METROLOGIC INSTRUMENTS, INC. HoloTrak ® IS8000 Series Holographic Scanners Installation and User’s Guide.
Copyright © 2006 by Metrologic Instruments, Inc. All rights reserved. No part of this work may be reproduced, transmitted, or stored in an y form or by any means without prior written consent, except by revie wer, who may quote brief passages in a review, or provided for in the Copyright Act of 1976.
ii T ABLE OF C ONTENTS Introducti on........................................................................................................... 1 Scanner and Accessori es ................................................................................
.
1 I NTRODUCTION Metrologic's IS8000 series scanners bring th e benefits of holography to long- range industrial bar code scanning. Enclosed in rug ged NEMA-12 cases for industrial environments, the IS8000 seri es provides omnidirection al scanning with optional no code/ wrong code warning.
2 S CANNER AND A CCESSORIES The following is a list of parts included in a st andard IS8000 kit: IS8000 Series HoloTrak ® Holographic Scanner Power Supply [ MLPN 46-46207-US] 12VDC @ 4.16 Amps, 220VAC or [ MLPN 46-46207-UK] 12VDC @ 4.
3 I NSTALLING THE S CANNER TO THE H OST (S TANDARD F EATURES ) Single HoloTrak ® Installation To avoid potential problems, do not po wer up the scanner until the communication cable is secur ed to the host. 1. Turn off the host system. 2. Connect the communications cable to the HoloTrak ® and to the host.
4 I NSTALLING THE S CANNER TO THE H OST (S TANDARD F EATURES ) Primary/Secondary setup c ontinued from previous page. All equipment must be connec ted before po wer-up. After both units complete the power-up diagnostics, the operator will need to program the primary unit with the MetroSet Program provided.
5 I NSTALLING THE S CANNER TO THE H OST (O PTIONAL F EATURES ) Industrial Controller MX001 Interface Option The MX001 Industrial Control Interface performs three functions. Through an object sensor inpu t, the scanner can be alerted to any package present in the scan field.
6 I NSTALLING THE S CANNER TO THE H OST (O PTIONAL F EATURES ) External Speaker Box Option When attached, beeper tones from the scanner are emitted f rom both the scanner speaker and externa l speaker to serve as an audible ind ication of the operation of the scanner.
7 I NSTALLING THE S CANNER TO THE H OST (O PTIONAL F EATURES ) Hand-held Port Option A Non-Decode Hand-Hel d Scanner can be connected to an IS8000 series fixed scanner. To avoid potential problems, do not po wer up the scanner until the communication cable is secur ed to the host.
8 C ONFIGURATION TO THE H OST S YSTEM The scanner is shipped from the factor y c onfigured to a set of default conditions noted in the Default Settings section of this guide. In order for the scanner to communicate with the host system, it wi ll need to be proper ly configured.
9 I NSTALLING AND R UNNING THE M ETRO S ET ® 2 P ROGRAM Installation 1. Close all open applications. 2. Utilizing a web bro wser application go to www.metrologic .com . Click on the Corporate Headquarters link t o go to the Metrologic home page. 3. Click on the Guides, Manuals, & Software link (see picture above).
10 I NSTALLING AND R UNNING THE M ETRO S ET ® 2 P ROGRAM 4. From the Download Center use the drop down menu to select the Metroset 2 Configuration & Utility soft ware. 5. A prompt will appear. Select save and choose a locati on for the Metroset 2 install file.
11 I NSTALLING AND R UNNING THE M ETRO S ET ® 2 P ROGRAM 6. A prompt will appear to acknowledge that the do wnload is complete. • Click on Run to begin installation. G o to step 8. • Click on Open Folder to open the current folder location of the MetroSet2Install.
12 I NSTALLING AND R UNNING THE M ETRO S ET ® 2 P ROGRAM 7. Locate the MetroSet2Install.exe file. Double click on the file to begi n the installation of Metroset 2. 8. The installation wizard will display the Welcome dialog box , select N ext to proceed to with the installation.
13 I NSTALLING AND R UNNING THE M ETRO S ET ® 2 P ROGRAM 9. To accept the default in stallation directory, select N ext . To change the destination folder, select Brow se to locate and choose the desired folder. 10. When the setup process is complete, select Yes , and choo se Finish to restart your computer.
14 I NSTALLING AND R UNNING THE M ETRO S ET ® 2 P ROGRAM How to Start To avoid potential problems, do not po wer up the scanner until the communication cable is secur ed to the host.
15 P ARTS OF THE S CANNER Green and Red LEDs The red LED is on when the unit is done po wer up, the VLD is emitting light and the unit is ready to scan.
16 M OUNTING B RACKET I NSTALLATION 1. Locate the mounting bolts on the sides of the scanner. 2. Align the bolts on the scanner with the channels on the mounting bracket. 3. Slide the unit into place until the scanner b olts sit securely in the bottom of the channels.
17 V ISUAL AND A UDIBLE I NDICATORS When the scanner is on, the acti vity of the red and gree n LEDs indicate the status of the scanner. Flashing Red, No Gre en During power-up and di agnostic mode. Steady Red, No Green Unit has completed power-up, the VLD is emitting light, and the unit is ready to scan.
18 L ABELS – A C D B.
19 D EPTH OF F IELD AND S CAN P ATTERN C HARACTERI STICS IS8300 Depth of Field and Scan Pattern The depth of field area defined is based conservatively on a 0.
20 D EPTH OF F IELD AND S CAN P ATTERN C HARACTERI STICS IS8400 Depth of Field and Scan Pattern The depth of field area defined is based conservatively on a 0.
21 D EPTH OF F IELD AND S CAN P ATTERN C HARACTERI STICS IS8500 Depth of Field and Scan Pattern The depth of field area defined is based conservatively on a 0.
22 D EPTH OF F IELD AND S CAN P ATTERN C HARACTERI STICS IS8800 Depth of Field and Scan Pattern The depth of field area defined is based conservatively on a 0.
23 M AINTENANCE Smudges and dirt can interfere with the pro per scanning of a bar code. Therefore, the output windo w will need occasional cleaning. 1. Spray optical qualit y cleaning fluid onto li nt free, nonabrasive cleanin g cloth. 2. Gently wipe the output window.
24 T ROUBLESHOOTING G UIDE P ROBLEM P OSSIBLE C AUS E ( S ) A CTION N EEDED No LEDs No Power No Scan Pattern No Power Check the power supply and outlet. Alternating Red and Green LEDs flash with a razz tone. Motor Failure Contact a Metrologic service representative.
25 A PPENDIX A Design Specifications IS8800 IS85 00 IS8400 IS8300 Optical Number of VLD 5 5 3 5 Scan pattern Omni Omni Omni Omni Scan lines per second 5,600 5,600 3,360 5,250 Focal planes 4 4 4 5 Scan l ines pe r focal plane 20 20 12 15 Total scan lines 80 80 48 75 Minimum bar width 0.
26 A PPENDIX A ( CONTINUED ) Design Specifications IS8800 IS85 00 IS8400 IS8300 Electrical Input Voltage 12 VDC 12 VDC 12 VDC 12 VDC Starting Pow er 69 watts 69 watt s 53 watts 53 watts Operating powe.
27 A PPENDIX B Default Settings Many functions of the scanner can be co nfigu red, or enabled/disabled. T he factory programs the scanner to a set of default parameter s. These defaults are marked with an asterisk (*) in the default column on the foll owing pages.
28 A PPENDIX B ( CONTINUED ) P ARAM ET E R D EFAULT RS-232 RS-422 L IGHT P EN 200 msec to find Supps Code 128 Coupon Option Code 128 Coupon Conv ersion 378/379 lock.
29 A PPENDIX B ( CONTINUED ) P ARAM ET E R D EFAULT RS-232 RS-422 L IGHT P EN Fire triac if good read/match Fire triac if bad match Fire triac if no read Triac Duration (sec.) Delay Time before firing (sec.
30 A PPENDIX B ( CONTINUED ) P ARAM ET E R D EFAULT RS-232 RS-422 L IGHT P EN Transmit SANYO ID Chr SNI Beetle Mode French PC Term not currently supported Xon/Xoff handshaki ng.
31 A PPENDIX B ( CONTINUED ) P ARAM ET E R D EFAULT RS-232 RS-422 L IGHT P EN Transmit Code 39 Start/S top Characters Transmit ITF Mod 10 check digi t Transmit Code 39 Mod 43 che ck di.
32 A PPENDIX C IS8000 Series Pinout Connections D ESCRIPTION P IN S IGNAL 1 Ground 2 RS-232 Receive Input 3 RS-232 Transmit Output 4 CTS Input 5 RTS Output 6 Reserved 7 Ground 8 Reserved 9 RS-422 R ec.
33 A PPENDIX C IS8000 Series Pinout Connections D ESCRIPTION P IN S IGNAL 1 12VDC Input Power to Scanner 2 12VDC Input Power to Scanner 3 Earth Ground 5 Power Ground 4 Power Ground Connector Type on S.
34 A PPENDIX D Limited Warranty The IS8000 series scanners are manufactured by Metrologic at its Blackwood, New Jerse y, U.S.A. facility. The IS8000 series scanners hav e a two (2) year limited warranty from the date of manufacture.
35 A PPENDIX E - R EGULATORY C OMPLIANCE EMC Emissions FCC Part 15, ICES-003, CISPR 22, EN 55022 Immunity CISPR 24, EN 55024 Changes or modifications not expressly approved by the party respo nsible for compliance could void the user’s authority to operate the equipment.
36 A PPENDIX F For patent information, please refer to www.honeywellaidc.com/patents ..
37 I NDEX A ac input/outlet ............. 3-7, 9, 21, 28 accessori es ................................... 2 approvals ......... 4, 13, 20, 21, 30, 31 assignments pin ................................ 10, 27, 28 audible ............................... 6, 10, 12 B bar code .
34 I NDEX specificati ons ......................... 20, 21 system interfaces .......... 1, 8, 20, 22 T tones ............................................. 6 transform ers .................................. 9 troubleshoot ing ..........................
.
December 2006.
Ein wichtiger Punkt beim Kauf des Geräts Metrologic Instruments IS8000 (oder sogar vor seinem Kauf) ist das durchlesen seiner Bedienungsanleitung. Dies sollten wir wegen ein paar einfacher Gründe machen:
Wenn Sie Metrologic Instruments IS8000 noch nicht gekauft haben, ist jetzt ein guter Moment, um sich mit den grundliegenden Daten des Produkts bekannt zu machen. Schauen Sie zuerst die ersten Seiten der Anleitung durch, die Sie oben finden. Dort finden Sie die wichtigsten technischen Daten für Metrologic Instruments IS8000 - auf diese Weise prüfen Sie, ob das Gerät Ihren Wünschen entspricht. Wenn Sie tiefer in die Benutzeranleitung von Metrologic Instruments IS8000 reinschauen, lernen Sie alle zugänglichen Produktfunktionen kennen, sowie erhalten Informationen über die Nutzung. Die Informationen, die Sie über Metrologic Instruments IS8000 erhalten, werden Ihnen bestimmt bei der Kaufentscheidung helfen.
Wenn Sie aber schon Metrologic Instruments IS8000 besitzen, und noch keine Gelegenheit dazu hatten, die Bedienungsanleitung zu lesen, sollten Sie es aufgrund der oben beschriebenen Gründe machen. Sie erfahren dann, ob Sie die zugänglichen Funktionen richtig genutzt haben, aber auch, ob Sie keine Fehler begangen haben, die den Nutzungszeitraum von Metrologic Instruments IS8000 verkürzen könnten.
Jedoch ist die eine der wichtigsten Rollen, die eine Bedienungsanleitung für den Nutzer spielt, die Hilfe bei der Lösung von Problemen mit Metrologic Instruments IS8000. Sie finden dort fast immer Troubleshooting, also die am häufigsten auftauchenden Störungen und Mängel bei Metrologic Instruments IS8000 gemeinsam mit Hinweisen bezüglich der Arten ihrer Lösung. Sogar wenn es Ihnen nicht gelingen sollte das Problem alleine zu bewältigen, die Anleitung zeigt Ihnen die weitere Vorgehensweise – den Kontakt zur Kundenberatung oder dem naheliegenden Service.