www .omega.com e-mail: info@omega.com omega.com ® OMEGA ® U ser ’ s Guide TX82B Panel Mounted Current Loop Indicator Shop on line at.
Ser vicing Nor th America: USA: One Omega Drive, P .O. Box 4047 ISO 9001 Certified Stamford CT 06907-0047 TEL: (203) 359-1660 F AX: (203) 359-7700 e-mail: info@omega.com Canada: 976 Bergar Laval (Quebec) H7L 5A1 TEL: (514) 856-6928 F AX: (514) 856-6886 e-mail: info@omega.
T ABLE OF CONTENTS 1.0 GENERAL INFORMA TION . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1 2.0 SPECIFICA TIONS . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
UNP ACKING INSTRUCTIONS Remove the Packing List and verify that you have received all equipment, including the following: QTY DESCRIPTION 1 Current Loop Meter Indicator with all applicable connnectors attached.
1 1.0 GENERAL INFORMA TION Model TX82B two-wire current-loop indicator accepts 1-5 mA, 4-20 mA, or 10-50 mA process signals and digitally displays the process variable in percentage or engineering units of such measurements as pressure, flow , temperature and level.
2 2.0 SPECIFICA TIONS 2.1 INPUT Current 1-5 mA, 4-20 mA or 10-50 mA Protection 200 mA max forward and 1000 mA max reverse V oltage Drop 2.5 V max forward and 1.
3 2.4 DISPLA Y T ype 7-segment LCD Color Black digits with white background Symbols -1.8.8.8.0.0, 12.7 mm (0.5 in) height Polarity Minus sign Overrange Three least-significant digits blank Extra digit.
4 3.0 MECHANICAL ASSEMBL Y AND INST ALLATION 3.1 SAFETY CONSIDERA TIONS T o ensure safe operation, follow the guidelines below: VISUAL INSPECTION: Do not attempt to operate the instrument if damage is found. SIGNAL WIRING: Insert the proper plus and minus signal wires into the plug-in screw-clamp connector terminals marked plus and minus.
3.3 INST ALLA TION/REMOVAL The TX82B is housed in a 1/8 DIN case. The electronic circuitry can be installed or removed from the front and is attached to the case with two M4 screws through the rear panel. Panel Mounting 1. Remove the two thumbnuts on the rear of the case.
6 Figure 3-2. DIN Case Dimensions 4.0 SIGNAL INPUT CONNECTIONS (P3) The signal input connections for all meters are made at connector P3 as follows (see Figure 3-1): P3 Connection Signal 1 Signal Hi 2 Signal Lo 3 No Connection 4.15 [105.4] SLEEVE 3.50 [88.
7 5.0 CONFIGURA TION The standard TX82B meter is factory-configured for an input of 4-20 mA to display 00.0 to 100.0. Field configuration for input current range, decimal point location, dummy right-h.
8 Figure 5-1 Main Board Jumper Locations A AH J F C D E B G BJ E H G R4 U3 U2 P3 VR1 S1 S2 S2 SP AN ADJ. ZERO ADJ. VR2 R30 S1 F C D.
9 6.0 CUSTOMER CONFIGURA TION AND CALIBRA TION Use this procedure to determine the configuration of the TX82B customized setup. The procedure is general; customers can specify any two current inputs and their corresponding digital readings. Pin-groups are shown in Figure 5-1.
10 6.2 CONFIGURA TION PROCEDURES 1. Remove all push-on jumpers. 2. For an input current range of 1-5 mA, no jumper is required. For 4-20 mA input, install a push-on jumper at S1-H. For 10-50 mA input, install a push-on jumper at S1-J. 3. If N2 (Section 6.
11 Figure 7-1 Wiring Diagram for TX82B Usage for FM (Factory Mutual) SAFE AREA APPARATUS 2 EARTHED BUS BAR + - 3 1 4 INTER- CONNECTING CABLES SHUNT ZENER DIODE SAFETY BARRIER SAFE AREA HAZARDOUS AREA .
.
.
.
WARRANTY / DISCLAIMER OMEGA ENGINEERING, INC. warrants this unit to be free of defects in materials and workmanship for a period of 13 months from date of purchase. OMEGA Warranty adds an additional one (1) month grace period to the normal one (1) year product warranty to cover handling and shipping time.
M1060/0303 10579ML-99 J Where Do I Find Everything I Need for Process Measurement and Contr ol? OM EGA…Of Course! Shop on line at www .omega.com TEMPERA TURE Thermocouple, RTD & Thermist.
Ein wichtiger Punkt beim Kauf des Geräts Omega TX82B (oder sogar vor seinem Kauf) ist das durchlesen seiner Bedienungsanleitung. Dies sollten wir wegen ein paar einfacher Gründe machen:
Wenn Sie Omega TX82B noch nicht gekauft haben, ist jetzt ein guter Moment, um sich mit den grundliegenden Daten des Produkts bekannt zu machen. Schauen Sie zuerst die ersten Seiten der Anleitung durch, die Sie oben finden. Dort finden Sie die wichtigsten technischen Daten für Omega TX82B - auf diese Weise prüfen Sie, ob das Gerät Ihren Wünschen entspricht. Wenn Sie tiefer in die Benutzeranleitung von Omega TX82B reinschauen, lernen Sie alle zugänglichen Produktfunktionen kennen, sowie erhalten Informationen über die Nutzung. Die Informationen, die Sie über Omega TX82B erhalten, werden Ihnen bestimmt bei der Kaufentscheidung helfen.
Wenn Sie aber schon Omega TX82B besitzen, und noch keine Gelegenheit dazu hatten, die Bedienungsanleitung zu lesen, sollten Sie es aufgrund der oben beschriebenen Gründe machen. Sie erfahren dann, ob Sie die zugänglichen Funktionen richtig genutzt haben, aber auch, ob Sie keine Fehler begangen haben, die den Nutzungszeitraum von Omega TX82B verkürzen könnten.
Jedoch ist die eine der wichtigsten Rollen, die eine Bedienungsanleitung für den Nutzer spielt, die Hilfe bei der Lösung von Problemen mit Omega TX82B. Sie finden dort fast immer Troubleshooting, also die am häufigsten auftauchenden Störungen und Mängel bei Omega TX82B gemeinsam mit Hinweisen bezüglich der Arten ihrer Lösung. Sogar wenn es Ihnen nicht gelingen sollte das Problem alleine zu bewältigen, die Anleitung zeigt Ihnen die weitere Vorgehensweise – den Kontakt zur Kundenberatung oder dem naheliegenden Service.