Benutzeranleitung / Produktwartung 760004-A des Produzenten Emerson Process Management
Zur Seite of 118
Instruction Manual 760004-A February 2002 http://www.processanalytic.com Model NGA 2000 HFID Heated Flame Ionization Detector Module.
Emerson Process Management Rosemount Analytical Inc. Process Analytic Division 1201 N. Main St. Orrville, OH 44667-0901 T (330) 682-9010 F (330) 684-4434 e-mail: gas.
Instruction Manual 760004-A February 2002 Rosemount Analytical Inc. A Division of Emerson Process Management Contents i Model NGA 2000 HFID TABLE OF CONTENTS PREFACE......................................................................................
Instruction Manual 760004-A February 2002 ii Contents Rosemount Anal ytical Inc. A Division of Emerson Process Management Model NGA 2000 HFID 4-4 Burner Internal Components ..............................................................................
Instruction Manual 760004-A February 2002 Rosemount Analytical Inc. A Division of Emerson Process Management Contents iii Model NGA 2000 HFID LIST OF ILLUSTRATIONS Figure 1-1. Flame Ionization Detection Technol ogy ....................................
Instruction Manual 760004-A February 2002 iv Contents Rosemount Anal ytical Inc. A Division of Emerson Process Management Model NGA 2000 HFID.
Instruction Manual 760004-A February 2002 Rosemount Analytical Inc. A Division of Emerson Process Management Preface P-1 Model NGA 2000 HFID PREFACE The purpose of this manual is to provide information con cerning the components, functions, installation and maintenance of the NGA 20 00 HFID and the System Accessories of the NGA 2000 System.
Instruction Manual 760004-A February 2002 P-2 Preface Rosemou nt Analytical Inc. A Division of Emerson Process Management Model NGA 2000 HFID SAFETY SUMMARY If this equipment is used in a manner not specifi ed in the s e instructions, protective system s may be impaired.
Instruction Manual 760004-A February 2002 Rosemount Analytical Inc. A Division of Emerson Process Management Preface P-3 Model NGA 2000 HFID WARNING . HIGH TEMPERATURE This equipment is used in the analy s is of sample gases at temperatures o f up to 250 C.
Instruction Manual 760004-A February 2002 P-4 Preface Rosemou nt Analytical Inc. A Division of Emerson Process Management Model NGA 2000 HFID WARNING . POSSIBLE EXPLOSION HAZARD Ensure that all gas connections are made as lab eled and are leak free. Improper gas connections could result in explosion or death.
Instruction Manual 760004-A February 2002 Rosemount Analytical Inc. A Division of Emerson Process Management Preface P-5 Model NGA 2000 HFID GENERAL PRECAUTIONS FOR HANDLING AND STORING HIGH PRESSURE .
Instruction Manual 760004-A February 2002 P-6 Preface Rosemou nt Analytical Inc. A Division of Emerson Process Management Model NGA 2000 HFID NAMUR 97-C219 N 9 6 DOCUMENTATION The following NGA 2000 HFID instruction material s are av ailable. Contact Customer Service Center or the local representative to order.
Instruction Manual 760004-A February 2002 Rosemount Analytical Inc. A Division of Emerson Process Management Preface P-7 Model NGA 2000 HFID GLOSSARY OF TERMS Analyzer Module The module that contains all sensor/detecto r compon ents for development of a Primary Variable sign al; in- cludes all signal conditioning and temperature control circuitry.
Instruction Manual 760004-A February 2002 P-8 Preface Rosemou nt Analytical Inc. A Division of Emerson Process Management Model NGA 2000 HFID Softkey s The five function softkeys located below the front panel di splay; they assum e the function displayed directly above each on the display, a function dictated by software.
Instruction Manual 760004-A February 2002 Rosemount Analytical Inc. A Division of Emerson Process Management Description and Spe cifications 1-1 Model NGA 2000 HFID SECTION 1 DESCRIPTION AND SPECIFICA.
Instruction Manual 760004-A February 2002 1-2 Description and Specifications Rosemount Anal ytical Inc. A Division of Emerson Process Management Model NGA 2000 HFID concentration of hydrocarbons in the sample. This measure of concentration i s placed on the network, where it can be shown on a data acquisition device.
Instruction Manual 760004-A February 2002 Rosemount Analytical Inc. A Division of Emerson Process Management Description and Spe cifications 1-3 Model NGA 2000 HFID Detect or (Burner) Purge Air Outlet.
Instruction Manual 760004-A February 2002 1-4 Description and Specifications Rosemount Anal ytical Inc. A Division of Emerson Process Management Model NGA 2000 HFID 1-5 SPECIFICATIONS a. General Measurement Species .................... Total hydrocarbons Ranges (H2/He Fuel) Low range .
Instruction Manual 760004-A February 2002 Rosemount Analytical Inc. A Division of Emerson Process Management Description and Spe cifications 1-5 Model NGA 2000 HFID c. Gas Requirements Sample .......................................... Non-flammable, below 100% of LEL Flow rate .
Instruction Manual 760004-A February 2002 1-6 Description and Specifications Rosemount Anal ytical Inc. A Division of Emerson Process Management Model NGA 2000 HFID d. Gas Connections Sample In ...................................... 1/4" O.D. tube fitting, stainless steel Regulated Air In .
Instruction Manual 760004-A February 2002 Rosemount Analytical Inc. A Division of Emerson Process Management Description and Spe cifications 2-1 Model NGA 2000 HFID SECTION 2 INSTALLATION 2-1 UNPACKING If the HFID Analyzer Module is received as a separate unit, carefully examine the shipping carton and contents for sign s of damage.
Instruction Manual 760004-A February 2002 2-2 Description and Specifications Rosemount Anal ytical Inc. A Division of Emerson Process Management Model NGA 2000 HFID graph-grade tubing.
Instruction Manual 760004-A February 2002 Rosemount Analytical Inc. A Division of Emerson Process Management Description and Spe cifications 2-3 Model NGA 2000 HFID b. Pneumatic Connections WARNING HIGH TEMPERATURES The Sample In, Sample Bypass Out, and Burner Exhaust Out gases and fittings can reach temperatures of up to 250 ° C.
Instruction Manual 760004-A February 2002 2-4 Description and Specifications Rosemount Anal ytical Inc. A Division of Emerson Process Management Model NGA 2000 HFID introduced through the sampl e inlet at the rear of the module. Zero Gas - Analysis is affected by the background gas of the sample.
Instruction Manual 760004-A February 2002 Rosemount Analytical Inc. A Division of Emerson Process Management Description and Spe cifications 2-5 Model NGA 2000 HFID DIFFUSE R BULKHEAD REDUCER 1/4T - 1.
Instruction Manual 760004-A February 2002 2-6 Description and Specifications Rosemount Anal ytical Inc. A Division of Emerson Process Management Model NGA 2000 HFID 2-6 ELECTRICAL CONNECTIONS Two electrical conne ctions are required on the Analyzer Module: POWER and NETW ORK (See Figure 2-3 and Figure 2-4).
Instruction Manual 760004-A February 2002 Rosemount Analytical Inc. A Division of Emerson Process Management Description and Spe cifications 2-7 Model NGA 2000 HFID 6.0 [152] .7 [18] 1.9 [48] 6.8 [174] 8.2 [208] .6 [15] 2.8 [71] 4.3 [ 110 ] 1.1 [28] 2.
Instruction Manual 760004-A February 2002 2-8 Description and Specifications Rosemount Anal ytical Inc. A Division of Emerson Process Management Model NGA 2000 HFID HARNESS , GLOW PLUG J3 P3 J4 P4 J5 .
Instruction Manual 760004-A February 2002 Rosemount Analytical Inc. A Division of Emerson Process Management Description and Spe cifications 2-9 Model NGA 2000 HFID 2-7 INSTALLATION CONSIDERATIO NS CH.
Instruction Manual 760004-A February 2002 2-10 Description and Specifications Rosemount Anal ytical Inc. A Division of Emerson Process Management Model NGA 2000 HFID.
Instruction Manual 760004-A February 2002 Rosemount Analytical Inc. A Division of Emerson Process Management Operation 3-1 Model NGA 2000 HFID SECTION 3 OPERATION 3-1 OVERVIEW Prior to initial startup, the user should leak test the module as outlined in Section 2.
Instruction Manual 760004-A February 2002 3-2 Operation Rosemount Analytical Inc. A Divisi on of Emerson Process Management Model NGA 2000 HFID • Power supply test - Verifies that all internal DC voltages are within the required tolerances. • Network test - Checks the inte rnal network interface.
Instruction Manual 760004-A February 2002 Rosemount Analytical Inc. A Division of Emerson Process Management Operation 3-3 Model NGA 2000 HFID and the Purge Gas variable does not read "ON," contact factory.
Instruction Manual 760004-A February 2002 3-4 Operation Rosemount Analytical Inc. A Divisi on of Emerson Process Management Model NGA 2000 HFID 3-4 CALIBRATION Calibration gas setup is as follo ws: 1. Set oven temperature setpoi nt. 2. Apply regulated air at a pressure b etween 10 and 25 psig.
Instruction Manual 760004-A February 2002 Rosemount Analytical Inc. A Division of Emerson Process Management Operation 3-5 Model NGA 2000 HFID 1. Verify that correct zero and span calibra- tion gases are being used properly.
Instruction Manual 760004-A February 2002 3-6 Operation Rosemount Analytical Inc. A Divisi on of Emerson Process Management Model NGA 2000 HFID than 0.5 inches of water. (Monitor display message, Purge Gas (ON) in Mi scellane- ous Control Parameters menu for correct state.
Instruction Manual 760004-A February 2002 Rosemount Analytical Inc. A Division of Emerson Process Management Operation 3-7 Model NGA 2000 HFID Figure 3-1. Typical Curves of Module Respon se vs. Pressure Setting on Fuel Pressure Regul ator Figure 3-2. Typical Curves of Module Respon se vs.
Instruction Manual 760004-A February 2002 3-8 Operation Rosemount Analytical Inc. A Divisi on of Emerson Process Management Model NGA 2000 HFID 614 2 35 Torque Se quence: Screw #1, 4 to 5 turns Screw .
Instruction Manual 760004-A February 2002 Rosemount Analytical Inc. A Division of Emerson Pr ocess Management Maintenance and Service 4-1 Model NGA 2000 HFID Cover Insulation Shield Securing Screw (4).
Instruction Manual 760004-A February 2002 4-2 Maintenance and Service Rosemount Anal ytical Inc. A Division of Emerson Process Management Model NGA 2000 HFID Oven Assembly 659551 Fan Assembly 657414 R.
Instruction Manual 760004-A February 2002 Rosemount Analytical Inc. A Division of Emerson Pr ocess Management Maintenance and Service 4-3 Model NGA 2000 HFID Oven Assembly 659551 4-2 OVEN Though the oven can be replaced as a com- plete unit, all internal components are field re- placeable.
Instruction Manual 760004-A February 2002 4-4 Maintenance and Service Rosemount Anal ytical Inc. A Division of Emerson Process Management Model NGA 2000 HFID b. Disassembly 1. Refer to Figure 4-4A. Remove the four retaining screws on the oven cover, remove cover.
Instruction Manual 760004-A February 2002 Rosemount Analytical Inc. A Division of Emerson Pr ocess Management Maintenance and Service 4-5 Model NGA 2000 HFID Figure 4-4. Oven Assembly A B C Inner Oven Front Panel 657352 Preamp connector nut, washer (typ 2) Exhaust Oven Cover 657105 (with insulation 657346) Nuts and Fer- rules (2 ea.
Instruction Manual 760004-A February 2002 4-6 Maintenance and Service Rosemount Anal ytical Inc. A Division of Emerson Process Management Model NGA 2000 HFID 4-3 BURNER This section covers burner com ponents which can be replaced without rem oval of oven from the chassis.
Instruction Manual 760004-A February 2002 Rosemount Analytical Inc. A Division of Emerson Pr ocess Management Maintenance and Service 4-7 Model NGA 2000 HFID Set Screw M3X0.
Instruction Manual 760004-A February 2002 4-8 Maintenance and Service Rosemount Anal ytical Inc. A Division of Emerson Process Management Model NGA 2000 HFID 4-4 BURNER INTERNAL COMPO N ENTS WARNING BURNER CONTAMIN ATION Do not handle internal parts of the burn er with bare hands.
Instruction Manual 760004-A February 2002 Rosemount Analytical Inc. A Division of Emerson Pr ocess Management Maintenance and Service 4-9 Model NGA 2000 HFID Burner Cap Burner Cap Retainer Gasket 656931 Burner Base Burner Base Combustion Chamber A B O-Ring 904373 b.
Instruction Manual 760004-A February 2002 4-10 Maintenance and Service Rosemount Analytical Inc. A Division of Emerson Process Management Model NGA 2000 HFID Air Baffle (Ref) 102260 See Below Sample Jet 657005 Gasket 102256 Jet Assembly 657012 Gasket 102273 Nut, Jet 657016 Air Baffle ≈ .
Instruction Manual 760004-A February 2002 Rosemount Analytical Inc. A Division of Emerson Pr ocess Management Maintenance and Service 4-11 Model NGA 2000 HFID c. Burner Jet Installation WARNING BURNER CONTAMIN ATION Do not handle internal parts of the burn er with bare hands.
Instruction Manual 760004-A February 2002 4-12 Maintenance and Service Rosemount Analytical Inc. A Division of Emerson Process Management Model NGA 2000 HFID Sample RTD 657061 Cartridge Heater 659643 Thermostat 657065 4-5 THERMAL BLOCK The sample RTD can be repl aced with the thermal block attached to burner and mounte d in oven.
Instruction Manual 760004-A February 2002 Rosemount Analytical Inc. A Division of Emerson Pr ocess Management Maintenance and Service 4-13 Model NGA 2000 HFID b. Cartridge Heater 1. Remove oven from analyzer mod ule per Section 4-2a. 2. Remove burner/thermal bl ock from oven per Section 4-2b.
Instruction Manual 760004-A February 2002 4-14 Maintenance and Service Rosemount Analytical Inc. A Division of Emerson Process Management Model NGA 2000 HFID 4-6 ELECTRONICS ASSEMBLY The electronics a.
Instruction Manual 760004-A February 2002 Rosemount Analytical Inc. A Division of Emerson Pr ocess Management Maintenance and Service 4-15 Model NGA 2000 HFID Figure 4-12.
Instruction Manual 760004-A February 2002 4-16 Maintenance and Service Rosemount Analytical Inc. A Division of Emerson Process Management Model NGA 2000 HFID b. Case Temperature Sensor 1. Per Section 4-6, remove securin g hardware from electronics asse mbly and lay on bench.
Instruction Manual 760004-A February 2002 Rosemount Analytical Inc. A Division of Emerson Pr ocess Management Maintenance and Service 4-17 Model NGA 2000 HFID c. Case Pressure Purge S w itch 1. Per Section 4-6, remove securin g hardware from electronics asse mbly and lay on bench.
Instruction Manual 760004-A February 2002 4-18 Maintenance and Service Rosemount Analytical Inc. A Division of Emerson Process Management Model NGA 2000 HFID d. Preamp Assembly 1. Per Section 4-6, remove securin g hardware from electronics asse mbly and lay on bench.
Instruction Manual 760004-A February 2002 Rosemount Analytical Inc. A Division of Emerson Pr ocess Management Maintenance and Service 4-19 Model NGA 2000 HFID 4-7 FAN ASSEMBLY 1. Disconnect and n ote location of cables. 2. Remove the two hex nuts secu ring the fan to the chassis, lift fan assembly out.
Instruction Manual 760004-A February 2002 4-20 Maintenance and Service Rosemount Analytical Inc. A Division of Emerson Process Management Model NGA 2000 HFID 4-8 FLOW CONTROLLER 1. Disconnect the all tubing and wiring con- nectors, note locations. 2. Remove the four hex nuts se curing the flow controller assembly to the analyzer module chassis.
Instruction Manual 760004-A February 2002 Rosemount Analytical Inc. A Division of Emerson Pr ocess Management Maintenance and Service 4-21 Model NGA 2000 HFID Figure 4-18. Flow Controller Assembly Remove and discard bracket supplied with regulator, assembly as shown.
Instruction Manual 760004-A February 2002 4-22 Maintenance and Service Rosemount Analytical Inc. A Division of Emerson Process Management Model NGA 2000 HFID 4-9 DC POWER SUPPLY MODULE Disconnect and note location of all wiri ng to DC power supply module.
Instruction Manual 760004-A February 2002 Rosemount Analytical Inc. A Division of Emerson Pr ocess Management Maintenance and Service 4-23 Model NGA 2000 HFID LED Indicator Assembly 657029 Purge Air F.
Instruction Manual 760004-A February 2002 4-24 Maintenance and Service Rosemount Analytical Inc. A Division of Emerson Process Management Model NGA 2000 HFID Replacing Front Panel Components 1. To access compon ents, remove the four front panel mounting screws (two on front, one on each side).
Instruction Manual 760004-A February 2002 Rosemount Analytical Inc. A Division of Emerson Pr ocess Management Maintenance and Service 4-25 Model NGA 2000 HFID a. LON/Power Module 1. Disconnect wiri ng connectors, note locations. 2. Refer to Figure 4-20.
Instruction Manual 760004-A February 2002 4-26 Maintenance and Service Rosemount Analytical Inc. A Division of Emerson Process Management Model NGA 2000 HFID 5. Inspect o-ring for damage, repl ace if necessary. Install o-ring onto regu- lator threaded shaft.
Instruction Manual 760004-A February 2002 Rosemount Analytical Inc. A Division of Emerson Pr ocess Management Maintenance and Service 4-27 Model NGA 2000 HFID 4-11 REAR PANEL COMPONENTS The following .
Instruction Manual 760004-A February 2002 4-28 Maintenance and Service Rosemount Analytical Inc. A Division of Emerson Process Management Model NGA 2000 HFID a. Fuel In 2-Way Solenoid Valv e 1. Disconnect wiri ng solenoid valve wir- ing connector, note location.
Instruction Manual 760004-A February 2002 Rosemount Analytical Inc. A Division of Emers on Process Management Troubleshooting 5-1 Model NGA 2000 HFID SECTION 5 TROUBLESHOOTING 5-1 TROUBLESHOOTING CHECK LIST a. Safety System 1. Verify purge supply pressure at bulk- head is between 10 and 20 psig.
Instruction Manual 760004-A February 2002 5-2 Troubleshooting Rosemount Analytical Inc. A Division of Emerson Process Mana gement Model NGA 2000 HFID 18. Verify that there is +10VDC to the reference thermistor. c. Drift 1. Verify that the sample, burner air, and fuel supply pressures are constant.
Instruction Manual 760004-A February 2002 Rosemount Analytical Inc. A Division of Emerson Process Management Replacement Parts 6-1 Model NGA 2000 HFID SECTION 6 REPLACEMENT PARTS WARNING PARTS INTEGRITY Tampering with or unauthorized substitution of components may ad versely affect safety of this product.
Instruction Manual 760004-A February 2002 6-2 Replacement Parts Rosemount A nalytical Inc. A Divisi on of Emerson Process Management Model NGA 2000 HFID 6-2 GENERAL 813344 Fuse, 6A 903107 Fuse, Therma.
Instruction Manual 760004-A February 2002 Rosemount Analytical Inc. A Division of Emerson Pr ocess Management Replacement Parts 6-3 Model NGA 2000 HFID 6-4 OVEN COMPONENTS 659551 Oven Assembly 657359 Burner Assembly 657205 Igniter Assembly 903736 O-Ring (Igniter Assembly) 657063 RTD Detector 903125 Set Screw M3X0.
Instruction Manual 760004-A February 2002 6-4 Replacement Parts Rosemount A nalytical Inc. A Divisi on of Emerson Process Management Model NGA 2000 HFID.
Instruction Manual 760004-A February 2002 Rosemount Analytical Inc. A Division of Emerson Proc ess Management Return of Material 7-1 Model NGA 2000 HFID SECTION 7 RETURN OF MATERIAL 7-1 RETURN OF MATERIAL If factory repair of defective equipment is re- quired, proceed as follows: 1.
Instruction Manual 760004-A February 2002 7-2 Return of Material Rosemount A nalytical Inc. A Division of Emerson Process Management Model NGA 2000 HFID.
Instruction Manual 760004-A February 2002 Rosemount Analytical Inc. A Division of Emerson Pr ocess Management Appendix A – Menu Displays 8-1 Model NGA 2000 HFID SECTION 8 APPENDIX A - MENU DISPLAYS .
Instruction Manual 760004-A February 2002 8-2 Appendix A - Menu Displays Rosemount Analytical Inc. A Division of Emerson Proc ess Management Model NGA 2000 HFID MENU: 3 FLOCHEKI1 MENU: 4 ACALSET MENU:.
Instruction Manual 760004-A February 2002 Rosemount Analytical Inc. A Division of Emerson Pr ocess Management Appendix A – Menu Displays 8-3 Model NGA 2000 HFID MENU: 6 ANALSETI1 MENU: 7 CALLIST MEN.
Instruction Manual 760004-A February 2002 8-4 Appendix A - Menu Displays Rosemount Analytical Inc. A Division of Emerson Proc ess Management Model NGA 2000 HFID MENU: 9 ACALSETI1 MENU: 10 APARLSTI1 ME.
Instruction Manual 760004-A February 2002 Rosemount Analytical Inc. A Division of Emerson Pr ocess Management Appendix A – Menu Displays 8-5 Model NGA 2000 HFID MENU: 12 AMMANI1 MENU: 13 AMSVC MENU:.
Instruction Manual 760004-A February 2002 8-6 Appendix A - Menu Displays Rosemount Analytical Inc. A Division of Emerson Proc ess Management Model NGA 2000 HFID MENU: 15 ADIAG MENU: 16 AMPWR MENU: 17 .
Instruction Manual 760004-A February 2002 Rosemount Analytical Inc. A Division of Emerson Pr ocess Management Appendix A – Menu Displays 8-7 Model NGA 2000 HFID MENU: 18 AMTEMP MENU: 19 AMMISC MENU:.
Instruction Manual 760004-A February 2002 8-8 Appendix A - Menu Displays Rosemount Analytical Inc. A Division of Emerson Proc ess Management Model NGA 2000 HFID MENU: 21 ADIAGI1 MENU: 22 RANGESETAM ME.
Instruction Manual 760004-A February 2002 Rosemount Analytical Inc. A Division of Emerson Pr ocess Management Appendix A – Menu Displays 8-9 Model NGA 2000 HFID MENU: 24 LINRANGE1 MENU: 25 LINRANGE2.
Instruction Manual 760004-A February 2002 8-10 Appendix A - Menu Displays Rosemount Anal ytical Inc. A Division of Emerson Pro cess Management Model NGA 2000 HFID MENU: 27 LINRANGE4 MENU: 28 LINRANGE0.
Instruction Manual 760004-A February 2002 Rosemount Analytical Inc. A Division of Emerson Pr ocess Management Appendix A – Menu Displays 8-11 Model NGA 2000 HFID MENU: 30 FLOCHEK1I1 MENU: 31 FILTER .
Instruction Manual 760004-A February 2002 8-12 Appendix A - Menu Displays Rosemount Anal ytical Inc. A Division of Emerson Pro cess Management Model NGA 2000 HFID MENU: 33 AMTEMPI1 MENU: 34 AM2VA MENU.
Instruction Manual 760004-A February 2002 Rosemount Analytical Inc. A Division of Emerson Pr ocess Management Appendix A – Menu Displays 8-13 Model NGA 2000 HFID MENU: 36 TLIMITSA MENU: 37 AMMISCI1 .
Instruction Manual 760004-A February 2002 8-14 Appendix A - Menu Displays Rosemount Anal ytical Inc. A Division of Emerson Pro cess Management Model NGA 2000 HFID MENU: 39 FILTERI1 MENU: 40 LINRANGE0I.
Instruction Manual 760004-A February 2002 Rosemount Analytical Inc. A Division of Emerson Pr ocess Management Appendix A – Menu Displays 8-15 Model NGA 2000 HFID MENU: 42 CALFACTORS MENU: 43 R1FACTO.
Instruction Manual 760004-A February 2002 8-16 Appendix A - Menu Displays Rosemount Anal ytical Inc. A Division of Emerson Pro cess Management Model NGA 2000 HFID MENU: 45 RN3FACTORS MENU: 46 RN4FACTO.
Instruction Manual 760004-A February 2002 Rosemount Analytical Inc. A Division of Emerson Pr ocess Management Appendix A – Menu Displays 8-17 Model NGA 2000 HFID MENU: 48 LINRANGE1I1 MENU: 49 CALFAC.
Instruction Manual 760004-A February 2002 8-18 Appendix A - Menu Displays Rosemount Anal ytical Inc. A Division of Emerson Pro cess Management Model NGA 2000 HFID MENU: 51 APARLST4 MENU: 52 APARLST5 M.
Instruction Manual 760004-A February 2002 Rosemount Analytical Inc. A Division of Emerson Pr ocess Management Appendix A – Menu Displays 8-19 Model NGA 2000 HFID MENU: 54 IGNITION MENU: 55 LISTNOTES.
Instruction Manual 760004-A February 2002 8-20 Appendix A - Menu Displays Rosemount Anal ytical Inc. A Division of Emerson Pro cess Management Model NGA 2000 HFID MENU: 57 AUTOFLAMEI1 MENU: 58 DISPLAY.
Instruction Manual 760004-A February 2002 Rosemount Analytical Inc. A Division of Emerson Pr ocess Management Appendix A – Menu Displays 8-21 Model NGA 2000 HFID MENU: 60 AMTOPINFO MENU: 61 ANALSET2.
Instruction Manual 760004-A February 2002 8-22 Appendix A - Menu Displays Rosemount Anal ytical Inc. A Division of Emerson Pro cess Management Model NGA 2000 HFID MENU: 63 TLIMITSAI1 MENU: 64 IGNITION.
Instruction Manual 760004-A February 2002 Rosemount Analytical Inc. A Division of Emerson Pr ocess Management Appendix A – Menu Displays 8-23 Model NGA 2000 HFID MENU: 66 AMHELPINDEX2 MENU: 67 SOFT_.
Instruction Manual 760004-A February 2002 8-24 Appendix A - Menu Displays Rosemount Anal ytical Inc. A Division of Emerson Pro cess Management Model NGA 2000 HFID MENU: 69 AM2VC MENU: 70 AM2VD MENU: 7.
Instruction Manual 760004-A February 2002 Rosemount Analytical Inc. A Division of Emerson Pr ocess Management Appendix A – Menu Displays 8-25 Model NGA 2000 HFID MENU: 72 LINFUNCT MENU: 73 POLYSETUP.
Instruction Manual 760004-A February 2002 8-26 Appendix A - Menu Displays Rosemount Anal ytical Inc. A Division of Emerson Pro cess Management Model NGA 2000 HFID MENU: 75 POLYGAS1 MENU: 76 POLYSETI1 .
Instruction Manual 760004-A February 2002 Rosemount Analytical Inc. A Division of Emerson Pr ocess Management Appendix A – Menu Displays 8-27 Model NGA 2000 HFID MENU: 78 POLYGAS3 MENU: 79 POLYGAS4 .
Instruction Manual 760004-A February 2002 8-28 Appendix A - Menu Displays Rosemount Anal ytical Inc. A Division of Emerson Pro cess Management Model NGA 2000 HFID MENU: 81 POLYGAS6 MENU: 82 POLYGAS7 M.
Instruction Manual 760004-A February 2002 Rosemount Analytical Inc. A Division of Emerson Pr ocess Management Appendix A – Menu Displays 8-29 Model NGA 2000 HFID MENU: 84 POLYGAS9 MENU: 85 POLYGAS0 .
Instruction Manual 760004-A February 2002 8-30 Appendix A - Menu Displays Rosemount Anal ytical Inc. A Division of Emerson Pro cess Management Model NGA 2000 HFID MENU: 87 MIDPOINT3 MENU: 88 MIDPOINT4.
Instruction Manual 760004-A February 2002 Rosemount Analytical Inc. A Division of Emerson Pr ocess Management Appendix A – Menu Displays 8-31 Model NGA 2000 HFID MENU: 90 EXP_CAL MENU: 91 ZEROI2 MEN.
Instruction Manual 760004-A February 2002 8-32 Appendix A - Menu Displays Rosemount Anal ytical Inc. A Division of Emerson Pro cess Management Model NGA 2000 HFID MENU: 93 EXP_CAL_DATI MENU: 94 UNITS .
Instruction Manual 760004-A February 2002 Rosemount Analytical Inc. A Division of Emerson Pr ocess Management Appendix A – Menu Displays 8-33 Model NGA 2000 HFID MENU: 96 POLYSETI2 MENU: 97 POLYSETI.
Instruction Manual 760004-A February 2002 8-34 Appendix A - Menu Displays Rosemount Anal ytical Inc. A Division of Emerson Pro cess Management Model NGA 2000 HFID MENU: 99 STORE MENU: 100 ANALOPI1A ME.
Instruction Manual 760004-A February 2002 Rosemount Analytical Inc. A Division of Emerson Pr ocess Management Appendix A – Menu Displays 8-35 Model NGA 2000 HFID MENU: 102 RFACTORSIA MENU: 103 RFHIS.
Instruction Manual 760004-A February 2002 8-36 Appendix A - Menu Displays Rosemount Anal ytical Inc. A Division of Emerson Pro cess Management Model NGA 2000 HFID MENU: 105 RFHIST4A MENU: 106 SW_DIAGI.
Instruction Manual 760004-A February 2002 Rosemount Analytical Inc. A Division of Emerson Pr ocess Management Appendix A – Menu Displays 8-37 Model NGA 2000 HFID MENU: 108 ANALSETI3 MENU: 109 STORED.
Instruction Manual 760004-A February 2002 8-38 Appendix A - Menu Displays Rosemount Anal ytical Inc. A Division of Emerson Pro cess Management Model NGA 2000 HFID MENU: 111 ZEROI1A MENU: 112 ZERO_NOW2.
Instruction Manual 760004-A February 2002 Rosemount Analytical Inc. A Division of Emerson Pr ocess Management Appendix A – Menu Displays 8-39 Model NGA 2000 HFID MENU: 114 AMMISC2 MENU: 115 MPARMS2 .
Instruction Manual 760004-A February 2002 8-40 Appendix A - Menu Displays Rosemount Anal ytical Inc. A Division of Emerson Pro cess Management Model NGA 2000 HFID MENU: 117 REBOOT MENU: 118 ABOUT MENU.
Instruction Manual 760004-A February 2002 Rosemount Analytical Inc. A Division of Emerson Pr ocess Management Appendix A – Menu Displays 8-41 Model NGA 2000 HFID MENU: 120 ALARM1 MENU: 121 MANDATA.
Instruction Manual 760004-A February 2002 8-42 Appendix A - Menu Displays Rosemount Anal ytical Inc. A Division of Emerson Pro cess Management Model NGA 2000 HFID.
WARRANTY Goods and part(s) (ex cluding consumables) ma nufactured by Seller are warranted to be free from defects in workmanship and material un der normal use and service for a peri od of twelve (12) months from the date of shipment by Seller. Consumables, glass elect rodes, membranes, liquid junctions, electrolyte, o-rings, etc.
Instruction Manual 760004-A February 2002 © Rosemount Analytical Inc. 2001 Emerson Process Management Rosemount Analytical Inc. Process Analytic Division 1201 N. Main St. Orrville, OH 44667-0901 T (330) 682-9010 F (330) 684-4434 E gas.csc@emersonprocess.
Ein wichtiger Punkt beim Kauf des Geräts Emerson Process Management 760004-A (oder sogar vor seinem Kauf) ist das durchlesen seiner Bedienungsanleitung. Dies sollten wir wegen ein paar einfacher Gründe machen:
Wenn Sie Emerson Process Management 760004-A noch nicht gekauft haben, ist jetzt ein guter Moment, um sich mit den grundliegenden Daten des Produkts bekannt zu machen. Schauen Sie zuerst die ersten Seiten der Anleitung durch, die Sie oben finden. Dort finden Sie die wichtigsten technischen Daten für Emerson Process Management 760004-A - auf diese Weise prüfen Sie, ob das Gerät Ihren Wünschen entspricht. Wenn Sie tiefer in die Benutzeranleitung von Emerson Process Management 760004-A reinschauen, lernen Sie alle zugänglichen Produktfunktionen kennen, sowie erhalten Informationen über die Nutzung. Die Informationen, die Sie über Emerson Process Management 760004-A erhalten, werden Ihnen bestimmt bei der Kaufentscheidung helfen.
Wenn Sie aber schon Emerson Process Management 760004-A besitzen, und noch keine Gelegenheit dazu hatten, die Bedienungsanleitung zu lesen, sollten Sie es aufgrund der oben beschriebenen Gründe machen. Sie erfahren dann, ob Sie die zugänglichen Funktionen richtig genutzt haben, aber auch, ob Sie keine Fehler begangen haben, die den Nutzungszeitraum von Emerson Process Management 760004-A verkürzen könnten.
Jedoch ist die eine der wichtigsten Rollen, die eine Bedienungsanleitung für den Nutzer spielt, die Hilfe bei der Lösung von Problemen mit Emerson Process Management 760004-A. Sie finden dort fast immer Troubleshooting, also die am häufigsten auftauchenden Störungen und Mängel bei Emerson Process Management 760004-A gemeinsam mit Hinweisen bezüglich der Arten ihrer Lösung. Sogar wenn es Ihnen nicht gelingen sollte das Problem alleine zu bewältigen, die Anleitung zeigt Ihnen die weitere Vorgehensweise – den Kontakt zur Kundenberatung oder dem naheliegenden Service.