© 1998–2003 National Instruments Corp. All rights reserved. LabVIEW™, National Instru ments™, NI™, ni.com™, NI-DAQ ™, and SCXI™ are t rademarks of National Instrument s Corporation. Product and com pany names mentioned herein are trademarks or trade names of their respective companies.
SCXI-1321 T erminal Block Installation Guide 2 ni.com italic Italic text denotes variables, emphasis, a cross reference, or an int roduction to a key concept. This font also denote s te xt that is a placeholder for a word or v alue that you must supply .
© National Instruments Cor poration 3 SCXI-1321 T erminal Block Installa tion Guide Connecting the Signals Note Refer to the Read Me First: Safety and Radio-Frequency Interference document before removing equipment covers, or connecting or disconnecting any signal wires.
SCXI-1321 T erminal Block Installation Guide 4 ni.com Figure 1. SCXI-132 1 Parts Locator Diagram 1 Strain-Relief Bar 2 Strain-Relief Screws 3 Safety-Ground Lug 4 Mating Connector 5 Thumbscrew 6 Top-Co.
© National Instruments Cor poration 5 SCXI-1321 T erminal Block Installa tion Guide Figure 2. SCXI-1321 Circuit Parts Locator Diagram Installing the T erminal Block To connect the terminal block to the SCXI mo dule front connector, complete the follow ing steps: 1.
SCXI-1321 T erminal Block Installation Guide 6 ni.com Note T o minimize the temperature gradient inside the termin al block, move the SCXI chassis aw ay from any extr eme temperature dif ferential. Specifications All specifications are typical at 25 °C unless otherwise specified .
© National Instruments Cor poration 7 SCXI-1321 T erminal Block Installa tion Guide Environmental Operating temperature........................ .... 0 to 50 °C Storage temperature ............. ................. . –20 to 70 °C Humidity ..........
SCXI-1321 T erminal Block Installation Guide 8 ni.com Note Refer to the Declaration of Conformity (DoC) for this product fo r any additional regulatory compliance in formation. T o obtain th e DoC for this p roduct, click Declarations of Conformity Information at ni.
© National Instruments Cor poration 9 SCXI-1321 T erminal Block Installa tion Guide The factory default v alue of all the nulling resistors on the terminal blo ck is 39 k Ω . Note These resistors are socketed for easy re placement. These sockets best f it a 1/4 W resistor lead size.
SCXI-1321 T erminal Block Installation Guide 10 ni.com Assuming a strain-gauge range with a gauge factor of GF = 2 and a quarter-bridge conf iguration, this range corresponds to ±1,498 µε , as given by the strain formula for a quarter-bridge strain-gauge configuration: where Figure 3.
© National Instruments Cor poration 11 SCXI-1321 T erminal Bloc k Installa tion Guide Figure 4. Half-Bridge Nulling Circuit CH+ CH+ CH– EX+ EX– + – + – R 1 R 2 4.
SCXI-1321 T erminal Block Installation Guide 12 ni.com Figure 5. Full-Bridge Nulling Cir cuit Shunt Calibration Shunt calibration prov ides a method of adjusti ng for gain error to improve accuracy.
© National Instruments Cor poration 13 SCXI-1321 T erminal Bloc k Installa tion Guide z is the module slot of the SCXI-1121 w is the channel o f the module that you want to engage th e shunt and take.
SCXI-1321 T erminal Block Installation Guide 14 ni.com gauge with a gauge factor of GF = 2, V exc = 3.3 V , and R = 120 Ω prod uces the following: V change = 0.3321 mV Replacing the strained voltage by V change in the quarter-bridge strain equation produces an equ i valent 199 µε of change.
© National Instruments Cor poration 15 SCXI-1321 T erminal Bloc k Installa tion Guide Y ou can read a temperature at the begi nning of the test and use that value with the data that follows. Using this method assumes that there are no temperature v ariations during the measurement period.
321924C-01 J an03 *321924C-01* W3 Nulling circuit of Channel 2 is enabled (factory default setting) Nulling circuit of Channel 2 is disabled W4 Nulling circuit of Channel 3 is enabled (factory default setting) Nulling circuit of Channel 3 is disabled Table 4.
Ein wichtiger Punkt beim Kauf des Geräts National Instruments SCXI-1321 (oder sogar vor seinem Kauf) ist das durchlesen seiner Bedienungsanleitung. Dies sollten wir wegen ein paar einfacher Gründe machen:
Wenn Sie National Instruments SCXI-1321 noch nicht gekauft haben, ist jetzt ein guter Moment, um sich mit den grundliegenden Daten des Produkts bekannt zu machen. Schauen Sie zuerst die ersten Seiten der Anleitung durch, die Sie oben finden. Dort finden Sie die wichtigsten technischen Daten für National Instruments SCXI-1321 - auf diese Weise prüfen Sie, ob das Gerät Ihren Wünschen entspricht. Wenn Sie tiefer in die Benutzeranleitung von National Instruments SCXI-1321 reinschauen, lernen Sie alle zugänglichen Produktfunktionen kennen, sowie erhalten Informationen über die Nutzung. Die Informationen, die Sie über National Instruments SCXI-1321 erhalten, werden Ihnen bestimmt bei der Kaufentscheidung helfen.
Wenn Sie aber schon National Instruments SCXI-1321 besitzen, und noch keine Gelegenheit dazu hatten, die Bedienungsanleitung zu lesen, sollten Sie es aufgrund der oben beschriebenen Gründe machen. Sie erfahren dann, ob Sie die zugänglichen Funktionen richtig genutzt haben, aber auch, ob Sie keine Fehler begangen haben, die den Nutzungszeitraum von National Instruments SCXI-1321 verkürzen könnten.
Jedoch ist die eine der wichtigsten Rollen, die eine Bedienungsanleitung für den Nutzer spielt, die Hilfe bei der Lösung von Problemen mit National Instruments SCXI-1321. Sie finden dort fast immer Troubleshooting, also die am häufigsten auftauchenden Störungen und Mängel bei National Instruments SCXI-1321 gemeinsam mit Hinweisen bezüglich der Arten ihrer Lösung. Sogar wenn es Ihnen nicht gelingen sollte das Problem alleine zu bewältigen, die Anleitung zeigt Ihnen die weitere Vorgehensweise – den Kontakt zur Kundenberatung oder dem naheliegenden Service.