Benutzeranleitung / Produktwartung NI 9233 des Produzenten National Instruments
Zur Seite of 34
OPERATING INSTRUCTIONS AND S PECIFICATIONS NI 9233 4-Channel, ± 5 V, 24-Bit IEPE Analog Input Module ni.com/manuals Deutsch Français.
NI 9233 Operating I nstruction s and Specificati ons 2 ni.com This doc u ment describes ho w to u se the National Instr u ments 9 233 and incl u des specif ications and con nector assignments for the NI 9233. V isit ni.com/info and enter rd s oftwarever s ion to determine which software yo u need for the mod u les yo u are u sing.
© National Inst ruments Corp. 3 NI 9233 Operating Instructi ons and Specific ations Safety Guidelines Operate the NI 9233 only as described in these operating instr u ctions. Hot Surface This icon denotes that the component may be hot. T o u ching this component may res u lt in bodily inj u ry .
NI 9233 Operating I nstruction s and Specificati ons 4 ni.com Caution S u bstit u tion of components may impair s u itability for Class I, Division 2. Caution For Zone 2 applications, install the system in an enclos u re rated to at least IP 54 as defined b y IEC 60529 an d EN 60529.
© National Inst ruments Corp. 5 NI 9233 Operating Instructi ons and Specific ations Special Conditions fo r Marine Appli cations Some mod u les are Lloyd’ s Register (LR) T ype Approv ed for marine applications. T o verify Llo yd’ s Register cert if ication, visit ni.
NI 9233 Operating I nstruction s and Specificati ons 6 ni.com Connecting the NI 9233 The NI 9233 has fo u r BNC connectors that provide connections for fo u r sim u ltaneo u sly sampled analog inp u t channels.
© National Inst ruments Corp. 7 NI 9233 Operating Instructi ons and Specific ations Each channel has a BNC connector to which yo u can connect an Integrat ed Electronic Piez oelectri c (IEPE) sensor . The center pin of the connector , AI+, provides the DC excitation and the A C signal connection.
NI 9233 Operating I nstruction s and Specificati ons 8 ni.com Figure 2. Connecting a Grou nded IEPE Sensor to the NI 9233 Figure 3. Connecting a Floating IEPE Sensor to the NI 9233 NI 9233 IEPE Sensor.
© National Inst ruments Corp. 9 NI 9233 Operating Instructi ons and Specific ations The NI 9233 analog inp u t channels are referenced to the chassis gro u nd thro u gh a 50 Ω resistor . T o minimize gro u nd noise, make s u re that the chassis gro u nd is connected to the earth gro u nd.
NI 9233 Operating I nstruction s and Specificati ons 10 ni.com The NI 9233 also has TEDS circ u itry . For more information abo u t TEDS, visit ni.com/info and enter rdted s .
© National Inst ruments Corp. 11 NI 9233 Operating Instructi ons and Specific ations freq u ency depends on th e data rate. Fig u re 5 sho ws typical passband flatness for data rates above 25.65 kS/s and less than or eq u al to 25.65 kS/s. Figure 5. T ypical Passba nd Flat ness for the NI 9233 F requency/Data Rate 0.
NI 9233 Operating I nstruction s and Specificati ons 12 ni.com The relati ve phases of th ese signals also ha ve a freq u ency-dependent delay . The vari ation in the phase delay with freq u ency is called the phase nonlinearity . Fig u re 6 shows the phase nonlinearity for data rates abo ve 2 5.
© National Inst ruments Corp. 13 NI 9233 Operating Instructi ons and Specific ations Figure 6. Phase Nonlinearity of the NI 9233 0.45 0.15 4 2 0 –2 –4 0.10 0.05 0.00 0.40 0.35 0.30 0.25 0.20 Data Rate ≤ 25.65 kS/s Phase Error (Degrees) Data Rate > 25.
NI 9233 Operating I nstruction s and Specificati ons 14 ni.com Stopband The filter signif icantly atten u ates all signals abo ve the stop band freq u ency . The primary goal of the filter is to pre vent aliasing. Therefore, the stopband freq u ency scales precisel y with the data rate.
© National Inst ruments Corp. 15 NI 9233 Operating Instructi ons and Specific ations share a single ma ster timebase so u rce. Refer to the software help for information abo u t conf ig u ring the master timebase so u rce for the NI 9233. V isit ni.com/info and enter c s erie s doc for information abo u t C Series doc u mentation.
NI 9233 Operating I nstruction s and Specificati ons 16 ni.com Note The cRIO-9151 R Series Expansion chassis does not s u pport sharing timebases between mod u les. Sleep Mode This mod u le s u pports a low-po wer sleep mode. S u pport for sleep mode at the system le vel depe nds on the chassis that the mod u le is pl u gged into.
© National Inst ruments Corp. 17 NI 9233 Operating Instructi ons and Specific ations Specifications The follo wing specifications a re typical for the range – 40 to 70 °C u nless otherwise noted. Input Characteristics N u mber of channels ..... ..
NI 9233 Operating I nstruction s and Specificati ons 18 ni.com Data rates 1 ( f s ) f s ≤ 25.65 kS/s .......... .............. ..... , n = 2, 3, ..., 25 f s > 25.65 kS /s .............. ........... ... , n = 2, 3 Inp u t co u pling . ........... .
© National Inst ruments Corp. 19 NI 9233 Operating Instructi ons and Specific ations AC c u toff freq u ency response Inp u t range . ............. ............ ............. . ±5 V AC v o l t a g e f u ll-scale range Minim u m ........ ............
NI 9233 Operating I nstruction s and Specificati ons 20 ni.com Common-mode vo ltage range (AI– to earth gro u nd) .............. .......... ±2 V max IEPE excitation c u rrent Minim u m ........ ........... ............ ..... 2.0 mA T ypica l ......
© National Inst ruments Corp. 21 NI 9233 Operating Instructi ons and Specific ations Overv oltage protection (with respect to chassis gro u nd) For an IEPE se nsor connected to AI+ and AI– .... ............. .......... ±30 V For a lo w-impedance so u rce connected to AI+ and AI – .
NI 9233 Operating I nstruction s and Specificati ons 22 ni.com Channel-to-channel matching Gain T ypical ...... .............. ............. . 0.07 dB Maxim u m ...... ............. .......... 0.27 dB Phase ( f in in kHz) .. ........... .......... f in · 0.
© National Inst ruments Corp. 23 NI 9233 Operating Instructi ons and Specific ations Crosstalk ( f in = 1 kHz) Paired channels (0 and 1, 2 and 3)...... .......... ........ –100 dB Nonpaired channels ...... .............. –110 dB CMRR ( f in ≤ 1 kHz) Minim u m .
NI 9233 Operating I nstruction s and Specificati ons 24 ni.com T otal harmonic distortion (THD) Intermod u lation distortion (–1 dBFS) DIN 250 Hz/8 kHz 4:1 amplit u de ratio ......... ............ –80 dB CCIF 11 kHz/12 kHz 1:1 amplit u de ratio ..
© National Inst ruments Corp. 25 NI 9233 Operating Instructi ons and Specific ations Power Requirements Power cons u mption fr om chassis Active mode .............. ............ ..... 620 mW max Sleep mode .......... ........... ......... ... 0.5 mW max Thermal dissipation (at 70 °C) Active mode .
NI 9233 Operating I nstruction s and Specificati ons 26 ni.com Hazardous Locations U.S. (UL) .......... ........... ........... .......... Cla ss I, Division 2, Gro u ps A, B, C, D, T4; Class I, Zone 2, AEx nC IIC T4 Canada (C-UL) .......... .........
© National Inst ruments Corp. 27 NI 9233 Operating Instructi ons and Specific ations Electromagnetic Compatibility This prod u ct meets the req u irements of the follo wing EMC standards for electric.
NI 9233 Operating I nstruction s and Specificati ons 28 ni.com CE Compliance This prod u ct meets the essent ial req u irem ents of app licable E u ropean Directi ves as follows: • 2006/95/EC; Low-V.
© National Inst ruments Corp. 29 NI 9233 Operating Instructi ons and Specific ations Operating shock (IEC 60068-2-27)........... ........... ........ 30 g, 11 ms half sine, 50 g, 3 ms half sine, 18 s.
NI 9233 Operating I nstruction s and Specificati ons 30 ni.com Maxim u m altit u de.......... ......... .......... 2,000 m Poll u tion Degree (IEC 60664) .......... 2 Environmental Mana gement NI is committed to designing and man u fact u ring prod u cts in an en vironmentally responsibl e manner .
© National Inst ruments Corp. 31 NI 9233 Operating Instructi ons and Specific ations Calibration Yo u can obtain the calibration cert ificate and information abo u t calibration services for the NI 9233 at ni.com/calibration . Calibration interval ..
NI 9233 Operating I nstruction s and Specificati ons 32 ni.com Where to Go for Support The National Instr u ments W eb site is yo u r complete reso u rce for technical s u pport.
© National Inst ruments Corp. 33 NI 9233 Operating Instructi ons and Specific ations Korea 82 02 3451 3400, Lebanon 961 (0) 1 33 28 28, Malaysia 1800 887710, Mexico 01 800 010 0793, Netherlands 31 (0.
© 2004–2008 Nati onal Instruments Corp. All rights reserved. 373784F-01 Sep08 National Instruments, NI, n i.com, and LabVIEW are trademarks of National Instrument s Corporation. Refer to the Terms of Use section on ni.com/legal for more information about National Instruments trademarks.
Ein wichtiger Punkt beim Kauf des Geräts National Instruments NI 9233 (oder sogar vor seinem Kauf) ist das durchlesen seiner Bedienungsanleitung. Dies sollten wir wegen ein paar einfacher Gründe machen:
Wenn Sie National Instruments NI 9233 noch nicht gekauft haben, ist jetzt ein guter Moment, um sich mit den grundliegenden Daten des Produkts bekannt zu machen. Schauen Sie zuerst die ersten Seiten der Anleitung durch, die Sie oben finden. Dort finden Sie die wichtigsten technischen Daten für National Instruments NI 9233 - auf diese Weise prüfen Sie, ob das Gerät Ihren Wünschen entspricht. Wenn Sie tiefer in die Benutzeranleitung von National Instruments NI 9233 reinschauen, lernen Sie alle zugänglichen Produktfunktionen kennen, sowie erhalten Informationen über die Nutzung. Die Informationen, die Sie über National Instruments NI 9233 erhalten, werden Ihnen bestimmt bei der Kaufentscheidung helfen.
Wenn Sie aber schon National Instruments NI 9233 besitzen, und noch keine Gelegenheit dazu hatten, die Bedienungsanleitung zu lesen, sollten Sie es aufgrund der oben beschriebenen Gründe machen. Sie erfahren dann, ob Sie die zugänglichen Funktionen richtig genutzt haben, aber auch, ob Sie keine Fehler begangen haben, die den Nutzungszeitraum von National Instruments NI 9233 verkürzen könnten.
Jedoch ist die eine der wichtigsten Rollen, die eine Bedienungsanleitung für den Nutzer spielt, die Hilfe bei der Lösung von Problemen mit National Instruments NI 9233. Sie finden dort fast immer Troubleshooting, also die am häufigsten auftauchenden Störungen und Mängel bei National Instruments NI 9233 gemeinsam mit Hinweisen bezüglich der Arten ihrer Lösung. Sogar wenn es Ihnen nicht gelingen sollte das Problem alleine zu bewältigen, die Anleitung zeigt Ihnen die weitere Vorgehensweise – den Kontakt zur Kundenberatung oder dem naheliegenden Service.