Operating and Maintenance Guide N3970A OPTICAL POWER METER.
.
■ N3970A OPTICAL POWER METER SOURCE QUICK REFERENCE GUIDE ■ T o remove interchangeable connector , move interface to mid position, and pull off adaptor . ■ T o defeat auto power-off , hold POWER for 3 seconds at turn on until ON and perm are displayed.
1 O PTICAL P OWER M ETER N3970A Congratulations on your purchase of this instrument, which has been engineered to provide the best possible r eliability, convenience and performance. To get the best use from your equipment and ensure its safe operation, please spend a few minutes to read this manual.
2 ■ C O N T E N T S BEFORE READING FURTHER, OPEN THE CONTROL PANEL PICTURE AT THE BACK OF THE MANUAL Service and Suppor t 3 Introduction & applications 4-5 Safety Summary 6-7 Battery and externa.
3 N3970A UM-1 ■ S E R VICE AND SUPPOR T Any adjustment, maintenance, or repair of this product must be p e rf o r med by q ualifie d personn el. Conta ct your c ustomer engine er through you r local Agile nt T echnologies Ser v i c e Center. You can find a list of local service representatives on the W eb at: http://www.
4 ■ I N T R O D U C T I O N A N D A P P L I C AT I O N S The N3970A Optical Power Meter is used for field or laboratory testing of fiber optic systems at typically 850, 1300, 1310 and 1550 nm. It features superb measurement confidence, ease of use and reduced cost of ownership, and is for use by installers, technicians and engineers.
5 N3970A UM-1 The tone detector is very sensitive, and is useful for fiber identification, and calibrating the modulation frequency of test tone sources. The instru ment ha s shock absorbe nt corn ers, an d a toug h pol ycarbonate housing which ha s passed ext ensive drop t esting.
6 ■ SAFE TY S U MMAR Y The following general safety precautions must be observed during all phases of operation, service, and repair of this instrument. Failure to comply with these precautions or with specific warnings elsewhere in this manual violates safety standards of design, manufacture, and intended use of the instrument.
7 N3970A UM-1 ■ SAFET Y SUM MAR Y Initial Inspection Inspect the shipping container for damage. If there is damage to the container or Cushioning, keep them until you have checked the contents of the shipment for completeness and verified the instrument both mechanically and electrically .
8 ■ B A TTERY AND E XTERNAL P OWER T he i nst ru men t i s po wer e d by t wo 1. 5V dry b att e ri es (A l kal in e C) for an ope rating period of t ypica lly 250 ho urs or two 1.2V r echargeable batteries for an operating period of approximately 125 hours.
9 N3970A UM-1 ■ O PTICAL CONN EC TO R The optical port is mounted on a swivel, which allows the connector to be angled outwards for accessibility, and then pushed back and covered with the snap cover to provide dirt and drop protection. The other port can be used to store a spare interchangeable optical connector .
10 ■ OP TICAL C ONNECT OR Cleaning The glass power meter interface does not make contact with the inserted connector: there is a slight air gap. Therefore it should not wear out, and only needs occasional cleaning.
11 N3970A UM-1 ■ GETTI NG S T A R TE D AND TURNING ON Thi s and following sections s how you how to use your instrument: Check that the cor r ect accessories have been supplied, and an optical connector adaptor has been fitted. If you have any queries, call your supplier to resolve the situation.
12 ■ GETTING ST A R TE D AND TU R NING ON T o disabl e or enabl e the bu zzer , h old dow n SHIFT fo r 3 second s. T o add the car r y strap: slip the end of the strap through the slit on the curved section at the back of the instrument, and secure the buckle.
13 N3970A UM-1 ■ AUTOT EST OPERA T I O N General Autotest features For Autotest to function, all instruments involved in the test mu st be Ag il ent Ha ndhe ld i ns trum en ts w it h a n A utote st function.
14 ■ AU TOTEST OPERA T I O N meter with a test lead. Set the source to Autotest mode, and then push SET REF on the meter for 3 seconds. This will store the reference at all relevant wavelengths. Either a ‘local’ or ‘remote’ reference can be stored.
15 N3970A UM-1 ■ P O W E R M E T E R O P E R AT I O N After instrument turn on, the meter will be calibrated at the last used λ . To change parameters, push POWER METER, then: T o select a calibrated λ , push +/-. λ is displayed on the top right side of the display.
16 ■ S TORING & R ECALLING R EADINGS There are various ways of storing and recording data, giving the user flexible ways of working. They are summarised here for your convenience. Storage of reference values: A separate reference is stored for each λ and can be recalled when the instrument is next turned on.
17 N3970A UM-1 ■ S TORING & R ECALLING R EADINGS The memory location is shown at the top right of the display . When storing data, the memory automatically increments to facilitate the collection of data on a job. Y ou can also go back and over-write particular records.
18 ■ RS 232 INT ERF A C E The connection details of the instrument, and connecting cables are as follows : Some (out of spec) serial ports need a 10 K resistor in the D connector across wires 1 & 2.
19 N3970A UM-1 ■ C ARE OF Y OUR I NSTRUMENT ■ Follow the directions in this manual on cleaning optical connectors. ■ Limit the input optical power to no more than 5 dB over - range. ■ During storage and transport, keep the instrument in its carr y case to protect against crushing, vibration, dust and moisture.
20 ■ A CCURACY C ONSIDERATIONS All measurements Optical connectors should be kept clean and in good condition. Due to emitter center λ uncertainty (eg 1310 +/- 20 nm), fiber attenuation may vary with different sources. This λ uncertainty can also affect power meter calibration.
N3970A UM-1 ■ ACCURACY CONSIDERATIONS 21.
22 ■ C ALIBRATION & M AINTENANCE Enabling / disabling ‘shift’ keypad & re-booting the microprocessor: See section on ‘getting started & turning on.
23 N3970A UM-1 ■ C ALIBRATION & M AINTENANCE ■ Measure and record the source power using the reference meter ■ T ransfer the same power level to the meter to be calibrated, set MAX/MIN to display power, then +/- to adjust the r eading to match the noted reference reading.
24 ■ C ALIBRATION , M AINTENANCE & INSTRUMENT RETURNS ■ Hinged display cover removal can be done at this point. ■ The ribbon connector can be disengaged to completely separate the instrument halves. ■ Further dis-assembly from this stage should be easily apparent to a technician.
25 N3970A UM-1 ■ SPECIFICATIONS - DEFINITION OF TERMS Power Range The range of input powers for which the instrument can be used. Maximum Input Power The input power is not to be exceeded to avoid destroying the instrument.
26 ■ S PECIFICATIONS & O RDERING I NFORMATION Sensor Element InGaAs W avelength Range 850-1650 nm Calibration Wavelengths 850 1 , 1300, 1310, 1550 nm Power Range +5 to –60 dBm (850 nm) +5 to –70 dBm (other wavelengths) Maximum Input Power +15 dBm Uncertainty at reference conditions 2 ±0.
27 N3970A UM-1 ■ S PECIFICATIONS & O RDERING I NFORMATION 1: All specifications at 850 nm are typical. 2: 23 ± 3 o K, -10 to –30 dBm, calibra tion wavelen gt hs ± 2nm, f iber typ e standa rd SMF , c onnector type ce ramic F C / PC, at d ay of calibration.
28 ■ S PECIFICATIONS & O RDERING I NFORMATION ORDERING INFORMATION Instrument N3970A Optical Power Meter Standard accessories: Included with instrument: FC, ST, SC optical connector, manual, batteries, NA T A traceable calibration certificate, car r y strap, pouch, protective holster & RS 232 cable.
29 N3970A UM-1 ■ PERFO RMANCE TEST The procedures in this section test the performance of the instrument. The complete specifications to which the Agilent N39 70A Opt ical P ower Me ter is teste d are g iven o n pages 26-28 . All tests can be performed without access to the interior of the instrument.
30 ■ PER FORMANCE TEST T able 1 Required Equipment Instrument/Accessory Recommended Required Alternative Model Characteristics Models Lightwave Multimeter 8163A 8164A, 8166A Interface Module (one 81.
31 N3970A UM-1 ■ P ERFORMANC E TEST T est Record Results of the performance test may be tabulated on the Test R e c o r d p rovided at the e nd of the t est pr o c e d u r e. It is r ecommended that you fill out the Test Record and refer to it while doing the test.
32 ■ PER FORMANCE TEST Performance Test The performance test given in this section includes the T otal Uncertainty/Accuracy Test. Perform each step in the order given, using the cor responding test equipment. Note Make sure that all optical connections are dry and clean.
33 N3970A UM-1 ■ P ERFORMANC E TEST 4. Switch on the instruments. a. Switch on the lightwave multimeter. b. Configure the laser source. ■ Set the laser source attenuation to 0 dB ■ Set the wavelength to 1310 nm ■ N ote th e ac tu al ( di sp la yed ) w a vel en gth of the la se r s our ce i n t he t es t r e p o r t c.
34 ■ PER FORMANCE TEST 6. Measure the DUT. a. Make sure the N3970A (DUT) has warmed up b. Connect the attenuator output cable to the DUT as shown in figure below . T est Setup for the DUT Measurement c. Select the cor rect wavelength on the DUT. d. Set the attenuator to its value for setting 1.
35 N3970A UM-1 ■ P ERFORMANC E TEST Model N3970A Optical Power Meter Serial No. Options Firmware Rev . T est Facility Performed by Special Notes Date Ambient Temperatur e o C Relative Humidity % Lin.
36 ■ PER FORMANCE TEST Description Model No. 1. Lightwave Multimeter 8163A 2. Light Source 1310/1550 nm 81654A 3. Optical Head 81624A #C01 4. Optical Head Interface Module 81618A 5.
37 N3970A UM-1 Model N3970A Optical Power Meter T otal Uncertainty/Accuracy Test T est Wavelengths = 1300 nm and 1310 nm Source Wavelength = nm Report No. Date ■ P ERFORMANC E TEST Setting Number 1 2 3 4 -3.00 dBm -20.00 dBm -30.00 dBm -50.00 dBm -3.
38 ■ PER FORMANCE TEST Model N3970A Optical Power Meter T otal Uncertainty/Accuracy Test T est Wavelengths = 1550 nm Source Wavelength = nm Report No. Date Setting Number 1 2 3 4 -3.00 dBm -20.00 dBm -30.00 dBm -50.00 dBm -3.30 dBm -20.30 dBm -30.30 dBm -50.
39 N3970A UM-1 ■ CLEA NING I N F O RM A T I O N The following Cleaning Instructions contain some general safety precautions, which must be observed during all phases of cleaning. Consult your specific optical device manuals or guides for full information on safety matters.
40 ■ C LE ANING INFOR MA T I O N dampened with water, to clean the external case parts. Do not attempt to clean internally. Do not install parts or perform any unauthorised modification to optical devices. * Refer servicing only to qualified and authorised personnel.
41 N3970A UM-1 ■ CLEA NING PROCEDU RES Keep your dust and shutter caps on when your instrument is not in use. This should prevent it from getting too dirty. If you must clean such instruments, please refer the instrument to the skilled personnel of Agilent's service team.
42 ■ DIS CLAIMER & W A R R A N T Y Information in this manual is given in good faith for the benefit of the user. It cannot be used as the basis for claims against Ag ilen t T ec hn ol ogie s o r i ts r e p r es en tati ve s, if ac cide nt al damage or inconvenience results from use or attempted repair of the equipment.
43 N3970A UM-1 ■ N O T E S.
44 ■ N O T E S.
Ein wichtiger Punkt beim Kauf des Geräts Agilent Technologies N3970A (oder sogar vor seinem Kauf) ist das durchlesen seiner Bedienungsanleitung. Dies sollten wir wegen ein paar einfacher Gründe machen:
Wenn Sie Agilent Technologies N3970A noch nicht gekauft haben, ist jetzt ein guter Moment, um sich mit den grundliegenden Daten des Produkts bekannt zu machen. Schauen Sie zuerst die ersten Seiten der Anleitung durch, die Sie oben finden. Dort finden Sie die wichtigsten technischen Daten für Agilent Technologies N3970A - auf diese Weise prüfen Sie, ob das Gerät Ihren Wünschen entspricht. Wenn Sie tiefer in die Benutzeranleitung von Agilent Technologies N3970A reinschauen, lernen Sie alle zugänglichen Produktfunktionen kennen, sowie erhalten Informationen über die Nutzung. Die Informationen, die Sie über Agilent Technologies N3970A erhalten, werden Ihnen bestimmt bei der Kaufentscheidung helfen.
Wenn Sie aber schon Agilent Technologies N3970A besitzen, und noch keine Gelegenheit dazu hatten, die Bedienungsanleitung zu lesen, sollten Sie es aufgrund der oben beschriebenen Gründe machen. Sie erfahren dann, ob Sie die zugänglichen Funktionen richtig genutzt haben, aber auch, ob Sie keine Fehler begangen haben, die den Nutzungszeitraum von Agilent Technologies N3970A verkürzen könnten.
Jedoch ist die eine der wichtigsten Rollen, die eine Bedienungsanleitung für den Nutzer spielt, die Hilfe bei der Lösung von Problemen mit Agilent Technologies N3970A. Sie finden dort fast immer Troubleshooting, also die am häufigsten auftauchenden Störungen und Mängel bei Agilent Technologies N3970A gemeinsam mit Hinweisen bezüglich der Arten ihrer Lösung. Sogar wenn es Ihnen nicht gelingen sollte das Problem alleine zu bewältigen, die Anleitung zeigt Ihnen die weitere Vorgehensweise – den Kontakt zur Kundenberatung oder dem naheliegenden Service.