Benutzeranleitung / Produktwartung 76-907 des Produzenten Fluke
Zur Seite of 26
Nuclear Associates 76-907 and 76-908 AAPM MRI Phantoms Users Manual March 2005 Manual No. 38616 Rev. 3 © 2003, 2005 Fluke Corporat i on, All rights reserved .
Fluke Biomedical Radiation Management Services 6045 Cochran Road Cleveland, Ohio 44139 440.498.2564 www.flukebiomedical.com/rms.
i Table of Contents Section 1: General Info rmation ................................................................................... 1-1 1.1 Introduc tion .............................................................................................
(Blank page).
General Information Introduction 1 1-1 Section 1 General Information 1.1 Introduction This user's manual descr ibes a set of MRI phantoms that we re manufactured ba sed on the AAPM recommendations and thei r use in measuring t he MRI system performan ce.
Nuclear Associates 76-908 & 76-907 Operators Manual 1-2 1.2.1 3D Resolution and Slice (3DRAS) Phantom (Model 76-908) The 3DRAS phantom has outer dim ensions of 6" x 6" x 5" (Figure 1-1).
General Information Phantom Description 1 1-3 Figure 1-3. Resolution Sections in Three Imaging Pl anes The slice-thickness phanto m consists of two crossed thin ramps (Figure 1-4 ). Figure 1-4. Slice Thickness Phantom Triangular blocks make 4" x 4" sq uare with diagonal gaps of 1 or 2 mm.
Nuclear Associates 76-908 & 76-907 Operators Manual 1-4 A ramp-crossing angle of 90° yields an a ngle of 45° between the ramp and the imaging slice plane. Ea ch slice thickness section con sists of four triangular blocks arrang ed to form (signal producing) h ot ramps filling the gaps of 1 mm or 2 mm (Figure 1-5).
General Information Phantom Description 1 1-5 MR Phantom for Spatial Linearity, Signal-To-Noise, and Image Artifact Figure 1-6. UAL Phantom MRI Phantom for Spatial Linearity, Signal-To-Noise, and Image Artifact Figure 1- 7.
Nuclear Associates 76-908 & 76-907 Operators Manual 1-6 The flood phantom has a uniform solution, the image of which can be used to measure the uniformity in signal intensity in the images. The grid section con s ists of grids that can be used to assess geometri c linearity.
Operation Phantom Preparation 2 2-1 Section 2 Operation 2.1 Phantom Preparation The phantoms can be shipped filled with solutions or solutions can be shipped separately.
Nuclear Associates 76-908 & 76-907 Operators Manual 2-2 2.2 Preparation for Scanning 2.2.1 Positioning the Phantom The 3DRAS phantom can be pla ced in a head coil or a body coil. The center of the phantom should coincide approximately with the center of the RF coil.
Operation Tests with 3D Resolution and Slice (3DRAS) Phantom 2 2-3 resonance frequency adju stment and some may be completely automated. Resonance frequency shoul d be recorded daily for trend analysis. Values of resonance frequency should gene rally not deviate by more tha n 50 ppm between successive daily measurements.
Nuclear Associates 76-908 & 76-907 Operators Manual 2-4 • The high contrast resolution should rem ain constant for repeated measurements und er the same scan conditions and should be equ al to the pixel size. For example, for a 25.6 cm field of view with a 256 x 256 acquisitions m atrix, the resolution should be 1 mm.
Operation Tests with 3D Resolution and Slice (3DRAS) Phantom 2 2-5 direction) is measured and relate d to the slice posi tion. For a 45° ramp, the distance from a centered reference pin to the slice profile midpoint will be equ al to the point of the ramps located at the isocenter.
Nuclear Associates 76-908 & 76-907 Operators Manual 2-6 Figures 2-2 and 2-3 represent two adja cent sagital vi ews of the phantom. The resolution hole s are well resolved in vertical and ho rizontal direction and the two slices are 6 mm apart (SP = 14.
Operation Tests with 3D Resolution and Slice (3DRAS) Phantom 2 2-7 Figures 2-4 and 2-5 show two adj acent coronal view s of the phantom. The slice thickne ss lines of 1 mm ramps on the right and bottom of the images are too faint to be seen . Figure 2-5 Figure 2-4.
Nuclear Associates 76-908 & 76-907 Operators Manual 2-8 Figure 2-6 shows a resolution hole imag e against a background le ss opaque than in previou s images, indicating that the position of the resolution blo ck is su ch that it is partially covered by the slice.
Operation Tests with 3D Resolution and Slice (3DRAS) Phantom 2 2-9 Figures 2-7 and 2-8 show an exampl e of the slice thickness measureme nt analysis. A signal intensity profile is drawn vertically through a set of slice thi c kness ramp Images. The profile is the sli ce thickness profile.
Nuclear Associates 76-908 & 76-907 Operators Manual 2-10 2.4 Tests With Uniformity and Linearity (UAL) Phantom 2.4.1 Image Uniformity Image uniformity refers to the ability of the MR imag ing syste m to produce a constant signal re sponse throughout the scanned volume when th e object bei ng imaged has homogeneou s MR characteristics.
Operation Tests with Uniformity and Linea rity (UAL) Phantom 2 2-11 (2) Non-linear m agnetic field gradients Variability is best observed over the largest field-of-view. The phantom should occupy at least 60% of the largest field-of-view. Figu re 1-1 provides an illustration of a patte rn that is used to evaluate spatial linearity.
Nuclear Associates 76-908 & 76-907 Operators Manual 2-12 DC-Offset Errors DC-offset errors typically appear a s a single bright pixel (sometimes a s a dark pixel if overflow or processing has occurred) at the center of the image matrix. The existence of this error is a ssessed visually.
Operation Tests with Uniformity and Linea rity (UAL) Phantom 2 2-13 Figure 2-9.
Nuclear Associates 76-908 & 76-907 Operators Manual 2-14 Figure 2-10 is a flood image showing the uniformity o f si gnal intensity. Field non-uniformit y is also shown on the outer edges. Figure 2-10 Figure 2-11 shows a horizont al inten sity profile with a narrow window setting.
Operation Tests with Uniformity and Linea rity (UAL) Phantom 2 2-15 Figure 2-12 is an image of the signal bubble for quadr ature e rror. If there were any quadrature phase error, another ghost circle woul d be visible in the lo wer right-hand corne r of the image.
Fluke Corporation Radiation Management Services 6045 Cochran Road Cleveland, Ohio 44139 440.498.2564 www.flukebiomedical.com/rms.
Ein wichtiger Punkt beim Kauf des Geräts Fluke 76-907 (oder sogar vor seinem Kauf) ist das durchlesen seiner Bedienungsanleitung. Dies sollten wir wegen ein paar einfacher Gründe machen:
Wenn Sie Fluke 76-907 noch nicht gekauft haben, ist jetzt ein guter Moment, um sich mit den grundliegenden Daten des Produkts bekannt zu machen. Schauen Sie zuerst die ersten Seiten der Anleitung durch, die Sie oben finden. Dort finden Sie die wichtigsten technischen Daten für Fluke 76-907 - auf diese Weise prüfen Sie, ob das Gerät Ihren Wünschen entspricht. Wenn Sie tiefer in die Benutzeranleitung von Fluke 76-907 reinschauen, lernen Sie alle zugänglichen Produktfunktionen kennen, sowie erhalten Informationen über die Nutzung. Die Informationen, die Sie über Fluke 76-907 erhalten, werden Ihnen bestimmt bei der Kaufentscheidung helfen.
Wenn Sie aber schon Fluke 76-907 besitzen, und noch keine Gelegenheit dazu hatten, die Bedienungsanleitung zu lesen, sollten Sie es aufgrund der oben beschriebenen Gründe machen. Sie erfahren dann, ob Sie die zugänglichen Funktionen richtig genutzt haben, aber auch, ob Sie keine Fehler begangen haben, die den Nutzungszeitraum von Fluke 76-907 verkürzen könnten.
Jedoch ist die eine der wichtigsten Rollen, die eine Bedienungsanleitung für den Nutzer spielt, die Hilfe bei der Lösung von Problemen mit Fluke 76-907. Sie finden dort fast immer Troubleshooting, also die am häufigsten auftauchenden Störungen und Mängel bei Fluke 76-907 gemeinsam mit Hinweisen bezüglich der Arten ihrer Lösung. Sogar wenn es Ihnen nicht gelingen sollte das Problem alleine zu bewältigen, die Anleitung zeigt Ihnen die weitere Vorgehensweise – den Kontakt zur Kundenberatung oder dem naheliegenden Service.