Leica Stereo- Fluorescence Systems User Manual.
.
3 Contents Page Leica MZ10 F fluorescence stereomicroscope Description .............................................................. 4 Safety concept .......................................................... 4 Fluorescence applications .............
4 Description Fluorescence technique Some substances fluoresce when irradiated with short-wave light. This property is utilized in the fluorescence technique, in which certain structures and features which do not fluo- rescence can be tagged with a fluorescing dye.
5 Servicing and repairs – Repairs may only be carried out by Leica-trained service technicians or by technical personnel authorized to do so by the person in charge of the instrument – Only original Leica spare parts may be used – Unplug the power cable before opening the supply unit.
6 Applications Natural science Application Fluorescence filter set Anatomy •Monitoring of capillaryflow •GFP 1, GFP2 Biology •Gene expression inchick.
7 Industry Application Fluorescence filter set Electronics •Solder paste onSMDs •GFP 1, GFP2 •Epoxy resin onSMD plates •GFP 1, GFP.
8 Overview Leica MZ10 F: Its components and controls 1. Choice of stand and of focusing drive (manual or motor -driven) 2. Microscope carrier for Leica MZ10 F 3a. Leica MZ10 F optics carrier with third beam path 3b. Integrated FLUOIII filter system 3c.
9 Filter sets Designation Excitation filter Barrier filter GFP fluorescence GFP1 425/60 nm 480 nm GFP Plus fluorescence GFP2 480/40 nm 510 nm GFP plant fluorescence GFP3 470/40 nm 525/50 UV fluorescen.
10 Leica MZ10 F Assembly 1. Incident- or transmitted-light base 2a. Manual focusing drive, or 2b. Motorized focusing drive 3. Microscope carrier for Leica MZ10 F 4. Leica MZ10 F optics carrier with FLUOIII™ filter system 5a. Lamp housing 106Z 5b. Stray-light protection 6.
11 Leica MZ10 F Focusing drive ➜ base of stand ▶ Connect manually-operated or motor -driven focusing drive, complete with column, to base, as described in user manual M2-105-0EN. Before using the motor focus, you must read the safety directions in the accompanying user manual M2-267- 104.
12 Assembly Lamp housing The Leica 106Z lamp housing is equipped. For information about the supply units for the 106Z lamp housing, see page 14. Model 105Z lamp housings, if available, can also be used.
13 ▶ Using the focusing knob, displace the collector: • Duringthis movement, the collector must not touch the flexible power cable. If necessary , bend the cable away ▶ Carefully .
14 Assembly Supply unit for lamp housing 106Z with 50W mercury-vapour burner 1. On/off switch 2. Running-time meter 3. Selector for L1/L2 burners 4. Fuse 5. Selectorfor 50Hz/60Hz 6. Connector for lamp 7. Power cable The supply unit is intended for power supplies above 220V .
15 3 1 2 4 5 6 Hg50W Hg 100W Adjusting the mercury vapour lamp This adjustment is important for obtaining a uniform light spot and good-quality fluorescence.
Dummy filter carrier In the dummy filter carrier there are two openings for the observation beam paths instead of the barrier filters. The third path has been closed. Use the dummy filter carrier: – if you are using less than four sets of filters – if you are working without the fluorescence illuminator for periods of up to 15 seconds.
17 Use MZ10 F: Bringing the specimen into focus and observing it Observe the safety directions on page 4 and the recom- mendations on page 5 regarding the use of the mercu- ry-vapour lamp. Changing filters In the rapid filter changer , all four positions must be occupied, either with filter sets or with dummy filter carriers (see page 16).
18 Overview Outfit with stereo-fluorescence module The components 1 Choice of stand and of focusing drive (manual or motor -driven) 2 Microscope carrier 3 Leica MS5, MZ6, MZ7.
19 Filter sets Excitation Dichromatic Barrier filters beam splitters filters GFP fluorescence 425/60nm 470nm GG475 GFP-Plus fluorescence 480/40nm 505nm LP 510nm LP GFP-Plant 470/40nm 495nm 525/50nm UV.
20 Assembly Stereo-fluorescence module 1 Incident- or transmitted-light base 2a Manual focusing drive, or 2b Motorized focusing drive 3 Microscope carrier 4 Choice of optics carrier 5 Fluorescence mod.
21 Important: With the illuminator on, insert the filterhol- der with diaphragm into slide-in unit 3 to avoid being suddenly exposed to excessively bright light (page 22). ▶ Remove the binocular tube. ▶ Loosen the 3 socket-head screws. ▶ Remove the cover .
22 Assembly Stereo-fluorescence module UV-light excluder ➜ side-faced column Always position the UV-light excluder so that the user never looks directly at the light spot (page 5). ▶ Using the clamping screw , secure the clamp on the side- faced column, beneath the focusing drive.
23 Use Stereo-fluorescence module Focusing and observing Observe the safety regulations on page 4 and note the recommendations on page 5 for using the mercury- vapour burner .
L ei c a M i c r os y s te ms – th e b r a n d f o r o u ts tan d i n g p r od u c ts Leica Microsystems – an international company with a strong network of customer services. Australia: Gladesville T el. +61 2 9879 9700 Fax +61 2 9817 8358 Austria: Vienna T el.
Ein wichtiger Punkt beim Kauf des Geräts Leica 106Z (oder sogar vor seinem Kauf) ist das durchlesen seiner Bedienungsanleitung. Dies sollten wir wegen ein paar einfacher Gründe machen:
Wenn Sie Leica 106Z noch nicht gekauft haben, ist jetzt ein guter Moment, um sich mit den grundliegenden Daten des Produkts bekannt zu machen. Schauen Sie zuerst die ersten Seiten der Anleitung durch, die Sie oben finden. Dort finden Sie die wichtigsten technischen Daten für Leica 106Z - auf diese Weise prüfen Sie, ob das Gerät Ihren Wünschen entspricht. Wenn Sie tiefer in die Benutzeranleitung von Leica 106Z reinschauen, lernen Sie alle zugänglichen Produktfunktionen kennen, sowie erhalten Informationen über die Nutzung. Die Informationen, die Sie über Leica 106Z erhalten, werden Ihnen bestimmt bei der Kaufentscheidung helfen.
Wenn Sie aber schon Leica 106Z besitzen, und noch keine Gelegenheit dazu hatten, die Bedienungsanleitung zu lesen, sollten Sie es aufgrund der oben beschriebenen Gründe machen. Sie erfahren dann, ob Sie die zugänglichen Funktionen richtig genutzt haben, aber auch, ob Sie keine Fehler begangen haben, die den Nutzungszeitraum von Leica 106Z verkürzen könnten.
Jedoch ist die eine der wichtigsten Rollen, die eine Bedienungsanleitung für den Nutzer spielt, die Hilfe bei der Lösung von Problemen mit Leica 106Z. Sie finden dort fast immer Troubleshooting, also die am häufigsten auftauchenden Störungen und Mängel bei Leica 106Z gemeinsam mit Hinweisen bezüglich der Arten ihrer Lösung. Sogar wenn es Ihnen nicht gelingen sollte das Problem alleine zu bewältigen, die Anleitung zeigt Ihnen die weitere Vorgehensweise – den Kontakt zur Kundenberatung oder dem naheliegenden Service.