2 Leica lamp housing and mercury burner T able of contents Page 1. Safety concept . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3 1.1 Description . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
1.1 Description 1.2 Safety precautions For the Leica MZ16 FA 3 Leica lamp housing and mercury burner 1. Safety concept The Leica 106 z lamp housing is the light source for the fluorescence stereomicroscope Leica MZ16 F A as well as the fluorescence module for Leica stereomicroscopes.
Lamp housing The Leica MZ16 F A fluorescence stereomicroscope and the fluorescence module for Leica stereomicroscopes may only be used in combination with the Leica lamp housing for 100W or 50W mercury burners, their corresponding supply units and stray-light protection.
Supply units The 100W mercury burners are powered by the ebq 100 supply unit. Read the separate instruction manual provided with this supply unit. The supply unit for 50W mercury burners is described in this manual. For the ebq supply unit: only fuses of the specified type with the listed rated current may be used.
• Use the stray-light protection. Danger to eyes from UV radiation. • During assembly work on xenon burners, always wear the protective gloves and face protection provided. Risk of explosion • Electrical components must be assembled at least 10 cm away from the wall and from flammable substances.
The Leica Microsystems 106 z lamp housing is used. If the 105 z lamp housing model is available, it can also be used. Follow the safety instructions! The following gas discharge lamps can be used. They require different types of lamp mounts. Order No.
Unscrew the 2 hexagon-head screws. Carefully pull out the base with lamp mount and connecting cable. Unscrew the top Phillips screw . Pull up the cooling element and pull it to the side. Unscrew the bottom clamping system. Remove the transport anchor and store it for later transport.
Adjust the collector on a trial basis using the focusing knob. • During this process, the collector may not come into contact with the flexible power supply . If necessary , bend the power supply away from the collector . Carefully close the lamp housing and make sure that the contact plug is set firmly in the socket.
Unscrew the hexagon-head screw on the adapter part using an Allen key . Insert the lamp housing into the adapter part and tighten the hexagon-head screw .
Pull the focusing knob outward. Carefully replace the collector and push it to the left until the focusing knob comes to a stop in the guide groove. Adjust the collector on a trial basis. Reinstall the lamp base and close the lamp housing.
4.1 Supply unit for lamp housing 106 z with 50W mercury lamp 12 Leica lamp housing and mercury burner T ype 301-185.003/10372 Input: 220V/240V ±10% 220V ~ 50/60Hz Hg 50W L1/L2 Osram No. 0-958 P . Max. 320V A 1. On/off switch 2. Operating time meter 3.
T o minimize the risk of explosion, replace a discolored high-pressure mercury burner promptly when it has reached the end of its rated life. Follow the manufacturer’ s specifications and the minute meter on the supply unit! Switch on the supply unit.
14 Leica lamp housing and mercury burner After installing the mercury burner , adjust it exactly as de- scribed. Never let the bright tips of the electric arcs, the focal spots, cross when projected. Danger of explosion from overheating Adjustment is also important to obtain a homogenous light surface and good fluorescence.
15 Leica lamp housing and mercury burner 3 2 1 5 4 6 Hg 50W Hg 100W Fig. a: Focus the arc. Fig. b: Position the arc at the crosshairs. Fig. c: Focus and position the second arc.
Leica Microsystems – the brand for outstanding products Leica Microsystems – an international company with a strong network of customer services Australia: Gladesville, NSW T e l. +1 800 625 286 Fax +61 2 9817 8358 Austria: Vienna T e l. +43 1 486 80 50 0 Fax +43 1 486 80 50 30 Canada: Richmond Hill/Ontario Tel.
Ein wichtiger Punkt beim Kauf des Geräts Leica MZ16 FA (oder sogar vor seinem Kauf) ist das durchlesen seiner Bedienungsanleitung. Dies sollten wir wegen ein paar einfacher Gründe machen:
Wenn Sie Leica MZ16 FA noch nicht gekauft haben, ist jetzt ein guter Moment, um sich mit den grundliegenden Daten des Produkts bekannt zu machen. Schauen Sie zuerst die ersten Seiten der Anleitung durch, die Sie oben finden. Dort finden Sie die wichtigsten technischen Daten für Leica MZ16 FA - auf diese Weise prüfen Sie, ob das Gerät Ihren Wünschen entspricht. Wenn Sie tiefer in die Benutzeranleitung von Leica MZ16 FA reinschauen, lernen Sie alle zugänglichen Produktfunktionen kennen, sowie erhalten Informationen über die Nutzung. Die Informationen, die Sie über Leica MZ16 FA erhalten, werden Ihnen bestimmt bei der Kaufentscheidung helfen.
Wenn Sie aber schon Leica MZ16 FA besitzen, und noch keine Gelegenheit dazu hatten, die Bedienungsanleitung zu lesen, sollten Sie es aufgrund der oben beschriebenen Gründe machen. Sie erfahren dann, ob Sie die zugänglichen Funktionen richtig genutzt haben, aber auch, ob Sie keine Fehler begangen haben, die den Nutzungszeitraum von Leica MZ16 FA verkürzen könnten.
Jedoch ist die eine der wichtigsten Rollen, die eine Bedienungsanleitung für den Nutzer spielt, die Hilfe bei der Lösung von Problemen mit Leica MZ16 FA. Sie finden dort fast immer Troubleshooting, also die am häufigsten auftauchenden Störungen und Mängel bei Leica MZ16 FA gemeinsam mit Hinweisen bezüglich der Arten ihrer Lösung. Sogar wenn es Ihnen nicht gelingen sollte das Problem alleine zu bewältigen, die Anleitung zeigt Ihnen die weitere Vorgehensweise – den Kontakt zur Kundenberatung oder dem naheliegenden Service.