Benutzeranleitung / Produktwartung CFW 600 des Produzenten Adams
Zur Seite of 32
Adam Equipment CFW SERIES (P.N. 6162, Revision A5, May 2006) Software Version: V1.04 @Adam Equipment Company 2006.
@Adam Equipment Company 2006.
CONTENTS 1.0 INTRODUCTION ............................................................................................................ 3 2.0 SPECIFICATIONS ...............................................................................................
@Adam Equipment Company 2006 2.
1.0 INTRODUCTION • The CFW series of scales provide an accurate, fast, versatile series of general purpose weighing scales with counting and check-weighing functions. • There are 2 series of scales- CF W scales are kilogram only scales and the CFWa scales offer changeable unit from pounds to kilograms.
2.0 SPECIFICA TIONS CFW SERIES Model # CFW 60 CFW 150 CFW 300 CFW 600 Maximum Capacity 60 kg 150 kg 300 kg 600kg Readability 5 g 10 g 20 g 50g Repeatability (Std Dev) 5 g 10 g 20 g 50g Linearity ± 10.
Common Specifications Interface RS-232 bi-directional Interface (Optional) Stabilisation Time 2 seconds typical Operating Temperature 0°C - 40°C 32°F - 104°F Power supply 9 VDC, 800 mA from extern.
3.0 INST ALLA TION 3.1 LOCATING THE SCALES • The scales should not be placed in a location that will reduce the accuracy. • Avoid extremes of temperature. Do not place in direct sunlight or near air conditioning vents. • Avoid unsuitable tables.
3.3 SETTING UP THE SCALES • The pillar is attached to the base using a bracket that must be attached to the base frame first using the 4 bolts supplied. The pillar is secured to the bracket using 2 sets of screws. The cable from the base to the indicator module is run through the tube and taken out through the plastic support at the top.
4.0 KEY DESCRIPTIONS [Zero] or Sets the zero point for all subsequent weighing. The display shows zero. A secondary function is of "Enter" key used when setting up the value for the Parameters.
A secondary function is to move the active/flashing digit to the right when setting values for the Parameters. [Func] or [F] Selects the Functions of the scale. If the scale is weighing, it will select parts counting. If it is not in weighing mode, it will return the user to weighing.
6.0 OPERA TION 6.1 ZEROING THE DISPLAY • You can press the [Zero/Enter] key at any time to set the zero point. This will usually be necessary when the platform is empty.
tared a second time if another type of product was to be added to the first one. Again only the weight that is added after taring will be displayed. • When the container is removed a negat ive value will be shown.
• Press the [Zero/Enter] key again when the number matches the number of parts used as the sample. As more weight is added the display will show the number of parts (pcs).
6.5 CHECK-WEIGHING Check-weighing is a procedure to cause an alarm to sound when the weight on the scale meets or exceeds values st ored in memory. The memory holds values for a high and a low limit. Either or both can be used. See PARAMETERS Section for the procedure.
Checkweighing facility can be set up during Weighing or Parts Counting by entering values as Low or/and High Limits keyed in by the user. The limits are displayed in kg or pcs respectively.
• Remove the weight, allowing the scale to return to zero and put a second weight on. Press the [Print/M+] key, the display will show " ACC 2 ", then the new total and finally the value of the second weight. No. of accumulation of w eights Total w eight for 3 accumulations • Continue until all weights have been added.
7.0 USER P ARAMETERS • The scale has 7 User Parameters that can be set by the user. To set these Parameters press the [Func] key. The display will show the first parameter, " F0 H-L ". • Pressing the [Tare/ Ï ] will cycle through the other parameters.
SETTING OF USER PARAMETERS PARAMETER SUB- PARAMETER DESCRIPTION DEFAULT VALUE SEt Lo Sets a value for the Low Limit used for checkweighing. 000.000 F0 H-L SEt HI Sets a value for the High Limit for checkweighing. 000.000 to CLr Clears the Accumulation Total in the memory without printing the results.
F5 Prt Sets the RS-232 - P bAUd- Select from the desired baud rates. Options are 600, 1200, 2400, 4800 and 9600 P nodE - Set the printing mode to- P Prt (Print only when [Print] is pressed) P Cont (Print continuously) P Auto (Print automatically whenever there is any weight on the scale).
8.0 BA TTER Y OPERA TION • The scales can be operated from the batt ery if desired. The battery life is approximately 70 hours. • When the battery needs charging a symbol on the display will turn on. The battery should be charged when the symbol is on.
9.0 RS-232 INTERF ACE The CFW Series of scales can be ordered with an optional RS-232 bi-directional interface. Specifications: RS-232 output of weighing data ASCII code 4800 Baud (600-9600 selectable.
When parts counting- the weight, unit w eight and count w ill be printed: GS 12.340kg GS for Gross we ight, NT for net weight and a unit of weight UW 123.
10.0 CALIBRA TION The scale can be calibrated using the fo llowing procedure. For a more detailed method of calibrating the scale, it will be necessary to enter the secure Programming Menu.
1 1.0 SER VICE SECTION To set the Service Parameters it will be necessary to enter the User Parameters as before and then use a password code while entering the programming parameter “ PROG ”. 11.1 CFW FACTORY PROGRAMMING AND CALIBRATIONS • The parameter “ PROG ” allows access to the calibration section.
SETTING OF SERVICE PARAMETERS SUB- PARAMETER FUNCTIONS DESCRIPTION DEFAULT SETTING AZn 0 Auto Zero operating range. (Options: 0.5d, 1.0d, 2d, 4d) Sets the operational range of the auto zero function. 1d 0-Auto Initial Zero range at power on. (Options: 0%, 2%, 5%, 10% and 20%) Sets the range for setting a zero when power is applied.
dEC1 Selects the decimal point position (Options: 0, 0.0, 0.00, 0.000). C 0.00 InC Selects the increment size. (Options: 1, 2, 5 and 10). inC 10 CAP Sets the scale capacity. The maximum value for the scale is as specified for the scale. For example 60kg.
12.0 ERROR CODES ERROR CODES DESCRIPTION SUGGESTIONS - -oL - - Over-range Remove weight from the scale. If the problem persists contac t your dealer or Adam Equipment for assistance. Err 1 Date Setting Error Enter date using correct format and reasonable values.
13.0 REPLACEMENT P ARTS AND ACCESSORIES If you need to order any spare parts and accessories, contact y our supplier or Adam Equipment. A partial list of such items is as follows- • Power Supply Module • Main Power cord • Replacement Battery • Stainless Steel Pan • In use cover • RS-232 option • Printer, etc.
WARRANTY INFORMATION Adam Equipment offers Limited Wa rranty (Parts and Labour) for the components failed due to defects in ma terials or workmanship. Warranty starts from the date of delivery. During the warranty period, should any repairs be necessary, the purchaser must inform its supplier or Adam Equi pment Company.
Manufacturer’s Declaration of Conformity This product has been manufactured in accordance with the harmonised European standards, following the provisions of the below stated directives: Electro Magnetic Compatibility Directive 89/336/EEC Low Voltage Directive 73/23/EEC Adam Equipment Co.
@Adam Equipment Company 2006 ADAM EQUIPMENT is an ISO 9001:2000 certified global organisation with more than 30 years experience in the producti on and sale of electronic weighing equipments. Products are sold through a world wide distributor network -supported from our com pany locations in the UK, USA, SOUTH AFRICA and CHINA.
Ein wichtiger Punkt beim Kauf des Geräts Adams CFW 600 (oder sogar vor seinem Kauf) ist das durchlesen seiner Bedienungsanleitung. Dies sollten wir wegen ein paar einfacher Gründe machen:
Wenn Sie Adams CFW 600 noch nicht gekauft haben, ist jetzt ein guter Moment, um sich mit den grundliegenden Daten des Produkts bekannt zu machen. Schauen Sie zuerst die ersten Seiten der Anleitung durch, die Sie oben finden. Dort finden Sie die wichtigsten technischen Daten für Adams CFW 600 - auf diese Weise prüfen Sie, ob das Gerät Ihren Wünschen entspricht. Wenn Sie tiefer in die Benutzeranleitung von Adams CFW 600 reinschauen, lernen Sie alle zugänglichen Produktfunktionen kennen, sowie erhalten Informationen über die Nutzung. Die Informationen, die Sie über Adams CFW 600 erhalten, werden Ihnen bestimmt bei der Kaufentscheidung helfen.
Wenn Sie aber schon Adams CFW 600 besitzen, und noch keine Gelegenheit dazu hatten, die Bedienungsanleitung zu lesen, sollten Sie es aufgrund der oben beschriebenen Gründe machen. Sie erfahren dann, ob Sie die zugänglichen Funktionen richtig genutzt haben, aber auch, ob Sie keine Fehler begangen haben, die den Nutzungszeitraum von Adams CFW 600 verkürzen könnten.
Jedoch ist die eine der wichtigsten Rollen, die eine Bedienungsanleitung für den Nutzer spielt, die Hilfe bei der Lösung von Problemen mit Adams CFW 600. Sie finden dort fast immer Troubleshooting, also die am häufigsten auftauchenden Störungen und Mängel bei Adams CFW 600 gemeinsam mit Hinweisen bezüglich der Arten ihrer Lösung. Sogar wenn es Ihnen nicht gelingen sollte das Problem alleine zu bewältigen, die Anleitung zeigt Ihnen die weitere Vorgehensweise – den Kontakt zur Kundenberatung oder dem naheliegenden Service.