Model 399VP Combination pH/ORP Sensor Instruction Manual PN 51-399VP/rev .E March 2012.
CAUTION SENSOR/PROCESS APPLICA TION COMP A TIBILITY The wetted sensor materials may not be comp atible with process com position and operating conditions.
MODEL 399VP pH/ORP T ABLE OF CONTENTS MODEL 399VP pH/ORP SENSOR T ABLE OF CONTENTS SECTION TITLE P AGE 1.0 DESCRIPTION AND SPECIFICA TIONS .......................................................... 1 1.1 Features and Applications .....................
MODEL 399VP pH/ORP T ABLE OF CONTENTS MODEL 399VP pH/ORP SENSOR LIST OF FIGURES Figure No. Title Page 2-1 Sensor Dimensions ........................................................................................... 4 2-2 Model 399VP Shown in V arious Flow-Through Installations .
MODEL 399VP pH/ORP SENSOR SECTION 1.0 DESCRIPTION AND SPECIFICA TIONS SECTION 1.0. DESCRIPTION AND SPECIFICA TIONS 1.1 FEA TURES AND APPLICA TIONS 1.2 PHYSICAL SPECIFICA TIONS 1.
MODEL 399VP pH/ORP SENSOR SECTION 1.0 DESCRIPTION AND SPECIFICA TIONS 1.3 ORDERING INFORMA TION The Model 399VP pH/ORP Sensor is available with custom glass electrodes housed in a molded T efzel body with 1 in. MNPT threads suit able for insertion, submersion or flow through installation.
MODEL 399VP pH/ORP SENSOR SECTION 2.0 INST ALLA TION SECTION 2.0. INST ALLA TION 2.1 UNP ACKING AND INSPECTION 2.2 MOUNTING 2.3 FLOW -THROUGH INST ALLA TIONS 2.4 INSERTION INST ALLA TIONS 2.5 SUBMERSION INST ALLA TIONS 2.1 UNP ACKING AND INSPECTION. Inspect the carton for any damage.
MODEL 399VP pH/ORP SENSOR SECTION 2.0 INST ALLA TION 4 FIGURE 2-1. Sensor Dimensions INCH MILLIMETER.
MODEL 399VP pH/ORP SENSOR SECTION 2.0 INST ALLA TION 2.3 FLOW-THROUGH INST ALLA TIONS. The Model 399VP Sensor has a 1 in. MNPT process connection at the front and back end of the sensor for mount- ing directly into a 1 in. or 1½ in. tee. See Figure 2-2 for installation configurations.
6 MODEL 399VP pH/ORP SENSOR SECTION 2.0 INST ALLA TION 2.4 INSERTION INST ALLA TIONS. The Model 399VP sensor has forward and rear facing 1 in. MNPT process connections for pipeline installations (see Figure 2-4 or 2-5) and flange connections. The Model 399VP is also suitable for side-of-t ank installation.
MODEL 399VP pH/ORP SENSOR SECTION 2.0 INST ALLA TION 2.5 SUBMERSION INST ALLA TIONS. The Model 399VP Sensor has a 1 in. MNPT process connection at the back of the sensor (see Figure 2-1). Utilizing a standard 1 in. union (such as PN 9330022), the sensor may be mounted to a 1 in.
8 MODEL 399VP pH/ORP SENSOR SECTION 2.0 INST ALLA TION FIGURE 2-6. SENSOR SHOWN IN HANDRAIL MOUNTING ASSEMBL Y (PN 1 1275-01) The Handrail Mounting Assembly can be used in applications where Model 399VP needs to be placed in a pond, tank, aer- ation basin, or open channel.
MODEL 399VP pH/ORP SENSOR SECTION 2.0 INST ALLA TION FIGURE 2-8. JUNCTION BOX (PN 22719-02) AND PIPE MOUNT UNION ACCESSOR Y (PN 9330022) FIGURE 2-7. JET SPRA Y CLEANER (PN 12707-00) USED WITH MODEL 399VP The Jet S pray Cleaner eliminates routine manual sensor maintenance by cleaning Model 399VP with water or compressed air .
10 MODEL 399VP pH/ORP SENSOR SECTION 3.0 WIRING SECTION 3.0. WIRING Figures 3-2 thru 3-23 provide the diagrams for wiring the 399VP sensor to various analyzer/transmitters. 1. If the cable needs to be extended, use a high quali- ty cable available from Rosemount Analytical.
11 MODEL 399VP pH/ORP SENSOR SECTION 3.0 WIRING FIGURE 3-1. Wire and Connector Pin Functions FIGURE 3-2. Model 81 Wiring FIGURE 3-3. Model 1 181 Wiring FIGURE 3-4.
12 MODEL 399VP pH/ORP SENSOR SECTION 3.0 WIRING FIGURE 3-6. Model 2081 Wiring FIGURE 3-8. Model 2081 Wiring through Remote Junction Box FIGURE 3-7. Model 1 181, 1050/1060, and 1003/1023 Wiring through Remote Junction Box FIGURE 3-5.
13 MODEL 399VP pH/ORP SENSOR SECTION 3.0 WIRING FIGURE 3-9. Model 81 Wiring through Remote Junction Box FIGURE 3-10. Model 3081 & 4081 Wiring through Remote Junction Box FIGURE 3-12.
14 MODEL 399VP pH/ORP SENSOR SECTION 3.0 WIRING FIGURE 3-13. Model 1054 Wiring FIGURE 3-14. Model 1054A/B and 2054 Wiring FIGURE 3-16. Model 1054A/B & 2054 Wiring through a Remote Junction Box FIGURE 3-15.
MODEL 399VP pH/ORP SENSOR SECTION 3.0 WIRING FIGURE 3-18. Model 2700 Wiring FIGURE 3-17. Model SCL-(P/Q) Wiring FIGURE 3-19. Model 54epH Wiring FIGURE 3-20.
16 MODEL 399VP pH/ORP SENSOR SECTION 3.0 WIRING FIGURE 3-21. Wiring Model 399VP-09, 400, 401, 402, 403, and 404 to Model 1055 (Pipe/W all Mount). DWG .
17 MODEL 399VP pH/ORP SENSOR SECTION 3.0 WIRING FIGURE 3-22. Wiring Model 399VP-09, 400, 401, 402, 403, and 404 to Model 1055 (Panel Mount). DWG . NO. REV .
18 MODEL 399VP pH/ORP SENSOR SECTION 3.0 WIRING FIGURE 3-23. Prep aration of Extension Cable (PN 9200273) This cable is shipped as a raw cable. Cable must be prepared, as shown above, before inst allation. T o avoid preparing raw cable, PN 23646-01 can be used.
19 MODEL 399VP pH/ORP SENSOR SECTION 4.0 ST ART -UP AND CALIBRA TION SECTION 4.0. ST ART -UP AND CALIBRA TION 4.1 SENSOR PREP ARA TION Shake down the sensor to remove any air bubbles that may be present in the tip of the pH glass bulb. In most cases, the pH sensor can simply be installed as shipped, and readings with an accuracy of ± 0.
20 MODEL 399VP pH/ORP SENSOR SECTION 4.0 ST ART -UP AND CALIBRA TION 4.3 ORP CALIBRA TION Most industrial applications have a number of ORP reactions occurring in sequence or simult aneously . There can be several components that are oxidized or reduced by the reagent s that are used.
21 MODEL 399VP pH/ORP SENSOR SECTION 5.0 MAINTENANCE SECTION 5.0. MAINTENANCE 5.1 GENERAL The Model 399VP sensor is a disposable type sensor and therefore requires minimum maintenance. The sensor should be kept clean and free of debris and sediment at all times.
22 MODEL 399VP pH/ORP SENSOR SECTION 5.0 MAINTENANCE 5.3 pH ELECTRODE CLEANING If the electrode is coated or dirty , clean as follows: 1. Remove the sensor from process.
23 MODEL 399VP pH/ORP SENSOR SECTION 6.0 TROUBLESHOOTING SECTION 6.0. TROUBLESHOOTING TROUBLESHOOTING . The table below lists common problems, causes and remedies typically encountered in process measurement. Problem Probable Cause Remedy Meter reads off scale (display T .
MODEL 399VP pH/ORP SENSOR SECTION 7.0 RETURN OF MA TERIAL SECTION 7.0 RETURN OF MA TERIAL 7.1 GENERAL. T o expedite the repair and return of instruments, proper communication between the customer and the factory is important. Before returning a product for rep air , call 1-949-757-8500 for a Return Materials Authorization (RMA) number .
FROM: RETURN BILL T O: _____________________________ _____________________________ _____________________________ _____________________________ _____________________________ ___________________________.
A W orldwide Network of Sales and Service Emerson Process Management’s field sales of fices are your source for more information on the fill line of Rosemount Analytical products. Field sales personnel will work closely with you to supply technical dat a and application information.
W ARRANTY Seller warrants that the firmware will execute the programming instructions provided by Seller , and that the Goods manufactured or Services provided by Seller will be free from defects in materials or workmanship under normal use and care until the expira- tion of the applicable warranty period.
Credit Cards for U.S. Purchases Only . The right people, the right answers, right now . ON-LINE ORDERING NOW A V AILABLE ON OUR WEB SITE http://www .raihome.
Ein wichtiger Punkt beim Kauf des Geräts Emerson 399VP (oder sogar vor seinem Kauf) ist das durchlesen seiner Bedienungsanleitung. Dies sollten wir wegen ein paar einfacher Gründe machen:
Wenn Sie Emerson 399VP noch nicht gekauft haben, ist jetzt ein guter Moment, um sich mit den grundliegenden Daten des Produkts bekannt zu machen. Schauen Sie zuerst die ersten Seiten der Anleitung durch, die Sie oben finden. Dort finden Sie die wichtigsten technischen Daten für Emerson 399VP - auf diese Weise prüfen Sie, ob das Gerät Ihren Wünschen entspricht. Wenn Sie tiefer in die Benutzeranleitung von Emerson 399VP reinschauen, lernen Sie alle zugänglichen Produktfunktionen kennen, sowie erhalten Informationen über die Nutzung. Die Informationen, die Sie über Emerson 399VP erhalten, werden Ihnen bestimmt bei der Kaufentscheidung helfen.
Wenn Sie aber schon Emerson 399VP besitzen, und noch keine Gelegenheit dazu hatten, die Bedienungsanleitung zu lesen, sollten Sie es aufgrund der oben beschriebenen Gründe machen. Sie erfahren dann, ob Sie die zugänglichen Funktionen richtig genutzt haben, aber auch, ob Sie keine Fehler begangen haben, die den Nutzungszeitraum von Emerson 399VP verkürzen könnten.
Jedoch ist die eine der wichtigsten Rollen, die eine Bedienungsanleitung für den Nutzer spielt, die Hilfe bei der Lösung von Problemen mit Emerson 399VP. Sie finden dort fast immer Troubleshooting, also die am häufigsten auftauchenden Störungen und Mängel bei Emerson 399VP gemeinsam mit Hinweisen bezüglich der Arten ihrer Lösung. Sogar wenn es Ihnen nicht gelingen sollte das Problem alleine zu bewältigen, die Anleitung zeigt Ihnen die weitere Vorgehensweise – den Kontakt zur Kundenberatung oder dem naheliegenden Service.