Benutzeranleitung / Produktwartung 9046 des Produzenten Pressure Systems
Zur Seite of 72
Model 9046 Ethernet Intelligent Temperature Scanner User’s Manual 2 nd Edition August 2008 NetScanner™ System www. PressureSystems .com.
© This User’s Manual is a copyright product of Pressure Systems, Inc., 2008 Permission is hereby granted to make copies and distribute verbatim copies of this manual, provided the copyright notice and this permission notice are preserved on all copies.
Pressure Systems, Inc. Model 9046 User’s Manual Page i www.PressureSystems.com Table of Contents 1.0 Introduction ......................................................................................................... 1 1.1 General ...............
Pressure Systems, Inc. Model 9046 User’s Manual Page ii www.PressureSystems.com Our Company Pressure Systems Incorporated, (PSI) develops, manufactures, and services level and pressure measuring instruments where the highest level of traceable accuracy is required for aerospace, industrial, municipal, and environmental applic ations.
Pressure Systems, Inc. Model 9046 User’s Manual Page iii www.PressureSystems.com PSI will return warranty items prepaid via UPS GROUND. If the customer desires another method of return shipment, PSI will prepay and add the shipping charges to the repair bill.
Pressure Systems, Inc. Model 9046 User’s Manual Page iv www.PressureSystems.com This page left intentionally blank..
Pressure Systems, Inc. Model 9046 User’s Manual Page 1 www.PressureSystems.com 1.0 Introduction 1.1 General 1.1.1 Introduction The Model 9046 Intelligent Temperature Scanner is a 16 channel modular scanning system designed to operate in harsh environments, such as turbine engine test cells.
Pressure Systems, Inc. Model 9046 User’s Manual Page 2 www.PressureSystems.com 1.1.2 Physi cal Description Figure 1-1: Exploded View of Module The Model 9046 temperature scanner is housed in a rugged machined aluminum case. The case is splash proof and finished to allow installation in a tur bine engine test cell environment.
Pressure Systems, Inc. Model 9046 User’s Manual Page 3 www.PressureSystems.com For operation, the Model 9046 only requires 18 to 36 VDC (i=200mA), and the appropriate Ethernet interface. 1.1.3 Theory of Operation The Model 9046 is a 32 channel analog multiplexer, controlled by a microprocessor.
Pressure Systems, Inc. Model 9046 User’s Manual Page 4 www.PressureSystems.com Precision T e mperature Device Jumper sockets for thermocouple/resistance mode selection (jumper = thermocouple) Socket.
Pressure Systems, Inc. Model 9046 User’s Manual Page 5 www.PressureSystems.com 2.1 Equipment Required The following items will be needed in order to set-up a Temperature Scanning Module: • Model 9046 Temperature Scanning Module • PSI powered network solution such as 90DB, 90DC, 98RK or 9FC.
Pressure Systems, Inc. Model 9046 User’s Manual Page 6 www.PressureSystems.com 2.2 Computer Preparation The PC must have its IP address and subnet mask set to be compatible with the addressing existing in the module. For a new module this will be the factory default.
Pressure Systems, Inc. Model 9046 User’s Manual Page 7 www.PressureSystems.com Figure 2-2: TCP/IP Properties Dialog Box Note Be sure to record the subnet mask addressing settings prior to changing them to the values shown to facilitate restoring the computer to its previous network configuration.
Pressure Systems, Inc. Model 9046 User’s Manual Page 8 www.PressureSystems.com 2.3 Module Connection The module must now be connected to the PC via the 9IFC interface.
Pressure Systems, Inc. Model 9046 User’s Manual Page 9 www.PressureSystems.com After supplying power to the module, observe the front panel for the following indications: • PWR LED should remain ON. If this LED is not ON, all other LED’s will also likely be OFF.
Pressure Systems, Inc. Model 9046 User’s Manual Page 10 www.PressureSystems.com 2.4 IP Address Assignment The module IP address assignment can only be changed if your computer is able to communicate with the module: 1. Use the PSI program NUSS to determine the current module IP address.
Pressure Systems, Inc. Model 9046 User’s Manual Page 11 www.PressureSystems.com 3. Connect to the module. (When a connection has been established, the top of the module icon will turn yellow.) Follow these steps: a. Left-Click on the module’s icon in the Node Map.
Pressure Systems, Inc. Model 9046 User’s Manual Page 12 www.PressureSystems.com 4. Open the Network Options (Single Module) Dialog Box: a. Right-click on the module’s icon ( in the Description Box). b. Click on “Configure”. c. Click on “Network Options”.
Pressure Systems, Inc. Model 9046 User’s Manual Page 13 www.PressureSystems.com 5. Set the subnet mask and IP address if they are not correct: ● IP Address - use the address cal led for by the system. ● Set the subnet mask to 255.255.255.0 ● Verify that the TCP Connect Port (Listen Port) is set to 9000 .
Pressure Systems, Inc. Model 9046 User’s Manual Page 14 www.PressureSystems.com 2.5 Set Sensor Type 1. Confirm that the correct SCM is installed and jumpered correctly: a. For Non-Universal SCMs: Confirm that each of the 16 hardware channels contains the correct SCM for the intended application.
Pressure Systems, Inc. Model 9046 User’s Manual Page 15 www.PressureSystems.com Table 1: Sensor Type Values.
Pressure Systems, Inc. Model 9046 User’s Manual Page 16 www.PressureSystems.com 2.6 Read Sensor Type Each channel type can be read using a command of the form: uxxx51 . where xxx = the channel # in HEX . Example: • u00a51 will return 18, indicating the channel is configured for a type T thermocouple.
Pressure Systems, Inc. Model 9046 User’s Manual Page 17 www.PressureSystems.com Command: c05 0002 Action: Enable thermocouple reference temperature alarms in autonomous data stream.
Pressure Systems, Inc. Model 9046 User’s Manual Page 18 www.PressureSystems.com 2.8 Set Temperature Units Set temperature units for all temperature channel s. The ‘v’ command is used with array index aa = 11 (hex) and new coefficient index = 0b (hex).
Pressure Systems, Inc. Model 9046 User’s Manual Page 19 www.PressureSystems.com The following example sets channel 1 EU filter parameter to 10: Command: v00152 10 Normal : A Response: Error: N08.
Pressure Systems, Inc. Model 9046 User’s Manual Page 20 www.PressureSystems.com 2.11 Read EU Filter Parameter The EU filter parameter can be read for an individual channel using the ‘u’ command with coefficient index number 52 hex. The following ex ample reads the EU filter parameter for channel 1.
Pressure Systems, Inc. Model 9046 User’s Manual Page 21 www.PressureSystems.com 3.0 Calibration Calibration consists of offset correction. Implement offset correction (automatically calc ulate and set offset coefficient) on a per channel basis using the ‘h’ command.
Pressure Systems, Inc. Model 9046 User’s Manual Page 22 www.PressureSystems.com Examples: Correct the offset in channel 1 with channel 1 at 100 degree C : Command: h0001 100.0 Response: 0.294167 (offset coefficient returned) Manually set channel 1offset coefficient to 0.
Pressure Systems, Inc. Model 9046 User’s Manual Page 23 www.PressureSystems.com.
Pressure Systems, Inc. Model 9046 User’s Manual Page 24 www.PressureSystems.com 4.0 Firmware 4.1 Description of Operation Table 1 shows the different sensor ty pes available along with the gain settings and engineering units for each type. Setting channel sensor type automatically causes the gain for that channel to be set according to Table1.
Pressure Systems, Inc. Model 9046 User’s Manual Page 25 www.PressureSystems.com Voltage If a channel is configured for straight voltage then no additio nal EU conversion is needed. Va is returned as transducer voltage. Vb is returned as cold junction temperature.
Pressure Systems, Inc. Model 9046 User’s Manual Page 26 www.PressureSystems.com 4.2 Updating the Firmware Special precautions must be taken when updating NetScanner™ firmware. If care is not taken, the module program memory can be corrupted rendering the module inoperative until the program memory chip is physically replaced.
Pressure Systems, Inc. Model 9046 User’s Manual Page 27 www.PressureSystems.com This page intentionally blank..
Pressure Systems, Inc. Model 9046 User’s Manual Page 28 www.PressureSystems.com 5.0 Command Set Command Quick Reference Type Operation Code Command Function A No Operation B Reset V Read Primary Transducer Volts a Read Primary Transducer A/D Counts b Read Primary Transducer E.
Pressure Systems, Inc. Model 9046 User’s Manual Page 29 www.PressureSystems.com psireboot Reboot Specified Module Commands psirarp Change Specified Module’s IP Address Resolution Method & Reboot Note The commands in this section are quite detailed and involved to anyone wishing to only use the 9046 module.
Pressure Systems, Inc. Model 9046 User’s Manual Page 30 www.PressureSystems.com 5.1 Summary Information About All Commands The command descriptions that follow are listed in the same order as the Quick Reference table above. All TCP/IP commands are listed first (i n alphabetical order: first by upper-case letters, then by lower-case letters).
Pressure Systems, Inc. Model 9046 User’s Manual Page 31 www.PressureSystems.com "Read Data" command, the first character will not contain an A or N character, and e ach datum may be in a format specified by a parameter of the command itself, or be implied by the particular command operation code.
Pressure Systems, Inc. Model 9046 User’s Manual Page 32 www.PressureSystems.com The individual description of each command follows immediately. Each command has a functional title (and operation code), followed by a s hort text description.
Pressure Systems, Inc. Model 9046 User’s Manual Page 33 www.PressureSystems.com 5.2 No Operation (Command ’A’) Purpose : This command has no internal module effect. It is used as a simple method to verify proper communications to the Model 9046 module.
Pressure Systems, Inc. Model 9046 User’s Manual Page 34 www.PressureSystems.com 5.3 Reset (Command ’B’) Purpose : Instructs the module to reset internal operating parameters, and to set all internal control variables to their default "reset" state (see description below).
Pressure Systems, Inc. Model 9046 User’s Manual Page 35 www.PressureSystems.com 5.4 Read Primary Transdu cer Volts (Command ’V’) Purpose : Returns the most recently acquired primary transducer (i.e., temperature/ resistance) raw data for the specified channels.
Pressure Systems, Inc. Model 9046 User’s Manual Page 36 www.PressureSystems.com 8 single binary float to single binary float (little endian: lsb first) 4 Example : Command module to return decimal v.
Pressure Systems, Inc. Model 9046 User’s Manual Page 37 www.PressureSystems.com 5.5 Read Primary Transdu cer A/D Counts (Command ’a’) Purpose : Returns the most recently acquired primary transducer (i.e., temperature/ resistance) raw data for the specified channels.
Pressure Systems, Inc. Model 9046 User’s Manual Page 38 www.PressureSystems.com 7 single binary float to single binary float (big endian: msb first) 4 8 single binary float to single binary float (l.
Pressure Systems, Inc. Model 9046 User’s Manual Page 39 www.PressureSystems.com 5.6 Read Primary Transducer E. U. Data at High Speed (Command ’b’) Purpose : Returns the most recent Primary Transducer E. U. data from all channels of the module as fast as possible.
Pressure Systems, Inc. Model 9046 User’s Manual Page 40 www.PressureSystems.com Note NUSS does not use this command normally. You may execute it manually via the ’Execute Commands’ or ’Execute Script’ functions of NUSS - though this should be avoided if NUSS is currently operating the module via other NUSS functions.
Pressure Systems, Inc. Model 9046 User’s Manual Page 41 www.PressureSystems.com 5.7 Define/Control Autonomous Host Streams (Command ’c’) Purpose : Defines and controls the autonomous delivery of up to three concurrent high-speed autonomous data streams to the host computer.
Pressure Systems, Inc. Model 9046 User’s Manual Page 42 www.PressureSystems.com Description : The firmware of any module, once fully initialized after Reset (or power-on), continuously scans and converts data for all Prim ary (Temperature/Resistance) Transducers at the highest possible speed.
Pressure Systems, Inc. Model 9046 User’s Manual Page 43 www.PressureSystems.com Note When using a module’s hardware trigger inputs to synchronize data stream outputs, the frequency of the trigger source should be no more than 200Hz even if the requested output is 100Hz or less.
Pressure Systems, Inc. Model 9046 User’s Manual Page 44 www.PressureSystems.com 5.7.1 Command ’c’- Sub-command Index 00: Configure a Host Delivery Stream This sub-command is used to configure the principle parameters of each of the three possible concurrent host delivery streams, one at a time.
Pressure Systems, Inc. Model 9046 User’s Manual Page 45 www.PressureSystems.com When the hardware trigger is used to synchronize stream scans ( sync = 0), it is assumed that the user would prefer to also synchronize internal data acquisition cycle.
Pressure Systems, Inc. Model 9046 User’s Manual Page 46 www.PressureSystems.com The ‘ f ’ parameter identifies the format of each selected acquired datum in each data group of the stream scan, and is a single numeric digit. Valid format codes are listed in the following table: f converts each internal response datum value from: max.
Pressure Systems, Inc. Model 9046 User’s Manual Page 47 www.PressureSystems.com " A " " A " " A " Example 2: To similarly acquire data at "relative" rates (1,.
Pressure Systems, Inc. Model 9046 User’s Manual Page 48 www.PressureSystems.com Note NUSS automatically creates a default Run State 0 for every module that uses its Run function.
Pressure Systems, Inc. Model 9046 User’s Manual Page 49 www.PressureSystems.com 5.7.2 Command ’c’- Sub-command Index 01: Start Stream ( s ) This sub-command is used to start the delivery of any previously configured host stream in a module.
Pressure Systems, Inc. Model 9046 User’s Manual Page 50 www.PressureSystems.com module. Each of the three possible host streams will report its own unique sequence n umber. Note that if a previously stopped data stream is restarted, the returned se quence numbers will resume with the next number at the point of the stream’s termination.
Pressure Systems, Inc. Model 9046 User’s Manual Page 51 www.PressureSystems.com 5.7.3 Command ’c’- Sub-command Index 02: Stop Stream ( s ) This sub-command is used to stop (or temporar ily suspend) the delivery of any previously started host stream in a module, one at a time or all together, whether the stream was unbounded or bounded.
Pressure Systems, Inc. Model 9046 User’s Manual Page 52 www.PressureSystems.com Example: Stop all the streams configured in the previous example: " c 02 0 " Read response: " A " .
Pressure Systems, Inc. Model 9046 User’s Manual Page 53 www.PressureSystems.com 5.7.4 Command ’c’- Sub-command Index 03: Clear Stream ( s ) This sub-command is used to "undefine" any previously configured host stream in a module, one at a time, or all together.
Pressure Systems, Inc. Model 9046 User’s Manual Page 54 www.PressureSystems.com Note NUSS uses this sub-command to undefined data specific streams when its Run State Editor form is used to define "custom" stream configurations (called Run States 1-9) and all three streams are not to be generated concurrently.
Pressure Systems, Inc. Model 9046 User’s Manual Page 55 www.PressureSystems.com 5.7.5 Command ’c’ - Sub-command Index 04: Return Stream Information This sub-command returns current stream configuration information in its response.
Pressure Systems, Inc. Model 9046 User’s Manual Page 56 www.PressureSystems.com Description : This sub-command returns the current configuration information for a particular stre am.
Pressure Systems, Inc. Model 9046 User’s Manual Page 57 www.PressureSystems.com 5.7.6 Command ’c’ - Sub-command Index 05: Select Prefix/Data Groups in a Stream This sub-command sets additional options that cause an already defined stream to deliver a specific prefix and/or other groups of data to the host.
Pressure Systems, Inc. Model 9046 User’s Manual Page 58 www.PressureSystems.com Note Selecting too many data groups will reduce module performance. bbbb (hex) data selected for inclusion in each str.
Pressure Systems, Inc. Model 9046 User’s Manual Page 59 www.PressureSystems.com Bit# 15 14 13 12 11 10 9 8 7 6 5 4 3 2 1 0 Chan # 16 15 14 13 12 11 10 9 8 7 6 5 4 3 2 1 Binary 1 0 0 0 0 0 0 0 0 0 0 0 0 0 0 1 Hex 8 0 0 1 The above example indicates that Channels 1 and 16 are "in alarm".
Pressure Systems, Inc. Model 9046 User’s Manual Page 60 www.PressureSystems.com 5.7.7 Command ’c’ - Sub-command Index 06: Select Protocol For Stream Delivery Command “ c 06 st pro [ remport [ ipaddr ]] ” ‘c’ is the operation code. ‘ 06’ is the sub-command index ( ii ) for Select Protocol .
Pressure Systems, Inc. Model 9046 User’s Manual Page 61 www.PressureSystems.com The optional remport parameter may be any value in the range 1024 to 65535.
Pressure Systems, Inc. Model 9046 User’s Manual Page 62 www.PressureSystems.com Note NUSS does not normally use this command since it generates and processes TCP/IP streams exclusively with a unique Socket allocated for each "connected" module.
Pressure Systems, Inc. Model 9046 User’s Manual Page 63 www.PressureSystems.com This page intentionally left blank..
Pressure Systems, Inc. Model 9046 User’s Manual Page 64 www.PressureSystems.com Appendix A Cable Diagrams Model 9046 Ethernet Interface Cable P/N 9080.
Pressure Systems, Inc. Model 9046 User’s Manual Page 65 www.PressureSystems.com Model 9046 Ethernet Interface Cable P/N 9080.
Headquarters/Factory : Pressure Sy stems, Inc. 34 Research Drive Hamtpton, VA 23666 USA Phone: (757) 865-1243 Toll Free: (800) 328-3665 Fax: (757) 865-8744 E-mail: sales@Pres sureSystems. com European Office: PSI, Ltd. 124, Victoria Road Farnborough, Hants GU14 7PW United Kingdom Phone: +44 1252 510000 Fax: +44 1252 510099 E-mail: PSI@WestonAero.
Ein wichtiger Punkt beim Kauf des Geräts Pressure Systems 9046 (oder sogar vor seinem Kauf) ist das durchlesen seiner Bedienungsanleitung. Dies sollten wir wegen ein paar einfacher Gründe machen:
Wenn Sie Pressure Systems 9046 noch nicht gekauft haben, ist jetzt ein guter Moment, um sich mit den grundliegenden Daten des Produkts bekannt zu machen. Schauen Sie zuerst die ersten Seiten der Anleitung durch, die Sie oben finden. Dort finden Sie die wichtigsten technischen Daten für Pressure Systems 9046 - auf diese Weise prüfen Sie, ob das Gerät Ihren Wünschen entspricht. Wenn Sie tiefer in die Benutzeranleitung von Pressure Systems 9046 reinschauen, lernen Sie alle zugänglichen Produktfunktionen kennen, sowie erhalten Informationen über die Nutzung. Die Informationen, die Sie über Pressure Systems 9046 erhalten, werden Ihnen bestimmt bei der Kaufentscheidung helfen.
Wenn Sie aber schon Pressure Systems 9046 besitzen, und noch keine Gelegenheit dazu hatten, die Bedienungsanleitung zu lesen, sollten Sie es aufgrund der oben beschriebenen Gründe machen. Sie erfahren dann, ob Sie die zugänglichen Funktionen richtig genutzt haben, aber auch, ob Sie keine Fehler begangen haben, die den Nutzungszeitraum von Pressure Systems 9046 verkürzen könnten.
Jedoch ist die eine der wichtigsten Rollen, die eine Bedienungsanleitung für den Nutzer spielt, die Hilfe bei der Lösung von Problemen mit Pressure Systems 9046. Sie finden dort fast immer Troubleshooting, also die am häufigsten auftauchenden Störungen und Mängel bei Pressure Systems 9046 gemeinsam mit Hinweisen bezüglich der Arten ihrer Lösung. Sogar wenn es Ihnen nicht gelingen sollte das Problem alleine zu bewältigen, die Anleitung zeigt Ihnen die weitere Vorgehensweise – den Kontakt zur Kundenberatung oder dem naheliegenden Service.