Benutzeranleitung / Produktwartung HH505 des Produzenten Omega Engineering
Zur Seite of 15
OMEGA HH505 Digital Thermometer.
OMEGAnet SM On-Line Service http://www.omega.com Internet e-mail info@omega.com Servicing Nor th America: USA: ISO 9001 Certified Canada: One Omega Drive, Box 4047 976 Bergar Stamford, CT 06907-0047 Laval (Quebec) H7L5A1 Tel: (203) 359-1660 Tel: (514) 856-6928 FAX: (203)359-7700 FAX: (514) 856-6886 e-mail: info@omega.
1 The symbol on the instrument indicates that the operator must refer to an explanation in this manual. WARNING To avoid electrical shock, do not use this instrument when working voltages at the measurement surface over 24V AC or DC. WARNING To avoid damage or burns, do not make temperature measurement in microwave ovens.
2 SPECIFICATIONS ELECTRICAL Temperature Scale: Celsius or Fahrenheit user-selectable. Measurement Range: Pt385(100 W ) -200°C to 800°C, (-328°F to 1472°F) Pt3926(100 W ) -200°C to 630°C, (-328°F to 1166°F) Resolution: 0.
3 ENVIRONMENTAL Ambient Operating Ranges: 0°C to 50°C (32°F to 122°F) <80% R.H. Storage Temperature: -20°C to 60°C (-4°F to 140°F) <70% R.H. GENERAL Display: 4½ digit liquid crystal display (LCD) with maximum reading of 19999. Overload: "----.
OPERATING INSTRUCTIONS Power Switch The key turns the thermometer on or off. In the data SET mode, can not power off, must leave data SET mode then power off. °C/°F Selecting the Temperature Scale (Main display) Reading the main displayed in either degrees Celsius(°C) or degrees Fahrenheit(°F).
5 5 6 Pt385/Pt3926 T1 Input RTD Probe Select The Pt385/Pt3926 key switch the T1 input selects the Pt385 or Pt3926 RTD probe as input. When the thermometer is turned on, it is set to the probe selected that was in use when the thermometer was last turned off.
6 REL Relative mode Pressing REL key to enter the Relative mode, zero the display, and store the displayed Reading as a reference value and annunciator REL is displayed. Present temperature reading displayed on second display. Press REL key again to exit the relative mode.
7 9 SET mode (Relative value set, Time set and Hi/Lo Limits value set) 1. Press SET key to enter Relative valuse SET mode (Press ENTER key can escape relative valuse set mode), REL set mode.
8 10 11 12 Hi Limit value Recall display mode Press Hi RECALL key to display the Hi set Limit value on second display. Press Hi RECALL key again to exit this mode. Lo Limit value Recall display mode Press Lo RECALL key to display the Hi set Limit value on second display.
9 WARNING To avoid possible electrical shock, disconnect the thermocouple connectors from the rmometer before removing the cover. Battery Replacement Power is supplied by a 9 volt "transistor" battery. The " " appears on the LCD display when replacement is needed.
10 TEMPERATURE VS RESISTANCE TABLE(ITS90) °C Pt385 Pt3926 -200°C 18.521 W 16.996 W -100°C 60.256 W 59.479 W 0°C 100.000 W 100.000 W 100°C 138.505 W 139.272 W 200°C 175.856 W 177.362 W 300°C 212.052 W 214.275 W 400°C 247.092 W 250.018 W 500°C 280.
1 HH 504 / HH 505 CA LIBRA TION PROCEDURE Note: The following calibration procedure s hould perform only by qualified technicians who have access to the items as fol lowing. Equipment: The class of calibrator ha d better 10 times g reater than the measur ed meter.
WARRANTY / DISCLAIMER OMEGA ENGINEERING, INC . warrants this unit to be free of defects in materials and workmanship for a period of 13 months from date of purchase. OMEGA W arranty adds an additional one (1) month grace period to the normal one (1) year product warranty to cover handling and shipping time.
Where Do I Find Everything I Need for Process Measurement and Control? OMEGA...Of Course! TEMPERATURE þ Thermocouple, RTD & Thermistor Probes, Connectors, Panels & Assemblies þ Wire: Thermoc.
Ein wichtiger Punkt beim Kauf des Geräts Omega Engineering HH505 (oder sogar vor seinem Kauf) ist das durchlesen seiner Bedienungsanleitung. Dies sollten wir wegen ein paar einfacher Gründe machen:
Wenn Sie Omega Engineering HH505 noch nicht gekauft haben, ist jetzt ein guter Moment, um sich mit den grundliegenden Daten des Produkts bekannt zu machen. Schauen Sie zuerst die ersten Seiten der Anleitung durch, die Sie oben finden. Dort finden Sie die wichtigsten technischen Daten für Omega Engineering HH505 - auf diese Weise prüfen Sie, ob das Gerät Ihren Wünschen entspricht. Wenn Sie tiefer in die Benutzeranleitung von Omega Engineering HH505 reinschauen, lernen Sie alle zugänglichen Produktfunktionen kennen, sowie erhalten Informationen über die Nutzung. Die Informationen, die Sie über Omega Engineering HH505 erhalten, werden Ihnen bestimmt bei der Kaufentscheidung helfen.
Wenn Sie aber schon Omega Engineering HH505 besitzen, und noch keine Gelegenheit dazu hatten, die Bedienungsanleitung zu lesen, sollten Sie es aufgrund der oben beschriebenen Gründe machen. Sie erfahren dann, ob Sie die zugänglichen Funktionen richtig genutzt haben, aber auch, ob Sie keine Fehler begangen haben, die den Nutzungszeitraum von Omega Engineering HH505 verkürzen könnten.
Jedoch ist die eine der wichtigsten Rollen, die eine Bedienungsanleitung für den Nutzer spielt, die Hilfe bei der Lösung von Problemen mit Omega Engineering HH505. Sie finden dort fast immer Troubleshooting, also die am häufigsten auftauchenden Störungen und Mängel bei Omega Engineering HH505 gemeinsam mit Hinweisen bezüglich der Arten ihrer Lösung. Sogar wenn es Ihnen nicht gelingen sollte das Problem alleine zu bewältigen, die Anleitung zeigt Ihnen die weitere Vorgehensweise – den Kontakt zur Kundenberatung oder dem naheliegenden Service.