Page 1 of the Philips M1191AL user manual — B-Series Sensors M1191B M1191BL
Seite 1 · Zum Vergrößern klicken
Seite 1

Reusable SpO 2 Sensors

Instruction for Use

A-Series Sensors M1191A M1191AL M1192A M1193A M1194A M1195A

Philips M1191AL — Reusable SpO 2 Sensors diagram (page 1)
Abb. 1 · Reusable SpO 2 Sensors
Philips M1191AL — Reusable SpO 2 Sensors diagram (page 1)
Abb. 2 · Reusable SpO 2 Sensors

Special Edition A-Series Sensors M1191ANL M1192AN M1193AN M1194AN M1195AN

Philips M1191AL — Special Edition A-Series Sensors M1191ANL M1192AN M1193AN M1194AN M1195AN diagram (page 1)
Abb. 3 · Special Edition A-Series Sensors M1191ANL M1192AN M1…

B-Series Sensors M1191B M1191BL

philips

Page 2 of the Philips M1191AL user manual — Reusable SpO 2 Sensors
Seite 2 · Zum Vergrößern klicken
Seite 2
Page 2 of the Philips M1191AL user manual — Reusable SpO 2 Sensors
Page 3 of the Philips M1191AL user manual — Product Overview
Seite 3 · Zum Vergrößern klicken
Seite 3

Product Overview

Three series of Philips reusable SpO2 sensors offer monitoring capability for a broad range of patient types as shown in the table below. Reference Specifications later in this IFU for sensor accuracy.

  • A-Series Sensors : For use with any Philips/Agilent/HP instrument that lists them as an accessory. Refer to your instrument’s Instructions for Use for A-Series sensor compatibility information.
  • B-Series Sensors: For use with the same Philips/Agilent/HP instruments that list A-Series M1191A and M1191AL sensors as compatible accessories; i.e. M1191B can replace M1191A and M1191BL can replace M1191AL.
  • Special Edition A-Series Sensors (AN-Series): For use with Philips brand OxiMax® compatible instruments M1020B or M3001A, Option A02. Also for use with the same Philips/Agilent/HP instruments that list A-Series sensors as compatible accessories. Each Special Edition ASeries sensor can replace its similarly numbered A-series sensor; i.e. M1191ANL can replace M1191AL; M1192AN can replace M1192A; M1193AN can replace M1193A; M1194AN can replace M1194A; and M1195AN can replace M1195A.
A-Series SensorsB-Series Sensors2Special Edition A-Series SensorsAdultPediatricInfantNeonate FingerEarFingerEar3FingerToe3HandFoot M1191AM1191BN/A3 1M1191AL1M1191BL1M1191ANL3 M1192AN/AM1192AN3 M1193AN/AM1193AN33 M1194AN/AM1194AN33 M1195AN/AM1195AN33

1These sensors have a Longer 3m (versus standard 2m) attached cable. 2These sensors extend compatibility to Philips brand OxiMax® compatible instruments. 3Preferred application sites for M1193A and M1195A sensors.

Page 4 of the Philips M1191AL user manual — Definition of Product Symbols
Seite 4 · Zum Vergrößern klicken
Seite 4

Intended Use Philips reusable SpO2 sensors are for multi-patient use, when continuous non-invasive arterial oxygen saturation and pulse rate monitoring are required.

Accessories Supplied with Sensors • M1627A Adult Wristband: Supplied only with adult finger sensors to secure sensor cable to back of adult patient’s hand and wrist.

  • Instructions for Use : Describes sensor models, patient applications, Warnings against improper use, and specifications.

Accessories Available for Separate Purchase • M1940A Adapter Cable: 8-pin (round female) to 12-pin (round male) adapter cable. Adapts 8-pin sensor connector to 12-pin instrument jack. Extends sensor cable length by 2m. DO NOT USE with M1191AL, M1191ANL, M1191BL sensors.

  • M1941A Extension Cable : 8-pin (round female) to 8-pin (round male) extension cable. Extends sensor cable length by 2m. DO NOT USE with M1191AL, M1191ANL, M1191BL sensors.

Definition of Product Symbols

AdultPediatricAdult FingerInfant/Neo HandAdult/Pediatric Ear InfantNeonate (Neo)Pediatric/Infant FingerInfant ToeInfant/Neo Foot Caution, See DocumentationPrescription Use OnlyStorage TempContains no Latex
Page 5 of the Philips M1191AL user manual — Warnings
Seite 5 · Zum Vergrößern klicken
Seite 5

Warnings

  • These sensors are for use only with Philips/Agilent/HP instruments. Before using, verify the combination of instrument/sensor is specified in the instrument’s user documentation (e.g. Instructions for Use); otherwise patient injury may result.
  • Pulse oximetry measurements are statistically distributed. Two-thirds of all pulse oximetry measurements can be expected to fall within the stated accuracy (refer to Specifications later in this IFU for stated sensor accuracy).
  • Connect the sensor only to the SpO 2 connection or the SpO 2 adapter cable of the oximeter.
  • Do not reuse the sensor on a different patient until it has been disinfected. The sensor can be reused on the same patient for the patient’s entire stay.
  • At elevated ambient temperatures, patient skin could be severely burned after prolonged sensor application at sites that are not well perfused. To prevent this condition, be sure to check patient application sites frequently. All listed sensors operate without risk of exceeding 41 o C on the skin if the initial skin temperature does not exceed 35 o C.
  • Be sure to apply the sensor to the patient at the preferred or alternative application site, while following application instructions provided later in this document. Failure to do so can cause inaccurate measurements.
  • To avoid venous pulsation, obstructed circulation, pressure marks, pressure necrosis, artifacts and inaccurate measurements, make sure you are using a sensor of the correct size and that the sensor is not too tight. If the sensor is too tight, because the application site is too large or becomes too large due to edema, the excessive pressure applied may result in venous congestion distal from the application site, leading to interstitial edema and tissue ischemia.
  • If the sensor is applied too loose it may fall off or compromise proper alignment of sensor optics and result in inaccurate readings.
Page 6 of the Philips M1191AL user manual — Warnings
Seite 6 · Zum Vergrößern klicken
Seite 6
  • Where possible, the application site for the sensor should be an extremity free of arterial catheters, blood pressure cuffs, or intravascular infusion lines.
  • Avoid sites subject to excessive motion. Try to keep the patient still, or move the sensor to a site with less motion.
  • Dysfunctional hemoglobin or intravascular dyes can cause inaccurate measurements.
  • Make sure the application site for the sensor is not deeply pigmented, or deeply colored. For example, nail polish, artificial nails, dye or pigmented cream may cause inaccurate measurements. In any of these cases reposition the sensor or choose an alternative sensor for use at a different site.
  • Cover the sensor with opaque material under conditions of strong or excessive light (infrared lamps, OR lamps, photo therapy). Failure to do so can result in inaccurate measurements.
  • Protect the connector from contact with any liquid.
  • Inspect the sensor application site every 2 to 3 hours to ensure skin integrity, correct optical alignment, and circulation distal to the sensor site. Skin irritation or ulceration may occur if the sensor is attached to one location for too long. Move the sensor application site every 4 hours, or more often if circulation or skin integrity is compromised. If the light source is not directly opposite the light detector, reapply the sensor, or choose an alternative sensor for use at a different site.
  • Do not use a sensor during MRI scanning. This may cause burns or inaccurate measurements.
  • M1191AL/M1191ANL/M1191BL only: Do not use these sensors with an extension or adapter cable.
  • M1195A/M1195AN only: Infant patients are prone to movement; therefore, when applying a sensor to an infant, secure the sensor cable with tape to prevent it from coming off the infant’s finger. If necessary, position the sensor on a different finger, or an alternative site (toe).
Page 7 of the Philips M1191AL user manual — Change Application Site Periodically
Seite 7 · Zum Vergrößern klicken
Seite 7

Before Applying the Sensor

Be sure to read and understand all Warnings listed in your SpO2 monitoring instrument IFU, as well as all Warnings described in this sensor IFU. Use sensors only with validated Instruments and at a recommended patient application area.

Inspect Sensor for Damage • Inspect the sensor outside and inside. To check the inside, gently open the sensor cavity and check for splits on or next to the transparent silicone that covers the optical elements. Be sure there are no blisters on the silicone, and no liquid leaking from sensor optics.

  • Any sensor showing signs of damage or alteration must not be used for further patient monitoring; instead, dispose of it using proper disposal procedures (see below).

Sensor Disposal

Dispose Deteriorated Sensors Any sensor showing signs of physical or electrical deterioration or failure should be disinfected, decontaminated, and disposed of according to local laws regarding the disposal of hospital waste.

Change Application Site Periodically

Move the sensor application site every 4 hours, or more often if circulation or skin integrity is compromised.

Page 8 of the Philips M1191AL user manual — Applying Finger (or Toe) Sensors
Seite 8 · Zum Vergrößern klicken
Seite 8

Applying Finger (or Toe) Sensors

Philips M1191AL — Applying Finger (or Toe) Sensors diagram (page 8)
Abb. 4 · Applying Finger (or Toe) Sensors
  • Adult Finger Sensors Any finger except thumb on patients more than 50kg.
  • A-Series : M1191A or M1191AL
  • Special Edition (SE) A-Series : M1191ANL
  • B-Series: M1191B or M1191BL
  • Pediatric Finger Sensors Any finger except thumb for patients between 15kg and 50kg
  • A-Series : M1192A
  • Special Edition (SE) A-Series : M1192AN
  • Infant Finger (or Toe) Sensors Any finger or toe (except thumb) for patients between 4kg and 15kg. Finger or toe diameter should measure from 7-8mm (0.27-0.31”).
  • A-Series : M1195A, or
  • Special Edition (SE) A-Series : M1195AN
StepApplying the Finger (or Toe) Sensor 1Select appropriate sensor for patient size (defined above). 2Position sensor over patient’s finger. If applying an adult finger sensor, be sure the sensor cable is positioned ON TOP of the finger/hand as shown above. 3Patient fingertip should touch but not protrude from end of sensor. If necessary, trim the fingernail to correctly position the sensor. 4If applying an adult finger sensor, secure the cable to BACK OF HAND with M1627A wristband (supplied only with adult finger sensors.) 5Plug sensor into instrument (or into adapter cable if needed). 6Inspect and change the sensor application site periodically.

adult application · shown

Page 9 of the Philips M1191AL user manual — Applying Hand/Foot Sensors
Seite 9 · Zum Vergrößern klicken
Seite 9

Applying Hand/Foot Sensors

  • Neonatal hand or foot, for patients between 1kg and 4kg.
  • A-Series : M1193A
  • Special Edition (SE) A-Series : M1193AN
StepApplying the Hand/Foot Sensor 1Position the sensor over the hand or foot with optical components opposite each other. 2Holding sensor, stretch strap slightly (not more than 2.5cm (1.0”). 3Insert stretched strap into slot, hold it there while threading end through latch. If strap is too long, thread it through second latch. 4Plug sensor into instrument (or into adapter cable if needed). 5Inspect and change the sensor application site periodically.
Applying Ear-Clip Sensors Adult or pediatric ear, for patients more than 40kg. • A-Series: M1194A, or • Special Edition (SE) A-Series: M1194AN StepApplying the Ear-Clip Sensor 1To improve perfusion, massage or warm the ear lobe. 2Clip probe onto fleshy part of the lobe. Plastic fixing mechanism minimizes artifact generated by patient motion. Do not position probe on cartilage or pressed against the head. 3Plug sensor into instrument (or into adapter cable if needed). 4Inspect and change the sensor application site periodically.
Page 10 of the Philips M1191AL user manual — Validated Disinfectants
Seite 10 · Zum Vergrößern klicken
Seite 10

Cleaning and Low-Level Disinfection

The reusable sensors should be cleaned and disinfected, but never sterilized. Follow the procedure outlined below.

  • Warnings
  • Use only validated cleaning agents and disinfectants listed below; do not use any others. Failure to do so may damage the sensor or its connecting wires and shorten product lifetime or cause safety hazards.
  • Select disinfectants carefully as some have very similar names but completely different compositions.
  • Do not immerse the sensor connector in any of the cleaning solutions, disinfectant, or other liquid (only the sensor and cable housing may be immersed, not the connector).
  • Do not soak sensors in disinfectants longer than specified by the disinfectant manufacturer.
  • Do not sterilize the sensors.

Validated Cleaning Agents

- Mild Detergent - Salt Solution (1%)

Validated Disinfectants

Metricide® 28 Cidex® Formula 7 Kohrsolin® (2%)

Metricide® Plus 30 Cidex® OPA Mucapur®-CD (1%)

Terralin® Liquid Cidex® Plus Isopropanol (70%) or Isopropanol Wipe (70%) Incidin® Liquid Omnicide® 28

StepCleaning and Low-Level Disinfection 1Clean sensor following instructions supplied with cleaning agent. 2Disinfect sensor following instructions supplied with disinfectant. 3Rinse sensor in water, wiping it dry with a clean cloth, leaving to dry completely. If you notice any deterioration or damage to sensor or cable, dispose the sensor immediately.
Page 11 of the Philips M1191AL user manual — Specifications
Seite 11 · Zum Vergrößern klicken
Seite 11

Specifications

Sensor Accuracy A-Series Sensors: For A-Series sensor SpO2 and Pulse Rate accuracy specifications, reference your Philips/Agilent/HP instrument’s Instructions for Use.

B-Series Sensors: SpO2 and Pulse Rate accuracy specifications for M1191B and M1191BL sensors are the same as indicated for M1191A and M1191AL sensors in your Philips/Agilent/HP instrument’s Instructions for Use.

Special Edition A-Series Sensors (AN-Series): SpO2 accuracy (functional saturation) for Special Edition A-Series sensors used in combination with M1020B or M3001A, Option A02, specified as the root-mean-square (RMS) difference between the measured values and the reference values, in the range 70 to 100% is: 2% for M1191ANL and M1192AN; 3% for M1193AN, M1194AN, and M1195AN.

Neonatal Applications: The claimed accuracy numbers for neonatal application are already increased by an additional 1% to account for affect on oximeter measurements due to fetal hemoglobin in neonatal blood, as predicted by the literature.

Light Emitting Diode Wavelength Ranges Wavelength ranges for the light emitting diodes used in these sensors are within 600nm-1000nm, with an optical output power of less than 15mW. Knowing the wavelength range can be useful to clinicians performing photodynamic therapy.

Measurement Validation The SpO2 accuracy has been validated in human studies against arterial blood sample reference measured with a CO-oximeter. In a controlled desaturation study, healthy adult volunteers with saturation levels between 70% and 100% SaO2 were studied. The population characteristics for those studies were:

  • Approximately 50% female and 50% male ranging in age from 18-45
  • Skin tone: from light to black Because pulse oximeter equipment measurements are statistically distributed, only approximately 2/3 of pulse oximeter equipment measurements can be expected to fall within the + Arms value measured by a CO-oximeter. Functional testers, such as an SpO 2 simulator cannot be used to assess the accuracy of pulse oximeter Sensors.

Frage zu Philips M1191AL stellen

Besitzer und unser Team antworten in der Regel innerhalb eines Tages – und Ihre Frage hilft auch anderen weiter.

Add your email to get notified when it’s answered — no spam, ever.

0 questions
Noch keine Fragen zu Philips M1191AL

Stellen Sie die erste Frage — andere Nutzer oder unser Team antworten Ihnen.

Originaldokument

Philips M1191AL PDF-Anleitung

247 Seiten·2.2 MB·Balkensucher·PDF
Klicken Sie oben auf eine Seite, um sie hier zu öffnen
Seite 1 / …
100%
Seiten werden geladen…