Benutzeranleitung / Produktwartung MPD1508 des Produzenten Powerex
Zur Seite of 27
Please read and save these instructions. Read carefully before attempting to assemble, install, operate or maintain the product described. Protect yourself and others by observing all safety information. Failure to comply with instructions could result in personal injury and/or property damage! Retain instructions for future reference.
To help recognize this information, observe the following symbols. Danger indicates an imminently hazardous situation which, if not avoided, will result in death or serious injury. Warning indicates a potentially hazardous situation which, if not avoided, could result in death or serious injury.
3 Medical Package System compressor to avoid stresses caused by changes in hot and cold conditions. 3. Piping supports should be anchored separately from the compressor to reduce noise and vibration. 4. Never use any piping smaller than the compressor connection.
Medical Package System 4 Specifications Medical Package Duplex – Model MPD Medical Package Triplex – Model MPT Medical Package Quadplex – Model MPQ MODEL HP PHASE RPM VOLTAGE MPD0508 5(2) 3 33.4 36.2 870 208 / 230 / 460 17.4 / 16.5 / 8.2 120 64 x 65 x 76 1288 MPD0758 7 .
Tank Drain moisture Daily 2500 Inlet air filter Replace ●▲ (Every 2,500 hrs or less) Blower fan Clean ●●●● Fan Duct Clean ●●●● Compressor Fins Clean ● (Every 2,500 hrs or less) B.
Medical Package System 6 Ref. No. Description Part Number Qty. 1 Compressor Pump: 5 HP OPT050 1 7-1/2 HP OPT100 1 10 HP OPT100 1 15 HP OPT150 1 2 Corrosion Resistant Vertical Tank 120 Gal. Consult factory for number 1 200 Gal. Consult factory for number 1 240 Gal.
Medical Package System 7 Ref. No. Description Part Number 1 Refrigerated Dryer: 20 CFM ACO200020 25 CFM ACO200025 35 CFM ACO300035 45 CFM ACO000045 55 CFM ACO200055 60 CFM ACO200060 70 CFM ACO200070 1.
Please read and save these instructions. Read carefully before attempting to assemble, install, operate or maintain the product described. Protect yourself and others by observing all safety information. Failure to comply with instructions could result in personal injury and/or prop- erty damage! Retain instructions for future reference.
2 Oilless Reciprocating Air Compr essor Pumps 4. Clearance must allow for safe, effec- tive inspection and maintenance. 5. If necessary, use metal shims or level- ing pads to level the compressor.
3 Oilless Reciprocating Air Compr essor Pumps GH Model No. HP A B C D E F (Diameter) (O.D. Flywheel) Dimensions (inches) OPS 010 1 - 1 1 / 2 6.7 3.2 11.7 16.2 7.5 4.2 .38 Ø11.2 Ax1 OPS 030 2 - 3 8.3 3.7 13.4 17.8 18.9 6.4 .43 Ø13.8 Bx1 OPT 050 5 11.
Oilless Reciprocating Air Compr essor Pumps Maintenance Schedule Operating Hours Item Action needed 500 2500 5000 10,000 15,000 20,000 Remarks Tank Drain moisture Daily Drain tank daily Inlet Air Filt.
5 Oilless Reciprocating Air Compr essor Pumps V alve Inspection/ Replacement (Inspect Every 5,000 Hours) INSPECT VALVE SET 1. Remove head bolts from cylinder head. Head bolts for some models are used also as cylinder bolts, which penetrates crankcase (See Figure 7).
Figure 10 6 Oilless Reciprocating Air Compr essor Pumps OPS010 OPT050, OPT150 OPS030 OPT100 Cylinder Bolt T orque 156 in. lb. 295 in. lb. OPS010, OPS030, OPT050, OPT100, OPT150 Head Bolt T orque 156 in. lb. Cylinder Inspection/Replacement 1. Remove cylinder head and valve set.
7 Oilless Reciprocating Air Compr essor Pumps 1. Remove cylinder head and cylinder. 2. Do not use tool to remove the ring. NOTE: Expanding ring too much and deforming can cause wear and leakage. (See Figure 12). Mark upper surface (not lower surface) of removed ring for easy distinction.
Pump No. of Piston Mark on Compressor No. Rings per Top of Model Bore Cylinders Piston Piston OPS010 65 mm 1 1 Flywheel side Fan side OPS030 65 mm 1 1 Flywheel side Fan side OPT050 LP 105 mm 1 1 Flywh.
9 Oilless Reciprocating Air Compr essor Pumps WRIST PIN BEARING SEAL REPLACEMENT (REPLACE EVER Y 10,000 HOURS) Replacing Oil Seal 1. Insert the plastic retention sleeve which protects the needle bearing from dropping out. 2. Remove the two oil seals by using a screwdriver.
10 Oilless Reciprocating Air Compr essor Pumps Inspect Connecting Rod, Wrist Pin and Crankshaft REPLACE CONNECTING ROD SET AND CRANKSHAFT AS A SET Pressing bearing into connecting rod and connecting rod set into crankshaft requires special fixtures. Without such fixtures, the squareness and paral- lelism of each part will be affected.
11 Oilless Reciprocating Air Compr essor Pumps Notes.
12 Oilless Reciprocating Air Compr essor Pumps Figure 24 - Replacement Parts Illustration for OPS010.
13 Oilless Reciprocating Air Compr essor Pumps Ref. Part Number for model No. Description OPS010 Qty 1 Cylinder 91000560 1 6 Cylinder head 91002630 1 16 Elbow (Unloader) 06800261 1 18 Unloader cap 010.
14 Oilless Reciprocating Air Compr essor Pumps Figure 25 - Replacement Parts Illustration for OPS030.
15 Oilless Reciprocating Air Compr essor Pumps Ref. Part Number for model No. Description OPS030 Qty 1 Cylinder 91000560 2 6 Cylinder head 91002640 2 16 Elbow (unloader) 06800261 1 17 Tee union assemb.
16 Oilless Reciprocating Air Compr essor Pumps Figure 26 - Replacement Parts Illustration for OPT050 197 197.
17 Oilless Reciprocating Air Compr essor Pumps Part Number Ref. for model No. Description OPT050 Qty 1 Cylinder (LP) 91000680 1 2 Cylinder (HP) 91947680 1 3 Liner 91177680 2 6 Cylinder Head (LP) 91002.
Oilless Reciprocating Air Compr essor Pumps 18 Figure 27 - Replacement Parts Illustration for OPT100 and OPT150 197 197 171 172.
1 Cylinder (LP) 91000670 91000680 2 2 Cylinder (HP) 91947680 91947680 1 6 Cylinder head (LP) 91002691 91002691 2 7 Cylinder head (HP) 91002680 91002680 1 14 Elbow (Intercooler) 01415690 01415690 4 17 .
Oilless Reciprocating Air Compr essor Pumps Powerex Limited W arranty POWEREX 3 YEAR / 10,000 HOUR EXTENDED PARTS LIMITED WARRANTY - Powerex warrants each Compressor Pump or Scroll Air-End against defects in material or workmanship from the date of purchase for a period of Three years or 10,000 hours , whichever may occur first.
Ein wichtiger Punkt beim Kauf des Geräts Powerex MPD1508 (oder sogar vor seinem Kauf) ist das durchlesen seiner Bedienungsanleitung. Dies sollten wir wegen ein paar einfacher Gründe machen:
Wenn Sie Powerex MPD1508 noch nicht gekauft haben, ist jetzt ein guter Moment, um sich mit den grundliegenden Daten des Produkts bekannt zu machen. Schauen Sie zuerst die ersten Seiten der Anleitung durch, die Sie oben finden. Dort finden Sie die wichtigsten technischen Daten für Powerex MPD1508 - auf diese Weise prüfen Sie, ob das Gerät Ihren Wünschen entspricht. Wenn Sie tiefer in die Benutzeranleitung von Powerex MPD1508 reinschauen, lernen Sie alle zugänglichen Produktfunktionen kennen, sowie erhalten Informationen über die Nutzung. Die Informationen, die Sie über Powerex MPD1508 erhalten, werden Ihnen bestimmt bei der Kaufentscheidung helfen.
Wenn Sie aber schon Powerex MPD1508 besitzen, und noch keine Gelegenheit dazu hatten, die Bedienungsanleitung zu lesen, sollten Sie es aufgrund der oben beschriebenen Gründe machen. Sie erfahren dann, ob Sie die zugänglichen Funktionen richtig genutzt haben, aber auch, ob Sie keine Fehler begangen haben, die den Nutzungszeitraum von Powerex MPD1508 verkürzen könnten.
Jedoch ist die eine der wichtigsten Rollen, die eine Bedienungsanleitung für den Nutzer spielt, die Hilfe bei der Lösung von Problemen mit Powerex MPD1508. Sie finden dort fast immer Troubleshooting, also die am häufigsten auftauchenden Störungen und Mängel bei Powerex MPD1508 gemeinsam mit Hinweisen bezüglich der Arten ihrer Lösung. Sogar wenn es Ihnen nicht gelingen sollte das Problem alleine zu bewältigen, die Anleitung zeigt Ihnen die weitere Vorgehensweise – den Kontakt zur Kundenberatung oder dem naheliegenden Service.