CVI ™ , LabVIEW ™ , Natio nal Instru ments ™ ,n i . c o m ™ ,N I - D A Q ™ , and SCXI ™ are trademarks of National Instruments Corporation. Product a nd company names mentioned herein are tr ademarks or trade names of their respective companies.
SCXI-1125 Calibration Procedure 2 ni.com Equipment and Other T est Requirements This section describes the equipment, software, documentation, and test conditions required for calibrat ing the SCXI-1125.
© National Instruments Corpor ation 3 SCXI-1125 Calibration Procedur e Documentation The NI-DAQ Function Refe rence Online Help fi le and NI-DAQ User Manual for PC Compa tibles c o n t a i nd e t a i l e di n f o r m a t i o no nu s i n gt h e NI-DAQ driver.
SCXI-1125 Calibration Procedure 4 ni.com Setting Up Y our Module Complete the following steps to set up yo ur module for testing : 1. Install the SCXI-1125 in sl ot 1 of the SCXI chassis. 2. Connect a 68-to-68-pin cable between the S CXI module and the E Series D A Q de vice installed in the host computer .
© National Instruments Corpor ation 5 SCXI-1125 Calibration Procedur e 7. Connect the calibrator to the appro priate analog input, starting with channel 0. Refer to Figure 1 to determine which pins on the 96-pin front conn ector correspond to t he positiv e and negati ve inputs of the specified channel.
SCXI-1125 Calibration Procedure 6 ni.com 13. Call SCXI_Scale . Set the following parameters: • SCXIChassisID — The v alue assigned b y MAX • moduleSlot — 1 • channel — The same channel num.
© National Instruments Corpor ation 7 SCXI-1125 Calibration Procedur e • calOp — ND_SELF_CALIBRATE • s etOfCalConstants — ND_USER_EEPROM_AREA • calRefVolts — 0.
SCXI-1125 Calibration Procedure 8 ni.com • filterMode — 1 • cutoffFrequency — 10,000 9. Connect the calibrator to the analog input you w ant to adjust. Refer to Figure 1 to determine the pins on the 96-pin f ront connector that correspond to the positiv e and neg ati ve inputs of the specif ied channel.
© National Instruments Corpor ation 9 SCXI-1125 Calibration Procedur e 16. Record binary1 with volt1 and binary2 with volt2 for later use. 17. Repeat steps 7 through 16 for the remaining gain v alues you want to adjust on this channel. 18. Repeat steps 6 through 17 for the remaining chan nels you want to adjust.
SCXI-1125 Calibration Procedure 10 ni.com V erifying Adjusted V alues After completing the adjustment s to your SCXI-1125 module, it is important that you verify the analog input operation by repeating t he steps listed in the Veri fying Gain and Offset Errors section of this docu ment.
© National Instruments Corpor ation 11 SCXI-1125 Calibration Procedur e Front and Rear Signal Connector Diagrams Figure 1. Front Signal Connector Pin Assignments CHSGND +5 V RSVD CH7+ CH6+ CH5+ CH4+ .
SCXI-1125 Calibration Procedure 12 ni.com Figure 2. Rear Signal Connector Pin Assignments SERCLK DIGGND SLOT0SEL* DAQD*/A SERDATIN OUTREF MCH7+ MCH6+ MCH5+ MCH4+ MCH3+ MCH2+ MCH1+ MCH0+ AOGND SCANCLK .
© National Instruments Corpor ation 13 SCXI-1125 Calibration Procedur e Specifications The following table contains the sp ecifications you should use w hen verifying and calibrating y our SCXI-1125 module. If you r device has been calibrated within the last year, the test point value should fall between the Upper Limit and Lower Limit valu es.
SCXI-1125 Calibration Procedure 14 ni.com 200 – 0.025 – 0.024750 – 0.02468642 – 0 .02481358 250 0.02 0.019800 0.01985126 0.019 74874 250 0 0.000000 0.000022 – 0.000022 250 – 0.02 – 0.019800 – 0.01974874 – 0.0 1985126 500 0.01 0.009900 0.
Ein wichtiger Punkt beim Kauf des Geräts National Instruments SCXI -1125 (oder sogar vor seinem Kauf) ist das durchlesen seiner Bedienungsanleitung. Dies sollten wir wegen ein paar einfacher Gründe machen:
Wenn Sie National Instruments SCXI -1125 noch nicht gekauft haben, ist jetzt ein guter Moment, um sich mit den grundliegenden Daten des Produkts bekannt zu machen. Schauen Sie zuerst die ersten Seiten der Anleitung durch, die Sie oben finden. Dort finden Sie die wichtigsten technischen Daten für National Instruments SCXI -1125 - auf diese Weise prüfen Sie, ob das Gerät Ihren Wünschen entspricht. Wenn Sie tiefer in die Benutzeranleitung von National Instruments SCXI -1125 reinschauen, lernen Sie alle zugänglichen Produktfunktionen kennen, sowie erhalten Informationen über die Nutzung. Die Informationen, die Sie über National Instruments SCXI -1125 erhalten, werden Ihnen bestimmt bei der Kaufentscheidung helfen.
Wenn Sie aber schon National Instruments SCXI -1125 besitzen, und noch keine Gelegenheit dazu hatten, die Bedienungsanleitung zu lesen, sollten Sie es aufgrund der oben beschriebenen Gründe machen. Sie erfahren dann, ob Sie die zugänglichen Funktionen richtig genutzt haben, aber auch, ob Sie keine Fehler begangen haben, die den Nutzungszeitraum von National Instruments SCXI -1125 verkürzen könnten.
Jedoch ist die eine der wichtigsten Rollen, die eine Bedienungsanleitung für den Nutzer spielt, die Hilfe bei der Lösung von Problemen mit National Instruments SCXI -1125. Sie finden dort fast immer Troubleshooting, also die am häufigsten auftauchenden Störungen und Mängel bei National Instruments SCXI -1125 gemeinsam mit Hinweisen bezüglich der Arten ihrer Lösung. Sogar wenn es Ihnen nicht gelingen sollte das Problem alleine zu bewältigen, die Anleitung zeigt Ihnen die weitere Vorgehensweise – den Kontakt zur Kundenberatung oder dem naheliegenden Service.