Muon Liquid Handling System User Guide C. Johnson, S.P. Cottrell et al Version 0.2 - 1 -.
Contents 1 Getting Started 3 1.1 Layout of the liquid handling system 3 1.2 Layout of the in situ sample stick 4 1.3 Layout of the pump 4 2 General procedures 6 2.1 Sample loading 6 2.1.1 Loading Vessel 1 6 2.1.2 Loading Vessel 2 6 2.2 Evacuation of the system 6 2.
1 Getting S t arted This manual describes the liquid handling system and the in situ sample cell as use d on the DEVA instrument with the “RF” spect rometer, for either norm al muon spins relaxation or RF resonance experiments.
1.2 Layout of the in situ sample stick The liquid sample stick is designed to fit into the DEVA f low cryostat, details of which can be found in the DEVA manual. It consists of a shapol target cell 30 mm x 30 mm with a mylar window upon which may be mounted an RF coil.
Roughing/ Turbo switch Off/Run switch Pressure reading Turbo control panel Turbo Pump Rotary Pump Figure 2 The lay out of the fr ont panel o n the vacuum pump use d with the muon liqui d handlin g system.
2 General procedures 2.1 Sample loading At this point, the system will be open to the atm osphere, ensure that valve V2 is closed to prevent air cond ensation in the cold trap. 2.1.1 Loading Vessel 1 • Open the tap on the right-hand side of Vessel 1.
2.2.1 Removing air from Vessel 1 Once the system has been pressuris ed, tap A on Vessel 1 can be opened. It is possible to bubbled helium gas through the sample liqui d against the non-retu rn valve, V13. If this is required, first ensure th at valves V 5, V7 and V8 are closed.
2.3 Freeze-Pump-Thaw cycle Using this system it is possible to o perat e freeze-pump-thaw cycles on sam ple liquids held in any of the three vessels. Procedur es appropriate to ea ch vessel are outlined below. 2.3.1 Sample in Vessel 1 • All taps and valves should be closed.
2.3.3 Sample in Vessel 3 • All taps and valves should be closed. • Open valves V2 and V3 along with tap C on Vessel 3 and freeze the sample liquid by slowly raising a dewa r of liquid nitrogen around it. • When it has been determined that the liquid ha s completely frozen switch the pump to roughing mode and open valve V1.
2.4.2 Vessel 3 to sample cell The transfer of liquid from Vessel 3 to the sample cell is brought about by the difference in pressure between that in the Vessel and a vacuum in the cell. However, in order to control the transfer of liquid th e pressure difference shou ld be as small as possible but not so low that the liquid starts to boil.
2.4.4 Recovery of liquid in sample cell to Vessel 1 The liquid in the sample cell cannot be recovered to Vessel 1 using the liquid handling rig as presently configured. 2.4.5 Recovery of liquid in sample cell to Vessel 2 The liquid in the sample cell can be r ecovered to Vessel 2 using the following procedure.
• All the liquid may not return in a single pass and repea ting the evacuation procedure may be required. However, some liquid m ay be present in the system and care should be taken during evacuation that th e trap does not become blocked. The liquid may also be encouraged to leave the sample cell by using a small amount of gas.
3 Example Experiment The paramagnetic signal in liquid n-hexane is difficult to observe because the muonium polarisation is found to decay on a microsecond time scale, which is further reduced by the presence of dissolved oxygen.
0.0 0.5 1.0 1.5 2.0 2.5 3.0 11.0 11.5 12.0 12.5 13.0 13.5 Initial asymmetry (%) Time ( µ s) 0.0 0.5 1.0 1.5 2.0 2.5 3.0 11.0 11.5 12.0 12.5 13.0 Initial asymmetry (%) Time ( µ s) 1.
Ein wichtiger Punkt beim Kauf des Geräts Boca Research None (oder sogar vor seinem Kauf) ist das durchlesen seiner Bedienungsanleitung. Dies sollten wir wegen ein paar einfacher Gründe machen:
Wenn Sie Boca Research None noch nicht gekauft haben, ist jetzt ein guter Moment, um sich mit den grundliegenden Daten des Produkts bekannt zu machen. Schauen Sie zuerst die ersten Seiten der Anleitung durch, die Sie oben finden. Dort finden Sie die wichtigsten technischen Daten für Boca Research None - auf diese Weise prüfen Sie, ob das Gerät Ihren Wünschen entspricht. Wenn Sie tiefer in die Benutzeranleitung von Boca Research None reinschauen, lernen Sie alle zugänglichen Produktfunktionen kennen, sowie erhalten Informationen über die Nutzung. Die Informationen, die Sie über Boca Research None erhalten, werden Ihnen bestimmt bei der Kaufentscheidung helfen.
Wenn Sie aber schon Boca Research None besitzen, und noch keine Gelegenheit dazu hatten, die Bedienungsanleitung zu lesen, sollten Sie es aufgrund der oben beschriebenen Gründe machen. Sie erfahren dann, ob Sie die zugänglichen Funktionen richtig genutzt haben, aber auch, ob Sie keine Fehler begangen haben, die den Nutzungszeitraum von Boca Research None verkürzen könnten.
Jedoch ist die eine der wichtigsten Rollen, die eine Bedienungsanleitung für den Nutzer spielt, die Hilfe bei der Lösung von Problemen mit Boca Research None. Sie finden dort fast immer Troubleshooting, also die am häufigsten auftauchenden Störungen und Mängel bei Boca Research None gemeinsam mit Hinweisen bezüglich der Arten ihrer Lösung. Sogar wenn es Ihnen nicht gelingen sollte das Problem alleine zu bewältigen, die Anleitung zeigt Ihnen die weitere Vorgehensweise – den Kontakt zur Kundenberatung oder dem naheliegenden Service.