Benutzeranleitung / Produktwartung 561s des Produzenten Fluke
Zur Seite of 16
® 561 Infrared Thermometer Users Manual PN 2562924 February 2006 Rev. 3, 4/08 © 2006-2008 Fluke C orporation, All rights reser ved. Printed in Chi na. Speci fications are su bject to change without notice. Al l product names ar e trademarks o f their respectiv e companies.
LIMITED WARRANTY AND LIMITATION OF LIABILIT Y This Fluke product will be free from defects in material and workmanship for two years from the date of purch ase. T his warranty does not cover fuses, disposable batte ries, or damage from accident, neglect, misuse, alteration, contamination, or abnorm al conditions of operation or handling.
i Table of Contents Title Page Introducti on ................................................................................ 1 Contacting Fl uke ........................................................................ 1 Safety Info rmation ..........
561 Users Manual ii.
1 561 Infrared Thermometer Introduction The Fluke 561 Infrared Thermometer (hereafter, the Ther mometer) can deter mine the surface temperature by measuring the amo unt of infrared energy radiated by the targ et’s surface or by contact using a thermocouple probe.
561 Users Manual 2 Safety Information XW Warning A Warning identifies conditions and actions that pose hazards to the user. To avoid electrical shock or pe rsonal in jury , follow these guidelines: • * Do not point laser directly at eye or indirec tly off reflect ive surfaces.
Infrared Thermometer Features 3 A V O ID E X P O S U R E - L A S E R R A D I A T IO N IS E M IT T E D F R O M T H IS A P E R T U R E CA UTION L A S E R R A D IA T IO N - D O N O T S T A R E IN T O B E.
561 Users Manual 4 AA Batteries (2) (+) Display Function Buttons Laser T rigger Battery Cover efh007f.eps Figure 2. Infrared Thermometer Display The primary temperature di splay reports the cu rre nt or last IR temperature read until the 7- second hold time elapses.
Buttons and Connector 5 3 4 5 6 2 1 8 7 efh01f.eps A Laser “On” symbol B SCAN or HOLD C ° C/ ° F sy mbol (Celsius/Fahrenheit) D Primary temperature display E Secondary temperature display F Emiss ivity LO, MED, HI G Temperature v alues for the MIN, MAX, DIF, KTC.
561 Users Manual 6 How the Thermometer Works Infrared thermometers measure the surface temperature of an opaq ue object. The Thermometer’s optics sense in frared energy, wh ich is collecte d and focused onto a dete ctor.
Operating the Thermometer 7 D:S=12:1 38 mm @ 300 mm 75 mm @ 900 mm 132 mm @ 1500 mm 1.5 " @ 12 " 3 " @ 36 " 5.3 " @ 60 " S D efh005f.eps Figure 5. Distance and Spot Size Field of View Make sure that the target is larger than the spot size.
561 Users Manual 8 Table 2. Surf ace Emis sivit y Measured Surface Switch Setting Measured Surface Switch Setting Aluminum Iron, Cast Oxidized Low Oxidized High, Medium Alloy A3003 Unoxidized Low Oxid.
Operating the Thermometer 9 Switch efh012f.eps Figure 7. Switching Between ° C and ° F Using the Contact Temperature Probe XW Warning To avoid electrical shock or personal inju ry, do not connect the optional external probe to live electrica l circuits.
561 Users Manual 10 Table 3 lists recommended Flu k e temperature probes for use w ith the Thermometer: Table 3. Recommended Temperature Probes Probe Usage 80PK-25 The piercing probe is the most versatile option.
Troubleshooting 11 Troubleshooting Symptom Problem Action --- (on display) Target temperature is over or under range Select target within speci ficat ions B Low battery Replace battery Blank display Possible dead battery Check and/or replace battery Laser does not work 1.
561 Users Manual 12 Wrap Thermocouple Probe ( model-specific) Type.............................................................................. Ty pe K with miniconnec tor and Velcro strap, ASTM E230-03 Standard Tolerance Measurement Rang e .........
Ein wichtiger Punkt beim Kauf des Geräts Fluke 561s (oder sogar vor seinem Kauf) ist das durchlesen seiner Bedienungsanleitung. Dies sollten wir wegen ein paar einfacher Gründe machen:
Wenn Sie Fluke 561s noch nicht gekauft haben, ist jetzt ein guter Moment, um sich mit den grundliegenden Daten des Produkts bekannt zu machen. Schauen Sie zuerst die ersten Seiten der Anleitung durch, die Sie oben finden. Dort finden Sie die wichtigsten technischen Daten für Fluke 561s - auf diese Weise prüfen Sie, ob das Gerät Ihren Wünschen entspricht. Wenn Sie tiefer in die Benutzeranleitung von Fluke 561s reinschauen, lernen Sie alle zugänglichen Produktfunktionen kennen, sowie erhalten Informationen über die Nutzung. Die Informationen, die Sie über Fluke 561s erhalten, werden Ihnen bestimmt bei der Kaufentscheidung helfen.
Wenn Sie aber schon Fluke 561s besitzen, und noch keine Gelegenheit dazu hatten, die Bedienungsanleitung zu lesen, sollten Sie es aufgrund der oben beschriebenen Gründe machen. Sie erfahren dann, ob Sie die zugänglichen Funktionen richtig genutzt haben, aber auch, ob Sie keine Fehler begangen haben, die den Nutzungszeitraum von Fluke 561s verkürzen könnten.
Jedoch ist die eine der wichtigsten Rollen, die eine Bedienungsanleitung für den Nutzer spielt, die Hilfe bei der Lösung von Problemen mit Fluke 561s. Sie finden dort fast immer Troubleshooting, also die am häufigsten auftauchenden Störungen und Mängel bei Fluke 561s gemeinsam mit Hinweisen bezüglich der Arten ihrer Lösung. Sogar wenn es Ihnen nicht gelingen sollte das Problem alleine zu bewältigen, die Anleitung zeigt Ihnen die weitere Vorgehensweise – den Kontakt zur Kundenberatung oder dem naheliegenden Service.