aufgrund der oben beschriebenen Gründe machen. Sie erfahren dann, ob Sie die zugänglichen Funktionen richtig genutzt haben, aber auch, ob Sie keine Fehler begangen haben, die den Nutzungszeitraum von ResMed P10 verkürzen könnten.
Jedoch ist die eine der wichtigsten Rollen, die eine Bedienungsanleitung für den Nutzer spielt, die Hilfe bei der Lösung von Problemen mit ResMed P10. Sie finden dort fast immer Troubleshooting, also die am häufigsten auftauchenden Störungen und Mängel bei ResMed P10 gemeinsam mit Hinweisen bezüglich der Arten ihrer Lösung. Sogar wenn es Ihnen nicht gelingen sollte das Problem alleine zu bewältigen, die Anleitung zeigt Ihnen die weitere Vorgehensweise – den Kontakt zur Kundenberatung oder dem naheliegenden Service.